FlowLine EchoBeam LR85 Series User manual

Flowline, Inc. | 10500 Humbolt Street, Los Alamitos, CA 90720 p 562.598.3015 f562.431.8507 wflowline.comAD310320 Rev A
LR83 & LR85 Series Display Manual (Addendum to LevelTap™ Manual)
©2023 Flowline, Inc.
All Rights Reserved
EchoBeam
®
80 GHz Radar Liquid & Solid Level Transmitter

|2 AD310320 Rev A
IntroductionSection One
TABLE OF CONTENTS
Section One | Introduction .............................................................................................4
Sensor Models ......................................................................................................4
Section Two | Getting Started .......................................................................................5
Setup Overview ....................................................................................................5
Section Three | Configuration Using Display ...............................................................6
Using the Display...................................................................................................6
Changing Display Values .......................................................................................7
Measure the Tank (Basic Dimensions) .................................................................8
Confirm the Tank Measurements ..........................................................................9
Basic Configuration Overview Using Push Button Display ................................10
Step 1 – Set the Units of Measurement ..................................................11
Step 2 – Set the Application Medium .....................................................12
Step 3 – Set the Fail-Safe Option ...........................................................13
Step 4 – Set the Sensor Height (4mA) ....................................................14
Step 5 – Set the Fill-Height (20mA) .........................................................15
Step 6 – Set the Display Value ................................................................16
Step 7 – Set the Sensor Name ...............................................................17
Step 8 – Check the EchoCurve ...............................................................18
Section Four | Advanced Configuration Using Display ..............................................19
Range Adjust .......................................................................................................19
Advanced Adjustments Overview............................................................19
Dead Band ...............................................................................................20
Maximum Range .....................................................................................22
Reverse mA .............................................................................................23
Distance Adjust........................................................................................24
HART .................................................................................................................25
Advanced Adjustments Overview............................................................25
HART........................................................................................................26
Level Variables.....................................................................................................27
Process Adjustments Overview ...............................................................27
Low Dielectric .........................................................................................28
Turbulent or Agitated Surface..................................................................29
Fast Filling or Emptying ..........................................................................30
Multiwave ................................................................................................31
Signal Properties .................................................................................................32
Process Adjustments Overview...............................................................32
First Echo.................................................................................................33
Echo Lost Timeout...................................................................................34
False Echo Memory .................................................................................35

AD310320 Rev A 3 |
IntroductionSection One
Echo Threshold........................................................................................36
Envelope Level ........................................................................................37
Damping Time .........................................................................................38
False Echo Edit .......................................................................................39
Near Band Detect ...................................................................................41
Section Five | Diagnostics ............................................................................................42
Entering Diagnostics Menu .................................................................................42
Choose Curve .....................................................................................................43
What is a Good EchoCurve ................................................................................45
What is a Bad EchoCurve ...................................................................................45
Simulation ...........................................................................................................47
Sensor Status .....................................................................................................48
Section Six | Data Sampling .........................................................................................49
Entering Data Sampling Menu ............................................................................50
Enable Log ..........................................................................................................50
Distance Log Interval ..........................................................................................51
Curve Log Interval ..............................................................................................52
View Distance Log ..............................................................................................52
View Curve Log ...................................................................................................53
Reset Log Data ...................................................................................................53
Section Seven | Information .........................................................................................54
Entering Information Menu .................................................................................55
Information ..........................................................................................................55
Software Version ................................................................................................55
Write Protect .......................................................................................................56
Language ............................................................................................................56
Reset .................................................................................................................57
BLE Password .....................................................................................................57
Section Eight | Troubleshooting ..................................................................................58
Troubleshooting Overview ..................................................................................58
Error Codes: 4-20mA/HART with Display ...........................................................59
Error Codes: ModBus .........................................................................................59
Section Nine | Appendix ...............................................................................................60
Factory Settings ..................................................................................................60
User Configuration ..............................................................................................60
ModBus Output Definitions ................................................................................61
Section Ten | Warranty, Returns and Limitations ......................................................63
Warranty .............................................................................................................63

