manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Fourtec
  6. •
  7. Other
  8. •
  9. Fourtec MicroLite User manual

Fourtec MicroLite User manual

Other manuals for MicroLite

1

Popular Other manuals by other brands

Opticon EduMic Compatibility guide

Opticon

Opticon EduMic Compatibility guide

Kessil A360W owner's manual

Kessil

Kessil A360W owner's manual

TEFAL ULTRA COMPACT MIXER manual

TEFAL

TEFAL ULTRA COMPACT MIXER manual

Martin Audio WS18X - SCHEMATICS manual

Martin Audio

Martin Audio WS18X - SCHEMATICS manual

Hendrickson AR2 series Service instructions

Hendrickson

Hendrickson AR2 series Service instructions

GO Industries 38017 installation instructions

GO Industries

GO Industries 38017 installation instructions

Cissell 30LBCMAN2 manual

Cissell

Cissell 30LBCMAN2 manual

RIDGID K-60SP manual

RIDGID

RIDGID K-60SP manual

Gage Bilt GB751SH-5 Original instructions

Gage Bilt

Gage Bilt GB751SH-5 Original instructions

OVE SIERRA 60 installation manual

OVE

OVE SIERRA 60 installation manual

ReelCraft 7800 OLB21 operating instructions

ReelCraft

ReelCraft 7800 OLB21 operating instructions

Body Armor 4x4 ECO-SERIES user manual

Body Armor 4x4

Body Armor 4x4 ECO-SERIES user manual

ROBLIN DROP Dimensions

ROBLIN

ROBLIN DROP Dimensions

EMG EMG-ACB BARREL Installation information

EMG

EMG EMG-ACB BARREL Installation information

MiniSolar MT-150 Assembling Guide

MiniSolar

MiniSolar MT-150 Assembling Guide

Parkside PSZ 250 B2 Operation and safety notes

Parkside

Parkside PSZ 250 B2 Operation and safety notes

ABB REM 610 Technical reference manual

ABB

ABB REM 610 Technical reference manual

Tyent ACE-13 user manual

Tyent

Tyent ACE-13 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Installing the DataSuite Software
1. DataSuite software is used to configure the MicroLite as well as download and
analyze the recorded data. Download the free DataSuite software from the
www.fourtec.com homepage.
2. Run the setup file to install the Windows ® software on Windows XP, Vista or
7 (32 or 64-bit). Follow the instructions to complete the installation.
3. Double-click the DataSuite desktop shortcut to launch the software. After the
configuration screens are completed, connect the MicroLite to the computer.
Connecting the MicroLite to the Computer
1. Only once the DataSuite software has been installed, connect the MicroLite
data logger via the USB interface. Following detection by the PC, the MicroLite
icon will appear in the main DataSuite window: .
2. If the logger icon does not appear, ensure the device is properly connected to
the USB port and go to Devices > Detect Device from the main menu.
Configuring the MicroLite from the DataSuite Software
1. Right-click on the MicroLite icon in the main DataSuite window and select
Setup from the context menu to launch the Setup window.
2. In the Properties tab, configure the relevant parameters e.g. sampling rate,
temperature scale, operation options, cyclic mode, unit comment, etc.
3. In the Sensors tab, if the MicroLite must be configured with alarm levels, select
the alarm checkbox and enter the specific Low and High alarm values.
4. In the Boomerang Report tab, enable the Boomerang checkbox to configure
this feature for automatically sending PDF data report via email when MicroLite
is connected to another computer. Refer to the Boomerang section for more
details.
5. From any tabs, click Setup and Run to start logging or Setup to run the logger
later. The MicroLite icon will appear as follows when in Run mode: .
6. To stop the MicroLite when connected to the computer, right-click the
MicroLite icon and select Stop from the context menu.
MicroLite Layout
General Specifications (of all MicroLite Models)
LITE5008P/LITE5032P (Internal Temp)
Range: -40 to 80 °C Accuracy: ± 0.3 °C
Memory: 8,000 or 32,000 samples
(cyclic)
Sampling Rate: Once per second to
once every 18 hours
Operating Mode: Timer Run,
Immediate Start, Push to Run (with
magnet)
Battery Life: Up to 2 years
depending on sampling rate,
operating conditions
Power Supply: Replaceable 3V
CR2032 battery
LITE5032P (Temp/RH)
Temp Range: -40 to 80 °C Accuracy: ± 0.3 °C
RH Range: 5 to 95 % Accuracy: ± 2 %RH
LITE5032P-EXT (External NTC Temp input)
Range: -50 to 150 °C Accuracy: ± 0.3 °C
LITE5032P-4/20 (External Current input)
