Masterflex E/S 07580-00 User manual

OPERATING MANUAL
Composite Sampler
07580-00
& accessories
Model No.
Cole-Parmer Instrument Co.
1-800-MASTERFLEX (627-8373) (U.S. and Canada only)
11 (847) 549-7600 (Outside U.S.) • (847) 549-7600 (Local) • www.masterflex.com
Barnant Company
1-800-637-3739 (U.S. and Canada only)
11 (847) 381-7050 (Outside U.S.) • (847) 381-7050 (Local) • www.barnant.com
A-1299-1015
Edition 01-11-11-02

TABLE OF CONTENTS
Title Page
SAFETY PRECAUTIONS......................................................3
INTRODUCTION
General..................................................................................4
Suction Lift............................................................................4
Scope ....................................................................................4
SETUP
Unpacking ............................................................................5
Assembly ..............................................................................5
Tubing Selection ..................................................................5
INSTALLATION
Site Requirements................................................................5
Power Requirements ..........................................................5
DESCRIPTION
Functional Description ........................................................6
Wetted Parts ........................................................................6
Controls, Indicators and Connectors ................................6
Familiarization System Features ........................................7
INITIAL OPERATION
System Boot-up Internal Battery........................................8
System Boot-up External Power ........................................9
Battery Charging................................................................10
Setting Time and Date ......................................................10
Security ..............................................................................10
OPERATION
Description of Programming ............................................11
Starting & Stopping Program............................................11
Menu....................................................................................12
Navigation ..........................................................................14
GLOBAL SETTINGS
General................................................................................14
Global Settings Tour ..........................................................14
VIEW/EDIT PROGRAMS
General................................................................................19
Program Menu Tour ..........................................................19
SAMPLING PROGRAM MENU OPTIONS
General................................................................................22
Program Start ....................................................................22
Sample Volume ..................................................................23
# of Contact Closures........................................................24
Default Program Settings..................................................25
Save Program ....................................................................25
Hot Keys..............................................................................26
CALIBRATING THE SYSTEM
General................................................................................26
Minimum Sample Size ......................................................26
Required Equipment..........................................................26
Calibration Procedure........................................................26
Previous Calibrations ........................................................27
CYCLING THE SYSTEM ....................................................27
MAINTENANCE
Sample Reservoir ..............................................................27
Tubing..................................................................................27
TROUBLESHOOTING
General................................................................................28
Error/Status Messages & Solutions................................28
USER REPLACEMENT PARTS SERVICING
Internal Battery Replacement ..........................................29
Motor Brush Replacement/Check....................................30
USER REPLACEMENT PARTS..........................................31
SPECIFICATIONS ..............................................................32
WARRANTY ........................................................................33
PRODUCT RETURN ..........................................................33
TECHNICAL ASSISTANCE ................................................33
ACCESSORY OPERATING INSTRUCTIONS
07571-50 Automotive Power Adapter ..............................34
07571-52 & -54 Auxiliary Power-Pak ................................35
07580-60 Remote Contact Closure Cable ......................38
07580-65 RS-232 Adapter..................................................38
2
NORPRENE, PHARMED, TYGON—Reg TM Norton Co.
WINDOWS—Reg TM Microsoft Corp
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries

