Flowmeter SVTU-11B User manual

WatermeterSVTU-11B
Operatingmanual
SMP.407251.010 Э
(Part1)
September 2010


3
Documentation:
1. OperatingmanualSMP407251.010 OM part1.
2. OperatingmanualSMP.407251.010 OM part2. Verificationmethodology.
Tableofcontents
1Preface.......................................................................................................................6
2Assignment.................................................................................................................6
3Technicalspecifications.............................................................................................6
4Packagecontents......................................................................................................12
5Meter’s structureandfunctioning.............................................................................15
6Markingand sealing.................................................................................................18
7Packingandmarking................................................................................................18
8Safetyprecautions....................................................................................................19
9Installation................................................................................................................20
9.1 Unpackingand degreasing.............................................................................20
9.2 Installationrequirements...............................................................................20
9.3 Requirements toFSinstallationandarrangement..........................................21
9.4 Installationofmetercomponentry.................................................................22
10 Setting-upProcedures...............................................................................................32
11 Theoperatingprocedure...........................................................................................35
12 Servicing..................................................................................................................38
13 Typicalfaultinesses andmethods oftheirelimination ..............................................42
14 Storing......................................................................................................................45
15 Transportation..........................................................................................................45
16 Guarantee ofmanufacturer.......................................................................................45
Appendix Meterdesignation......................................................................................47
AppendixB Schematiccircuits formeterinstallations..................................................48
AppendixC Overallandconnectingdimensions ofcalculator......................................49
AppendixD Menuofcontrolofthemeter....................................................................50
AppendixE Examples ofreports, creatingby“SempalDM”software..........................58
AppendixF Schematicofdevice cable.........................................................................60
AppendixK Theschemeofpressuresensormounting..................................................63
AppendixL OverallandconnectingdimensionsforFS...............................................64
AppendixM Howtosethydrauliczero.........................................................................67

4
Abbreviationlist
OM –operatingmanual.
FS –flowmeasurementsectionwithultrasonicflowsensors FlS.
RТD–resistivetemperaturedetector.
RTD–T–platinumresistivetemperaturedetectormanufacturedbySEMPALCo.
NSCRТD–nominalstaticcharacteristicofRТD.
FlS–flowsensor.
TS–temperaturesensor.
PT–pressuretransducer.
DN –nominaldiameter.
N–nominaloverpressure.
PC–personalcomputer.
RDU –datareader.
–digitonthedevice display.

5
Informationforcustomers
SVTU-11Bwatermeters(hereinafterreferredtoasmeters)arecomplexmeasuringde-
vices, whichshouldundergo thestarting-upand adjustmentworksbyqualifiedpersonnel
duringcommissioning.
Manufacturer's guarantees (48 months sinceshipment)extendonthemeters which
wereputintooperationbythespecializedenterprises having correspondingcredentials from
company-manufacturer. Moredetailedinformationis resultedinsection16 ‘Manufacturer’s
Guarantee’.
Arecalibrationintervalis no morethan4years.
TheQualitySystemofSEMPALis certificatedunder ISO9001:2000.
Ifyouhaveanyquestionsaboutpurchase,maintenance,operationandservice ofme-
ters, contactus orourauthorizedregionalrepresentatives.
“SEMPALCoLТD”contacts:
3KulibinaStreet,Kyiv, 03062 Ukraine
Phone/fax:(+38044)239-2197, (+38044)239-2198.
http://www.sempal.com

