DOS & DON’TS
¬Hold syringe vertically when filling disks.
¬Do not fill disk while in the meter. Fill disk on clean,
dry surface.
¬The disk should not contain any large air bubbles.
Air bubbles will result in erroneous results.
¬Only the Universal Disk Cover (Code 1719) can be
used with the WaterLink®Spin Touch™.
¬Empty syringe of previous sample before filling with
next sample.
¬Remove filled disk from meter after testing.
¬Store disks at 70°-80°F/21°-27°C.
BLUETOOTH CONNECTION
The WaterLink®Spin Touch™is capable of connecting to
a Bluetooth enabled device, such as a phone or tablet,
and the Mobile Bluetooth Printer. The Spin Touch will
only connect to and print from the LaMotte Bluetooth
Printer [Code 5-0066]. Other Bluetooth printers are not
supported.
Whenever the button appears on the touch
screen the Spin Touch is capable of connecting to a
Bluetooth enabled phone or tablet. The will become
highlighted when connected and be dim when not
connected. Whenever the button appears on the
touch screen the Spin Touch is capable of connecting to
the Mobile Bluetooth Printer. The will be highlighted
when connected and dim when not connected. The Spin
Touch cannot connect to both a phone or tablet and the
printer at one time.
410-778-3100 | 800-344-3100 | www.lamotte.com
Disk US Patent No. 8,734,734
FCl US Patent No. 8,987,000
FCl EU Patent No. EP2784503 A1
TCl US Patent No. 8,993,337
3580-QG 10.19.18
FILLING: Multiple-Use Disks
When the syringe is placed in
the pool or spa sample, and
the plunger is pulled all the
way up, the syringe will hold
more than enough sample
water to adequately fill one
section of the disk. The syringe
is held vertically and the tip is
inserted into the fill hole on the
disk. The plunger is pressed
slowly and smoothly to fill the
disk.
1. Fill any unused section
of the disk using slow,
even pressure. Add
the sample water until
the Blank Chamber is
full and the Over Flow
Chamber is partially
filled. Correct Fill
Blank
Chamber
Over Flow
Chamber
2. Do not overfill the disk.
If the disk is overfilled,
sample water will flow
out of the vent hole in the
center of the disk. The disk
is not leaking. Dry the disk
and run the test.
Over
Filled
(Dry disk before placing in meter)
Blank
Chamber
Over Flow
Chamber
3. Do not under fill the
disk. If the disk is
under filled, the blank
chamber will not be full.
The reagent chambers
will not fill entirely
and results will be
inaccurate. Under Filled
Blank
Chamber
4. Do not introduce air
bubbles into the disk. The
reagent chambers will not
fill entirely and results will
be inaccurate. As soon as
a bubble starts to form,
pull back on the plunger
to draw the bubble out of
the disk. Begin the filling
process again.
Air Bubble
5. Wet disks should
be dried thoroughly
with a lint free wipe.
Handle the disk by the
edges. Between tests
store the disk on a flat
surface. Do not agitate
a partially filled disk
to avoid leaking and
contamination. Partially
used disks should be
discarded at the end of
the day.
6. Disks should be filled and used within 10 minutes. They
cannot be filled ahead of time.
For filling and troubleshooting tips go to: www.waterlinkspintouch.com
See the WaterLink®Spin Touch™manual for additional information.