
measure some of the vacuum units relative to your current atmospheric pressure. (mmHg, inHg,
%Vacuum)
In this mode move the “-” cursor with the ▲ or ▼ button to be next to
the item to be set. Now press ► “Enter”.
On the screen you will now see the “Live Pressure” and the last
calibrated setting.
Before you calibrate the pressure make sure the vacuum line from the
unit is NOT connected to your vacuum system or pump and that there
are no valves that may affect the pressure to the sensor. You need an
open line to the AutoVH 1 units vacuum chamber to calibrate accurately.
Press ► “Enter” to set the Calibrated pressure equal to the atmospheric
pressure.
Now press the ◄“Left” or “Back” button to exit this calibration menu
and press the ◄“Left” or “Back” button again to exit the “SETTINGS”
menu.
To Set the Pressure range for the Vacuum Pump
Press the ▲ or ▼ button to move the “>” pointer to the “Pressure
Settings” line
Press ► “Enter”
The unit will now enter the Pressure settings menu.
The basic unit of pressure is mBar and the unit measures the
absolute pressure in mBar and this is converted to the other units
selectable on the controller. With the unit’s vacuum line not attached
the atmospheric pressure would be seen on the display. For
Gauteng, South Africa, it is about 860mBar and should be close to
1000mBar at the coast. The reset function in the “Settings” menu will
allow you to set vacuum levels relative to the calibrated pressure.
In this mode move the “-” cursor next to “U”, “L” and “Units” with the ▲ or ▼ button to be next to the
unit. Now press ► “Enter”.
“U”: upper pressure limit (least vacuum)
“L”: lower pressure limit (highest vacuum)
Press ▲ or ▼ button to select the desired unit.
(mBar, %Vacuum, inches mercury (inHg) or mmHg)
See the table at the back of the manual for the definitions and relative values between units and
some conversion calculation.
It is important to note the Upper and Lower definitions used here. The upper limit is the least vacuum
and the Lower limit is the higher vacuum setting. In other words your vacuum pump will start at the
Upper limit of pressure and then pump air out of the system, lower the
pressure until it reaches the Lower limit.