3
SAFETY PRECAUTIONS
DANGERS: NEVER apply AC voltages directly to the EXTERNAL POWER INPUT receptacle on the front
panel. The application of AC voltages can result in injury and death of the operator and
destruction of the unit. Use only the AC/DC Universal Power Supply/Converter supplied
with the unit to power unit from an AC source.
NEVER short or connect the terminals of the battery terminals together. Shorting of the
battery terminals causes rapid internal heating of the battery resulting in the explosion of
the battery and severe injury to or the death of the operator.
The AC/DC Universal Power Supply/Converter is rated for INDOOR USE ONLY. DO NOT use
the AC/DC Power Supply/Converter in an outdoor environment to either charge the battery
or power the drive. Electrical shock, severe injury and/or death are possible if this warning
is ignored.
DO NOT BURN OR INCINERATE THE BATTERY. THE BATTERY MAY EXPLODE CAUSING
SEVERE INJURY OR DEATH OF THE PERSONNEL IN THE AREA. (Dispose of the old battery
by recycling.)
CAUTIONS: Do not reverse the connections to the battery. If the battery connections are reversed,
damage to the unit will occur.
Fully charge the unit before using for the first time. Damage to the internal battery can result
if the battery is in a fully discharged state and operation of the unit is attempted.
Immersion or submersion of the unit will result in improper operation and possible damage to
the unit.
Use of pump heads, tubing sizes and formulations other than those specified in this manual,
or the mounting and use of two or more pump heads concurrently will result in improper
operation and possible damage to the unit.
WARNINGS: Tubing breakage may result in fluid being sprayed from the pump. Use appropriate measures
to protect the operator and equipment.
Turn drive off, remove all the power to the unit, including the AC/DC Power Supply/Converter,
Automotive Power Adapter or other external power source if present before removing or
installing pump head or tubing. This will help prevent accidental activation of the drive
mechanism so fingers or loose clothing will not get caught in the pump drive mechanism.
WARNING: PRODUCT USE LIMITATION
This product is not designed for, nor intended for use in, patient-connected applications, including, but not
limited to, medical and dental use and, accordingly, has not been submitted for FDA approval.

4
INTRODUCTION AND GENERAL DESCRIPTION
GENERAL
The Masterflex®E/S™Composite Sampler is a DSP controlled,
portable, self-contained, programmable peristaltic pump
system designed to meet EPA, NPDES guidelines for the
sampling of storm and waste water.
The Masterflex®E/S™Composite Sampler as shown in Figure
1, consists of an exterior housing, a controls console which
contains the internal power source, pump drive system and
system controls, a Masterflex®L/S®Easy-Load®peristaltic
pump head and a sealed and vented sample reservoir. The
drive system is capable of operating one Masterflex®L/S®
Standard®, L/S®Easy-Load® or L/S® PTFE pump head.
The Masterflex®E/S™Composite Sampler is current limited,
protected from reverse power supply polarity and transient
voltages.
You may operate the Masterflex®E/S™Composite Sampler on
the internal battery, an external 12V DC source or indoors
using the 115V AC or 230V AC with the AC/DC Universal
Power Supply/Converter supplied with the unit.
The system features a built-in spill-proof gel battery and
recharging system allowing the unit to be recharged from an
AC power source using the supplied AC/DC Universal Power
Supply/Converter or from an automotive 12V DC electrical
system if used with the Automotive Power Adapter accessory.
It is designed and programmed to accept, control and power
one Masterflex®L/S®Easy-Load®, Standard or PTFE Tubing
pump head for fluid transfer and sampling in the field.
The unit is housed in a high visibility, protective polyethylene
housing. The control panel and the drive controls have an
environmental protection rating of IP56.
When closed and latched, the drive will float for a minimum of 30
minutes if dropped into a lake or stream to allow recovery of the unit.
SUCTION LIFT
The Masterflex®E/S™ Composite Sampler is capable of lifting
a sample of water up to twenty-five (25) feet vertically. This
capability is called suction lift and is measured from the
surface of the liquid to the pump head.
If the liquid to be sampled has a viscosity or density greater
than water, this lifting capability is reduced or diminished.
This lifting capability is also diminished as altitude above
sea level increases.
SCOPE OF THE MANUAL
This manual contains instructions for installing, setting up,
programming, operating and maintaining the sampling
system. It is intended for use by technicians and operators
as well as maintenance personnel. The TROUBLESHOOTING
section contains a list of possible problems, their probable
causes, and actions to take to remedy each problem.
Instructions for replacing user-serviceable parts and a list of
parts that can be ordered are included.
The ACCESSORIES section lists all available accessories for
the Masterflex®E/S™Composite Sampler.
EXTERIOR
HOUSING
LEVEL
SENSOR
KEYPAD
ON/OFF 0-1
STATUS
SWITCH
EXTERNAL
POWER IN
RECEPTACLE
RS-232
CONNECTOR
SAMPLE
RESERVOIR
FLOW
DETECTOR
MASTERFLEX
EASY-LOAD
PERISTALTIC PUMP
CONTROL
CONSOLE
FIGURE 1