|4 AD310320 Rev A
Introduction Section One
SENSOR MODELS
Offered in two different models, EchoBeam®is an FMCW radar level sensor that provides a continuous output
proportional to the liquid or solids level in a tank or sump. Available outputs include 4-20mA, 4-20mA w/
HART and ModBus RTU. Make sure that the model purchased is appropriate for your application.
Series
Max
Range
Liquid
Max
Range
Solids
Beam
Angle Material Mounting Application
LR83
98.42’
(30m)
49.21’
(15m)
6° PVDF
1½” NPT
(1½” G)
Liquids: Measures large bulk storage
tanks up to 30m or difficult process tank
applications or open channel flow.
Solids: Measures small to medium silos
or stockpiles up to 15m.
LR85
393.7’
(120m)
198.8’
(60m)
4° PFA
3” NPT
(M80 x 3)
or
Gimbaled
3” ANSI
(DN80)
Liquids: Measures large bulk storage
tanks up to 120m or difficult process
tank applications.
Solids: Measures medium to large silos
up to 60m, including those with
significant dust or very limited
measurement space.

AD310320 Rev A 5 |
Getting Started Section Two
SETUP OVERVIEW
The below list highlights the initial steps required to set up your transmitter for operation.
Step 1: Set the Units of Measurement
Select the Units of Measurement (inches, feet, millimeters, centimeters or meters) that will be used by the
sensor and displayed by the App (and sensor push button display module if applicable to your model).
Step 2: Set the Application
Select the type of Application media (liquid or solids) that the sensor will measure.
Step 3: Set the Fail-Safe Option
Select the Fail-Safe Output (3.9mA, 20.5mA, 22mA or Hold Last Value) that the sensor will use to indicate
that it cannot measure a valid signal return or level.
Step 4: Set the Sensor Height (4mA)
Input the Sensor Height distance or 4mA set point which corresponds to the tank empty setting.
Step 5: Set the Fill-Height (20mA)
Input the Fill Height distance or 20mA set point which corresponds to the tank full setting.
Step 6: Set the Display Value
Select the level Display Value (Height, Distance, Current, Percent) that will be indicated by the App (and
sensor push button display module if applicable to your model).
Step 7: Set the Sensor Name
Input a unique Sensor Name to identify the device (location, tank and/or media) within 36 characters.
Step 8: Check the EchoCurve
This is a quick diagnostic tool to determine if the transmitter is reading the correct level.

|6 AD310320 Rev A
Configuration Using Display Section Three
USING THE DISPLAY
The display module features a dot matrix LCD display with 4 push buttons on
a removable puck. Out of the box, the display indicates level in feet and
depicts the level on a bar graph at the right side of the display. The four
buttons perform the following functions:
• ESC
o Exit configuration mode
o Return to a higher menu level
o Display EchoCurve
• UP Arrow
o Modify parameter values
o Choose display mode
• RIGHT Arrow
o Choose configuration options
o Choose parameter digits to edit
o Display contents of
parameters
• OK
o Enter Menu and Options
o Confirm configuration options
o Confirm changes to
parameters
Menu Introduction
1. To enter the Main Menu (from the Main Screen), press the OK
button.
2. Use the Right Arrow button to scroll through the Main Menu
options.
1. Configuration – Below are the configuration menu functions:
a. Basic
b. Advanced
2. Diagnostics – Below are the diagnostic menu functions:
a. Choose Curve
b. Simulation
c. Sensor Status
3. Data Logging – Below are the data logging menu functions:
a. Enable
b. View Distance Data
c. View EchoCurve Data
d. Clear Memory
4. Information – Below are the information menu functions:
a. Information
b. Software Version
c. Write Protect
d. Language
e. Factory Reset
3. To select one of the functions, press OK.
4. To exit the programming mode, press ESC.