6
1Preface
Thisoperating manual(hereinafterreferredtoasOM)containsthefollowing informa-
tion:theassignment,fieldofapplication,performance capabilityandcompleteness, aprin-
cipleofoperationandadesign, themethod ofinstallationandcommissioning, themethodof
operationandmaintenance service ofmeters.
Whilemeters areinexploitation,itis necessarytobestrictlyguidedbypresentOM.
Constantlyworkingatfunctionalenhancement,improvementofperformance capability
andincreaseofmeterreliability,manufacturingcompany SEMPALCoLTDreservesthe
righttoitselftochangethemeterdesignwithoutclaimingit inpresentOM.
2Assignment
2.1 WatermetersSVTU-11Bareintendedformeasurementofthewatervolumeand
mass.
Metersalsomeasurethevolume,mass, temperatureandoverpressureofthewateror
fluidswithsolidparticlesofno morethan200 µmand residuemass ofno morethan
500 mg/l(hereinafterreferredtoas awater).
2.2 Metersareappliedformetering(including fiscalmetering)ofwatervolumeaccord-
ingtoactingrulesforthecontrolofsupplyandconsumptionofwateronindustrialobjects
andobjectsofmunicipalservices. Functionalfeaturesofmeters’configurationsaregivenin
3.5.
3Technicalspecifications
3.1 Watermeters canhavenextconfigurations:
−1–1-channelwatermeter;
−3–2-channelwatermeter(twoindependentwatermeters);
3.2 Metersdependingontheirpermissibleerrorlimitswhilemeasuring volumeand
mass ofthewaterarereleasedinthefollowingmodifications: М1and М2.
3.3 Metersofallmodificationsinaccordance withenvironmentalconditionsmeetclass
СofEN14341. Themeters canworkunderthefollowingconditions:
−atmosphericpressurecanvaryfrom84.0up to106.7kPa;
−relativeairhumidityis up to95 %.
3.4 Metersmeettherequirements ofEN1434.
3.5 Themeterconsists ofthefollowingfunctionalunits:
−flowmeasurementsections withultrasonicflowsensors (FS);
−resistivetemperaturedetectors (RТD);
−SVTU-11Bcalculator.
Highestpossiblenumberofmeasurementchannelsdependingonmeter’sconfiguration
isgiveninthetable3.1
Table3.1
Channel
Number of channels
Ultrasonic
flow measurement channel
2
Temperature measurement channel
3
Pressure measurement channel
2

7
Distinctivefunctionalfeaturesofmeterconfigurationsandbasicfunctionalunitsare
representedinthetable3.2(meters’configurationsindetails aregivenintheappendixB).
Table3.2
Distinctivestructural and functional features
Configuration
1
3
1
Number of FS
1
2
2
Number of RТD
1
2
Additionallythemetercanhaveoneortwooverpressuretransducers(furtherunderthe
text-pressuretransducersorPT),whichareusedfortransformationofthewateroverpres-
sureinaproportionalelectricsignalwithcurrentfrom4upto20 mA.
Metrologicalperformance ofpressuretransducers PTfits totheindividualorder.
3.6 Meters indicatetheresults ofmeasurements inSI unitsystem.
3.7 Measuredresults aredisplayedon2-sectionalLCD.
Thefirstsectionhas8-bit7-segmentindicatorwithspecialsymbols. Themeasuringda-
taarealways displayedinthis section.
Thesecondsectionhasonerow(onthetopoftheindicator)fordisplaying textinfor-
mation. This sectionis switchedononlyduringoperationwithkeyboard.
−Thecalculatorindicates thefollowingquantities:
−volume(mass)ofthewater,m3(ton);
−volumetricflowrateofthewater,m3/h;
−watertemperature, С;
−overpressureofthewater, МPа;
−worktimeandnon-worktime, hour;
−currenttime(hours, minutes, seconds)anddate.
3.8Thecalculatorprovidesstorageofsucharchivedata asmeasuredvaluesofvolume
(mass)ofthewater, worktimeand non-worktimeand alsoaveragemeasuredtemperatures:
perhour -during70 precedingdays (hourlyarchive);
perday -within1precedingyear(dailyarchive).
All storedinformationand measurableparameterscanbereadbystandardinterface
RS-232C.
3.9 Thenumberofdisplaydigits:
−forvolume(mass)ofthewateris 8;
−forvolumetric(mass)flowrateofthewateris 5;
−forwatertemperatureis 5;
−foroverpressureofthewateris 3;
−forworktime, non-worktime, andcurrenttimeis 7.
3.10 Theminimumbitvalueofdigitaldisplaywhileindicating:
−volume(mass)ofthewatercanvaryfrom10-7 upto1m3(from10-7 upto1ton);
−volumetric(mass)flowrateofthewatercanvaryfrom0.001 up to0.1m3/h(from
0.001 upto0.1ton/h);
−watertemperatureis 0.01 °C;
−overpressureofthewateris 0.01 МPа;

