Additional information?
If you would like further assistance with
using our Neoprobe, or need advice on how to
order, please call us on: 0114 268 8880.
ukmedical.com
Manufactured by
Supplied by UK Medical Ltd.
Press POWER button on console; wait until
console screen displays ‘no LInc’
Press POWER button on probe, Bluetooth will
reconnect within approximately 5 seconds
Replacement Batteries
1. CR123 and RCR123 Li-Ion Batteries
2. RCR123 (Rechargeable) batteries have at least 5 hours of continuous use
3. CR123 (Disposable) batteries have at least 10 hours of continuous use
After several seconds the console screen
will display numbers in the target count and
background count rows – Bluetooth pairing is
now complete.
• Probe will automatically enter sleep mode after 5
minutes of inactivity or if no radiation is detected.
Console screen will display ‘no LInc’.
• Probe can be reconnected (synchronised) from
within the sterile field by pressing the probe
button, BT reconnection will be instantaneous.
• Console screen displays ‘Lo bAtt’ upon power up
- Replace probe battery before procedure
• Console screen displays ‘Lo bAtt’ once during case
- Probe has enough power to complete the current procedure, but
battery should be changed immediately after surgery is complete
• Flashing ‘no probe connected’ symbol with
chime after indicates low battery warning
- Change battery immediately
• Remove battery from probe when stored to avoid slow battery drain
• Probe sleep mode (BT de-synchronisation) can
be initiated manually by pressing and holding
the background count button on the console for
approximately 5 seconds or until a beep is heard.
It’s Neoprobe /
Steps for use
Setting up & initial bluetooth pairing continued
Plug and Play (initial Bluetooth pairing must have been established)
Step 5. Bluetooth Operation.
Probe Battery Management
Step 1.
Step 2.
These instructions are to be used as a reference guide only. Please refer to the full instructions for use prior to using this product.
Volume
Power
Probe
Input
Target
Count
Background
Count
Mute
Count Range
10,000
50,000
1,000
100
Neoprobe
Gamma Detection System
Volume
Power
Probe
Input
Target
Count
Background
Count
Mute
Count Range
10,000
50,000
1,000
100
Neoprobe
Gamma Detection System
Volume
Power
Probe
Input
Target
Count
Background
Count
Mute
Count Range
10,000
50,000
1,000
100
Neoprobe
Gamma Detection System
Volume
Power
Probe
Input
Target
Count
Background
Count
Mute
Count Range
10,000
50,000
1,000
100
Neoprobe
Gamma Detection System