Guangzhou ZHIYUAN Electronics PA5000 Series User manual

PA5000 Series Power Analyzer
Start Guide
Date: 2016/12/08
V1.00
Guangzhou ZHIYUAN Electronics Co., Ltd.
www.zlg.cn
User
Manual
UM01010101

ii
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
C
Co
on
nt
te
en
nt
ts
s
Chapter 1: Product Introduction.................................................................8
1.1 Introduction........................................................................................................ 8
1.2 Features.............................................................................................................. 8
1.3 Application System.......................................................................................... 10
Chapter 2: Function Description...............................................................11
2.1 Main Functions................................................................................................. 11
2.2 Measurement Functions.................................................................................. 11
2.3 Operating Mode................................................................................................ 12
2.4 Voltage/Current Mode...................................................................................... 13
2.5 Communication Interface................................................................................ 13
2.6 Display Screen ................................................................................................. 14
2.7 Measurement and Analysis Function............................................................. 18
2.7.1 Waveform display ..................................................................................................18
2.7.2 Trend Analysis........................................................................................................19
2.7.3 Harmonic analysis function....................................................................................20
2.7.4 FFT Function..........................................................................................................21
2.7.5 IEC Harmonics Measurement................................................................................22
2.7.6 Vector Display........................................................................................................23
2.7.7 Flicker Analysis......................................................................................................24
2.7.8 Cycle-by-cycle Analysis.........................................................................................25
2.7.9 Normal Analysis Mode ..........................................................................................26
2.7.10 Waveform Operation............................................................................................26
2.7.11 Motor Measurement.............................................................................................27
2.8 Supplemental Information............................................................................... 28
2.8.1 Input Element.........................................................................................................28
2.8.2 Wiring Unit ............................................................................................................30
Chapter 3: Panel Introduction...................................................................31
3.1 Front Panel....................................................................................................... 31
3.1.1 Panel components...................................................................................................31
3.1.2 Display Screen .......................................................................................................32
3.1.3 Function Key area..................................................................................................35
3.2 Rear Panel ........................................................................................................ 40
3.3 Side Panel......................................................................................................... 42
Chapter 4: Circuit Connection ..................................................................44
4.1 Connect with Measurement Circuit................................................................ 44
4.1.1 Wiring System........................................................................................................44
4.1.2 Diagram of Connecting Test Accessories...............................................................48

iii
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
4.2 Use PT/CT to Measure Circuit......................................................................... 49
4.2.1 PT and CT..............................................................................................................49
4.2.2 Wiring Method.......................................................................................................49
4.2.3 Diagram of Connecting Test Accessories...............................................................52
4.3 Connect Measuring Circuit by Current Sensor............................................. 52
4.3.1 Shunt Current Sensor and Voltage Output Clamp Sensor......................................53
4.3.2 Wiring Method.......................................................................................................53
4.3.3 Diagram of Connecting Test Accessories...............................................................54
Chapter 5: Before Starting Measurement ................................................55
5.1 Overview........................................................................................................... 55
5.2 System Settings............................................................................................... 55
5.3 Zero Calibration ............................................................................................... 56
5.3.1 Function Introduction.............................................................................................56
5.3.2 Steps.......................................................................................................................57
5.4 Configuration Wizard....................................................................................... 57
Chapter 6: Measurement Condition..........................................................60
6.1 Select wiring mode .......................................................................................... 60
6.2 Set measurement range .................................................................................. 60
6.2.1 Set the measurement range for direct input............................................................60
6.2.2 How to set the range when using the external current sensor ................................63
6.2.3 Ratio Factor............................................................................................................63
6.3 Measurement Interval Setting......................................................................... 65
6.3.1 Set synchronous source..........................................................................................65
6.3.2 Set data update rate ................................................................................................66
6.4 Filter Configuration.......................................................................................... 67
6.5 Power Measurement Parameter Initialization................................................ 69
6.5.1 Set calculation formulas for apparent power, reactive power and corrected power
.........................................................................................................................................70
6.5.2 Set the formula for efficiency.................................................................................73
6.6 Averaging Function.......................................................................................... 75
6.6.1 Function Introduction.............................................................................................75
6.6.2 Steps.......................................................................................................................76
6.7 Synchronous Measurement............................................................................ 76
6.8 U-I Phase........................................................................................................... 77
Chapter 7: Exception Handling and Maintenance...................................78
7.1 Exception Handling ......................................................................................... 78
7.2 Recommended Accessory Replace Period ................................................... 79
Chapter 8: Specifications..........................................................................80
8.1 Input Parameters.............................................................................................. 80
8.1.1 Input terminal Type................................................................................................80
8.1.2 Input Type ..............................................................................................................80
8.1.3 Number of Input Elements.....................................................................................80

