
3
Elastocon Manual – Cell Ageing Oven EB 19, EB 20
2.2.1 Temperature setting
Tap on the ”SV” box (Set value) to set the test temperature
(Fig 2.2.1a).
A numeric keyboard will show. Type in the set temperature
withthenumerickeyboardandnishwithataponthe
return button (Fig 2.2.1b).
Or use the panel keys to mark the SV display and
adjust the temperature setting.
Press the arrow down or up to mark the SV box and
then use the arrow key left and right to adjust the value.
The ”PV” box is an indicator and shows the present
temperature in the oven.
2.2.2 Test name (Fig 2.2.2)
Press the grey box labeled Test name, on the main screen.
Type the test name from the touch screen keyboard. One test
name row on the screen can hold up to 40 characters.
1. Use the white arrow keys on the upper left corner to swap
between different keyboard character layouts.
2. TapontheEnterkeytonalizethetestnameinput.
3. To re-enter any character on the same test name row,
it is possible to correct any character in the test name with
a tap on any of the white left or right arrow buttons in the
lower left and right corner on the screen.
4. To cancel any input changes, tap on the ESC key and
return to the previous screen.
2.2.3 Test time
In the ”Time” box (h:m left) to the right of each cell a count-
down time can be typed in (Fig 2.2.3). Tap the box, type in the
time and press Return. The time will now start countdown.
This time can only be changed when the time symbol to the
right is disabled. Press the symbol for at least 2 seconds and
a red line is shown. The time can now be changed.
This function can also be used if the user don’t want
the countdown to start immediately. Press the symbol
for at least 2 seconds until the red line is shown. Click the
“Time” box, type in the time and press enter. When the
countdown should start press the symbol (again >0 sec) until
the red line disappears.
The countdown will now start. Time can also be changed by
marking hours and use the arrows left and right.
Fig 2.2.1a Temperature setting
Fig 2.2.1b Temperature setting
Fig 2.2.2 Test name
Fig 2.2.3 Test time