QUICK
START GUIDE
BEFORE FIRST USE
¬Fully charge battery using
the Spin Touch USB cable and
adapter plug in AC outlet.
¬Ensure your Spin Touch has the
latest firmware:
1. Download and install
the WaterLink®Connect
application for Windows®
at http://softwarecenter.
lamotte.com/.
2. Plug meter into the computer
with the provided USB cable
and launch WaterLink®
Connect application from the
Start Menu.
3. Any firmware updates will
automatically be applied to
the meter.
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the
most accurate result, wait until water sample is below 90 °F (32 °C) to test.
TESTING
1. Press and hold until the meter turns on.
2. Tap . Select a sanitizer type. Tap to confirm.
3. Tap . Select a disk series (found on disk packaging). Tap to
confirm.
4. Remove disk from packaging.
5. Use syringe (1189) to fill disk with pool or spa water sample.
6. Insert disk. Cover disk with Universal Disk Cover. Close lid.
7. Tap to start test. Tap to cancel the test.
8. Results will be displayed.
9. Select an option.
- Tap to return to the Test Screen.
- Tap highlighted to save test results to test log if Auto Save is
not enabled.
- Tap the hightlighted to send the results to the enabled
Mobile Bluetooth Printer.
- Tap highlighted to send results to Bluetooth enabled device.
10. Press and hold for 2 seconds to turn meter off.
FILLING: Single-Use Disks
1. Fill the disk using slow,
even pressure. The sample
water will fill the spaces
between the baffles in a
counterclockwise order.
Each space will fill from the
bottom to the top. Sample
water should be added
until the sample water in
the fourth chamber fills
to the top of the chamber
slightly past the embossed
fill line. It is OK to fill
slightly past the fill line.
Correct Fill
(Fill line)
2. Do not overfill the disk.
If the disk is overfilled,
sample water will flow
out of the overflow hole
in the center of the disk.
The disk is not leaking.
Dry the disk and run the
test.
O
ver
F
illed
(Dry disk before placing in meter)
3. Do not under fill the disk. If
the disk is under filled, the
reagent chambers will not
fill entirely and results will
be inaccurate. Under
Filled
4. Do not introduce air
bubbles into the disk.
The reagent chambers
will not fill entirely
and results will be
inaccurate. As soon as
a bubble starts to form,
pull back on the plunger
to draw the bubble out
of the disk. Begin the
filling process again.
Air
Bubble
5. Wet disks should be dried
thoroughly with a lint free
wipe. The disk should be
handled by the edges.
6. Disks should be filled and used within 10 minutes.
They cannot be filled ahead of time.
For filling and troubleshooting tips go to: www.lamotte.com/spin
Compatible with iPhone 6 Plus, iPhone 6, iPhone 5s,
iPhone 5c, iPhone 5, iPad Air 2, iPad Air, iPad mini 3,
iPad mini 2, iPad mini, and iPad (4th generation).
Compatible with Android devices.
DOS & DON’TS
¬Hold syringe vertically when filling disks.
¬Do not fill disk while in the meter. Fill disk on clean,
dry surface.
¬The disk should not contain any large air bubbles.
Air bubbles will result in erroneous results.
¬Only the Universal Disk Cover (Code 1719) can be
used with the WaterLink®Spin Touch™.
¬Empty syringe of previous sample before filling with
next sample.
¬Remove filled disk from meter after testing.
¬Store disks at 70°-80°F/21°-27°C.
BLUETOOTH CONNECTION
The WaterLink®Spin Touch™is capable of connecting to
a Bluetooth enabled device, such as a phone or tablet,
and the Mobile Bluetooth Printer. The Spin Touch will
only connect to and print from the LaMotte Bluetooth
Printer [Code 5-0066]. Other Bluetooth printers are not
supported.
Whenever the button appears on the touch
screen the Spin Touch is capable of connecting to a
Bluetooth enabled phone or tablet. The will become
highlighted when connected and be dim when not
connected. Whenever the button appears on the
touch screen the Spin Touch is capable of connecting to
the Mobile Bluetooth Printer. The will be highlighted
when connected and dim when not connected. The Spin
Touch cannot connect to both a phone or tablet and the
printer at one time.
[email protected]
410-778-3100 | 800-344-3100 | www.lamotte.com
Disk US Patent No. 8,734,734
FCl US Patent No. 8,987,000
FCl EU Patent No. EP2784503 A1
TCl US Patent No. 8,993,337
3580-QG 10.19.18
FILLING: Multiple-Use Disks
When the syringe is placed in
the pool or spa sample, and
the plunger is pulled all the
way up, the syringe will hold
more than enough sample
water to adequately fill one
section of the disk. The syringe
is held vertically and the tip is
inserted into the fill hole on the
disk. The plunger is pressed
slowly and smoothly to fill the
disk.
1. Fill any unused section
of the disk using slow,
even pressure. Add
the sample water until
the Blank Chamber is
full and the Over Flow
Chamber is partially
filled. Correct Fill
Blank
Chamber
Over Flow
Chamber
2. Do not overfill the disk.
If the disk is overfilled,
sample water will flow
out of the vent hole in the
center of the disk. The disk
is not leaking. Dry the disk
and run the test.
Over
Filled
(Dry disk before placing in meter)
Blank
Chamber
Over Flow
Chamber
3. Do not under fill the
disk. If the disk is
under filled, the blank
chamber will not be full.
The reagent chambers
will not fill entirely
and results will be
inaccurate. Under Filled
Blank
Chamber
4. Do not introduce air
bubbles into the disk. The
reagent chambers will not
fill entirely and results will
be inaccurate. As soon as
a bubble starts to form,
pull back on the plunger
to draw the bubble out of
the disk. Begin the filling
process again.
Air Bubble
5. Wet disks should
be dried thoroughly
with a lint free wipe.
Handle the disk by the
edges. Between tests
store the disk on a flat
surface. Do not agitate
a partially filled disk
to avoid leaking and
contamination. Partially
used disks should be
discarded at the end of
the day.
6. Disks should be filled and used within 10 minutes. They
cannot be filled ahead of time.
For filling and troubleshooting tips go to: www.waterlinkspintouch.com
See the WaterLink®Spin Touch™manual for additional information.

