【3-2 Contamination of the Piping System】
Be sure to use well-cleaned pipes and joints for piping. A standard filter has been installed. However, if
a large amount of dust adheres, the gas flow will be disturbed. In particular, if air from a compressor or
blower fan is used, it is likely that a large amount of oil mist and water may enter the meter. In such a case,
always install an oil filter or water removing filter in front of the meter.
【3-3 Installation Method】
1) For installation, utilize the mounting threaded holes (M4) on the bottom of the meter block.
2) The standard piping connection is VCR. Use suitable joints and be sure that they will not leak.
【3-4 Wiring】
Be sure to use suitable connecting cables.
4. Operation
【4-1 Warming Up】
After the power was turned on, warm up the meter for 15 minutes or longer with the supply of gas stopped.
If the meter is not warmed up, the measuring accuracy will deteriorate.
【4-2 Zero/Span Calibration】
If the user has the reference flow meter, the span may be calibrated. If no reference meter is available,
conduct the zero calibration only.
1) Zero calibration
When the indication has become stable with gas shut off after the meter was warmed up, using a small
screwdriver, adjust the “ZERO” variable resistor installed on the top of the case to zero. Be sure that
the indicated value or the output becomes zero.
2) Span calibration
If the reference meter is not available, never tinker with the “SPAN” variable resistor.
In case of Model 5100, make the gas flow in full scale flow rate, and adjust “SPAN”volume as output
flow signal voltage becomes DC5V.
In case of Model5100, set flow-setting command to 10.00 (or adjust flow set voltage to 5V), and
adjust “SPAN”volume as the reference flow meter shows full scale.
If the user does not have the reference flow meter, never tinker with the “SPAN” variable resistor. Please avoid
calibration with user as much as possible, and please send it back to Kofloc.