AD310320 Rev A 7 |
Configuration Using Display Section Three
CHANGING DISPLAY VALUES
The numeric values are set using the Right Arrow and Up Arrow buttons. Press the Right Arrow button to
select the next digit and the Up Arrow button to increment the digit value. The digit being changed is
highlighted. Press the OK button to accept a setting or the ESC button to exit without saving changes. The
below exercise illustrates how to change the value of an Empty Configuration. Follow the steps to change the
setting from 39.370 ft to 12.000 ft. This example applies to all functional settings starting from the Main Menu.
1) From the Sensor Height Screen, press OK to begin changing the display value.
a) The first digit “0” will now be highlighted.
2) Press Right Arrow to move from the first digit to the second digit.
a) The second digit “3” will now be highlighted.
3) Press Up Arrow repeatedly until the second digit scrolls from “3” to “1”.
a) Each press of the Up Arrow will increase the digit by 1.
i) When the value reaches 9, it will drop to 0 and begin again.
ii) Stop pressing Up Arrow when the second digit is highlighted as “1”.
b) Press Right Arrow to move one digit to the right.
i) The third digit “9” will now be highlighted.
4) Press Up Arrow repeatedly until the third digit scrolls from “9” to “2”.
a) Stop pressing Up Arrow when the third digit is highlighted as “2”.
b) Press Right Arrow to move one digit to the right.
i) The fourth digit “3” will now be highlighted.
5) Press Up Arrow repeatedly until the fourth digit scrolls from “3” to “0”.
a) Stop pressing Up Arrow when the fourth digit is highlighted as “0”.
b) Press Right Arrow to move one digit to the right.
i) The Fifth digit “7” will now be highlighted.
6) Press Up Arrow repeatedly until the fifth digit scrolls from “7” to “0”.
a) Stop pressing Up Arrow when the fifth digit is highlighted as “0”.
b) Press Right Arrow to move one digit to the right.
i) The Sixth digit “0” will now be highlighted.
7) Press OK to accept the setting as 12.000.

|8 AD310320 Rev A
Configuration Using Display Section Three
MEASURE THE TANK (BASIC DIMENSIONS)
EchoBeam®measures the distance between the
sensor and the liquid or solids surface below.
Measuring the tank is one of the most important
aspects in configuring the transmitter. When
measuring the tank, consider the location of the
transmitter with respect to fittings, risers, dome
tops and bottoms, and identify where the
measurements are taken from the transmitter. The
basic measurements for configuration are
described to the right:
• Sensor Height
• Fill-Height
All measurements are made from the bottom of the
tank up to the bottom of the sensor (Sensor Height)
or to the operational full for the tank (Fill-Height).
Example: Mounting in a Solid
Example: Mounting in a Liquid

AD310320 Rev A 9 |
Configuration Using Display Section Three
CONFIRM THE TANK MEASUREMENTS
All measurements with EchoBeam®are referenced from the bottom of the transmitter. Once the transmitter
has been installed, you may need to adjust your original measurements due to the location of the EchoBeam®
in the application. Shown below are the offsets from the bottom of the EchoBeam®to the top of the
mounting threads. Use the offsets to calculate the exact location of the transmitter’s point of measurement.
LR83 Series
LR85 Series

|10 AD310320 Rev A
Configuration Using Display Section Three
BASIC CONFIGURATION OVERVIEW USING PUSH BUTTON DISPLAY
Below are the 8 basic steps to configure the sensor for operation. Each step is described in detail on the
following pages:
Step 1: Set the Units of Measurement
Select the Units of Measurement (inches, feet, millimeters, centimeters or meters) that will be used by the
sensor and displayed by the App (and sensor push button display module if applicable to your model).
Step 2: Set the Application Medium
Select the type of Application media (liquid or solids) that the sensor will measure.
Step 3: Set the Fail-Safe Option
Select the Fail-Safe Output (3.9mA, 20.5mA, 22mA or Hold Last Value) that the sensor will use to indicate
that it cannot measure a valid signal return or level.
Step 4: Set the Sensor Height (4mA)
Input the Sensor Height distance or 4mA set point which corresponds to the tank empty setting.
Step 5: Set the Fill-Height (20mA)
Input the Fill Height distance or 20mA set point which corresponds to the tank full setting.
Step 6: Set the Display Value
Select the level Display Value (Height, Distance, Current, Percent) that will be indicated by the App (and
sensor push button display module if applicable to your model).
Step 7: Set the Sensor Name
Input a unique Sensor Name to identify the device (location, tank and/or media) within 36 characters.
Step 8: Check the EchoCurve
This is a quick diagnostic tool to determine if the transmitter is reading the correct level.