iv
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
8.1.4 Voltage Measurement Range..................................................................................80
8.1.5 Current Measurement Range..................................................................................81
8.1.6 Input Bandwidth.....................................................................................................82
8.1.7 Common Mode Voltage .........................................................................................82
8.1.8 Filter.......................................................................................................................82
8.1.9 Range Switching ....................................................................................................82
8.1.10 A/D Converter......................................................................................................82
8.2 Monitor.............................................................................................................. 83
8.3 Accuracy........................................................................................................... 83
8.3.1 Basic Accuracy.......................................................................................................83
8.3.2 Input Range............................................................................................................87
8.3.3 Input Display Value................................................................................................87
8.3.4 Influence of Line Filter ..........................................................................................87
8.3.5 Angle Error.............................................................................................................88
8.3.6 Temperature Coefficient.........................................................................................88
8.3.7 Accuracy of 12 months...........................................................................................88
8.4 Measurement Modes ....................................................................................... 88
8.5 Measurement Items ......................................................................................... 89
8.6 Measurement Functions/Measurement Conditions...................................... 90
8.7 Motor Function................................................................................................. 91
8.7.1 Analog Input Parameters........................................................................................91
8.7.2 Pulse Frequency Input Parameters.........................................................................92
8.8 Waveform Sampling Data Storage Function ................................................. 92
8.9 Storage.............................................................................................................. 92
8.10 General Specifications .................................................................................. 93
8.11 External Dimensions...................................................................................... 94

1
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
Safety Precautions
The instrument may be used in the high voltage environment. In order to prevent
personnel injury or death caused by electrical shock or other hazards, please
make sure to thoroughly read and understand the “Safety Precautions” chapter.
The following precautions described herein must be observed during the usage. If the
instrument is used in a manner not specified in this manual, the protection provided by
the instrument may be impaired. Guangzhou ZHIYUAN Electronics Co., Ltd assumes
no liability for the customer’s failure to comply with these requirements.
General Precautions
In order to avoid injury to operators and damage to the instrument or the devices
connected with the instrument, make sure to comply with the precautions below.
Use correct power cord
Only the dedicated power cord allowed by the country where the instrument is used can
be used.
Check all ratings before use the instrument
In order to avoid fire and high current impulse, please check all ratings and instructions
of the instrument before use the instrument. For the detail information about all ratings,
please refer to the related user manual.
Use appropriate over voltage protection
In order to avoid the electric shock to operators, make sure an over voltage (such as the
voltage caused by thunderstorm) will not occur.
Do not operate the instrument when the chassis is open
Do not operate the instrument when the chassis is open.
Replace power fuse
If the power fuse is necessary to be changed, please contact ZHIYUAN Electronics.
When any fault occurs during the use of instrument
When there may be a fault during the use of instrument, please stop operating the
instrument immediately and contact ZHIYUAN Electronics in time. ZHIYUAN
Electronics will send engineers to inspect, maintain and adjust it or replace parts.
Keep instrument ventilated
Inadequate ventilation will cause instrument temperature increase, which may result in
damage to the instrument. Please inspect vent and fan constantly to keep the
instrument ventilated.
Do not use the instrument in a wet environment
In order to avoid the risk of short circuit and electric shock, do not operate the
instrument in a wet environment.
Do not use the instrument in an explosive environment

