EMS HFG01 User manual

Products User Manual Template – Issue 2
Document number Y0321/DOC/038 Rev. 2
Author: ...............................................................
D Cullen BEng (Hons), MSc, CEng, CMgr, MIET, MIEEE, MCMI
Test Instrumentation Manager
Checked: .....................................................................................
M S Anslow MEng, CEng, MIET MIEEE
Senior Engineer (Products)
Approved: ................................................................
D Cullen BEng (Hons), MSc, CEng, CMgr, MIET, MIEEE, MCMI
Test Instrumentation Manager
HFG01
Operation
Manual

HFG01 Operation Manual
York EMC Services Page 2 of 21 Doc. No. Y0321/DOC/038 Rev. 2
HFG01
Operation Manual
Document Number
Y0321/DOC/038
C
HANGE
H
ISTORY
S
HEET
Date Issue Pag/Para
Description
30/03/2017 1 - 1
st
Issue
11/11/17 2 Safety instructions updated
UNCONTROLLED ELECTRONIC COPY
Users should satisfy themselves that they are in possession of the most up to date
version of this manual by calling +44 (0) 330 4303456 or by emailing

HFG01 Operation Manual
York EMC Services Page 3 of 21 Doc. No. Y0321/DOC/038 Rev. 2
Limited Warranty
York EMC Services warrants each HFG01 to be free
from defects in material and workmanship for a period of
one year from the date of shipment to the purchaser.
This warranty extends to the original purchaser only. It
does not apply to any product or parts subject to misuse,
neglect, accident, unauthorised service or abnormal
conditions of operation.
In the event a product covered by this warranty fails,
York EMC Services, will, without charge, repair and
recalibrate the instrument if returned to their test
laboratory within one year of the original purchase –
provided that York EMC Services examination discloses,
to its satisfaction, that the product is defective. York EMC
Services may, at its option, replace the product in lieu of
repair. If the defect was caused by misuse, neglect,
accident, unauthorised service or abnormal conditions of
operation, repairs will be billed at a nominal cost. In such
cases, an estimate will be provided before work is
started, if requested by the purchaser.
For warranty service, contact York EMC Services.
Provide the serial number of the equipment and
complete details regarding the failure. You will then be
given shipping instructions. Return the instrument to the
factory. Repairs will be made and the equipment will be
returned to you. York EMC Services assumes no
responsibility for loss of, or damage to, products in
transit.

HFG01 Operation Manual
York EMC Services Page 4 of 21 Doc. No. Y0321/DOC/038 Rev. 2
DISCLAIMER
THE HFG01 HAS BEEN DESIGNED TO PRODUCE HARMONIC AND FLICKER
DISTURBANCES TO THE POWER SUPPLY TO WHICH THE TEST PORT HAS
BEEN CONNECTED, AND AS SUCH IT IS ONLY INTENDED TO BE
CONNECTED TO SUITABLE TEST EQUIPMENT. CONNECTION TO THE
PUBLIC MAINS DISTRIBUTION NETWORK IS NOT PERMITTED.
YORK EMC SERVICES LTD ASSUMES NO LIABILITY FOR ANY DAMAGE
THAT MAY RESULT FROM THE USE OF THIS EQUIPMENT OUTSIDE ITS
RECOMMENDED OPERATING CONDITIONS.
PRECAUTIONS
THE HFG01 MUST NOT BE OPERATED WITH THE TEST SUPPLY INPUT
PLUGGED DIRECTLY INTO A PUBLIC MAINS SUPPLY SYSTEM.
WHEN OPERATED, THE HFG01 WILL CAUSE HARMONIC AND FLICKER
EMISSIONS TO THE SUPPLY CONNECTED TO THE TEST SUPPLY INPUT.