AD310320 Rev A 11 |
Configuration Using Display Section Three
STEP 1 - SET THE UNITS OF MEASUREMENT
This function sets the units for all measurement values to be entered into the transmitter.Options for units of
measurement are inches, feet, millimeters, centimeters or meters.
1) From the Main Screen, press OK to advance into
the Main Menu.
2) Press OK to advance into the Configuration Menu.
3) Press OK to advance into Basic Menu (Application
screen will appear).
4) Press Right Arrow to scroll to Units of
Measurement.
5) Press OK to change the setting for Units of
Measurement.
a) Press Right Arrow to move the triangle to the
desired setting.
b) When the setting (unit of measurement) is
correct, press OK to save the setting.
6) When done, press ESC to return to the Basic
Menu, press ESC again to return to the
Configuration Menu and press ESC a third time to
return to the Main Screen, or if you want to
advance directly into Application Medium, press
Right Arrow.

|12 AD310320 Rev A
Configuration Using Display Section Three
STEP 2 - SET THE APPLICATION MEDIUM
This function informs the transmitter of the media that is being measured.
1) From the Main Screen, press OK to advance into
the Main Menu.
2) Press OK to advance into the Configuration Menu.
3) Press OK to advance into Basic Menu (Application
screen will appear).
4) Press OK to change the setting for Application
Medium.
a) Press Up Arrow or Right Arrow to move the
triangle to the desired setting.
b) Options for Application Medium are Liquids or
Solids.
c) When the setting (Application Medium) is
correct, press OK to save the setting.
5) When done, press ESC to return to the Basic
Menu, press ESC again to return to the
Configuration Menu and press ESC a third time to
return to the Main Screen, or if you want to
advance directly into Fail-Safe, press Right Arrow.

AD310320 Rev A 13 |
Configuration Using Display Section Three
STEP 3 - SET THE FAIL-SAFE OPTION
This function sets the default output when EchoBeam®is not able to acquire a valid return signal.
1) From the Main Screen, press OK to
advance into the Main Menu.
2) Press OK to advance into the
Configuration Menu.
3) Press OK to advance into Basic
Menu.
4) Press Right Arrow (at least two
times) to scroll to Fail-Safe.
5) Press OK to change the value for Fail-
Safe.
a) Press Up Arrow or Right Arrow
to move the triangle to the desired
setting.
b) Options for Fail-Safe are Hold
Last Value, 21.5mA, 20.5mA or
3.9mA.
c) When the value is correct, press
OK to save the setting.
6) When done, press ESC to return to
the Basic Menu, press ESC again to
return to the Configuration Menu and
press ESC a third time to return to the
Main Screen, or if you want to
advance directly into Sensor Height,
press Right Arrow.

|14 AD310320 Rev A
Configuration Using Display Section Three
STEP 4 - SET THE SENSOR HEIGHT (4MA)
This function sets the Sensor Height set point for the transmitter corresponding 4mA to empty.
1) From the Main Screen, press OK to
advance into the Main Menu.
2) Press OK to advance into the
Configuration Menu.
3) Press OK to advance into Basic
Menu.
4) Press Right Arrow (at least three
times) to scroll to Sensor Height.
5) Press OK to change the value for
Sensor Height.
a) Press Up Arrow to increase the
value of the number highlighted.
Up Arrow will scroll from 0 to 9
and back again.
b) Press Right Arrow to move one
segment to the right. Right Arrow
will scroll left to right and then
back to the first segment.
c) When the value is correct, press
OK to save the setting.
6) When done, press ESC to return to
the Basic Menu, press ESC again to
return to the Configuration Menu and
press ESC a third time to return to the
Main Screen, or if you want to
advance directly into Fill-Height,
press Right Arrow.
Note: If Fill-Height is greater than Sensor Height, the Sensor Height change will not be saved.
Note: The Dead Band and Maximum Range are automatically set based on the Sensor Height and Fill-Height.
After writing to unit, the Dead Band is set to 0.9 x (Sensor Height – Fill-Height) and the Maximum range is set
to 1.05 x (Sensor Height).