2
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
In order to avoid instrument damage or personal injury, do not operate the instrument in
an explosive environment.
Keep instrument surface clean and dry
To prevent dust or moisture in the air affecting instrument performance, please clean
and dry the surface of instrument.
Electrostatic protection
Static electricity will cause damage to the instrument. The instrument should be tested
in the anti-static area. The internal and external conductors of cables should be
connected to ground to discharge the static electricity before they are connected to the
instrument.
Transportation safety
To avoid damage to the parts (such as keys, knobs or interfaces on the panel) due to
falling of instrument during transportation, please note transportation safety.
Warning signs
CAUTION symbol indicates a hazard. Improper handling or execution in a
specified process, operation and so on can lead to loss of important data or
damage to the instrument. Please completely read and fully understand the
related notes before continue the operation.
WARNING symbol indicates a serious hazard. Improper handling or execution
in a specified process, operation and so on can lead to injury to the user or
damage to the instrument. Please completely read and fully understand the
related notes before continue the operation.
Safety Information
The safety symbols of Power analyzer are shown below.
Be careful,
Dangerous!
CE Authenticate
Grounding
Be careful,
Electric shock!
Don’t throw the
used batteries
into the garbage.
Recyclable

3
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
Measurement Category
PA Power Analyzer is at the category of CAT II (1000V). Its input can be connected to
the power supply within the CAT II condition (Max. 1000VAC). Do not use in the
measurement category above CAT II (1000V).
Measurement CategoryⅥ
It is applicable to measure for the terminal of the
low-voltage facilities
For example, electricity meters, measurement of
primary over current protection devices and ripple
control units, overhead line and cable system.
Measurement CategoryⅢ
It is applicable to measure for building facilities
For example, measured for the output socket,
industrial device and other devices of the distribution
board, circuit breaker and wirings include cables, bus
bars, junction boxes, switches and fixed facilities,
such as those stationary motors permanently
attached to fixed facilities.
Measurement CategoryⅡ
It is applicable to measure for circuits directly connected
to the low-voltage device
For example, measured for household appliances,
portable tools and other similar devices.
Measurement Category 0
It is applicable to measure for the circuit not directly
connected with main supply
For example, measured for the circuit which is not
supplied by main supply or measured for the circuit
which is supplied by main supply especially with
internal protection. In the latter case, the transient
stress is not the same. If the instrument is used for
this measurement category, pleasure ensure that the
transient voltage is less than or equal to 3000V peak
value.

4
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
General Precautions
In order to avoid injury, death of operators or damage to the instrument, make sure to
comply with the precautions below.
Do not operate the instrument when the protective grounding or fuse might be
defective. Make sure to check the protective grounding or fuse before using the
instrument. Do not operate the instrument if the protective earth or fuse might be
defective.
Do not remove the case from the instrument. Some parts of the instrument use
high voltages, which are extremely dangerous. For internal inspection and
adjustment, please consult ZHIYUAN Electronics.
Unplug if abnormal behavior occurs. If you notice smoke or unusual odors
coming from the instrument, immediately turn off the power and unplug the
power cord. In addition, turn off the power to the circuit under measurement that
is connected to the input terminals.
Do not operate in an explosive atmosphere. Do not operate the instrument in the
presence of flammable liquids or vapors.
Do not damage the power cord. Nothing should be placed on top of the power
cord. The power cord should also be kept away from any heat sources. When
unplugging the power cord from the outlet, never pull by the cord itself. Always
hold and pull by the plug. If the power cord is damaged, check the part number
and purchase a replacement from the distributor.
Keep electrically charged objects away from the instrument. Keep electrically
charged objects away from the input terminals. They may damage the internal
circuits.
Unplug during extended non-use. If the instrument will not be used in a long time,
turn off the power to the circuit under measurement and the instrument, and
remove the power cord from the outlet.
Do not place objects on top of the instrument. Never stack the instrument or
place other instruments or any objects containing water on top of the instrument.
This may cause the damage to the instrument.
Do not damage the LCD. Since the LCD screen is very vulnerable and can be
easily scratched, do not allow any sharp objects near it. In addition, do not apply
vibration or shock to it.
Carry the instrument properly. First, turn off the circuit under measurement and
remove the measurement cables. Then, turn off the instrument and remove
power cord and other cables. When moving the instrument, hold the handle on
each side of the instrument.
Take a measurement. Before measuring, the power supply of the measured
circuit must be shut down before wiring; when measuring, the connecting
terminal of the measured circuit and instrument must be far away; after
measuring, the power supply of the measured circuit must be cut off, and then
disconnect the wiring and turn off the instrument.
When wiping off dirt. When cleaning the case or the operation panel, first turn off

