Quick Start Instructions
VIP 4400 Vessel Washer (continued)
DISTEK, INC. Rev: A 3844-0001
Quick Start Instructions: VIP 4400 Rev Date: 01/17/05 Page 2 of 2
In order to remove most of the residual media from
the wash cycle, it may be necessary to remove the
UWM from the vessel, gently shake UWM above the
vessel, reinsert it into the vessel and execute a
manual drain.
Setting the Heater (Optional)
The optional heating module allows the rinse media to be heated up to 60ºC prior to
cleaning the vessels. This allows for a more efficient wash. Pressing the
ON/OFF/SET button will toggle the heater on and off.
1. Ensure the unit is powered on.
2. Press and hold the ON/OFF/SET button on the main keypad.
3. The temperature display will begin blinking. Using the Up or Down Arrow
Keys change the blinking display to the desired set temperature.
4. Press and hold the ON/OFF/SET to confirm the new set temperature.
5. Press the ON/OFF/SET button to activate the heater. The Heater LED should
be illuminated illustrating that the unit is now heating.
6. Once the set temperature is reached (the display will update as the tank heats)
the unit is ready for operation.
Emptying the Waste Tank
The tubing supplied with the unit MUST be
connected to the quick-connect port located on
the rear cover prior to trying to empty the Waste
Tank.
Prolonged storage of acidic media in the Waste Tank
can cause irreversible damage to the tank. It is
recommended to empty the Waste Tank everyday.
1. Connect the supplied tubing to the quick-connect fitting located on the rear
cover.
2. Place the other side (outlet) of this tubing into the waste containment area
(waste drum, sink, etc…).
3. Power the unit on.
4. Press the Empty Waste Tank button located on the rear cover for 2-3 seconds
to activate the waste pump.
5. When the Waste Tank is empty the draining will automatically stop. If
necessary press and hold the Empty Waste Tank button to ensure the tank is
completely empty