AD310320 Rev A 15 |
Configuration Using Display Section Three
STEP 5 - SET THE FILL-HEIGHT (20MA)
This function sets the Fill-Height set point for the transmitter corresponding 20mA to full.
1) From the Main Screen, press OK to
advance into the Main Menu.
2) Press OK to advance into the
Configuration Menu.
3) Press OK to advance into Basic Menu
(Application screen will appear).
4) Press Right Arrow (at least four
times) to scroll to Fill-Height.
5) Press OK to change the value for Fill-
Height.
a) Press Up Arrow to increase the
value of the number highlighted.
Up Arrow will scroll from 0 to 9
and back again.
b) Press Right Arrow to move one
segment to the right. Right Arrow
will scroll left to right and then
back to the first segment.
c) When the value is correct, press
OK to save the setting.
6) When done, press ESC to return to
the Basic Menu, press ESC again to
return to the Configuration Menu and
press ESC a third time to return to the
Main Screen, or if you want to
advance directly into Display, press
Right Arrow.
Note: If Fill-Height is greater than Sensor Height, the Fill-Height change will not be saved.
Note: The Dead Band and Maximum Range are automatically set based on the Sensor Height and Fill-Height.
After writing to unit, the Dead Band is set to 0.9 x (Sensor Height – Fill-Height) and the Maximum range is set
to 1.05 x (Sensor Height).

|16 AD310320 Rev A
Configuration Using Display Section Three
STEP 6 - SET THE DISPLAY VALUE
EchoBeam®can display (on the local display or within the LevelTap™ App) either the liquid/solids level or the
air gap above the liquid/solids.
1) From the Main Screen, press OK to
advance into the Main Menu.
2) Press OK to advance into the
Configuration Menu.
3) Press OK to advance into Basic Menu
(Application screen will appear).
4) Press Right Arrow (at least five times)
to scroll to Display.
5) Press OK to change the setting for
Display.
a) Press Up Arrow or Right Arrow
to move the triangle to the desired
setting.
b) When the setting is correct, press
OK to save the setting.
6) When done, press ESC to return to
the Basic Menu, press ESC again to
return to the Configuration Menu and
press ESC a third time to return to the
Main Screen, or if you want to
advance directly into Sensor Name,
press Right Arrow.

AD310320 Rev A 17 |
Configuration Using Display Section Three
STEP 7 - SET THE SENSOR NAME
This enables you to provide a unique name or identifier for each transmitter. This is the name of the
transmitter that appears as the main identifier when using the LevelTap™ App.
Note: It is easier to use the LevelTap™ App to change this setting than with this display. Please refer to the
LR80, LR81, LR83 & LR85 Series Manual for instructions to set the Sensor Name.
1) From the Main Screen, press OK to
advance into the Main Menu.
2) Press OK to advance into the
Configuration Menu.
3) Press OK to advance into Basic
Menu (Application screen will
appear).
4) Press Right Arrow (at least six
times) to scroll to Sensor Name.
5) Press OK to change the setting for
Sensor Name.
a) Press Up Arrow to increase the
value of the character
highlighted. Up Arrow will scroll
from A to Z, then 0 to 9 and back
again.
b) Press Right Arrow to move one
segment to the right. Right
Arrow will scroll left to right and
then back to the first segment.
c) When the setting is correct,
press OK to save the setting.
6) When done, press ESC to return to
the Basic Menu, press ESC again to
return to the Configuration Menu
and press ESC a third time to return
to the Main Screen.
Note: Only Capital / Upper Case letters are available when using the display.