5
SETUP
UNPACKING
Care has been taken in the packaging of the Masterflex®E/S™
Composite Sampler to protect it in transit to its final destination.
After opening the carton, save the packing material until proper
product operation has been verified.
The Masterflex®E/S™Composite Sampler consists of the
following items:
(1) Sampling Unit
(1) AC/DC Universal Power Supply/Converter
(1) Sample Line Tubing Connector
(2) Keys for the locking latches
(1) 25 feet of Masterflex®L/S®24 size peroxide-cured
silicone tubing
(1) Operator's Manual CD
(1) Warranty Card
ASSEMBLING
Unit is shipped fully assembled and plumbed with Masterflex®
L/S® 24 size silicone peroxide-cured tubing. The operator only
needs to connect an external sample line using the supplied
tubing connector into Sample Line Inlet bulkhead connector
shown in Figure 2.
TUBING SELECTION
The flow rate at which a sample can be taken is dependent
on the tubing size, tubing material and the height the sample
must be raised to enter the sample reservoir. The purity of
the sample is dependent on the formulation of the tubing
material. The Composite Sampler is supplied with L/S®24
size tubing made of peroxide cured silicone resins. The tubing
supplied with the unit has been selected based on its purity
versus flow and lift performance. Optimum lift and flow
performance can be had by using Tygon®LFL or Lab tubing.
See Table 1 for other tubing materials and sizes available for
operation with the Composite Sampler.
INSTALLATION
SITE REQUIREMENTS
The E/S Composite Sampler is designed to operate in most
outdoor conditions. Ensure that the work area temperature
range is between 1°C to 50°C (39°F to 122°F) and that the
relative humidity remains between 10% and 90%.
Position the sampler in an upright position with the carrying
handle at the top of the case (as shown), preferably on a firm
level surface, adjacent to the source to be sampled. The
sampler may also be suspended or attached to a vertical
support using the D-rings at the top of the exterior housing.
POWER REQUIREMENTS:
The Masterflex®Composite Sampler is internally powered by a
12 V sealed gel battery. The unit may also be operated or
charged in an outdoor environment using the Automotive
Power Adapter Accessory 07571-50 Auxiliary Power Pack
07571-52 or 07571-54, or any DC source supplying 11-18 volts.
Sampler can be operated or charged indoors from a 115
(90-130) VAC or 230 (180-260) VAC 50/60 Hz incoming power
using the AC/DC Universal Power Supply/Converter (supplied).
DANGER: NEVER apply AC voltages directly
to the EXT PWR IN receptacle on
the front panel. The application of
AC voltages can result in injury
and death of the operator and
destruction of the unit. Use only
the AC/DC Universal Power Supply/
Converter supplied with the unit to
power unit from an indoor AC
source.Fully charge the unit before
using for the first time. Damage to
the internal battery can result if the
battery is fully discharged and
operation of the unit is attempted.
Tubing formulation L/S®15 L/S®24
Silicone, peroxide-cured 96400-15 96400-24
Tygon, Lab®06409-15 06409-24
Tygon LFL®06429-15 06429-24
PharMed®06485-15 06485-24
TABLE 1 Tubing Materials