8
−worktimeandnon-worktimecanvaryfrom10-5 tо1hour;
−currenttimeis 1sec.
3.11 Metersprovidemeasurementsofvolume(mass)ofthewaterinrangesofvolume-
tricflowrates resultedinthetable3.3.
Thetablecontainsvaluesofthelowerlimitoftheflowrate(qi),permanentflowrate
(qp)and theupperlimit oftheflowrate(qs).
Table3.3
Flowmeasurement
sectionFS
Measurementrangeofthewater, m
3
/hour
qiqpqs
FS-20 0.05 56.5
FS-32 0.22 22 30
FS-50 0.7 70 90
FS-65 1.2 120 160
FS-80 1.8 180 230
FS-100 2.8 280 360
FS-125 4.5 450 580
FS-150 6.5 650 850
Thelowestmeasurablevolumetricflowrateisequalto0.5·qi.Iftheflowrateislower
thanqiflowratemeasurementerrorisnotstandardized.
3.12 Thetemperatureofthewatercanvaryinarangefrom0upto150 °С.
Themeasurementrangeofsecondary(notformass calculation)temperaturescanvary
from-49 °Сto150 °С.
Numberofchannelsfortemperaturemeasurementcanvaryfrom1tо3inaccordance
withtheorder.
3.13 Themaximalpressuremeasuredbythemeteris 4 МPа.
Therangeoftheelectricsignalsfrompressuretransducervariesfrom4upto20 mA.
3.14 Meters haveoutputsignalfordatacommunicationbystandardinterface RS-232C.
3.15 Meters’powersupplyisrealizedbybuilt-inlithiumbatteryof3.6V.Batterylife-
cycleis 5-6years. Batteryreplacementcanbedonedirectlyon-sitewithoutmetershutdown.
WARNING!Meteris deliveredwithswitchedoffbattery.
Whiledeliveringbyairtransportthebatteryshouldbeswitchedoff andremovedfrom
themeter.
3.16 Nominaldiameters(DN),overalldimensions andweightofFS, аndalsolengthand
weightofRТDdependingontheirtypeareshownintables 3.4, 3.5 andonFigure3.1.
Thenotice
1ItispossibletoincreasethetotallengthofFS duetothelengthofstraightsec-
tions beforeandafterplaces forinstallationofultrasonicflowsensors.
2DN isthedesignationofnominalbore,whichnumericalvalue isapproximately
equal tointernaldiameter ofattached pipesections.
3.17 Forconfigurations1and3inmodification М1watermetermeetsaccuracyclass 1,
inmodification М2watermetermeets toaccuracyclass 2inaccordance withEN1434.

9
Permissibleerrorlimitsforconfigurations 1and3whilemeasuringwatervolume:
−±(1+0.01∙qp/q)%foraccuracyclass 1;
−±(2+0.02∙qp/q)%foraccuracyclass 2;
whereqismeasuredflowratevalue,m3/h.
3.18 Limitsofpermissibleabsoluteerrorforwatermeterswhilemeasuring watertem-
peratureare ±(0.2+0.002⋅Θ) °Сwhere Θisnumericalvalueoftemperature.
3.19 Limits ofpressuremeasurementerror:
−±0.5%whenPTfrommeter’s setareused;
−
22
PT
0.2
±+δ
whencustomer's PTareused,
where δPT is permissibleerrorlimit ofcustomer's PT.
Specificcoefficients ofPTcurveareenteredintocalculator’smemory.
3.20 Thecalculatorhastwopassivepulseoutputs. Pulseoutputsmeettoclass OD inac-
cordance withEN1434.
Maximumswitchedvoltageshouldbenomorethan10 Vwhileswitchedcurrentisno
morethan10 mА.
Maximumpulse-repetitionfrequencyis 100Hzwhilepulsedurationis 4ms ±0.5ms.
3.21 Pulsesettings (setbyuser)arefrom1to9999999 pulse/unit. Where«unit»is unit of
measurefortransformedphysicalquantity.Theusercanselectfromthefollowingphysical
quantities:volume(pulse/m3),mass (pulse/ton).
3.22 Limitsofabsolutemetererrorwhilemeasuringtheworktimeandnon-worktime
are ±1minuteper24 hours.
3.23 Measurementinformationaboutwatervolume,worktimeandnon-worktimeis
storedinnonvolatilemeter’smemorywithin8years withpoweroff.
3.24 Themaximalwateroverpressureinsideflowmeasurementsectionis1.6 МPа
(16 kgf/cm2).
3.25 Timeformetercommissioning afterbatteryinstallationisnomorethan10 minutes
(batteryis switchedoffindeliveryset).
3.26 Protectionclass ofcalculatorenclosureis IP65.
3.27 Calculatormass is no morethan750 gram.
3.28 Calculatoroveralldimensionsdon’texceed170×110×35 mm,andwithdevicecon-
nectorand wallmountingaccessories –250×110×60 mm(see Appendix С).
3.29 Nominaldiameters(DN),overalldimensions, andFS weight,aswellaslengthand
weightofRTDdependingontheirtypeareshownintables 3.7, 3.8andinfigures 3.1,9.9.
3.30 Meanerror-free worktimeformetersisnotless than50 000 hours, forcalculators –
100 000 hours.
3.31 Totalaveragemeterlifecycleis notless than12 years.