|18 AD310320 Rev A
Configuration Using Display Section Three
STEP 8 - CHECK THE ECHOCURVE
EchoCurve presents a physical representation of the return signal across the operational range for the radar
transmitter. From the Main Screen, press ESC to view the EchoCurve.
There are two components to observe when looking at the EchoCurve:
• Check the Peak(s) along the EchoCurve
o Ideally, there should be a significant peak which
represents the liquid or solids level in the tank.
o This peak should be above the Envelope Line
(Threshold Filter).
o There may be more, and smaller peaks located
before and after the large peak.
Echo Peaks along the EchoCurve
• Check the Location of Peak (Return)
o There are two values above EchoCurve.
o The bottom value represents the air gap
distance from the radar transmitter to the liquid
level.
o The top value represents the air gap distance
from the radar transmitter to the peak amplitude
location.
Location of Echo Return
o These two values should be close to one another (or the same value).
§ The example above indicates that the liquid level is located 7.208 feet away (Bottom
Value) and the peak amplitude is located 7.208 feet away (Top Value) from the radar
transmitter.
§ The liquid or solids level should be at this peak location.
• Change the Distance Scaling
o The distance scaling for the EchoCurve defaults to a scale of 0 to 118’ (0 to 36m) for the LR83
and 0 to 393’ (0 to 120m) for the LR85.
o Press Up Arrow to change the Scale
• Show the Envelope Curve
o The Envelope Curve does not appear when the EchoCurve is first viewed (default).
o To show the Envelope Curve, press Right Arrow.
o To remove the Envelope Curve, press Right Arrow.
Series
Default
Up Arrow
Up Arrow
Up Arrow
Up Arrow
LR83
118’ (36m)
59’ (18m)
30’ (9m)
9’ (3m)
118’ (36m)
LR85
393’ (120m)
197’ (60m)
98.4’ (30m)
32.8’ (10m)
393’ (120m)

AD310320 Rev A 19 |
Advanced Configuration Using Display (Range Adjust) Section Four
ADVANCED ADJUSTMENTS OVERVIEW
These optional functions are used to change the transmitter output characteristics.
Note: These adjustments should only be performed after having completed the configuration steps previously
described with the transmitter installed on the tank. Where so, perform the following applicable Advanced
Adjustments.
• Dead Band
• Maximum Range
• Reverse mA
• Distance Adjust

|20 AD310320 Rev A
Advanced Configuration Using Display (Range Adjust) Section Four
DEAD BAND
This setting defines the minimum distance that the EchoBeam®will detect valid echo returns. EchoBeam®
will not read a level above this setting. All values will be in the Units previous defined during configuration.
Default setting will be at 90% of the difference between the Sensor Height and Fill-Height values.
• If Sensor Height = 20 feet and Fill-Height = 19 feet, the difference between the two values is 1 foot.
• 90% of 1 foot is 0.9 feet (set Dead Band to 0.900 feet)
1) From the Main Screen, press OK to
advance into the Main Menu.
2) Press OK to advance into the
Configuration Menu.
3) Press Right Arrow to switch to
Advanced.
4) Press OK to advance into Advanced
Menu.
5) Press OK to advance into Range
Adjust. Menu.
6) Press OK to change the value for
Dead Band.
a) Press Up Arrow to increase the
value of the number highlighted.
Up Arrow will scroll from 0 to 9
and back again.
b) Press Right Arrow to move one
segment to the right. Right Arrow
will scroll left to right and then
back to the first segment.
c) When the value is correct, press
OK to save the setting.
7) When done, press ESC to return to
the Range Adjust Menu, press ESC to
return to the Advanced Menu, press
ESC again to return to the
Configuration Menu and press ESC a
fourth time to return to the Main
Screen, or If you want to advance
directly into Maximum Range, press
Right Arrow.
Note: Dead Band is completely independent from the 20mA setting (Fill-Height). If the Dead Band is set
below the location of the Fill-Height, then the transmitter will not measure above the Dead Band. The
maximum level will not correspond to the current at the Dead Band.
Other manuals for EchoBeam LR85 Series
1
This manual suits for next models
1
Table of contents