6
DESCRIPTION
FUNCTIONAL DESCRIPTION
The control console of the Composite Sampler provides the
operator interface for programming, control and charging of the
sampling system. The DSP in the control console scans the
keypad, communicates to the display, communicates with the
integral pump drive, assesses the state of and controls
charging of the battery.
The system is preprogrammed with one (1) fluid transfer
program and five (5) sampling programs. All of the factory-
supplied sampling programs can be modified by the operator
and saved without modifying or destroying the factory
supplied programming. The factory supplied programming
and operational parameters are stored in non-volatile flash
memory within the DSP controller.
Program options and instructions from the operator are
stored in non-volatile memory consisting of an EEPROM
which supplements the flash memory of the DSP.
CONTROLS, INDICATORS AND
CONNECTORS
All the controls, connectors, and indicators on the controls
console and exterior housing are shown in Figure 2 and 3.
Table 2 lists all of the operator controls and indicators.
WETTED PARTS
All parts in direct contact with the sample being taken are
considered wetted parts. The following is a listing of the
wetted parts and their associated materials.
Sample intake tubing: Unit is supplied with 1⁄4" ID 25 feet of
Masterflex®L/S®24 size peroxide-cured silicone tubing.
Sample intake tubing connectors: Delrin®acetal co-polymer
Tubing: Masterflex®L/S®24 size peroxide-cured Silicone
tubing.
Flow detector: PVC and stainless steel
Flow detector tubing connectors: polycarbonate and Viton®
Discharge tubing: Masterflex®L/S®24 size peroxide-cured
silicone tubing.
Discharge tubing fittings: Delrin®acetal co-polymer
Sample reservoir: HDPE (high density polyethylene)
Sample reservoir cap: polypropylene
Breather vents, sample reservoir: medical grade ABS/Teflon®
SAMPLE
LINE INLET REMOTE CONTACT
CLOSURE
CONNECTOR
RS232
CONNECTOR EXTERNAL POWER IN
RECEPTACLE
11-18VDC
ON-OFF
0-1/ STATUS
SWITCH
FIGURE 2

7
Control or indicator Description and function
0-1/ STATUS Use to turn the unit ON, OFF, check unit status or awaken unit from an inactive mode
DISPLAY 24 character by two-line LCD that displays all menus needed for programming and operating
the system.
MENU Use to access the menu.
Selection arrow keys Use to navigate menu options shown on the display.
Numeric keypad 0-9 Use for numeric entry.
START/STOP Use to initiate program or to stop a program which is running.
CANCEL Cancels any changes that were made to the current menu selection and returns to previous menu
ENTER Accepts the selected item and moves down to the next menu level. If there is not a lower menu
level, it will return to the top level for the selected item.
CLOCK Used to set date and time. The date is displayed in a MM:DD:YY format. The time format is a
24hour clock displayed in a HH:MM:SS..
TIME Used access the time settings in the loaded program.
SAMP. VOL Used to access the sample volume parameter in the loaded program without going through the
menu.
LIFT HEIGHT Used to access the lift height parameter in the loaded program without going through the menu
CYCLE TEST Used to verify unit operation in the field.
CAL Used to verify and calibrate unit in the field for accurate sample delivery.
TABLE 2 Operator Controls and Indicators
FIGURE 3
FAMILIARIZATION WITH
SYSTEM FEATURES
The system has unique features with which the operator needs to
become familiar during the course of system operation. Understanding
the features will enable the operator to know if the system is
operating properly and effectively. These features are as follows:
1. Automatic Power-Up: Any time an external power source is
connected to the EXT POWER IN receptacle located on the front
of the control console, the system will power up and go through a
boot up routine or display a system message as to system status.
If no action is taken, the system will charge the battery.
2. Automatic Power Conservation: If the system is not imple-
menting an internal command or the operator has not pressed a
key on the control console for 10 minutes the system to conserve
power will POWER DOWN. The LCD display will display a message
for 5 seconds:
POWERING DOWN

