Fukuda Denshi DSC-8510 User manual

Patient Monitor
●●●●●●●●●●●●●●●●●●●●●●●●●●●●
DS-8500 System
Service Manual
DSC-8510/8530
CC-82/83
zBefore setting up/maintenance, please read this “Service Manual” thoroughly.
zAfter reading, keep this manual for future reference.

Service manual Delivery Notice
Service Manual (No. ): Q’ty 1
Please confirm your receipt of the above service manual by filling in
and sending us back this sheet by return.
Delivery of Service Manual
Please note that this service manual is a confidential document and
needs to be kept with an utmost care under person in charge. If the
technical drawing in the service manual is unreadable, you can
request it to us by specifying a page or a part.
When you have received this service manual, verify that there is no
paging disorder or missing page. You are requested to kindly fill the
underlined area below in this sheet and send it back to us after
verification. Please be noted that photocopying of this manual is
strictly prohibited.
Also, for improvement of the future service manual, your comment and
request will be appreciated. If you have any comment or request on
usability, viewability, readability, or if you notice anything hard to
understand on this service manual, please inform it to us.
Fukuda Denshi Co., Ltd.
Development & Production Support Dept.
2-35-8 Hongo Bunkyo-ku, Tokyo, 113-8420 Japan
Receivedby
of (Company/Hospital)
onthedayof

Revision History
Model Name
DS-8500 System Service Manual
Edition Revised Items Reason of the Revision Revised Date
1
-
New Edition 2011.8

Blank Page

Preface
Thank you for purchasing this product.
Before using this product, read the following precautions to make sure the product is
used correctly and safely.
Safety Precautions·························································· ii
Labels Attached to the Unit ··········································ii
Measurement Unit for Each Parameter ······················· iv
Graphic Symbols ························································· vi
Precautions for Safe Operation of Medical Electrical
Equipment ·································································· vii
Precautions for Safe Operation of Medical Telemetry
(When using HLX-501) ·············································· viii
Precautions about the Maintenance ···························· ix
Precautions about Pacemaker ·····································x
Non-Explosion Proof·····················································x
Defibrillation Safety·······················································x
Electrosurgery Safety ·················································· xi
Precautions about Magnetic Resonance Imaging ······· xi
Precautions about Connections to
Peripheral Devices ····················································· xii
Precautions about the Fuse········································ xii
Accessories and Optional Accessories······················· xii
Precautions about the DS-8500 System···················· xiii
Precautions about the Wired Network System
(DS-LAN ΙΙ/DS-LAN ΙΙΙ)·············································· xx
Precautions about the Wireless Network System······ xxi
Precautions for Use of Bidirectional Wireless
Communications (TCON) ········································· xxii
Precautions about the Ventilator Monitoring·············xxiii
Precautions for Use of SpO2Sensor ······················· xxiv
Precautions for Use of NIBP Cuff ···························· xxiv
Disposing of Equipment, Accessories,
or Components ······················································ xxiv
Precautions about Transportation····························· xxv
Precautions about RTC or Data Backup··················· xxv
To Prepare for Emergency Use ································ xxv
Electromagnetic Compatibility···································xxvi
Precautions for Safe Operation under Electromagnetic
Influence·································································· xxvi
EMC Guidance ························································ xxvi
Compliance to the Electromagnetic Emissions ··xxvii
Compliance to the
Electromagnetic Immunity (1)·························xxvii
Compliance to the
Electromagnetic Immunity (2)························xxviii
Recommended Separation Distances between
Portable and Mobile RF Communications
Equipment and the DS-8500 System ············· xxix
i

Safety Precautions
yRead the “Safety Precautions” thoroughly before use to ensure correct and safe use of the
product.
yMake sure to follow the precautions indicated below, as these are important messages related to
safety.
DANGER Failure to follow this message may cause immediate threat of death or serious
injury, or complete failure of the equipment.
WARNING Failure to follow this message may result in death or serious injury, or
complete failure of the equipment.
CAUTION Failure to follow this message may cause injury or failure to the equipment.
NOTE
A note is not related to product safety, but provides information about the
correct use and operating procedures to prevent incorrect operation and
malfunction of the equipment.
Labels Attached to the Unit
Make sure to read the warning labels attached to the unit and comply with these requirements
while operating the unit.
CAUTION
Do not damage or erase the warning labels attached to the unit.
These warning labels contain descriptions important for handling and
operating the unit properly and safely. A damaged label may compromise
safe operation.
ii

