Keeler 2508-P-8026 User manual

Disposable Probe Adapter Part Number 2508-P-8026
Instructions for use
The system automatically initiates a purging
cycle of 90 seconds. During the purge cycle an
animated Wait Symbol is displayed alongside
the Probe Symbol.
Three short beeps signal that the purge cycle is
complete. The equipment is now ready for use
as indicated by the timer and the Ready Symbol.
Freezing cycles can now be controlled by the Cryomatic
footswitch as normal (see Cryomatic user manual).
Warnings and Precautions
• Only to be used in conjunction with Keeler
Cryomatic (2509-P-1000) and Disposable Cryo
Probe (2508-P-7022)
• Check probe before use – do not use or repair
if damaged or deformed
• Disconnect probe from adapter only when
system is depressurised
• Maximum pressure 5860kPa (850psi)
• Consult Cryomatic instruction manual if in doubt
• Unauthorised interference invalidates warranty
• This device should be used by trained staff
experienced in the use of ophthalmic equipment
• USA federal law restricts this device to sale or
on the order of a physician
1
Connect the Probe Adapter to the Cryomatic console
by aligning the coupling on the console spigots
and pushing to a positive click (It is impossible to
connect the coupling the wrong way round – correct
orientation is with the release button uppermost).
Connect the disposable probe by inserting the probe
plug into the adapter and pushing the probe firmly
and squarely against the collar(1).
The safety latch(2) will operate and the locking
ring(3) will spring towards the probe to confirm that
the probe has been successfully connected.
Operation
2
Before use ensure that the rear safety
clip(4) is in the fully down position
- under no circumstances should it
be used if the safety clip is not in
position (refer to image on page 2).
1
2
4
3
Sterilized using Ethylene Oxide
Manufacturer
Caution
The CE mark on this product indicates
it has been tested to and conforms
with the provisions noted within the
93/42/EEC Medical Device Directive
Do Not Reuse
Use By
Batch Code/Lot Number/Batch Number
Consult Instructions For Use
This Symbol on the Product or on its
Packaging and instructions indicates
that it was put on the market place
after August 2005 and that this
product shall not be treated as
Household Waste. EU only.
Symbols
3
0088
Continued overleaf
1
2
4
3
The disposable probe should only be removed from the adapter
when the system is depressurised as indicated by the ready symbol.
Disconnecting the disposable probe
4
2509-P-8019 Issue E
To disconnect the disposable probe from the
Probe Adapter move the locking ring(1) away
from the disposable probe and operate the
latch(2).
The disposable probe should be gently ejected
from the adapter. Lift the safety clip(3) and remove
the disposable probe plug from the adapter.
The Probe Adapter can itself be removed from
the Cryomatic Console by operating the release
button(4).
2 years against faulty workmanship, materials and labour. Warranty will be conditional on
routine maintenance and will not cover calibration or mechanical issues caused as a result of
lack of maintenance, poor use, incorrect transport or inappropriate storage conditions.
If necessary, the Probe Adapter can be cleaned using a
disposable cloth soaked in mild detergent and warm
water. Do not use abrasive compounds or pads. The user
should avoid getting electrical contacts wet during the
cleaning process.
There are no user serviceable parts in the Probe
Adapter and operator maintenance is restricted to
inspection of the adapter for signs of damage before
every use.
The Probe Adapter should be inspected annually by
Keeler trained personnel.
Repairs should be carried out by
Keeler trained personnel or their
representatives.
Ensure that this equipment is disposed of
in accordance with local regulations.
Please contact the supplier if in doubt.
Product warranty
6
Manufactured by
Keeler Limited, Clewer Hill Road, Windsor, Berkshire, SL4 4AA UK
Freephone: 0800 521251 Tel: +44 (0)1753 857177 Fax: +44 (0)1753 827145
Distributed by
USA Sales Office, Keeler Instruments Inc, 3222 Phoenixville Pike, Building #50, Malvern,
PA 19355, USA. Toll Free: 1 800 523 5620 Tel: 1 610 353 4350 Fax: 1 610 353 7814
Cleaning, maintenance and disposal
5
0088

Popular Adapter manuals by other brands

Amfeltec Corp. SKU-037 Hardware manual

Amfeltec Corp.

Amfeltec Corp. SKU-037 Hardware manual

Motorhead MH-03-BA-B1-USB-1 instruction manual

Motorhead

Motorhead MH-03-BA-B1-USB-1 instruction manual

Clarion BLT573 Installation & owner's manual

Clarion

Clarion BLT573 Installation & owner's manual

Goobay 65369 manual

Goobay

Goobay 65369 manual

ABB RCNA-01 user manual

ABB

ABB RCNA-01 user manual

Canon NA10 user manual

Canon

Canon NA10 user manual

VXI X100 manual

VXI

VXI X100 manual

Sony PS VITA PCH-ZAC1 E instruction manual

Sony

Sony PS VITA PCH-ZAC1 E instruction manual

tams elektronik S88-A manual

tams elektronik

tams elektronik S88-A manual

Kontron ADA-LVDS-DVI user guide

Kontron

Kontron ADA-LVDS-DVI user guide

3idee HP-HP24F instruction manual

3idee

3idee HP-HP24F instruction manual

Stahl 9411/24 Series operating instructions

Stahl

Stahl 9411/24 Series operating instructions

Digitus DA-10170 user manual

Digitus

Digitus DA-10170 user manual

Panasonic KX-TGA575S Instrucciones de operación

Panasonic

Panasonic KX-TGA575S Instrucciones de operación

Jabra A210 Specifications

Jabra

Jabra A210 Specifications

HighPoint HighPoint RocketRAID 2700 Series user guide

HighPoint

HighPoint HighPoint RocketRAID 2700 Series user guide

Belkin F8T016 Quick installation guide

Belkin

Belkin F8T016 Quick installation guide

Elecom EC-AC4465 user manual

Elecom

Elecom EC-AC4465 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.