flexicare provu 040-07-0035U User manual

PROVU 3.5” DISPLAY
REF No. 040-07-0035U
232mm
123mm

2
CONTENTS 02
PRODUCT KEY 03
TECHNICAL SPECIFICATIONS 04
INTENDED USE 05
PRODUCT DESCRIPTION 05
PRECAUTIONS 05
WARNINGS & CAUTIONS 05
SETTING UP THE CHARGING DOCK 06
CHARGING 06
USING WITH PROVU VL 06
AFTER USE 06
CLEANING 07
WARRANTY STATEMENT 07
ABUSE, MISUSE OR ACCIDENTAL DAMAGE 07
TROUBLESHOOTING 08
SOFTWARE 09
BATTERY INDICATOR 10
SYMBOLS GLOSSARY 10

3
PRODUCT KEY
1. Display
2. Charging Dock
3. USB Charging Cable
4. Power Button
5. Wall Plug
3
1
2
4
5

4
Technical Specications
DISPLAY ELECTRICAL POWER RATING
POWER INPUT Input: 5.0V 1.2A
BATTERY TYPE 3.7V, 1500mAh, Lithium-ion
TYPICAL BATTERY RUN TIME Minimum of 60 minutes
TYPICAL BATTERY CHARGE TIME 90 minutes – 2 hours (from zero charge)
TYPICAL STANDBY TIME Up to 48 hours
WALL PLUG ELECTRICAL POWER RATING
INPUT 100-240V AC - 0.5A 50-60Hz
OUTPUT Max 5.2V DC 1.2A
DIMENSIONS
WIDTH X HEIGHT X THICKNESS 69mm x 89mm x 15mm
WEIGHT 90g
DISPLAY
RESOLUTION (PIXELS) 480 x 320
ORIENTATION Landscape
DISPLAY TYPE 3.5”
START UP TIME (FROM STANDBY) 2 seconds
START UP TIME (FROM BOOT MODE) 20 seconds
SOFTWARE VERSION Refer to display screen while charging
IP RATING IP64
CONNECTIONS
USB CONNECTION Micro USB
STORAGE AND TRANSPORTATION
STORAGE TEMPERATURE 10℃ to 40℃ (50˚F to 104˚F)
TRANSPORT TEMPERATURE: 10℃ to 40℃ (50˚F to 104˚F)
RELATIVE HUMIDITY ≤85%, non-condensing
ATMOSPHERIC PRESSURE 700 to 1060 hPa
OPERATING CONDITIONS
TEMPERATURE 20 ± 10℃
RELATIVE HUMIDITY ≤85%, non-condensing
ATMOSPHERIC PRESSURE 700 to 1060 hPa

5
Intended Use
The ProVu 3.5” Display is a digital display intended to show live images from the ProVu Video Laryngoscope (VL)
only, via a micro-USB connection, as an aid to intubation for the hospital use.
Product Description
The ProVu 3.5” Display is a non-sterile, reusable digital display intended to show live images from the ProVu Video
Laryngoscope.
When connected to the ProVu device, a live video is shown on the display, which can be used to further aid
intubation of the patient.
The display can be recharged on the charging dock provided or with the ProVu charging cable and, when fully
charged, will provide a minimum of 1 hour of continuous use.
Flexicare does not recommend the modication of the display. Any attempts to do so, will result in the failure of
the display and voids warranty.
Precautions
-Ensure that you always have easy access to the power cord, plug or adapter to disconnect this display from
the power.
-Inspect the ProVu 3.5” Display and all accessories for damage prior to each use. Do not use if it appears
damaged.
-Charge the ProVu 3.5” Display fully before use. For shipping requirements, ProVu 3.5” Displays are shipped
with zero charge.
-Before charging, ensure that all components are completely dry, including the micro-USB port.
-If not connected to ProVu VL and left unmoved for more than 5 minutes, the ProVu 3.5” Display will switch into
“sleep mode” and the screen will go blank. This is normal. To reverse this, briey press the Power Button. A
prompt image should be seen after 2 seconds.
Warnings & Cautions
-ProVu 3.5” Display and Accessories should only be used by trained medical professionals.
-Do not use within Magnetic Resonance Imaging (MR/MRI) environments or with other high interference
sources.
-Do not sterilize the display.
-The use of devices and other accessories which are not approved by Flexicare may place the patient at risk. It is
recommended that the ProVu 3.5” Display should only be used with the ProVu Video Laryngoscope and ProVu
accessories.
-Always disconnect from ProVu 3.5” Display from the charging dock before cleaning.
-Ensure the voltage recommendations of the display wall plug are compatible with power supplies.
-Do not clean the display or display accessories using any abrasive agents or mineral acids.
-Do not use sterilants with caustic ingredients, such as surgical scrub solutions, peroxide solutions, bleaches,
or povidone-iodine solutions.
-Inadequate battery levels may result in sudden failure without warning. Check sucient battery level before
use.
-Do not disassemble display.
-Do not modify the display.
-Battery is non-replaceable and non-removable.
-There are no user serviceable parts in the device. Any faulty device should be replaced or returned within
warranty to the manufacturer.
-Do not attempt to charge while battery is in use.
-Do not attempt maintenance of the device during use.
-Do not submerge.
-If the display screen does not illuminate from “sleep mode”, it is likely the battery needs to be charged or the
device needs to be reset. Follow Troubleshooting steps below for proper steps.