5
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
the circuit under measurement and the instrument and remove the power cord
from the outlet; then use a clean, soft and dry cloth to wipe the instrument gently.
Do not use volatile chemicals, because they might cause discoloring and
deformation.
Clean the fan dust cover regularly. Check and clean the fan dust cover every
30 days to avoid heat or danger, blockage and overheating will affect the
accuracy of the instrument.
Connect Power Supply and Ground
Use correct power supply. Before connecting the power cord, ensure that the
power supply voltage matches the related voltage of the instrument and it does
not exceed the maximum related voltage of the provided power cord.
Use correct power cord and plug. To prevent the possibility of electric shock or
file, make sure to use the power cord supplied by ZHIYUAN Electronics. The
main power plug must be plugged into an outlet with a protective earth terminal.
Do not use an extension cord without protective grounding.
Connect the protective ground terminal. Be sure to connect the protective earth
to prevent electric shock before turning on the power. The power cord that comes
with the instrument is a three-pin power cord. So the power cord should be
connected to a properly grounded three-prong outlet.
Do not impair the protective grounding. Never cut off the internal or external
protective earth wire or disconnect the wiring of the protective earth terminal;
otherwise the electric shock might be caused.
Avoid electric shock. Power cord should be inserted into a socket with ground
protection on the wall or within visible area. The socket cannot be used in the
over current situation.
Instrument Installation Precautions
Far away from the harsh environment. Far away from direct sunlight, heat
sources, excessive amount of soot, steam, corrosive or flammable gas, strong
magnetic field sources, high voltage equipment or power lines, water, oil or
chemicals.
Flat, even surface. The instrument should be installed on a stable horizontal
surface; otherwise the measurement precision may be impaired and fall off.
Well-ventilated location. Inlet/Vent holes are located on the top and bottom of the
instrument. To prevent internal overheating, allow at least 20mm of space around
the inlet and vent holes. When connecting measurement wires and other various
cables, extra space should be reserved for operations.
Ambient temperature and humidity. Ambient temperature: 5〜40°C; Ambient
humidity: 20〜80%RH.

6
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
Wiring Precautions
To prevent the possibility of electric shock and damage to the instrument, please pay
attention to the precautions below.
Employ protective earth before connecting measurement cables. The power cord
that comes with the instrument is a three-pin power cord. Connect it to a properly
grounded three-pong outlet. Please cut off the power to the circuit and make sure
there is no voltage in it before touch a circuit.
Cut off the power to the circuit under measurement when wiring or disconnecting
the circuit. Connecting or removing measurement cables while the power is on is
dangerous.
Take special caution not to wire a current circuit to the voltage input terminal or a
voltage circuit to the current input terminal.
When strip the insulation cover of the measurement cables, make sure that the
conductive parts (bare wires) wired to the input terminal do not expose from the
terminal. In addition, tighten the screw of input terminal so that the cable does
not come loose.
Do not touch the input interface. When the measuring circuit is being applied to
the current input terminal, do not touch the current sensor input terminal.
Similarly, when the measuring circuit is being applied to the current sensor input
terminal, do not touch the current direct input terminal. Since these terminals are
electrically connected inside the instrument, the operation is very dangerous.
Withstand voltage of Potential Transformer and Current Transformer. When
using the external potential transformer (PT) or current transformer (CT), make
sure it has enough withstand voltage with respect to the voltage (U) being
measured. In addition, the secondary side of CT should not become an open
circuit while the power is on. If this happens, high voltage will appear at the
secondary side of CT, which is extremely dangerous.
For safety reasons, when mounting the instrument on a rack, a switch should be
furnished from the front side of the rack, which is used to cut off the circuit under
measurement.
Loop voltage and current. Ensure that the measurement circuit voltage and
current and common-mode voltage within the rated range of the instrument;
Connect external sensors. Do not use the bare metal BNC connector when
connecting external sensors. Remove all connections of the current direct input
terminals;
Current direct input. The connection of the sensor input terminal must be
disconnected;
Connect external shunts. The power supply of the measuring circuit must be cut
off. It is dangerous if not shut down the power or remove the shunt.