10
Table3.4Types, dimensions andweightofRTD-T
RTDtypes Lengthinmm, nomorethan Weight, kg,
nomorethan
L
RTD
L
4 58 86 0.06
2 80 108 0.08
3 150 178 0.1
5 310 346 0.25
6 360 396 0.3
Thenotice.Selectthetypedepending onDN accordingtoinstructionsmentionedintables9.2,9.3
andinfigures9.8, 9.9,9.10
Mountingdimensions forRTDoftypes 2, 3, 4aregiveninFigure3.1.
Figure3.1
7
34
6
(1 pcs.)
9.7*
L
RTD
forsealing
Screwwithahole
20.5 L
(screwlocation
isundefined)
1
8
Hexahedrons=17mm
Plug УЗНЦ 05-7
M10x1.5 13

11
Table3.5Overallandconnectingdimensions, FSweight (fordrawingsseeAppendix L)
FS Nominaldi-
ameter, mm
Nominaloverall andconnectingdimensionsofFS,
mm n,
pcs.
Weight, kg,not
morethan(nofasts)
DN
L
D
f
H
D
d
d1
ForFS
Forflanges
У
-
20
20
270
Pipe
. G1”
90
∅
20
-
-
-
2.3
2.7*
У
-
32
32
180
M48x2
79
∅
32
-
-
-
1.8
2.6*
У
-
50
50
180
∅
122
-
∅
50
∅
102
∅11
6
4.8
2.2
У
-
65
65
200
∅
144
-
∅
63
∅
124
5.8
2.9
У
-
80
80
210
∅
155
-
∅
79
∅
135
6.9
3.2
У
-
100
100
230
∅
184
230
∅
(95…105)**
∅
164
8
7.8
4.1
У
-
125
125
265
∅
210
270
∅
(119…131)**
∅
190
10
10.6
5.2
У
-
150
150
315
∅
236
296
∅
(143…156)**
∅
212
∅
13
20.0
7.7
*
Weightofstraightsectionswithscrews
.
** NominalboresD aremultipleof1mm

12
4Packagecontents
4.1 Themeterpackagecontents is representedinthetable4.1.
Table4.1 Nameand
reference designation Marking Number
Additional information
SVTU-11B,
including: SMP.407251.010 1pcs. Configuration and com-
pleteness in accordance
with the order (see items
1…8)
1SVТU-11BCalculator
withbattery SMP.408843.012 1pcs.While in delivery the bat-
tery is switched off
2Flowmeasurementsection
(FS)withscrews forflowsen-
sors fixing(noscrews incom-
pletesetforFS-20…FS-150)
Markingfor FS-20 to
FS-150 included –see
Table 4.2
See ad-
ditional
informa-
tion
Number, configuration
and dimension-type in
accordance with the order
(see Appendixes Aand
tables 4.2, 4.3)
3Ultrasonicflowsensor
(FlS) withfluoroplasticseal
ring(FlSforFS-20 aren’tdeli-
veredseparately.Flowmea-
surementsectionFS-20 is deli-
veredtogetherwithFlS)
SMP.407151.009
(forFS-32…80);
SMP.407151.011-01
(forFS-100);
See ad-
ditional
informa-
tion
Number of FlSfor one
FSis defined by number
of places for their instal-
lation in accordance with
the order (see table 4.2)
4Resistivetemperaturede-
tectorRTD-TSMP.405212.003-01
(−02, −03)
See ad-
ditional
informa-
tion
Number and configura-
tion (type) in accordance
with the order
5Overpressuresensors (PS) -
See ad-
ditional
informa-
tion
Number, type and com-
pleteness accordingto the
order. Complete set can
include elements indi-
cated in Appendix K
6Connectioncable(device
cable) SMP.685694.029 1pcs.Number of communica-
tion lines and their length
accordingto the order
(see Appendixes A)
7SVТU-11B. Watermeter
OperatingManual SМP.407251.010 Э 1 pcs.
8Packaging (set) SМP.323248.002 1set
9Modem Type is selected in coor-
dination with the custom-
er
Incoordination with the
customerwhile ordering
the МDМunit
10 DataReaderSpecial order
11 Adaptermodule МС-1 Special order