DS-8500 System Main Unit (DSC-8500 Series)
iii

Measurement Unit for Each Parameter
The measurement units for this equipment are as follows.
Detail Parameter Display Unit Default
ECG HR
bpm
(beats per minute)
Invasive Blood Pressure PR_IBP bpm
Heart Rate / Pulse Rate
SpO2PR_SpO2bpm
ST Level ECG ST mm, mv mv
VPC ECG VPC beat/minute
Impedance Respiration RR_IMP
Bpm
(breaths per
minute)
CO2RR_CO2Bpm
Ventilator RR_VENT Bpm
Respiration Rate
Gas Module RR_GAS Bpm
Impedance Respiration APNEA s (second)
CO2APNEA s (second)
Apnea
Ventilator APNEA s (second)
Invasive Blood Pressure Invasive Blood Pressure BP mmHg, kPa
cmH2O (CVP only) mmHg
Non-Invasive Blood
Pressure Non-Invasive Blood Pressure NIBP mmHg, kPa mmHg
Arterial Oxygen Saturation SpO2SpO2%
Temperature Temperature TEMP
°C
End-Tidal CO2
Concentration CO2EtCO2mmHg, kPa, % mmHg
Inspiratory CO2
Concentration CO2InspCO2mmHg, kPa, % mmHg
Cardiac Output Cardiac Output CO L/minute
Blood Temperature Blood Temperature Tb °C
Injectate Temperature Injectate Temperature Ti °C
Airway Flow Airway Flow AWF L/minute
Airway Pressure Airway Pressure AWP cmH2O
Expiratory Tidal Volume E-TV mL
Inspiratory Tidal Volume I-TV mL
Tidal Volume TV mL
Tidal Volume
Inspiratory/Expiratory Ratio I:E (none)
Minute Ventilation MV L/minute
Minute Ventilation Spontaneous Minute Ventilation SMV L/minute
Compliance COMP mL/cmH2O
Static Compliance S_COMP mL/cmH2O
Compliance
Dynamic Compliance D_COMP mL/cmH2O
Expiratory Resistance E-RES cmH2O/L/Sec
Inspiratory Resistance I-RES cmH2O/L/Sec
Static Airway Resistance S-RES cmH2O/L/Sec
Airway Resistance
Dynamic Airway Resistance D-RES cmH2O/L/Sec
Mean Airway Pressure MEAN cmH2O
Maximum Airway Pressure PEAK cmH2O
Pause Airway Pressure PAUSE cmH2O
Airway Pressure
Minimum Airway Pressure P_Min cmH2O
Spontaneous Respiration Spontaneous Respiration S_RR Bpm
Peak End Expiratory
Pressure Peak End Expiratory Pressure PEEP cmH2O
Fraction of Inspiratory
Oxygen Fraction of Inspiratory Oxygen FIO2%
iv