8
After which the display will blank. If an external power source
is connected at the time, the battery will continue to charge
while the display is blank. If a sampling program was running
or in process at the time of power down the system will
continue to run the program. The display panel will show a
message when the unit starts to take a sample and show
the time until the next sample is to be taken. The display will
remain active for an additional 5-10 seconds before returning
to a powered down state.
If the display panel is not active, press the 0-1/STATUS key
on the control console to awaken the system.
3. Automatic Battery Charging: System will automatically
charge the internal battery any time an external power source
such as the AC/DC Universal Power Supply/Converter or
Automotive Power Adapter is plugged into the EXT PWR IN
receptacle. The rate at which the internal battery is charged
will vary with other system operations. Once the battery is
fully charged, the integral battery charging circuit will place
the battery on a floating or maintenance charge to maintain
the battery in a fully charged state.
4. Automatic Purging: The system will automatically purge
the sample intake line prior to the taking of any sample or the
occurrence of a rinse cycle. This is to ensure that the sample
inlet line is free of any obstructions or debris.
5. Glow-In-Dark Keypad: The keypad is printed with
phosphorescent pigments. These pigments will glow for up
to 20 minutes if exposed to ultraviolet radiation from the sun
or a krypton flashlight. This feature is helpful in low ambient
light conditions. The LCD display is backlightable and when
used with keys that glow in the dark, this lessens the need for
external lighting in dim and dark areas.
INITIAL OPERATION
SYSTEM BOOT—INTERNAL BATTERY
Press the 0-1/STATUS button on the control console. The
LCD display will activate and display the following messages:
System will power up the drive system to place the internal
battery under load and verify its condition.
The LCD panel will display one of the following messages,
depending upon battery condition.
If the battery is fully charged or you pressed cancel when
notified the battery is low, the system will complete boot-up
and display the following on the LCD screen:
If the message displayed is:
The system should not be operated until the battery has been
charged or an external power source has been connected to
the unit. Proceed to the BATTERY CHARGING section of the
manual on page 10.
SAMPLER ON VERSION X.X
CHECKING BATTERY
PRESS CANCEL TO QUIT
BATTERY FULLY CHARGED
BATTERY IS LOW
CHARGING IS RECOMMENDED
SAMPLER SYSTEM READY
HH:MM:SS MM/DD/YYYY
BATTERY IS VERY LOW
MUST CHARGE BATTERY!

SYSTEM BOOT—
EXTERNAL POWER
Open the exterior housing, connect the AC/DC Universal
Power Supply/Converter into the EXT PWR IN receptacle on
the front. Then connect the AC plug into an AC receptacle.
The system will come on automatically. The first screen dis-
played on the LCD panel will be:
This will be followed by another screen:
Remove the plug from the EXT PWR IN receptacle on the
control console. LCD will change to read:
The system will power up the drive system to place the
internal battery under load and verify its condition. After the
drive system has stopped, you may reconnect the external
power to the EXT PWR IN receptacle. The LCD panel will
display one of the following messages, depending upon
battery condition.
If the battery is fully charged or you pressed cancel when
notified the battery is low, the system will complete boot-up
and display the following on the LCD screen:
If the message displayed is:
The system should not be operated until the battery has been
charged. Proceed to the BATTERY CHARGING section of the
manual on page 10.
SAMPLER ON VERSION X.X
REMOVE EXT PWR TO CHECK
BATTERY—CANCEL TO SKIP
CHECKING BATTERY
PRESS CANCEL TO QUIT
BATTERY IS FULLY CHARGED
BATTERY IS LOW
CHARGING IS RECOMMENDED
BATTERY IS VERY LOW
MUST CHARGE BATTERY!
SAMPLER SYSTEM READY
HH:MM:SS MM/DD/YYYY
9