13
Nameand
reference designation Marking Number
Additional information
12 Protectionenclosure SMP.301538.006 Special order
13 SVTU-11B. Operatingma-
nual. Part2«Verificationme-
thodology»
SМP. 407251.010 Э2 1 pcs.Special order
14 Reservebelongings SМP.305653.001 Completeness and num-
ber is ordered
Notes
1FS is deliveredwithflanges andfasteners see table4.3.
2FS canbedeliveredwithstraightpipesections(lengthisup to25internaldiameters
ofthepipeline).Thespecifiedsectionscanbeweldedtoflangesifnecessary.Whileor-
deringstraightpipesectionsallnecessarymaterialsforinstallation(forexample,elec-
trodesforwelding, paint, sealingmaterials,etc.)canbedeliveredadditionally
3Reservebelongingscanincludeelementslistedinthetable4.1,table4.3,enclosure
oftheSVТU-11Bcalculator, aboardoftheSVТU-11Bcalculatoras itwas ordered.
4DataReadercanbedeliveredasapocketPC,laptop orasadedicateddevice RDU-
02. ModelofpocketPCorlaptop is selectedbycustomer.

14
Designationand basicparameters ofFSincompletesetareindicatedinthetable4.2.
(ForoutlinedrawingofFS see AppendixL)
Table4.2
FS DN,
mm Markingon FS Number of places
for flow sensors Marking
Nominalbore
Nominal
pressure
Singlebeamflowmeasurementsection
FS -20 20 DN 20
PN16
- FS -20
FS -32 32 DN 32
2
FS -32
FS -50 50 DN 50 FS -50
FS -65 65 DN 65 FS -65
FS -80 80 DN 80 FS -80
FS -100 100 DN 100 FS -100
У-125 125 DN 125 FS -125
У-150 150 DN 150 FS -150
Anothercomponentryincludedindeliverysetinaccordance withtheobligatoryorad-
ditionalorderarelistedinthetable4.3.
4.2 Meter’sconfiguration,FS type,connectingcables, numberofcommunications
channels and theirlengtharespecifiedwhileordering.Theformatofmeter’smarking isgiv-
eninAppendixA.
DescriptionofdesignfeaturesforFSandalsodrawings, mountingdimensionsarere-
sultedinAppendixK.
Warning!!!
IdentificationofFS standardsize isprovidedbymarkingofnominaldiameterDN on
FS body.
ThusnumericalvalueinFSreference designationmeetstoanumericalvalueinade-
signationofanominaldiameterDN (see table4.2).
Example:
−Marking“DN 32”putontheflowmeasurementsectionwithreference designation
FS-32.Nextelementofmarking “РN16”meansthatthisflowmeasurementsectionisin-
tendedforuseinheat-orwater-supplysystems withoverpressure1.6 МPа(16 kgf/сm2);
Table 4.3Componentryincludedindeliveryset
# Name Marking Assignment Number Setofdelivery
oblig
a-
tory
Special
order
1
Pipenipple
SMP
.752291.001
Connectionof FS
-
20 with
straightpipesection
2pcs. for
1FS
+
2
Pipenipple
SMP
.302661.002
Connectionof FS
-
32рwith
straightpipesection
+
3
Sleevenut
SMP
.758422.006
Mounting of FS
-
20
+
4
Sleevenut
SMP
.758422.001
Mounting of FS
-
32
+
5
Flange
SMP
.711154.013
-
01,013
-
03,
013-05(inaccordance withDN
ofFS)
Mounting of FS
-
50…80
+