Detail Parameter Display Unit Default
Mixed Venous Oxygen Saturation SvO2%
Central Venous Oxygen
Saturation ScvO2%
Arterial Oxygen Saturation SaO2%
Oxygen Uptake Index O2EI %
Oxygen Transport DO2mL/minute
Oxygen Consumption VO2mL/minute
Stroke Volume SV mL
Stroke Volume
(STAT Mode) SV_STAT mL
Stroke Volume Index SVI mL/m2
Stroke Volume Index
(STAT Mode) SVI_STAT mL/m2
Heart Rate HR bpm
(beats per minute)
Mean Arterial Pressure MAP mmHg
Central Venous Pressure CVP mmHg
Continuous Cardiac Output CCO L/minute
Continuous Cardiac Output
(STAT Mode) CCO_STAT L/minute
Continuous Cardiac Index CCI L/minute/m2
Continuous Cardiac Index
(STAT Mode) CCI_STAT L/minute/m2
Systemic Vascular Resistance SVR dynes-sec/cm5
Systemic Vascular Resistance
Index SVRI dynes-sec/cm5
Blood Temperature BT °C
Ejection Fraction EF %
Ejection Fraction
(STAT Mode) EF_STAT %
End-Diastolic Volume EDV mL
End-Diastolic Volume
(STAT Mode) EDV_STAT mL
End-Diastolic Volume Index EDVI mL/m2
End-Diastolic Volume Index
(STAT Mode) EDVI_STAT mL/m2
End-Systolic Volume ESV mL
End-Systolic Volume Index ESVI mL
Vigilance Data
・Vigilance
・Vigilance CEDV
・VigilanceΙΙ
・Vigileo
Stroke Volume Variance SVV %
Et CO2 Concentration CO2-E mmHg, kPa, % mmHg
Insp CO2 Concentration CO2-I mmHg, kPa, % mmHg
End Tidal Oxygen O2-E %
Inspired Oxygen O2-I %
Expired Nitrous Oxide N2O-E %
Inspired Nitrous Oxide N2O-I %
End Tidal Anesthetic Gas AGT-E %
Multigas Unit
Expired Agent gas AGT-I %
Bispectral Index BIS (no unit)
Signal Quality Index SQI %
Electromyograph EMG dB
BIS Monitor Data
Suppression Ratio SR %
aepEX Monitor Data AEPIndex aepEX (no unit)
v

Graphic Symbols
Refer to the following for the meaning of the symbols indicated on the equipment.
DS-8500 System Main Unit
Symbol Description
Caution; refer to accompanying documents
Indicates the need to refer to related accompanying documents
before operation.
Potential Equalization Terminal
Indicates the terminal to equalize the potential difference when
interconnecting the devices.
Protective Earth
Indicates the protective earth inside the equipment.
Power ON
Indicates that the main power switch is in the ON position.
Power OFF
Indicates that the main power switch is in the OFF position.
Electrostatic Sensitive Part
Directly touching this connector part with hands should be avoided.
DS-8500 System Symbols displayed on the screen
Symbol Description
Alarm OFF
Indicates the alarm is OFF.
Heart Rate Synchronization Mark
This mark flashes synchronizing to the heartbeat.
Respiration Synchronization Mark
This mark flashes synchronizing to the inspiration.
Message Icon
This mark will be displayed inside the parameter key when an alarm
message is present for that parameter. Whether or not to display this
icon can be selected on the monitor setup menu.
TCON Mark
Displays the Bidirectional Wireless Communication (TCON)
connection status while in communication.
vi

Precautions for Safe Operation of Medical Electrical Equipment
CAUTION
Read the following precautions thoroughly to correctly operate the device.
zUsers should have a thorough knowledge of the operation before using this
system.
zPay attention to the following when installing and storing the equipment.
yDo not install or store in an area where the equipment will be subject to
splashing water.
yDo not install or store in an area where the environmental conditions,
such as atmospheric pressure, temperature, humidity, ventilation,
sunlight, dust, sodium, sulfur, will adversely affect the system.
yPlace the equipment on a stable surface where there is no inclination,
vibration, or shock (including during transportation).
yDo not install or store in an area where there are chemical or gasses
stored.
yVerify the power frequency, voltage and allowable current (or power
consumption).
yEnsure the grounding is proper by connecting the accompanying power
cable to the hospital grade outlet.
zBefore operating the system, verify the following items.
yVerify the power voltage.
yCheck the cable connection and polarity to ensure proper operation of
the equipment.
yMake sure the power system has adequate earth ground.
yEnsure that all cables are firmly and safely connected.
yPay special attention when the device is used in conjunction with other
equipment as it may cause erroneous judgment and danger.
yEnsure all patient connections are proper and secure.
zDuring operation of the system, verify the following items.
yAlways observe the system and patient to ensure safe operation of the
equipment.
yIf any abnormality is found on the equipment or patient, take appropriate
measures such as ceasing operation of the equipment in the safest way
for the patient.
yDo not allow the patient to come in contact with the device.
yMake sure not to touch all Signal Input/Output parts in the device with the
patient at the same time.
zAfter using the system, verify the following items.
yUnplug all the cables from the patient before turning off the power.
yWhen unplugging the cables, do not apply excessive force by pulling on
the cable. Pull from the connector part of the cable.
yClean the accessories and cables, and keep them together in one place.
yKeep the unit clean to ensure proper operation for the next use.
zIf the equipment is damaged and in need of repair, user should not attempt
service. Label the unit “OUT OF ORDER” and contact our service
representative.
zDo not remodel the equipment.
zMaintenance Check
yMake sure to periodically check the equipment, accessories and cables.
yBefore reusing the device that has been left unused for a while, make
sure that the device works normally and safely.
zTo prevent burn injury to the patient, verify proper attachment of patient
ground plate or ECG electrode type when using the electrosurgical knife,
and verify paste volume or output energy when using the defibrillator. Also,
verify that proper ground is selected.
vii