HFG01 Operation Manual
York EMC Services Page 5 of 21 Doc. No. Y0321/DOC/038 Rev. 2
1
INTRODUCTION................................................................................................. 6
2
DESCRIPTION.................................................................................................... 7
3
DESCRIPTION OF SYMBOLS ........................................................................... 8
4
SPECIFICATION................................................................................................. 9
4.1
O
PERATING MODES
......................................................................................... 9
4.2
H
ARMONICS AND FLICKER MEASUREMENT SYSTEM TEST SUPPLY
......................... 9
4.3
O
PERATOR INDICATORS
.................................................................................. 9
4.4
P
OWER REQUIREMENTS
.................................................................................. 9
4.5
E
NCLOSURE
..................................................................................................10
4.6
C
OMPLIANCE
.................................................................................................10
4.7
E
NVIRONMENT
...............................................................................................10
5
OPERATION......................................................................................................11
5.1
C
ONTROLS AND INDICATORS
...........................................................................11
5.2
S
UPPLYING POWER TO THE
HFG01.................................................................12
5.3
O
PERATING MODES
........................................................................................12
5.4
E
NABLING THE TEST
.......................................................................................13
5.5
P
ERFORMING A TEST
......................................................................................13
5.6
T
ROUBLESHOOTING A TEST
.............................................................................14
6
OUTPUT PERFORMANCE................................................................................16
6.1
F
LICKER
L
OAD
O
PERATION
.............................................................................16
6.2
H
ARMONICS
L
OAD
O
PERATION
.......................................................................16
6.3
C
ALIBRATION
.................................................................................................17
7
SAFETY.............................................................................................................18
8
ELECTROMAGNETIC COMPATIBILITY ..........................................................19
8.1
G
ENERAL
......................................................................................................19
8.2
E
MISSIONS
....................................................................................................19
8.3
I
MMUNITY
......................................................................................................19
8.4
N
ON
-A
PPLICABILITY
.......................................................................................19
9
MAINTENANCE AND CALIBRATION...............................................................20
9.1
D
ISPOSAL
.....................................................................................................20

HFG01 Operation Manual
York EMC Services Page 6 of 21 Doc. No. Y0321/DOC/038 Rev. 2
1 I
NTRODUCTION
The Harmonics & Flicker Generator 01 (HFG01) has been designed by York EMC
Services for the purpose of providing an easy and reliable way to externally check
the performance of measurement systems used for testing against the EN/IEC
61000-3-2 harmonics and interharmonics and EN/IEC 61000-3-3 flicker standards.
The HFG01 allows the user to periodically verify their test equipment and its
operation together, helping to ensure compliance with standards and laboratory
quality procedures.
The HFG01 provides a series of harmonic and flicker disturbances of a nominal
but stable level. The HFG01 may therefore be used to verify the stability of a
measurement system. Additionally, due to its own stability it may be used as a
transfer standard from a calibrated system, or be used as a repeatable source of
interference for diagnostic and/or development design work.

HFG01 Operation Manual
York EMC Services Page 7 of 21 Doc. No. Y0321/DOC/038 Rev. 2
2 D
ESCRIPTION
The HFG01 simulates equipment under test (EUT), generating known, repeatable
levels of harmonic and flicker disturbance in one of four modes of operation.
The flicker test modes generate a fixed level of mains disturbance at a rate of 1 Hz
and 8.33 Hz. These modes support verification of the test systems used for EUT
flicker disturbance measurements as described in the EN/IEC 61000-3-3 test
standards.
The harmonics test modes are designed to produce a variety of harmonic-rich
current waveforms. These current waveforms support verification of the test
systems used for mains-powered equipment load current testing as described in
the EN/IEC 61000-3-2 test standards.
The HFG01 can be used with test systems operating at 230 Vac (nominal) supply
voltages and 50 Hz supply frequency. Connection to the test system is by a captive
lead pre-fitted with a CEE 7/7 European style plug thatfits directly with power outlet
used by a wide range of harmonics and flicker equipment.
The HFG01 itself is powered by a wide-input range AC/DC power supply unit that
is powered separately from the Test Supply under test. This ensures that the
disturbance signals generated by theHFG01 are not affected by the need to derive
the HFG01’s own supply needs from the Test Supply under test.
For ease of access all connections, user controls, and indicators are located on
the front panel of the HFG01. The HFG01 is supplied with the external power unit
with mains cable, manufacturer’s test certificate, and a CDROM containing digital
copies of the manual, test certificate, and files of the tabulated certificate data.