7
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
If a measurement accessory is used in a manner not specified by
ZHIYUAN Electronics, the protection provided by the accessory will be
impaired. In order to avoid the injury of operators or damage to the
instrument, do not use a defective or damaged measurement accessory.

8
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
C
Ch
ha
ap
pt
te
er
r
1
1:
:
P
Pr
ro
od
du
uc
ct
t
I
In
nt
tr
ro
od
du
uc
ct
ti
io
on
n
1.1 Introduction
PA5000 power analyzer adopts advanced measurement techniques to ensure reliable
measurement accuracy. It is capable of measuring voltage parameters, current
parameters and power parameters of the multiple-phase voltage and current accurately.
It also supports sampling waveform display, spectrum analysis, harmonics & flicker
analysis and other functions. The appearance of PA5000 power analyzer is shown in
Figure 1-1.
Figure 1-1: The appearance of PA5000 power analyzer
1.2 Features
The features of PA5000 power analyzer are described as follows:
Phase synchronous acquisition and accurate measurement of AD/DC voltage,
current and power parameters.
Support for up to seven-phase power input. All input elements with up to 5kV
electrical isolation.
PA5000 power analyzer has 16-bit resolution ratio, 0.1% basic accuracy, DC
measurement bandwidth with 0.1Hz~5MHz, and 2MS/s sampling rate.
PA5000 power analyzer ensures the measurement of faster signals with higher
accuracy. It is applicable to the measurement with higher requirements on the
sampling rate and bandwidth such as inverter, frequency conversion appliances,

9
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
motors, switching power supplies and so on.
Adopts patented PLL frequency-doubling technology to achieve harmonics
measurement with higher speed and wide dynamic range. In the harmonics
mode, the test of voltage, current fundamental, power, phase, harmonics
component and total harmonic distortion factor (THD) can be implemented. Up to
5 kHz inverter fundamental signal can be measured to satisfy the measurement
requirements of high band width invert. In addition, the input and output elements
can take multi-harmonic measurement simultaneously, with the fundamental
frequency harmonics up to 500th order.
The motor input element can be selected to connect the external torque sensor
and encoder to measure the motor torque, speed and mechanical power, which
is suitable for the motor and drive applications.
Equipped with standard USB, Ethernet, GPIB and RS-232 interfaces. Users can
control the power analyzer remotely through these interfaces. In addition, it
supports audio input and output interfaces.
Supports rich measurement and analysis functions as well as waveforms, trend
graphs, bar graphs, FFT, vector diagram, harmonics analysis, flicker analysis,
IEC harmonics measurement, cycle-by-cycle analysis, waveform operation,
integral operation and other functions.
Large capacity storage (60G) support for long term data recording.
PAM (PC client management software) communications with PA 5000 through
USB, Ethernet and GPIB interfaces. It manages measurement functions of the
power analyzer, acquires real-time date for analysis and storage and provides
powerful report function.
12.1" color LCD display with 1280*800 resolutions, which can display more
parameters and more detailed waveforms.
Supports rich and quick function keys with mouse and keyboard control.