6
Using with ProVu VL
For more detailed instructions on the use of the ProVu VL, refer to the ProVu VL instructions (IP0000- 018) supplied
with the handles and blades.
1. Use currently accepted medical techniques to choose the correct size of ProVu VL blade.
2. Securely connect the ProVu VL handle and blade to the ProVu VL 3.5” Display (Fig. A).
3. Observe the Display screen to verify that a live image is being received from the ProVu VL blade (Fig. B).
4. If no image is seen, briey press the power button, a live image should be seen after 2 seconds.
5. Observe the display screen to verify that a live image in being received from the ProVu VL blade.
6. Once a live image is seen, the device is ready to be used.
7. If neccessary, when space is restricted, the display can be tilted, in order to show optimum view.
After Use
1. Disconnect the ProVu VL handle and blade from the ProVu 3.5” Display.
2. Dispose of all single-use components in accordance with local policies and / or as advised by your local
healthcare provider.
3. Clean and dry the ProVu 3.5” Display referring to the cleaning instructions.
4. Place the ProVu 3.5” Display onto the charging dock, when not in use.
A BA B C
Setting up the Charging Dock
1. Remove all components from the box.
2. Connect the USB cable to the charging dock.
3. Wrap the cable around the device in the dedicated path on the underside of the charging dock, leaving
enough free cable outside the charging dock, to be able to reach the mains power.
4. Using the plug adaptor provided, plug into mains power source.
5. The charging dock is now ready to use.
Charging
1. Connect the ProVu 3.5” Display to the charging dock or directly into the USB cable, if not using the charging
dock.
2. Press the Power Button for 3 seconds.
3. After 3 seconds, an indication of charge is shown on the screen. If this screen does not display, check the
security of the connections.
4. When fully charged, the display will provide a minimum of 1 hour of continuous use.
5. For optimal performance, re-charge after every use.

7
Device Disposal
When the ProVu 3.5” Display has reached the end of its product life, it should be disposed of in accordance to
procedures set out in local guidelines.
ProVu 3.5” Display is suitable for cleaning or disinfection. Cleaning or disinfection should be carried out according
to local hospital policies and cleaning/disinfector manufacturer’s instructions.
Cleaning
1. Wear gloves throughout the cleaning procedure.
2. After each use, clean the exterior of the ProVu 3.5” Display using a standard germicidal wipe with at least
55% IPA (e.g. Super Sani-Cloth®Germicidal Disposable Wipe or equivalent).
3. Inspect the ProVu 3.5” Display and reusable accessories for contamination. If any is present repeat steps 1 & 2.
4. Once the ProVu 3.5” Display and reusable accessories are free from contaminates, dry with a clean, dry cloth.
5. Check display for damage. If display does not function, refer to troubleshooting steps.
Disinfection
Cleaning or disinfection may be done for the ProVu Display and reusable accessories. Cleaning or disinfection
should be carried out according to local hospital policies and cleaning/disinfector manufacturer's instructions.
1. ProVu 3.5” Display and reusable accessories disinfection should be carried out according to local hospital
policies and disinfector manufacturer’s instructions and are compatible with the following solutions:
-Chlorine or Bleach Wipes (Containing ≥5400ppm Sodium Hyperchlorite; e.g. Sani-Cloth®Chlor/Sani-Cloth®
Bleach or equivalent).
-Clinell Universal Wipes (2 quaternary ammonium compounds and 1 polymeric biguanide)
2. Ensure that the device to be disinfected is free from any visible contaminants before disinfection.
3. Dry with a clean, dry cloth.
4. Once dry, it is recommended that the display be placed on charge until next use.
Any serious incident that occurs in relation to the device should be reported to the manufacturer and the
competent authority of the Member State of the user and/or patient.
Warranty Statement
Flexicare Medical Limited warrants that the product purchased meets the labelled specications of the product and
will be free from defects in materials and workmanship that occur within one (1) year from the date of purchase.
This warranty does not cover damage caused by:
-Handling during shipping.
-Use or maintenance contrary to labeling and instructions, including but not limited to incorrect disassembly,
reprocessing and reassembly.
-Modication or repair by anyone not authorized by Flexicare Medical Limited.
Abuse, misuse or accidental damage
If a product covered by this warranty is determined to be defective materials, components, or workmanship,
and the warrant claim is made within the warranty period described above, Flexicare Medical Limited will, at its
discretion, repair or replace the defective product free of charge. Written authorisation must be obtained from
Flexicare Medical Limited to return the product prior to sending it with postage paid to Flexicare Medical Limited’s
designated agent.
This warranty is in lieu of all other warranties, expressed or implied, including but not limited to, the implied
warranties of merchantability and tness for a particular purpose. Flexicare Medical Limited’s obligation under
this warranty is limited to repair or replacement of products containing a defect. Flexicare Medical Limited is not
responsible for any indirect or consequential damages resulting from a product defect covered by the warranty.
This warranty shall be governed by and construed in accordance with Flexicare Medical Limited’s Terms and
Conditions