HFG01 Operation Manual
York EMC Services Page 8 of 21 Doc. No. Y0321/DOC/038 Rev. 2
3 DESCRIPTION
OF
SYMBOLS
Symbols displayed on the product are listed below:
Caution
Protective Earth Terminal
Alternating current
HFG01 power supply on
HFG01 power supply off
Direct current
DC connector polarity
ESD handling precaution
Separate collection for WEEE-
Waste Electrical and Electronic
Equipment

HFG01 Operation Manual
York EMC Services Page 9 of 21 Doc. No. Y0321/DOC/038 Rev. 2
4 S
PECIFICATION
4.1 Operating modes
(See Section 5. Output Performance for details of typical measurement results)
4.1.1 Current harmonics
Frequency range 5 Hz to 2.0 kHz (50 Hz Test Supply)
Stability < 5% of the test limit in EN 61000-3-2 standard
4.1.2 Flicker disturbance
Flicker rate 1 Hz (50 Hz Test Supply only)
8.3 Hz (50 Hz Test Supply only)
Stability < 5% of the reading
4.2 Harmonics and flicker measurement system test supply
4.2.1 Test Supply ratings
Voltage 200 Vac to 253 Vac (230 Vac nominal input)
Current 3 A (peak)
Frequency 50 Hz
Power rating 500 W max
Connector CEE 7/7 European style plug for connecting the HFG01
to the Test Supply under test.
Insulation rating 3000 Vrms
4.3 Operator indicators
(See Section 4. Operation for a full description of the control and indicator functions)
4.3.1
Operation indicators
Test enabled / running
System alarm
4.4 Power requirements
HFG01
Input connector 2.1 mm DC power jack (inner conductor +ve)
Voltage input 24 Vdc
Power consumption 24 W max
Power supply unit
Input connector IEC plug

HFG01 Operation Manual
York EMC Services Page 10 of 21 Doc. No. Y0321/DOC/038 Rev. 2
Voltage input 100 Vac to 240 Vac (nominal)
Power consumption 50 W max
4.5 Enclosure
Overall dimensions (not including captive leads):
Length 330 mm
Width 320 mm
Height 170 mm
Weight 8 kg
4.6 Compliance
Safety Low Voltage Directive, 2014/35/EU
EMC EMC Directive, 2014/30/EU
4.7 Environment
4.7.1 Operating Temperature Range
Recommended Ambient 10ºC to 30ºC
Maximum Ambient 5ºC to 40ºC
4.7.2 Humidity
Relative humidity: 20% to 80% (non-condensing)
4.7.3 Protection
IP rating IP10
NOTE. Do not allow any obstructions to interfere with the operation of the fans, or
site the HFG01 such that airflow is restricted. Doing so will cause the equipment to
run at an increased temperature and in so doing may affect its operation.

HFG01 Operation Manual
York EMC Services Page 11 of 21 Doc. No. Y0321/DOC/038 Rev. 2
5 O
PERATION
5.1 Controls and indicators
Identifier Description
Start Test Pressing this switch enables the test mode, and applies the
chosen load profile to the measurement system, as appropriate
for the test supply detected.
Pressing this switch again whilst the test is running disables the
test mode.
Test enabled indicator Indicates that the test mode is running.
Test Mode selector This switch selects one of four load profile test modes available
to be used. Changing this setting whilst a test is running will
automatically disable the current test mode.
Alarm indicator This indicator flashes if the temperature within the enclosure
exceeds the preset limit. In this state the test mode will be
disabled and the HFG01 will wait until the temperature has
dropped sufficiently before the test can be resumed.
A steady indicator warns of a system failure, which will need to
be cleared by fully restarting the HFG01 (refer to Section 4.6).
HFG01 power switch Main On/Off switch for the HFG01.
HFG01 power indicator Indicates that the HFG01 is on.
HFG01 power connector
HFG01 power supply input socket (24 Vdc).
Test supply fuse Protective fuse for the Test Supply input
Test supply cable Captive lead connector for the Test Supply input
Chassis earth Additional earthing point for the HFG01 enclosure
Figure 1: HFG01 front panel and controls