10
SAMPLER SYSTEM READY
HH:MM:SS MM/DD/YYYY
BATTERY CHARGING
The Composite Sampler is equipped with an internal four
state battery charger capable of diagnosing and automatically
charging the internal battery. The operator only needs to
connect the AC/DC Universal Power Supply/Converter or
Automotive Power Adapter accessory. The sampler system
will automatically power up and begin charging the battery.
After 10 minutes the LCD display on the control console will
power down, going blank. The battery will continue to charge
until it has reached is full potential. Once the battery is fully
charged, the battery will be placed on a float or maintenance
charge until the external power source is disconnected.
The system may run or be programmed while the unit is
charging. Battery charging will continue in the background
while other operations are taking place. Allow a minimum of
8 hours to fully charge the battery from a fully discharged
state with no other operations occurring. During battery
charging all functions of the sampler are operational.
WARNING! The AC/DC Power Supply/Converter
is rated for INDOOR USE ONLY.
DO NOT use the AC/DC Power
Supply/Converter in an outdoor
environment to either charge the
battery or power the drive. Electrical
shock, severe injury and/or death is
possible if this warning is ignored.
SETTING THE TIME AND DATE
Press the keypad's CLOCK key. The display will read:
TIME
The clock used in the Composite Sampler is a 24-hour clock,
does not use an AM or PM notation and shows the time in
an HH:MM:SS format. The clock displays 12:00 AM midnight
as 00:00:00 (the starting point of the clock), 6:00:00 AM as
06:00:00, 12:00:00 PM (or noon) as 12:00:00 and 6:00 PM
as 18:00:00.
If a sample was taken at 3:00 PM the clock will read 15:00:00.
The Composite Sampler clock was set at the factory based
upon our local time. The time is expressed as hour, hour,
minute, minute, second, second. If the current time on the
clock is correct, bypass setting the time by pressing ENTER
or CANCEL and you will be taken to the screen where you
may change the date. See the screen below.
If you need to correct the time shown on the clock, use the
numeric keypad to enter the current time and press ENTER.
The display will change to the date correction screen as
shown below.
DATE:
The calendar date is expressed as month, month, day, day,
year, year.
If the date is correct, press ENTER or CANCEL and you will
be taken back to the system prompt.
If you need to correct the date shown, use the numeric
keypad, enter the date and press ENTER.
EXAMPLE: To enter January 1, 2003, press 10103 on the
numeric keypad and press ENTER.
To enter December 31, 2002, press 123102 on
the numeric keypad and press ENTER.
You will then be taken back to the system prompt screen
which will now display the current time and date.
SECURITY
The Composite Sampler will be exposed to an outdoor
environment in which the operator may or may not be
present or in attendance. To protect the unit from tampering
by unauthorized individuals the sampler housing may be
locked using the two locking latches on the exterior of
the housing.
In addition a security feature has been incorporated into
the Composite Sampler to lock the keypad.
SET TIME (HR:MIN:SEC)
20:30:57 (23:59:59)
SET DATE (MON/DAY/YR)
09/21/02 (MM/DD/YY)