10
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
1.3 Application System
Figure 1-2 shows the application system of the power analyzer.
Figure 1-2: Application system

11
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
C
Ch
ha
ap
pt
te
er
r
2
2:
:
F
Fu
un
nc
ct
ti
io
on
n
D
De
es
sc
cr
ri
ip
pt
ti
io
on
n
2.1 Main Functions
The power analyzer includes the following functions:
Power parameter measurement. For example, voltage RMS, current RMS,
average value, peak value, crest factor, power factor and fundamental
component.
Motor measurement. The torque and speed signals can be measured.
Measurement and Analysis. Perform further analysis and processing for the
measured results. For example, Fast Fourier Transform (FFT), harmonic
analysis, waveform display, vector diagram, integral function, trend analysis and
the cycle-by-cycle analysis.
Data storage and management. Save and manage the measured data, such as
sampled value.
2.2 Measurement Functions
Table 2-1 lists the measurement functions of the power analyzer.
Table 2-1: Measurement functions
Measurement
Functions of
input
elements[6]
Voltage
Voltage RMS (Urms[1])
Voltage mean (Rectified mean voltage calibrated to the
RMS value (Umn[2]), Linear average voltage(Udc[3]),
Rectified mean voltage (Urmn[4])
Voltage RMS removed DC component (Uac)
Voltage maximum/minimum (U+pk/U-pk)
Current
Current RMS (Irms[1])
Current mean (Rectified mean current calibrated to the
RMS value (Imn[2]), Linear average current (Idc[3]), Rectified
mean current (Irmn[4])
Current RMS removed DC component (Iac)
Current maximum/minimum (I+pk/I-pk)
Power
Active power(P), apparent power(S), reactive power(Q), power
factor(λ), corrected power(Pc)
Efficiency
Efficiency of six input elements (η1~η6)
Others
Phase difference(φ), voltage frequency/current frequency,
voltage crest factor[5]/current crest factor[5], voltage/current
waveform factor[6]
Motor related
functions
Speed, Torque, Pm(motor output or mechanical power), SyncSp
(synchronous speed) and Slip

12
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
Measurement
Functions of
wiring units
Voltage
Voltage mean UΣ(UrmsΣ, UmnΣ, UdcΣ, UrmnΣ, UacΣ)
Current
Current mean IΣ(IrmsΣ, ImnΣ, IdcΣ, IrmnΣ, IacΣ)
Efficiency
PΣ(sum of active power), SΣ (sum of apparent power), QΣ (sum
of reactive power), λΣ (average of power factor), PcΣ(sum of
corrected power)
Others
φΣ (average of phase difference)
User-defined
functions
User-defined measurement functions (F1~F20)
Notes: [1] Urms and Irms are true voltage and current rms value respectively. Each
instantaneous value in one period is squared and the result is averaged, then the resulting average
value is squared again. The equations are shown in Appendix B.
[2] Umn/Imn is the rectified mean value calibrated to the rms value. One period of the voltage
or current is rectified and the resulting average is multiplied by the factor that would yield the true
rms value if the applied signal were a sine wave. The factor is different from the value obtained in
RMS mode when distorted waves or DC waves are measured. The equations are shown in
Appendix B.
[3] Udc/Idc is defined as the average of one period of the voltage/current. This mode is useful
for determining the average value of a DC input signal or the value of a DC signal that is
superimposed on an AC signal.
[4] Urmn/Irmn is the rectified mean voltage/current. One period of the voltage or current is
rectified and the resulting average is obtained.
[5] Crest factor is defined as the ratio of the waveform peak value to the rms value.
[6] Waveform factor is defined as the ratio of the waveform rms value to the rectified mean
value.
PA5000 power analyzer can also use digital filtering to achieve an averaging process
when calculating the measured voltage or current values.
2.3 Operating Mode
The power analyzer includes several operating modes.
Normal Mode. This mode is used to measure voltage, current, torque, speed,
power, and implement integral operation and waveform calculation.
Normal Measurement and Analysis Mode. In this mode, the measured data
saved in the normal mode are replayed and analyzed so that users can observe
and analyze them conveniently.
Harmonics Mode. Harmonics measurement can be performed on a signal with
0.5Hz~5kHz fundamental frequency.
IEC Harmonics Mode. In this mode, users are allowed to perform harmonics
measurement conforming to the international standard IEC61000-3-2 and
IEC61000-4-7.
Voltage Fluctuation and Flicker Mode. This mode allows users to perform
voltage fluctuation and flicker measurements conforming to IEC61000-3-3 and
IEC61000-4-15. Users can also set measurement conditions themselves and
then create a report of measurement results.