HFG01 Operation Manual
York EMC Services Page 12 of 21 Doc. No. Y0321/DOC/038 Rev. 2
5.2 Supplying power to the HFG01
Figure 2: HFG01 power supply
The HFG01 is powered by an external 24 Vdc power supply unit which connects
to the input socket on the front panel . Pressing the power switch with the
24 Vdc unit connected will turn the HFG01 on, which will be confirmed by the
green indicator . Pressing the power switch again will turn the HFG01 off.
5.3 Operating modes
Figure 3: HFG01 mode selection
The operating mode is selected using the Test Mode switch . The operation of
each test mode is described in Table 1 below. Note that if the Test Mode switch is
changed whilst a test is running, the test is automatically disabled.
Mode Test Mode Setting Load Profile Description
Flicker
8.3Hz Produces Pst value of > 1
(above the pass limit defined in EN/IEC61000-3-3)
1 Hz Produces Pst value of < 1
(below the pass limit defined in EN/IEC61000-3-3).
Harmonics
Fixed Single repetitive harmonic-rich current waveform
Fluctuating Repeating current profile alternating between two distinct
waveforms over a 10 second cycle
Table 1: HFG01 load operation test modes

HFG01 Operation Manual
York EMC Services Page 13 of 21 Doc. No. Y0321/DOC/038 Rev. 2
5.4 Enabling the test
Figure 4: Test Enable switch and indicator
The selected test mode is enabled by pressing the Start Test switch and the
indicator displays the running test status.
5.5 Performing a test.
The HFG01 acts as the equipment under test in the harmonics and flicker test
system, and should be connected to the single-phase test power supply as per the
typical test setup used (Figure 5 for example).
Figure 5: Example of a typical test setup
Connect the HFG01 power supply and turn the HFG01 on by pressing the
power switch . The power supply indicator should be illuminated green to
indicate the HFG01 is operating correctly.
The Test Supply Input cable can be plugged into the Test Supply power output
socket. Note that if the Test Supply does not provide a safety earth connection as
one of the conductors (e.g. 2-pin Live and Neutral only) then the Chassis Earth
wiring stud on the front of the HFG01 must be connected to a suitable earth first
(not shown). A suitable earth point may be provided elsewhere on the Harmonics
and Flicker test system.

HFG01 Operation Manual
York EMC Services Page 14 of 21 Doc. No. Y0321/DOC/038 Rev. 2
Configure the Harmonics and Flicker Test System to measure harmonic or flicker
disturbance as per the system instructions, with the Test Supply configured to
provide the required voltage and frequency.
Once the Test Supply has been safely connected, the Harmonics and Flicker Test
Supply can be enabled.
Select the required test load operation using the Test Mode switch .
The chosen test load operation can be enabled before or after the Harmonics and
Flicker Test System has started the test routine. When ready to enable the load
operation, press the Start Test switch . The Test Enable display should
indicate that the test is running.
Pressing the Start Test switch again will disable the running test. Changing the
Test Mode selection switch will also disable the running test. The new selected
test will need to be started by pressing the Start Test switch again.
Note:
If the HFG01 test has been running for more than 2 hours continuously it is
recommended that it is allowed to cool down before power is finally removed, to
reduce any latent heat build-up. To do this, leave the HFG01 turned on but with no
running test for approximately 10 to 15 minutes before fully powering down.
5.6 Troubleshooting a test
The test load operation will be disabled if the HFG01 detects too many dropped
voltage cycles or interruptions from the Test Supply.
If the HFG01 load exceeds its recommended operating temperature, the test load
operation will be disabled (Test Enable display will be OFF) and the Alarm
indicator will flash at a steady rate. Once the temperature has returned to
normal the Alarm indicator will turn off and the test can be resumed.
If the HFG01 detects asystem error, the Alarm indicator will fully turn on without
flashing. The system error will need to be cleared by removing all power from the
HFG01 before fully restarting. If fully restarting the device fails to clear the error,
the HFG01 may require servicing.
A flowchart for troubleshooting if the test cannot be enabled is shown in Figure 6.
In the case of other problems, please contact your supplier for advice.