8
Troubleshooting
PROBLEM
No live image is seen
POSSIBLE EXPLANATION ACTION NEEDED
Incorrect connection to ProVu 3.5”
Display
Remove and reconnect the ProVu 3.5” Display to the handle and blade via
the micro-USB ports as shown in the instructions for use section.
ProVu 3.5” Display needs to be charged Charge display via the dock or USB cable. Note that the display cannot be
charged while in use.
Display may be in sleep mode Briey press the power button, a live image should be seen after 2
seconds.
ProVu VL may be damaged Replace with a new ProVu VL handle and blade. This is not a serviceable
device.
ProVu 3.5” Display is damaged Replace the ProVu VL Display with a new display.
PROBLEM
Low Image Quality
POSSIBLE EXPLANATION ACTION NEEDED
Incorrect connection to ProVu 3.5”
Display
Remove and reconnect the ProVu 3.5” Display to the device via the
micro-USB ports as shown in the instructions for use section.
Dirty screen Wipe with an IPA cloth.
Damp screen Wipe with a dry cloth.
Camera lens may have secretions on it Clean Camera lens with a clean microber towel or replace ProVu device.
Illumination failure Replace with a new ProVu device.
Incorrectly inserted cable Ensure all connections are secure.
PROBLEM
Error screen is shown
POSSIBLE EXPLANATION ACTION NEEDED
Software has malfunctioned Hold down the Power Button for 5 seconds to power o. Release power
button. Press power button for 3 seconds to power on.

9
Software
The ProVu 3.5” Display has several on-screen prompts to assist use.
0.0.1.1
20
%
0.0.1.1
100
%
0.0.1.1
5% 75%
75% 75%
Charged
ProVu 3.5” Display is fully charged.
Display Connected
ProVu 3.5” Display is connected and
camera image is loading.
Display is Charging
ProVu 3.5” Display battery is charging.
Percentage charge is indicated in the circle.
Connect Display
Connect ProVu 3.5" Display to ProVu
device to view camera image
Error 2
Camera error or crash.
Replace ProVu device.
Software Version
Software version is displayed in the
bottom right of the display when
connected to power.
Charge Display
ProVu 3.5” Display has insucient battery.
Please charge via USB or charging dock.
Error 1
Software error or crash. Hold the power
button down for 5 seconds. Release power
button. Press power button for 3 seconds
to power on.
Disconnect Display
If camera image fails to load, disconnect
the ProVu 3.5” Display.

10
100% 20% 19%
Battery Indicator
During use, the battery indicator appears on screen to show battery power display power remaining.
The icon will change to ash on and o when battery charge is less than 20%.
When the battery charge is 5% or less, the Charge Display screen will be shown. Please charge ProVu 3.5” Display via USB or charging dock.
Symbols Glossary
Catalogue Number Serial Number Product Conforms with
Directive 93/42/EEC
Refer to Instruction
Manual/Booklet Manufacturer Caution
Caution: US Federal Law
Restricts this device to sale
by, or on the order of, a
physician.
Do Not Use Blade
to Open Type BF Applied Part Class II Equipment Keep Away
From Sunlight
Country of origin
700 hPa
1060 hPa
IP64
Keep Dry Recycle Separate Collection Alternating Current Direct Current Temperature
Limit
Humidity
Limitation
Atmospheric Pressure
Limitation IP64 USB/Micro-USB USB/Amplier Cable Box Date of
Manufacture

11

Patent Pending
IP0000-384/4 | 082021
Flexicare Medical Limited
Cynon Valley Business Park, Mountain Ash, CF45 4ER, UK.
Distributed by Flexicare, Inc.
15281 Barranca Pkwy, Unit D, Irvine, CA 92618
+44 (0)1443 474647
949-450-9999
+44 (0)1443 474222
949-450-9902
enquiries@exicare.com
inquiries@exicare.com
www.exicare.com
www.exicare.com
Table of contents