15
6
Flange
SMP
.711154.004
-
03 …004
-
05
(inaccordance withDNof FS)
Mounting of FS
-
100
+
7
Gasket(paronite)
SMP
.754152.013
Sealingof flangesFS
-
20
2pcs. for
1FS
+
8
Gasket(paronite)
SMP
.754152.009
Sealingof flangesFS
-
32
+
9
Gasket(paronite)
SMP
.754152.007…007
-
16
Sealingof flanges
FS-50…100
+
10
RTDpocket
(L
RTD
=58mm,type4)
SMP
.753137.002
-
03
Protectionof theRТDfrom
hydraulicimpacts
1pcs. for
1RТD
+
11
RTDpocket
(L
RTD
=80mm,type2)
SMP
.753137.002
-
01
+
12
RTDpocket
(L
RTD
=150mm,type3)
SMP
.302634.002
+
13
Sealingring (fluoropla
s-
tic)
SMP
.754176.003
Sealingof RТDfortypes2,
3,4
+
14
Sealingring (fluoropla
s-
tic)
SMP
.754176.003
-
01
Sealingof RTDbushfor
types2, 3, 4
1pcs. for
1sleeve
+
15
Bush(forangle
α
=45º)
SMP
.723144.007
Installationof RТDfortypes
2,3,4withoutbush 1pcs. for
1RТD
+
16
Bush(forangle
α
=60º)
SMP
.723144.008
+
17
Bush(forangle
α
=90º)
SMP
.723144.009
+
18
Bush(forangle
α
=45º)
SMP
.723144.007
-
01
Installationof RТDbushfor
types2, 3, 4 1pcs. for
1sleeve
+
19
Bush(forangle
α
=60º)
SMP
.723144.008
-
01
+
20
Bush(forangle
α
=90º)
SMP
.723144.009
-
01
+
21
Connectingpipe(G1/2"x
1/4")
SMP
.716161.004
Forinstallationof pressure
sensor 1pcs. for
1PS
+
22
Gasket(paroniteorflu
o-
roplastic)
SMP
.754156.010
ForPS
+
23
Gasket(paroniteorflu
o-
roplastic)
SMP
.754156.010
-
01
Forthreaded
+
24
G
а
sket
SMP
.754156.001
Sealingof device
connector
1
pcs.
+
25
AB1000WLVSet:
-crampon n/a Mounting of theCalculator
2pcs.
+
-
corbel
2
pcs.
+
-
washer«star»
2
pcs.
+
-
screw М4(hex)
2
pcs.
+
Fixings:(threaddiameterdandbolt lengthLfittoholesinflangesand total
flange
thickness)
26
-
screw М3x10
Mounting of deviceconne
c-
tor
4
pcs.
+
27
-
washer
3
4
pcs.
+
28
Bolts А.(dxL).88.35.019
FS-50…100
In acco
r-
dance with
totalnumber
ofholesin
FS flanges
(see table
3.3)
+
29
Nuts А.(d).9.35.019
+
5Meter’sstructureandfunctioning
5.1 Theprincipleofflowmeasurementofwatervolumeisultrasonictime-of-flight. The
timeforthesoundtotravelbetweenatransmitterandareceiverismeasured.Thetimedif-
ference isproportionaltotheaveragefluidvelocityandflowratecorrespondingly.Theinte-
gratedmomentaryflowratevaluesgivetheinformationaboutwatervolumewhichhas
passedthrough FS.Themass iscalculatedasafunctionofvolumeanddensityoftheflow
depending onits temperature.
5.2 Watertemperatureismeasuredbyplatinumresistivetemperaturedetectorsinstalled
inappropriatepipes.

16
5.3 Flowratemeasuringprocess iscontinuous. Flowratemeasurementisoccurreda
fewtimespersecondand obtaineddataarestored.Reading ofstoreddataisoccurred1time
persecondtocalculatethequantityofwater.
Measurementcyclefortemperatureandpressureis onetimein10 sec.
Everymeasurementcyclelastsfor1sec andincludesmeasurementofwaterparame-
ters, as well as self-diagnostics ofthemeter.
5.4 Watermeters aredeliveredinfollowingconfigurations (see AppendixB):
−1–1-channelwatermeter;
−3–2-channelwatermeter;
SchemesforsensorsconnectionindifferentconfigurationsarerepresentedinAppen-
dixB.
5.5 Calculationand storageofaveragetemperaturevaluesthemetercarriesoutas
arithmeticmeanforsetperiod oftime(hourorday).
5.6 Themetercarriesoutflowmeasurementintherangefrom0.5qi(qmin)tillq
s
(see table3.2).
Forall types ofFSmeasurementerrors areprovidedinarange[qi (qmin);qs], butinsub-
ranges[0.5.qi(qmin);qi(qmin)[ thementionedmeasurementerrorsarenotstandardized,how-
everdevice workingcapacityis kept, andwatermass storingandcalculationarecarriedout.
Whilemeasuringinstantaneousflowratevalues .q <0.5qi (qmin)thedevice indicates
zeroflowand mass storing m=q⋅ρisn’tcarriedout.
5.7 Measurementofwateroverpressureiscarriedoutbycalculatorwhilemeasuringthe
currentofelectricsignalfromthepressuretransducer.Overpressurevalues op (MPa),
measuredand displayedbythecalculator,andcurrentIin (mA)onaninputofpressuremea-
suringchannel(onthecalculatorinput)areconnectedbyafollowingratio:
21
opmeas11
21
(PP)
P(II)P
(II)
−
=−⋅+
−(5.12)
whereP
1
andP2arepressurevaluesintwopointsofpressuretransducercharacteristic
(forexample,minimalandmaximumpressure);
I
1
andI2arecorrespondinglycurrentsonthePToutputsintheabovespecified
points;
U
meas is valueofmeasuredcurrentontheoutputofPT.
5.8 Whilemeasuringtimeparametersthemetercarriesoutmeasurementoffollowing
quantities:timeofcorrectwork(running orworktime),timeofincorrectwork(timeofer-
rors),powerontime,poweroff time(non-workoridletime),andalsodisplayscurrent
date/time(takingintoaccountdaylightsaving).
Timeofcorrectwork(runningtime Тrun orworktime) –device operating time
(poweron,noerrormessages).Timeofcorrectworkon1-stand 2-nd channelsisdisplayed
on thewatermeterindicatorinamode «Indicationofadditionalparameters» (see item
11.2.2, table11.3ofthis Manual, andalsoAppendixD).
Timeofincorrectwork (timeoferrorsTerr) –device operating time(poweron,
thereareerrormessages). Тerr valuesarepresentedinprintoutsofstoreddataarchivesand
errorarchives.