Precautions for Safe Operation of Medical Telemetry (When using
HLX-501)
CAUTION
To operate the device correctly, read the following precautions carefully.
zThe medical institution (hereinafter referred to as the “Institution”) must
decide the telemetry installation plan to prevent interference and
interference between transmitters (telemetry based on destination country’s
radio law). When telemetry has already been installed and been used, radio
format, frequency, and antenna power are required to be examined to
prevent interference.
zWhen using telemetry which requires zone location, the Institution is to set
up the zones as an operation unit for each transmitter to prevent electronic
interference between telemetry throughout the medical institution.
zWhen using telemetry which requires zone location, display and identify
each prepared zone in the equipment.
zWhen laying receiver antenna for each transmitter, the Institution has to
examine the installation so that electronic interference does not occur.
zBased on the above examination result, the Institution should place each
receiver antenna as required.
In managing, make sure to follow the precautions below.
zThe institution appoints a person to manage the wireless channels for the
whole medical institution. And when using telemetry which requires zone
location, the Institution should nominate a person to manage the wireless
channels in each zone (a “Zone Manager”). However, when using such
telemetry in a local medical institution, one person can perform both
functions.
zSelect a telemetry manager who understands the characteristics and
functionality of telemetry systems, and is skilled in operating telemetry.
zWhen installing telemetry, the Overall Manager and the Zone Manager
have to understand the precautions for use of the telemetry in advance.
zThe Overall Manager takes responsibility of wireless channel management
and transmitter storage for the whole medical institution by giving proper
instruction.
zThe Overall Manager creates a management log, list of wireless channels,
management status for the whole medical institution (hereinafter referred to
as the “management log”. When changing a wireless channel, register it
in the log and give proper instructions to the zone manager or to the user.
zThe Zone Manager assumes responsibility for managing the wireless
channels, storing, and managing telemetry.
zThe Zone Manager assigns the transmitter to the user, and provides
enough education for use inside the zone.
zThe telemetry user verifies operation of the transmitter/receiver before use.
zThe telemetry user, if using the telemetry in a zone location, follows the
instructions of the Zone Manager for the zone and gives instructions to the
patient if required.
zWhen interference or breakdown occurs in telemetry communication, the
user is required to inform the zone manager and the overall manager of the
problems. The zone manager and overall manager are to deal with the
problem properly and/or contact their nearest Fukuda Denshi
representative for service.
viii

Precautions about the Maintenance
Safety Inspection and Maintenance
For safe operation of the equipment, regular inspection and maintenance is required. Once a year,
check all cables, devices, and accessories for damage, earth impedance, earth and leakage
currents, and all alarm functions. Also, ensure that all safety labels are legible. Maintain a record of
these safety inspections.
Immediate maintenance has to be carried out if ;
ythe equipment was subjected to extreme mechanical stress, e.g. after a heavy fall.
ythe equipment was subjected to liquid spill.
ythe monitoring function is interrupted or disturbed.
yparts of the equipment enclosure are cracked, removed, or lost.
yany connector or cable shows signs of deterioration.
Reference Refer to “12. Periodic Check” for details.
WARNING Never open the housing while the equipment is in operation or connected to
hospital grade outlet as it may result in electric shock.
Maintenance, Modifications, and Repairs
Fukuda Denshi is liable for the safety, reliability, and performance of its equipment only if;
yMaintenance, modifications, and repairs are carried out by authorized personnel.
yComponents are used in accordance with Fukuda Denshi operating instructions.
ix

