Technical Data Sheet
Presure • Temperature • Humidity • Air Velocity • Air Flow
Flush-mount multi-channel display
ATE 300
• Alternating display of 1 to 3 parameters
•
•
•
•
• Configuration via software or remote control
• Electroluminescent f
3 analogue inputs : 3 x 4-20mA or 3 x 0-10V
1 RS 232 digital input for KIMO external transmitter (class 200 & 300)
1 RS 485 digital input/output (Modbus protocol integrated)
22 pre-set units and 3 configurable units
ront display, made of brushed stainless steel or white
lacquered
Display features
Display lectroluminescent alphanumeric (38 x 48mm)
protection screen made of PMMA
st
1 line (measurement)...........4 digits x 8 segments - 14 x 48 mm
nd
2 line (unit) ...........................4 digits x 14 segments - 14 x 48 mm
Comma position ....................configurable 0 / 0,0 / 0,00
Measurement value ...............from -9,99 to 99,99 and from -999 to 9999
Accuracy.................................±0,1 % of the measurement ±1 digit
Number of channels..............from 1 to 3 channels, alternatively (3 seconds)
Location of channels.............with 3 LED: green, orange, red
Units available .......................22 pre-set units (see chart)
3 configurable units
Response time.......................< 1 sec.
....................................e
.................................................
.................................................
References
Pre-set units
ATE-300
Front face
-
B
I
White lacquered stainless steel
Brushed stainless steel
Front face...........................................316L wire brushed stainless steel
............................................................or white lacquered
Back housing ....................................flushmount, 304L stainless steel
Protection factor ...............................IP 65 in front face
Dimensions........................................see drawing
Weight................................................600g
Housing features
*All accuracies indicated in this technical datasheet were stated in laboratory conditions, and can be guaranted for
measurements carried out in the same conditions, or carried out with calibration compensation.
Technical specifications
Power supply....................................24 Vac / Vdc ±10%
Galvanic isolation ............................between input and power supply
Analogue inputs ...............................3 x 4-20 mA (4 wires) or 3 x 0-10 V
Consumption ....................................5 VA
Electro-magnetical compatibility ....EN 61 326
Electrical connection bles Ø1.5 mm² max
RS 485 communication....................digital: RTU Modbus protocol
communication speed configurable
from 2400 to 115200 Bauds
RS 232 communication....................digital: ASCII, proprietary protocol
Working temperature .......................0 to +50°C
Storage temperature ........................-10 to +70°C
Environment .....................................air and neutral gases
.........................screw terminal block for ca
...........................................................
...........................................................
Air velociy m/s
fpm
3
Airflow m/h
l/s
cfm
3
m/s
Temperature °C
°F
Pressure Pa
mmH O
2
mbar
Kpa
mmHg
inWg
PSI
Humidity %RH
g/kg ñ)
°C
°F
°C
°F (humid temp. )
KJ/Kg
(absolute hygro
(dew point Td)
(dew point Td)
(humid temp. Tw)
Tw
(Enthalpy i)