17
Power ontime Тpower –totaltimewhenthedevice powersupplyvoltageison.Itis
indicatedonthemeterdisplayinamode “Indicationofadditionalparameters” and alsois
presentedinadailyarchiveprintout.
Power off time (idletime Тidleornon-worktime) –totaltimewhenthedevice pow-
ersupplyvoltageisoff.Itisdisplayedonthemeterindicatorinamode «Indicationofad-
ditionalparameters», andalsoincludedindailyreport.
Currentdate/time –currenttime(taking intoaccountdaylightsaving).Itisindicated
onthemeterdisplayinamode “Indicationofadditionalparameters”. Automaticswitch
todaylightsaving timemodecanbe on/off.
Measurement,indicationandregistrationofabove-listedparametersarecarriedoutin
hours. Infigure5.1thetiming sheetisrepresented.Itexplainshowtimeofcorrectwork
Тrun andtimeofnon-working condition Тnw foraccounting period Тacc areresulted. Dur-
ing Тrun authenticmeasurementofall parameterswasmade,during Тnw thereisnoregis-
trationofanyparameterorthereis no powersupplyvoltage(batterydisconnected).
Figure5.1
5.9 ToconnecttoexternaldevicesRS-232Cis applied.RxD,TxD,DTR,RTSlinesare
used.
5.10 Metershaveafewseparatemoduleswhichconnectedbetweenthemselves. They
arecalculator,temperaturesensors, flowmeasurementsectionwithultrasonicflowsensors,
pressuretransducers.
5.10.1 Thecalculatorhousing hasfourcontrolbuttons, thedigitaldisplay,slotforcon-
nectingcable.
Watermetercontrolbuttons aredescribedinchapter11 andAppendixD.
5.10.2 Meter’sdigitaldisplayhastwosections. Thefirstonecontainsa8-bit7-
segmentindicatorandspecificsymbols. Inthissectionthemeasuringdataandunitsaredis-
played.Thesecond sectionisarowfordisplaying textinformation.Itisarrangedintheup-
perpartofindicator.Thissectionisenablediftheuserneedstosee additionalinformation
andcompletesomekind ofactions indevicemenu.
Thebasicworking modeofthemeteris displaying measuring datafrommainmenu. At
this timethesecondsectionis switchedoff.
5.10.3 FS isaspoolpiece withflangesand bushesformountingultrasonicflowsen-
sors. FS sizes(linearandangular) haverigidtolerancestoachieverequiredmeasuringaccu-
racyandstabilityinoperation. Sizes areresultedinthetable4.2, andalsointheAppendixL.
Тrun
Тpower
Тidle
Тerr
Тnw
Тacc

18
5.10.4 RТDareinstalledinpipelinesinspecialthermometerpocketsorbushes(inac-
cordance withtheorder)ontheflowand returnpipes ofheatexchangesystem.
5.10.5 Thelengthofconnectingcablesisdeterminedaccording toalay-outofmeter
components and canbewithinthelimits:
−from2to10 m(max30 m)forultrasonicflowsensors,
−from2to10 m(max30 m)fortemperaturesensors,
−from2to10 m(max30 m)forpressuresensors,
−from2to30 m(max50 m)forRS-232 interface.
Extendingcablelengthup valuesinsquares possibleonlybyadditionalagreement.
6Markingand sealing
6.1 Metermarkingonthecalculatorcontains thefollowingdata:
−nameandreference designationofwatermeters;
−trademarkofthemanufacturer;
−accuracyclass (meter’smodifications М1or М2);
−environmentalclass;
−serialnumber;
−dateofmanufacture(twolastdigits ofayear, printedbeforeserialnumber).
Ontheindicating device ofwatermeters:
−flowratelimit values.
Ontheflowmeasurementsection:
−nominalbore;
−flowdirectionarrow;
−maximumoperatingpressure.
Serialnumbers ofRТDandFSareputontheirbodies.
6.2 FS marking containsnominaldiameterDNandmaximaloperationaloverpressure
value N.
6.3 Metercomponentsaresealedup topreventanon-authorizedaccess toadjustment
ofmeter’s readings.
6.4 Thecalculatorissealedupbytwomasticseals. Sealsareputonfixing screwson
lateralcovers. Formasticfixingsealingcups areprovidedunderscrewheads.
Forrealizationofadditionalsealingbymeansofaleadseal,thescrewswithholesin
headsonlateralcoversarestipulated.Oncustomer’sdemandthereplacementofthese
screws byadditionalsealingcups is possible.
Markingofconnectingcables areputonlabelslocatednearcorrespondingconnectors.
7Packingand marking
7.1 Markingoftransportcontainerhasspecificinstructions"CAUTIOUSLYFRA-
GILE", “PROTECTFROM MOISTURE”,"TOP".
7.2 Metercomponentsarepackedintotheboxesinaccordance withdrawings ofmanu-
facturer.