HFG01 Operation Manual
York EMC Services Page 15 of 21 Doc. No. Y0321/DOC/038 Rev. 2
Figure 6: Troubleshooting flowchart

HFG01 Operation Manual
York EMC Services Page 16 of 21 Doc. No. Y0321/DOC/038 Rev. 2
6 O
UTPUT
P
ERFORMANCE
Please refer to the test results supplied with the unit for its unique performance.
Typical performance results are shown in the sections below.
6.1 Flicker Load Operation
The typical flicker disturbance generated by the HFG01 in each of the steady flicker
modes when tested according to EN/IEC 61000-3-3 is shown in Table 3.
Flicker Mode
Setting Load Operation Pst Plt
8.3 Hz
8.33 Hz flicker 1.10 1.10
1 Hz
1 Hz flicker 0.45 0.45
Table 3: Typical Pst results measured to EN/IEC 61000-3-3
6.2 Harmonics Load Operation
The typical current harmonics disturbance generated by the HFG01 in each of the
harmonics modes when tested according to EN/IEC 61000-3-2 is shown below:
Figure 7: Disturbances generated in Fixed Harmonic mode

HFG01 Operation Manual
York EMC Services Page 17 of 21 Doc. No. Y0321/DOC/038 Rev. 2
Figure 8: Disturbances generated in Fluctuating Harmonics mode
6.3 Calibration
The HFG01 is supplied with a manufacturer’s calibration certificate.
It is recommended that the HFG01 should be periodically checked to ensure
damage or long-term changes have not altered the output. The periodicity of these
checks is typically determined by the quality assurance regime employed by the
user, however it is recommended that the HFG01 output be retested annually. This
service is available from York EMC Services and can be arranged through your
supplier.

HFG01 Operation Manual
York EMC Services Page 18 of 21 Doc. No. Y0321/DOC/038 Rev. 2
7 S
AFETY
WARNING! The HFG01 must be earthed.
Any interruption of the mains earth conductor inside or outside the HFG01 case
may make the instrument electrically dangerous. If the Test Supply does not
present an earth connection to the Test Supply Cable, then the external earth
connector on the front of the HFG01 must be connected to a safety earth.
The HFG01 is classified as IEC Safety Class 1. It has been designed to meet the
requirements of EN/IEC/UL 61010, Safety requirements for Electrical Equipment
for Measurement, Control and Laboratory Use.
The HFG01 has been tested in accordance with EN/IEC/UL 61010 and has been
supplied in a safe condition. This instruction manual contains information and
warnings that have to be followed by the user to ensure safe operation and to retain
the instrument in a safe condition.
The HFG01 has been designed for indoor use only in a Pollution Degree 2
environment within the temperature range 5
O
C to 40
O
C and 20% to 80% Relative
Humidity (non- condensing). Do not operate whilst condensation is present.
Use of the HFG01 in a manner not specified by these instructions may impair the
safety protection provided. Do not use the HFG01 outside its rated supply voltage
or environmental ranges.
The HFG01 must be positioned in a way that allows full access to any on/off switch
on the Test Supply or Mains Supply.
Ensure that fuses with the required rated current and specified type are used for
replacement. The use of makeshift fuses and the short-circuiting of fuse holders is
dangerous and must be prohibited.
The mains cord included with the Power Supply Unit must only be replaced with
an adequately-rated cord
If the HFG01 is clearly defective, has been subject to mechanical damage,
excessive moisture or chemical corrosion the safety protection may be impaired.
In this situation the HFG01 should be withdrawn from use and returned to York
EMC Services for assessment and repair.
This equipment cannot be repaired by the user. DO NOT attempt to remove the
covers, open the case or otherwise disassemble the equipment. Only York EMC
Services trained service personnel with knowledge of fire and shock hazards
should service this equipment. There are high-voltage parts in this equipment
presenting a risk of severe injury or fatal shock to untrained personnel.
Do not wet the case of the HFG01 when cleaning it, for example by using a wet
cloth or aerosol cleaning fluid. If the case has become wet through any
circumstances then it MUST be dried before being connected to a power supply.

