Jorvet J0563 User manual

JORVET HAND-HELD
MINI DOPPLER
J0563 |USER MANUAL
PRODUCT FEATURES
uSmall, Hand-held Size uAutomatic Power “O”
uGreat Sensitivity uTwo Cu Sizes: 2.5cm & 5cm
uConvenient Probe Activation Button (Turns Doppler On & O)

Thank you for choosing the JorVet Hand-held Doppler.
We want you to get the most out of the MiniDop and to have your experience
in using it proceed with ease. Detailed instructions are provided along with the
product’s operating instructions on how to correctly use it and how to take full
advantage of all controls and functions.

CONTENTS
MiniDop Controls
Operating Instructions
Directions for Use
Cleaning Instructions
Warranty
Cautions
Package Contents
Specications
Principles of Doppler Flow
04
06
08
09
10
11

4
MINIDOP CONTROLS
PROBE CONNECTION
To connect the probe to the Doppler
HEADSET
To connect the headset
VOLUME CONTROL
To adjust volume control
POWER INDICATOR
Indicates power On/O
SPEAKER
Outputs Doppler sounds
1
2
3
4
5
5
4
1
2
3
FRONT

5
PROBE HOLDER
For probe handle placement when not in use
STRAP HOLES
To connect the carry strap
BATTERY COVER
Battery compartment for battery replacement
FLAT PROBE
8 MHz – Vascular
PROBE BUTTON
Activates the ES-100VX (On probe handle)
6
7
8
9
10
8
6
BACK
910
7

6
OPERATING INSTRUCTIONS
uOpen the battery cover. Set a 9V alkaline square type battery in the unit ensuring that
the positive and negative electrodes correspond to the (+/-) marks on the battery box.
See Figure A
uConnect the probe with the small dot up on the probe connector (12 o’clock).
See Figure B
uDepress the probe button to turn the unit ON. See Figure C
uMake sure the power indicator on the
unit is ON. Turn the volume control to
minimum, then slowly turn the volume
up. See Figure D
Figure A
Figure B
Figure D
Figure C

7
uPut the ultrasonic gel on the probe top or on the patient’s skin.
uPut the probe on the measurement area and move it slowly to locate the point where
maximum Doppler sounds are heard. The at probe should be 90 degrees placed at
on the animal’s skin over vessel.
uConnect the headset when necessary. Headset can be used to listen to Doppler sounds.
It will cut o the MiniDop speaker to the room. See Figure E
uDepress the probe button again to turn the unit OFF. If the unit is left ON, the power
automatically shuts OFF in about 5 minutes. See Figure F
Note:Theauto-otimedependsonroomtemperature.
uIf the power indicator becomes dark, replace the battery with a new one. Use a 9V
alkaline battery only. See Figure G
Note:Anon-alkalinebatterymaycauseashort
Figure G
Figure E Figure F

8
DIRECTIONS FOR USE
IndirectArterialBloodPressureMonitoringwiththeDopplerUnit
uClip the palmer aspect of the foot just proximal to the metacarpal pad. It is important to clip
the fur as short as possible. Rub a small amount of ultrasound gel on the skin in the midline
area where the palmer arterial arch is located. It can be palpated on large dogs and is actually
just slightly medial to the midline.
uPlace a large “dollop” of ultrasound gel on the non-numbered side of the Doppler ow probe.
Place the probe in the location of the artery described in above and tap it in place tightly.
(Turn on the Doppler unit with the ow probe plugged into the unit. The ow probe should not
be frequently disconnected from the unit as this loosens the connectors and wires in the unit).
uEnsure that the unit is charged. A swishing sound should be heard each time blood ows
through the artery. If blood ow is not heard, use the probe as a locator and move it slightly
until the swishing sound is heard and then re-tape in place.
uMake a loop in the cord of the ow probe. Tape the cord to the foot/metacarpal loop right
over the tape used to xate the ow probe to the skin initially.
uThe blood pressure cu is now in place. The cu’s width should approximate the width and
be up to two times that of the forearm (mid-radial region). It is applied so that the mid-section
on the rubber blade inside the cu is directly over the posterior medial aspect (where the
radial artery is located). This can be located by digitally occluding the area and hearing the
swishing sounds of the Doppler stop indicating the artery is at that location because its ow
is now temporarily blocked.
uWhen the cu is applied the INDEX line on the end of the cu should fall within the RANGE
on the cu, which is diagrammed on the inside. If outside this range, the reading will be either
falsely low (with cus too big for the size of the limb it is surrounding) or falsely high (with the
cu too small). An estimate can be arrived and that will indicate how far o, in percentage,
the cu reading will be, taken by observing the amount of length of the indexed area.
uA small amount of adhesive tape is added around the cu to prevent it from popping the
Velcro and showing a falsely high reading.
uThe cu is inated by squeezing the rubber bulb with the air outlet nozzle closed until the
manometer reading is 20-30mm Hg above when the swishing sound stopped. The air-outlet
nozzle is slowly opened and the cu is allowed to slowly deate. When the swishing sound is
rst heard again, this marks the ow of blood through the artery corresponding to SYSTOLIC
PRESSURE on the sphygmomanometer. The sound should be a shorter choppier sound than
was heard when the cu was not inated.
uThe cu deation is continued slowly as the characteristics of the swishing sound are
carefully monitored. A light “backward” sound or diastolic “swish” sound is again heard (or
as it is again sounded prior to any cu ination) corresponding to DIASTOLIC PRESSURE
when a very mild amount of “backow” in the artery occurs. The diastolic pressure is not
always evident.