Precautions about Pacemaker
WARNING
zMinute ventilation rate-adaptive implantable pacemakers can
occasionally interact with certain cardiac monitoring and diagnostic
equipment, causing the pacemakers to pace at their maximum
programmed rate. The cardiac monitoring and diagnostic equipment
may possibly send wrong information.
If such event occurs, please disconnect the cardiac monitoring and
diagnostic equipment, or follow the procedures described in the
operation manual of the pacemaker.
For more details, contact FUKUDA DENSHI personnel, your institution’s
professionals, or your pacemaker distributors.
Reference “Minute Ventilation Rate-Adaptive Pacemakers”
FDA alerts health professionals that minute ventilation
rate-adaptive implantable pacemakers can occasionally interact
with certain cardiac monitoring and diagnostic equipment, causing
pacemakers to pace at their maximum programmed rate.
[October 14, 1998 (Letter: www.fda.gov/cdrh/safety.html) – FDA]
Non-Explosion Proof
DANGER
Never operate the equipment in the presence of flammable anesthetics,
high concentration of oxygen, or inside hyperbaric chamber. Also, do not
operate the equipment in an environment in which there is a risk of
explosion.
Explosion or fire may result.
Defibrillation Safety
WARNING
zWhen defibrillating, keep away from the electrodes or medicament
applied to the patient chest. If this is not possible, remove the electrodes
or medicament before defibrillating.
If the defibrillator paddles are directly in contact with the electrodes or
medicament, an electrical shock may result from the discharged energy.
zWhen defibrillating, make sure that the electrodes, sensor cables, or
relay cables are firmly connected to the device.
Contacting the metal part of the disconnected cable may result in
electrical shock from the discharged energy.
zWhen defibrillating, do not touch the patient and the metal part of the
device or cables. Electric shock may result from the discharged energy.
x

Electrosurgery Safety
WARNING
The monitoring system contains protection against interference generated
by electrosurgical instruments. However, operating conditions, surgery site
with respect to the location of ECG electrodes, or the type of instrument
used, may cause noise on the ECG. The noise is generated at the tip of the
electrical knife and is difficult to completely eliminate because of the
frequency components of the ECG. To reduce electrosurgical interference,
take the following precautions:
Location
Locate the electrosurgical unit as far as possible from this unit and the
patient cable. This will help reduce interference on the ECG through the
monitor or cables.
Power Supply
Connect the electrosurgical unit to a power supply that is different from
that of the monitor. This will help prevent interference through the power
cable.
Electrode Placement
The amount of interference is considerably different depending on the
electrode position and surgery site. Place the ECG electrodes as far
away as possible from the surgery site. Do not place electrodes in the
path between the surgery site and the ground plate. If the electrodes
are placed in this path, the amount of interference will be quite large.
Position (+) and (–) electrodes as close as possible to each other.
Ground Plate
When using electrosurgical instruments, make sure the contact between
the patient and the ground plate is secure. If the connection is
incomplete, the patient may suffer a burn at the electrode site.
Precautions about Magnetic Resonance Imaging
WARNING
zDo not operate this equipment in magnetic resonance imaging (MRI)
environments.
zWhen conducting MRI test, remove the electrodes and sensors
connected to the patient (test subject).
The local heating caused by the induced electromotive force may cause
burn injury to the patient (subject). For details, refer to the operation
manual for the MRI testing device.
xi