HFG01 Operation Manual
York EMC Services Page 19 of 21 Doc. No. Y0321/DOC/038 Rev. 2
8 ELECTROMAGNETIC
COMPATIBILITY
8.1 General
The HFG01 has been designed to meet the requirements of EU Directive
2014/30/EU. Compliance was demonstrated by testing to the requirements of the
following standards: -
8.2 Emissions
EN61326-1 (2013) consisting of
•EN55011 (2009) + A1 (2010) Radiated RF emissions (Enclosure) and
Conducted RF emissions (applies to the HFG01 supply adapter), Class A
8.3 Immunity
Enclosure:
EN61326-1 (2013) consisting of
•EN61000-4-2 (2009) Electrostatic Discharge immunity, 4 kV Air & Contact
Discharge.
•EN61000-4-3 (2006) + A1 (2008) + A2 (2010) Radiated RF immunity
(Enclosure), 3 V/m.
AC power:
These apply to the HFG01 supply adapter only:
EN61326-1 (2013) consisting of
•EN61000-4-11 (2004) Voltage dips/short interruptions
•EN61000-4-4 (2004) + A1 (2010) Fast Transient common mode (AC mains
port)
•EN61000-4-5 (2006) Surge (AC mains port)
•EN61000-4-6 (2009) Conducted RF immunity (AC mains port)
•EN61000-3-2 (2006) Harmonic emission (AC mains port)
•EN61000-3-3 (2013) Voltage fluctuation (AC mains port)
8.4 Non-Applicability
EN61326-1 (2013) consisting of
•EN61000-4-4 (2004) Fast Transient common mode (signal ports)
•EN61000-4-5 (2006) Surge (signal ports)
•EN61000-4-6 (2009) Conducted RF immunity (signal ports)
These phenomena are deemed non applicable on the basis that the HFG01 has
no relevant signal ports. The AC test port of the HFG01 may be looked upon as a
signal port but the captive lead is less than 3m in length hence the tests do not
apply.
The test port is not connected directly to the public low voltage supply network and
so tests relating to AC ports are only applicable to the AC/DC adapter used for the
HFG01 supply input.

HFG01 Operation Manual
York EMC Services Page 20 of 21 Doc. No. Y0321/DOC/038 Rev. 2
9 M
AINTENANCE AND
C
ALIBRATION
Ensure that the airflow through the fan and ventilation duct areas remains
unobstructed, and that the HFG01 is in good condition.
For all maintenance and calibration procedures, the HFG01 should be returned to
York EMC Services for servicing by authorized personnel.
This equipment cannot be repaired by the user. DO NOT attempt to open the case
or disassemble the equipment. Only York EMC Services trained service personnel
with knowledge of fire and shock hazards should service this equipment. There
are high-voltage parts in this equipment presenting a risk of severe injury or fatal
shock to untrained personnel.
9.1 Disposal
Should the equipment require disposal, please contact your supplier for details on
returning the equipment for disposal/recycling.
Table of contents
Other EMS Portable Generator manuals
Popular Portable Generator manuals by other brands

Tesla
Tesla TIMV3500 GPU-24 manual

Mecc Alte spa
Mecc Alte spa ECO3 Series Operating and maintenance instructions

AlphaGen
AlphaGen ACX3100i Operation manual

Champion Power Equipment
Champion Power Equipment 41331 Owner's Manual and Operating Instructions

Stryker
Stryker RF MultiGen 0406-900-000 Instructions for use

Rohde & Schwarz
Rohde & Schwarz SGT100A Getting started