9
uDiastolic pressure can be detected by closely watching the dial of the manometer. If it
oscillates reproducibly at certain times on the deation cycle, this corresponds to within
10mm Hg of the diastolic pressure (by sound). For accuracy, when the diastolic sound cannot
be heard, pressure is simply recorded as “systolic all the way down”. This is especially
frequent during periods of high stress when the vessels are under signicant catecholamine
inuence and the area is acting “sti”, or in very small or cold patients where the vessels are
simple to small to allow sound change detection.
Note:Normalsystolicvaluesofdogsandcatsrangefrom110-160mmHg.
CLEANING INSTRUCTIONS
WARRANTY
Probe
MainUnit
uRemove the Doppler gel from the probe head after each use.
uClean or disinfect the probe using a cotton ball soaked in alcohol.
uConsult manufacturer before using any other chemical disinfectants.
uDO NOT submerge probe in any liquid.
uThis equipment is guaranteed for a period of one year after the date of purchase when used
under normal conditions.
uIn the event of any trouble during the warranty period, please contact the dealer from whom
you purchased the unit.
uIn case the warranty period is over, please contact the dealer for a charged service.
uTo clean main unit, use a little water and wipe with a soft dry cloth. DO NOT submerge in liquid.
THE WARRANTY CARD MUST BE RETURNED WITHIN TWO WEEKS OF PURCHASE!
IMPORTANT

10
CAUTIONS
PACKAGE CONTENTS
Probe
uThe standard probe is for transcutaneous use ONLY. The probe transducer is a very thin and
delicate part. Please take care NOT to drop or hit the probe tip. DO NOT submerge probe in
any liquid.
UltrasoundGel
uALWAYS use an ultrasound gel. Using other materials such as baby oil or cream may not
produce the correct Doppler sounds and may damage the probe.
Battery
uWhen battery is low, the power indicator becomes dark. Also there will be no speaker sounds.
Replace the battery if these conditions exist (refer to Figure G on page 7).
TheJorVetHand-heldDopplerPackageIncludestheFollowingItems
uJ0563D1 Flat Probe (Includes probe tip holder)
uJ0563D2 Mini Doppler Detector Unit
uJ0563D3 Sphygmomanometer for Mini Doppler
uJ0563D4 2.5cm Cu
uJ0563D5 5.0cm Cu
uJ0563D8 Headset for Doppler
uHadeco Carrying Case
u9V Alkaline Battery
uAdjustable Strap
uUltrasonic Gel
uUser Manual
uUse of the Doppler for the Feline Patient Booklet
uWarranty Card

11
The JorVet Doppler is designed to receive blood ow velocity information by ultrasound. A
specic frequency is transmitted from the probe to the patient.
Technically, the transceiver amplies a high frequency oscillation output for transmission to the
transducer. The voltage is converted by a piezoelectric crystal (transducer) to ultrasound. The
ultrasound beam is transmitted to blood cells owing through the arteries or to beating fetal
hearts. The ultrasound beam is reected by the crystal in the transducer which converts the
ultrasound into a voltage. A Doppler shift occurs between the emission and reception of the
ultrasound beam.
Advantages of using Doppler ow detectors include ease of application, portability, applicability
to hypotensive animals, and usefulness as an audible monitor of the arterial pulse and peripheral
blood ow.
PRINCIPLES OF DOPPLER FLOW
SPECIFICATIONS
PROBE MODEL FREQUENCY
J0563D1 Flat Probe 8mHz
DETAILS
Battery DC 9V, Alkaline Square Type
Battery Life Approximately 3 hours (Alkaline)
Auto Shut-O Automatic after approximately 5 minutes of inactivity.
Varies with room temperature.
Probe Button Power On/O
Speaker Output 200 mW or more
External Outputs Headset (3.5mm jack) – Cuts o sound from speaker
Dimensions
Main Unit
75mm (W) x 140mm (D) x 25mm (H)
Probe
20mm (Diameter) x 105mm (L)
Weight Approximately 270 grams
(Includes battery & probe)

Jorgensen Laboratories, Inc. |Loveland, CO 80538
Professional Veterinary Equipment Supplier
Table of contents