Precautions about Connections to Peripheral Devices
For safety and good performance of this equipment, connection of other manufacturers’ equipment
to the monitor is not authorized, unless the connection is explicitly approved by Fukuda Denshi. It
is the user’s responsibility to contact Fukuda Denshi to determine the compatibility and warranty
status of any connection made to another manufacturer’s equipment.
WARNING
For the connector with mark, only the peripheral devices specified by
Fukuda Denshi should be connected with the given procedure. Use of an
unspecified device may cause electric shock to the patient and/or operator
due to excessive leakage current.
CAUTION
All the peripheral device connectors on the DS-8500 system are isolated
from the power supply, but the peripheral devices are not isolated. To
prevent danger of electric shock, always position the peripheral devices
away from the patient.
When connecting peripheral devices to DS-8500 system, it is the user’s responsibility to verify that
the overall system complies with IEC 60601-1-1, “Collateral Standard: Safety Requirements for
Medical Electrical Systems”.
Precautions about the Fuse
DANGER If the fuse blows, do not continue using it as internal damage to the
equipment may be considered.
Accessories and Optional Accessories
WARNING
Use only the cables specified by Fukuda Denshi.
Use of other cables may result in increase in emission (unnecessary
generation of electromagnetic energy from this unit) or decrease in
immunity (resistance characteristics to electromagnetic interference
generating from other device).
xii

Precautions about the DS-8500 System
DANGER Be cautious when connecting to other device.
Danger such as electric shock may result to the patient and operator.
WARNING
zDo not connect a unit or cable not authorized by Fukuda Denshi to any I/O
connector. If done so by mistake, the DS-8500 system cannot deliver its
maximum performance and the connected units may be damaged, resulting
in a safety hazard.
zIf the DS-8500 system is used under an environment not fulfilling the
specified condition, not only that the equipment cannot deliver its maximum
performance, the equipment may be damaged and safety cannot be
ensured.
zUse only the accompanying 3-way AC power cable. Use of other cables
may result in electric shock to the patient and the operator.
zThe power cable must be connected to hospital grade outlet.
zWhen using multiple ME equipment simultaneously, perform equipotential
grounding to prevent potential difference between the equipment. Even a
small potential difference may result in electric shock to the patient and the
operator.
zThe patient classification selection influences the precision of the QRS
detection and NIBP measurement. Make sure the correct selection is
made.
zThe pacemaker use selection influences the precision of the QRS detection
and arrhythmia analysis. Make sure the correct selection is made.
zIf the QRS pace pulse mask function is set to OFF , 10ms or 20ms ,
a decrease in heart rate may not generate HR or ASYSTOLE alarms due to
erroneously detected QRS. Set this function to OFF , 10ms or
20ms only if you are sure that pacing failure will not occur, or when the
patient can be constantly monitored.
zWhen measuring the SpO2of patient with high fever or peripheral
circulatory insufficiency, check the sensor attachment periodically and
change the attachment site. The temperature of attachment site will rise 2
to 3°C due to the sensor heat which may result in burn injury.
zFor the following case, accurate measurement may not be possible.
yPatient with excessive abnormal hemoglobin (HbCO, MetHb)
yPatient with the pigment injected to the blood
yPatient receiving CPR treatment
yWhen a sensor is applied to a limb with NIBP cuff, arterial catheter, or
intracatheter
yWhen measuring at site with venous pulse
yPatient with body motion
yPatient with small pulse
zBefore the measurement, make sure the patient classification ( Adult /
Child / Neonate ) is properly selected. Otherwise, correct
measurement cannot be performed, and congestion or other injury may
result.
zUse the specified sampling tube and nasal prong manufactured by Oridion.
zFor CO2concentration measurement, always consider the circumference of
the intubation tube when using the airway adapter. If inappropriate airway
adapter is used for a patient with low ventilation, CO2may mix in to the
inspired air resulting in incorrect measurement, or apnea detection may
become difficult.
zWhen measuring CO2concentration of a patient treated with
mouth-to-mouth resuscitation, Jackson-Rees circuit, Mapleson D circuit of
which CO2gas may mix in, the value may be displayed lower than the
actual value.
xiii