13
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
FFT Mode. In this mode, the frequency spectrums of input signals (for example,
voltage (U), current (I), power (P) and Q) are displayed through FFT (Fast
Fourier Transform).
Cycle-by-Cycle Measurement and Analysis Mode. In this mode, the voltage,
current, power and other parameters of each cycle for AC input signal are
measured.
2.4 Voltage/Current Mode
Users can specify the calculation mode to display the calculation results for the
sampling data of voltage or current. The voltage/current mode determines the
calculation mode of the sampling data. Some measurement functions of the power
analyzer are related to the selection of the voltage/current mode. The voltage/current
mode includes five values as follows:
RMS: True RMS;
MEAN: Rectified mean value calibrated to RMS;
DC: Linear average voltage;
AC: Effective value ofAC component;
RMEAN: Rectified mean value.
2.5 Communication Interface
The power analyzer comes with USB, Ethernet, GPIB and RS-232 interfaces. Users
can remotely control the power analyzer through these interfaces, and can also
analyze the sampling data acquired by the instrument in detail and then create
reports.
In addition, the power analyzer provides audio input interface and audio output
interface on the front panel, as shown in Figure 2-1 and Figure 2-2.
Figure 2-1: Interfaces on the rear panel
Figure 2-2: Interfaces on the front panel

14
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
2.6 Display Screen
The voltage, current, power and other measurement data can be displayed by setting
the screen to numeric display. By dividing the screen into top and bottom halves,
waveform×8, trend×16, bar graph×8, waveform operation×2 and FFT×8 can be
displayed simultaneously.
1. Multiple items display
For numeric data display, users can select the number of displayed items. The
number of items can be selected from 6, 12, 24 and All (Display all). In addition, each
display item can be set separately. It is convenient for users to quickly check each
parameter to improve the measurement efficiency, as shown in Table 2-2.
Table 2-2: Measurement functions
Display format
Picture
Description
6 items
Each screen displays the
measurement results of 6
measurement items. Users can
browse data on each screen by
turning page.
12 items
Each screen displays the
measurement results of 12
measurement items. Users can
browse data on each screen by
turning page.

15
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
24 items
Each screen displays the
measurement results of 24
measurement items. Users can
browse data on each screen by
turning page.
All items
Each screen displays the
measurement data as much as
possible. When there are more
than 2 pieces of motor cards,
the numeric value display can
reach ten pages at most. The
displayed items can be
assigned for each page. By
selecting a displayed item, the
numeric data value that is
displayed at the position can be
changed.
Single-column
The measurement data of a
harmonic measurement item
can be displayed in
single-column display format.
Odd harmonics and even
harmonics can be displayed
separately.

16
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
Dual-column
Two different harmonics
measurement can be displayed
at the same time in dual-column
display format.
Six-column
Two different harmonics
measurement can be displayed
at the same time in six-column
display format.
User-defined
Users can set and display
measurement items according
to their own measurement
requirements
2. Combined view
Users can use the combined view to observe the numeric value vs. waveform, the
numeric value vs. trend graph, and the numeric value vs. histogram simultaneously.
For example, the combined view of the numeric value vs. trend graph is shown in
Figure 2-3.
Table of contents