Other manuals for WaterLink Spin Touch

4

Popular Chemical Test manuals by other brands

Hanna Instruments HI 93754C-25 HR instruction manual

Hanna Instruments

Hanna Instruments HI 93754C-25 HR instruction manual

Hanna Instruments HI 38073 instruction manual

Hanna Instruments

Hanna Instruments HI 38073 instruction manual

Hanna Instruments HI 3875 instruction manual

Hanna Instruments

Hanna Instruments HI 3875 instruction manual

Hanna Instruments DiST 2 user manual

Hanna Instruments

Hanna Instruments DiST 2 user manual

Hanna Instruments HI 38086 instruction manual

Hanna Instruments

Hanna Instruments HI 38086 instruction manual

Hanna Instruments HI 3855 instruction manual

Hanna Instruments

Hanna Instruments HI 3855 instruction manual

Hanna Instruments HI 3850 instruction manual

Hanna Instruments

Hanna Instruments HI 3850 instruction manual

Hanna Instruments HI 38051 instruction manual

Hanna Instruments

Hanna Instruments HI 38051 instruction manual

Hanna Instruments HI 3847 instruction manual

Hanna Instruments

Hanna Instruments HI 3847 instruction manual

Hanna Instruments HI 38000 instruction manual

Hanna Instruments

Hanna Instruments HI 38000 instruction manual

Hanna Instruments HI 3824 instruction manual

Hanna Instruments

Hanna Instruments HI 3824 instruction manual

Hanna Instruments HI 38001 instruction manual

Hanna Instruments

Hanna Instruments HI 38001 instruction manual

Hanna Instruments HI 38016 instruction manual

Hanna Instruments

Hanna Instruments HI 38016 instruction manual

Hanna Instruments HI 3835 instruction manual

Hanna Instruments

Hanna Instruments HI 3835 instruction manual

JBL PRO AQUATEST pH 7.4-9.0 INFORMATION FOR USE

JBL

JBL PRO AQUATEST pH 7.4-9.0 INFORMATION FOR USE

INDICAID 2120100 quick start guide

INDICAID

INDICAID 2120100 quick start guide

Elitech ENVOY 500 HDL manual

Elitech

Elitech ENVOY 500 HDL manual

Hanna Instruments HI 38050 Handbook

Hanna Instruments

Hanna Instruments HI 38050 Handbook

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.