WARNING
zWhen the system alarm is suspended, all the alarm will be suspended even
if the parameter alarm is set to ON. Also, the alarm event will not be stored
as recall.
zIf the upper/lower alarm limit of the parameter is set to OFF, or arrhythmia
alarm is set to OFF, alarm will not function even if the system alarm is set
to ON. Pay attention when setting them OFF.
zObjective and constant arrhythmia detection is possible through the fixed
algorithm incorporated in this monitor. However, excessive waveform
morphology change, motion artifact, or the inability to determine the
waveform pattern may cause an error, or fail to make adequate detection.
Therefore, physicians should make final decisions using manual recording,
alarm recording and recall waveform for evaluation.
zThe HR/PR alarm will not be generated unless the parameter key
corresponded to the selected HR/PR source is displayed. Make sure to
display the parameter key for the HR/PR source.
zThe RR/APNEA alarm will not be generated unless the parameter key
corresponded to the selected RR/APNEA source is displayed. Make sure to
display the parameter key for the RR/APNEA source.
zWhen selecting Silence or Time Disp. Only for the night mode, pay
attention not to miss any important alarm by simultaneously monitoring the
bed on other monitors such as central monitor.
zWhen lifting this device, hold the bottom part of the main unit and the
display unit.
zWhen attaching the display unit to the DSC main unit, place the display unit
facing down and slowly attach the main unit using the guide on the side of
the display unit. Then, secure it with the specified screws.
zAbout the Trolley
yWhen integrating and mounting a display unit + a main unit + an
exclusive stand to a trolley, make sure that they are fixed using the
screws (x4). Otherwise, the equipment may fall from the trolley, resulting
in injury to the operator or damage to the equipment.
yUse the trolley only with the equipment specified by Fukuda Denshi.
Otherwise, the monitor and trolley may fall down, resulting in injury or
damage to the monitor.
yMake sure to lock both casters when using or storing the trolley.
The trolley may move or fall down, resulting in injury or damage to the
monitor.
yDo not use or store the trolley where it will be subject to inclination of 10
degrees or more. The trolley or equipment may fall resulting in injury or
damage to the equipment.
xiv

CAUTION
zSystems
yUse only the accessories specified for this device. Otherwise, proper
function cannot be executed.
yDo not use the touch panel with the film attached. Malfunction of the
touch panel or damage may result.
yDue to its material characteristic, the touch panel expands/contracts
depending on the temperature/humidity. When the touch panel is left
unused for a while, or when the ambient temperature is low, the surface
film of the touch panel may expand, but this is not an abnormal condition.
This expansion will be reduced in few hours or half a day after the power
is turned ON.
yFor quality improvement, specifications are subject to change without
prior notice.
yThe display panel utilizes exclusive fluorescent light for the backlight.
Since this fluorescent light deteriorates by the life cycle, the display may
become dark, scintillate, or may not light by the long term use. In such
case, contact your nearest service representative and replace the LCD
panel.
yThis equipment is intended to be used for only one patient.
yThe installation of this equipment and attaching the option unit should be
performed by a person who is well acquainted with this equipment.
yIf it is not used for a long period, make sure to turn OFF the power of the
main unit.
zECG Monitoring
yThe indication for continuous use of the electrode is about one day.
yReplace the electrode if the skin contact gets loose due to perspiring,
etc.
yWhen an electrode is attached at the same location for a long time, some
patients may develop a skin irritation. Check the patient’s skin condition
periodically and change the electrode site as required.
yFor stable arrhythmia detection and ECG monitoring, verify proper
electrode placement, lead, waveform size, and filter mode selection. If
not properly selected, it may cause erroneous detection.
yThe arrhythmia detection level is related with the displayed waveform
size. Set a proper waveform size for monitoring.
When the waveform size is ×1/4, ×1/2, or ×1, the detection threshold is
250µV. When the waveform size is ×2 or ×4, the detection threshold is
150µV.
yThe QRS detection leads, arrhythmia detection leads, monitoring leads
on the central monitor, recording leads are fixed as ECG1 and ECG2.
Especially for arrhythmia detection, set the most appropriate leads with
high QRS for ECG1 and ECG2.
yAutomatic size/position of the ECG is effective only at the time the
AUTO key is pressed. This does not continually adjust size and
position.
yThe ESIS mode can largely reduce the artifact such as electrosurgery
noise and EMG, but it may also reduce the QRS amplitude. Using the
ESIS mode may erroneously detect the pacemaker spike. The ESIS
mode should be selected only during electrosurgery.
yThere are some cases when pacemaker pulse can not be detected
depending on the pacemaker type, pulse voltage, pulse width, electrode
lead type (unipolar, bipolar), or electrode placement which causes the
pacemaker pulse amplitude to decrease and disables pacemaker pulse
detection.
yIf signals similar to a pacemaker pulse are present, such as electric
blanket noise or excessive AC frequency noise, these may be
erroneously detected and displayed as a pacemaker pulse.
yWhen spontaneous QRS and pacemaker pulse overlap (ex. fusion beat,
etc.), QRS detection cannot be performed properly. In this case, the
heart rate is degraded.
xv