19
IncoordinationwiththecustomerFS deliverywithouttransportcontainerorincustom-
er’s containeris possible.
8Safetyprecautions
8.1 Meter’sdesignmeetstosafetyprecautionsinaccordance withStateStandard
GOST12.2.003.
8.2 Ifmeterrunstogetherwithequipmentof220Vpowersupply,thensafetyrequire-
ments forworkwithelectricalinstallations(voltageupto1000V)shouldbeconsidered.

20
9Installation
9.1 Unpackingand degreasing
Unpacking anddegreasingofmetersarecarriedoutaftertheirbeingindoorswithin2
hours, whileambienttemperaturevariesfrom10 up to30 Сand relativehumidityis nomore
than80 %.
Tomakeunpacking, pleasefollowtheinstructions:
−openpackingboxes;
−takeapackagewiththecalculatorand operationaldocumentation;
−checkupcompleteness ofmetersinaccordance withtheorder;
−takemetercomponentsfrompackingboxes, makeanexternalsurveyand becon-
vincedthatthereisnomechanicaldamages, coatinginfringementsandisolationofconnect-
ingcables.
Batteryinstallationintocalculatorshouldbedoneas follows:
−unscrewfourscrews, whichfixacoverofbatterymodule;
−connectabatterytoacalculatorboardinaccordance withthefigure12.1;
−carefullyputabatteryintoamodule,batterycableshouldbelocatedrightinsidea
module;
−closeacoverofmoduleandtightenitupbyscrews.
Waitforappearance ofanotice “battery”ontheindicator.
Setcurrenttime. Proceduretosettimeis describedinAppendixD.
9.2 Installationrequirements
9.2.1 Installationofmetercomponentsiscarriedoutinaccordancewithchosencon-
figuration,necessityofapplicationoftheadditionalequipment, andalsoparameters ofwater
consumptionobject.
Schematiccircuitsformeterinstallationdepending onitsconfigurationaregivenin
AppendixB.
9.2.2 Climaticconditionsinaroom,wheremetercomponentsareinstalledshouldbe
as follows:
1)Inaplace forFS andRТDinstallation:
−ambienttemperaturecanvaryfrom −40 °Cupto+70 °C;
−ambienthumidityis upto95 %, whiletemperatureis 35 °С;
2)Inaplace forcalculatorinstallation:
−ambienttemperaturecanvaryfrom5 °Cupto55 C;
−ambienthumidityis upto93 %whiletemperatureis 25 °С.
Climaticconditionsinaplace forinstallationoftheadditionalequipmentshouldmeet
theconditions specifiedintheoperationaldocumentationforthis equipment
9.2.3 ProtectionofplacesforFS and RTDinstallationfromdirectingress ofmois-
ture, dirt, oils andaggressiveliquids shouldbeprovided.
Thecontentofacidand alkali fumeinairofpremises, wheremetercomponentsarein-
stalled, shouldbewithinthelimits ofsanitaryrules.
ForoutdoorFSinstallationitisrecommendedtoprovideprotection(canopy)against
directingress ofatmosphericprecipitates onultrasonicflowsensors.
Table of contents
Other Flowmeter Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Mid-Continent Instruments
Mid-Continent Instruments 4300-4 Series Installation manual and operating instructions

Sealey
Sealey TP101 instructions

International Light Technologies
International Light Technologies ILT2400 Operation manual

Extech Instruments
Extech Instruments 407736 user manual

GREISINGER
GREISINGER GDH 01 AN operating manual

Fluke
Fluke 8808A Calibration manual