Range: 4 to 20 mA Accuracy: ± 20 µA
LITE5032P-V (External Voltage input)
Range: 0 to 10 V Accuracy: ± 5 V
Page 1
MicroLite Quick Start Guide www.fourtec.com | sup[email protected] Rev B, Jan-14, PN: 50257, © Fourtec –Fourier Technologies
2012
Temp/RH
sensor
location
Battery
holder
Magnet
holder
LCD
screen
Logger
cap
LED
USB
interface
MicroLite with external input in cap
Cradle for wall
mounting
Page 1
6. When complete, click Setup and Run.
7. Ensure that the destination computer, where the MicroLite will be connected
to download the data, has DataSuite installed with an Internet connection. Go
to Tools > Options > Email Settings and use the Test Email function in the
Email Profile Manager to verify the Email settings are working correctly.
MicroLite Working Modes
STOP: MicroLite is idle and not recording. When stopping, the logger
immediately ceases to record data, displaying STOP on the LCD.
To stop the logger directly from the unit, select the Turn off on key press option
in the logger Setup window in DataSuite. While the logger is running, place the
magnet key in the magnet holder for at least 15 seconds. The logger will then
stop sampling.
RUN: MicroLite is recording data. It will stop recording automatically when its
memory has reached full capacity (8,000, or 32,000 samples depending on the
model). When full the LCD displays FULL.
Note: For External input models only, the logger will only start running when
the cap is closed, to avoid data sampling before the external sensor is
connected
CYCLIC RUN: Similar to Run mode, but MicroLite will continuously sample data,
recording over old data when the memory reaches its capacity, beginning with
the earliest data recording. Cyclic Run mode must be selected from the logger
Setup window in DataSuite.
TIMER RUN: MicroLite can be configured to start recording at a predetermined
time. When MicroLite is set to Timer Run, the logger LCD displays tRUN.
PUSH TO RUN: MicroLite will only begin logging data when the user operates
the logger’s reed switch with the MicroLite magnet key. The logger LCD displays
PUSH following software configuration.
MARKING TIME STAMP: While the logger is running, placing the magnet key in
the magnet holder for at least 2 seconds records a new sample and time stamp
to the logger memory. Useful for marking specific events.
Downloading and Viewing the Logger Data
1. Connect the MicroLite to the computer. The logger is immediately detected and
data download begins automatically, as indicated by this icon:
2. To manually download data, right-click the MicroLite icon and select Download
Data.
3. View the downloaded data by double-clicking the MicroLite icon. The online
Graph window opens, from which you can also display data in Table or
Statistics views.
4. To open archived data, go to File > Open, select the relevant loggers from the
list, use the calendar to select the data set and click OK.
Using the Boomerang Feature
MicroLite units configured with Boomerang are able to automatically email a PDF
data report, via email to predefined contacts when connected to a computer. The
report contains all recorded data in graph format, alarm levels, logger profile and
other statistics. This feature is especially useful in the cold chain shipment industry.
The destination computer requires DataSuite and an Internet connection. A default
fourtec email profile is already defined within DataSuite.
To enable Boomerang:
1. Right-click on the MicroLite icon and select Setup.
2. Configure the logger using the Properties and Sensor tabs, and then click on
the Boomerang Report tab. Select the Include Boomerang Report checkbox to
enable this feature on the connected logger.
3. Select the relevant temperature unit, Cor F.
4. Enter the Generated by and Comment fields.
5. To add the Boomerang report recipients, click Add Contacts and select from the
contact list or click to manually add an email address
MicroLite Quick Start Guide www.fourtec.com | support@fourtec.com Rev A, Nov-12, PN: 50257, © fourtec –Fourier Technologies
2012
MicroLite Quick Start Guide www.fourtec.com | sup[email protected] Rev B, Jan-14, PN: 50257, © Fourtec –Fourier Technologies
2012
For full instructions on using the MicroLite with the
DataSuite software, please download the DataSuite
User Guide from www.fourtec.com/downloadcenter.
Page 2