CAUTION
yWhen continuously detecting AC noise artifact as pacemaker pulses,
QRS detection stops and heart rate is extremely degraded. Also
arrhythmia cannot be detected.
zRespiration Monitoring
yWhen a defibrillator is used during respiration monitoring, a large offset
voltage will be placed on the ECG electrodes, which may cause
interruption of monitoring for a few seconds.
zSpO2Monitoring
yIf the nail is rough, dirty, or manicured, accurate measurement will not be
possible. Select a different finger or clean the nail before attaching the
sensor.
yIf irritation such as skin reddening or rash appears with the sensor use,
change the attachment site or stop using the sensor.
yWhen fixing the sensor with tape, do not wind the tape too tight. At the
same time, check the blood flow constantly so that congestion is not
generated at the peripheral.
yEven a short duration of attachment may inhibit the blood flow and
generate compression necrosis and burn injury.
yChange the sensor attachment site constantly (every 4 hours). As the
temperature of the sensor attachment site normally rises 2 to 3°C
yAs skin for neonate / low birth weight infant is immature, change the
sensor attachment site more frequently depending on the condition.
yDirect sunlight to the sensor area can cause a measurement error.
Place a black or dark cloth over the sensor.
yWhen not performing measurement, unplug the relay cable and sensor
from the SpO2connector. Otherwise, the measurement data may be
erroneously displayed by the ambient light.
yThe DS-100A is intended for use on finger of adults weighing over 40 kg
(approximate). Do not use them on children or neonates. Also do not
apply them on the thumb or foot.
yThe light-emitting part of the sensor should be over the root of the
fingernail. Do not insert the finger too far into the sensor as it may hurt the
patient.
yMeasuring on a limb with NIBP cuff, arterial catheter, or intracatheter may
result in incorrect measurement.
zNIBP Monitoring
yPay attention when measuring the NIBP of patient with bleeding disorders
or hyper coagulation. The cuff inflation may cause petechia or circulatory
failure by the blood clot.
yFor the following situation, measurements will be terminated.
When the measurement time has exceeded 160 seconds for adult and
child, 60 seconds for neonate.
When the inflation value has exceeded 300mmHg for adult, 210mmHg for
child, 150mmHg for neonate.
yIf used with the incorrect patient classification, it will not only cause
erroneous measurement, but the inflating level for the adult may be
applied to child or neonate causing dangerous situation to the patient.
yThe 1-minute interval measurement will always start from 00 second.
Pressing the 1min start key will start the measurement from the next 00
second.
yThe alarm function will be ineffective for the BP value measured by Quick
SYS regardless of the ON/OFF selection of NIBP alarm.
yIf the mean display is set to OFF, the mean BP will not be generated.
Also the mean BP will not be displayed inside the tabular trend or the
NIBP list function if the display is set to OFF.
zBP Monitoring
yWhen the main power is turned ON, the BP value will not be displayed
until zero balance is performed. However, if the power is turned ON within
5 minutes after the power is turned OFF, the previous zero balance
information will be maintained, and BP value will be displayed.
xvi
This manual suits for next models
3
Table of contents
Other Fukuda Denshi Monitor manuals