11
LOCKING THE KEYPAD
The Composite Sampler keypad may be locked to ensure
that running programs are not inadvertently terminated or to
restrict access to the system. The only key which is functional
when the keypad lockout is enabled is the 0-1/STATUS key.
Pressing this key will not disable or interfere with any program
which is running. To lock the keypad, perform the following
procedure either from a system prompt or after you have
pressed the START/STOP key to start a program. The keypad
lockout cannot be enabled or disabled during programming
or the editing of a program.
1. Ensure that the control console is in a system ready prompt
or a program is running.
2. Enter 847 on the numeric keypad and press ENTER.
3. The following display appears for approximately five
seconds. The display reappears anytime a non-numeric
key is pressed and the LCD display is active.
4. To unlock the keypad, enter the same code and press
ENTER. The following display appears:
OPERATION
The unit is operated through the use of a menu system and
various hot buttons on the keypad located on the control
console face. The menus and their options are shown as words
and numbers used in everyday oral communications or
speech without the need for decryption or an additional legend.
This equipment should not be operated on a "learn as you
go" basis. Make sure to become familiar with and practice all
safety procedures presented in the SAFETY section before
operating the Composite Sampling System.
DESCRIPTION OF PROGRAMMING
The Composite Sampler can be operated using any of the
six (6) factory supplied programs through the menu system.
The five (5) sampling programs can be modified or edited and
stored for future use. Editing the factory supplied programs
and saving them as modified programs will temporarily
overwrite the factory program. The factory programs can be
individually restored to their original default settings upon
operator demand through the menu system.
NOTE: One program is always present or selected to run
whenever the unit is on. When received from the factory the
default program selected to be run is PROGRAM #1.
STARTING/STOPPING A PROGRAM
To start or stop a program press the START/STOP key.
Programs can only be started from a SAMPLER SYSTEM
READY prompt. Sampling programs can be stopped at any
time the display is active by pressing CANCEL. If the display
is inactive, press the 0-1/STATUS key to awaken the system
and then press CANCEL. The CONTINUOUS PUMPING
program can be stopped at any time. Stopping a program is
equivalent to canceling the program. Programs that have
been stopped cannot be restarted from the point at which
they were stopped. When you start the program again, it will
begin at the beginning of the program.
HINT: If the program you stopped was taking a sample, it
may be necessary to empty or clear the sample reservoir
prior to starting the program again.
CANCELING A PROGRAM
To cancel a program that has started, press the CANCEL
key. Canceling a program is the same as stopping a program
as described above. Pressing the CANCEL key does not
deselect the program selected to be run. It only cancels the
program schedule that has started.
HINT: The program will show up on the data log as a cancelled
program. If you do not want this record to appear on the data
log, clear the data log prior to starting the program again.
ACCESSING THE MENU
Operation of the system and editing programs is menu driven.
To access the menu press the MENU button on the control
console. The LCD display will show the following.
KEYPAD LOCKOUT ENABLED
KEYPAD LOCKOUT DISABLED
gGLOBAL SETTINGS
cVIEW/EDIT PROGRAMS

12
MENU
The menu is divided into two main sections or categories: GLOBAL SETTINGS and VIEW/EDIT PROGRAMS.
GLOBAL SETTINGS
This section consists of operational parameters related to the
Composite Sampler system as follows:
• DISPLAY CONTRAST
• SAMPLING LOCATION NUMBER
• PUMP HEAD SELECTION
• TUBING SIZE SELECTION
• TUBING FORMULATION SELECTION
• RINSE CYCLES
• VOLUME UNITS
• DISTANCE UNITS
• LIFT HEIGHT
• TUBING LENGTH
• LCD BACKLIGHTING
• DATA LOGGING
• DEFAULT GLOBAL SETTINGS
VIEW/EDIT PROGRAMS
The VIEW/EDIT PROGRAMS section consists of parameters
related to the sampling programs as follows:
• SELECT HOW PROGRAM IS STARTED
• SAMPLE VOLUME
• NUMBER OF SAMPLES
• TIME INTERVAL BETWEEN SAMPLES
EQUAL OR UNEQUAL
• # OF CONTACT CLOSURES
• DEFAULT THE PROGRAM SETTING
• SAVE USER PROGRAM
See Table #3 for menu categories and selections.

TABLE #3 MENU SELECTIONS
Defaults are shown in BOLD type or noted
13
CATEGORY SELECTION CHOICES SUB SELECTIONS
DISPLAY CONTRAST UP & DOWN ARROWS TO ADJUST
LOCATION NUMBER ENTER 0-99999999
PUMP HEAD Easy-Load/Standard SIZE 24 TYGON LFL
TYGON LAB
SIZE 15 SILICONE TUBING
PHARMED TUBING
PTFE TUBING PUMP PTFE 4 MM
PTFE 6 MM
RINSE CYCLES ENTER NUMBER 0-5 Default is 0
VOLUME UNITS MILLILITERS
GALLONS
DISTANCE UNITS FEET
GLOBAL METERS
LIFT HEIGHT ENTER NUMBER Default is 6 feet or 1.8 meters
TUBING LENGTH ENTER NUMBER Default is 12 feet or 3.6 meters
LCD BACKLIGHTING
NONE—POWER SAVER
BACKLIGHT ON KEY PRESS
KEY PRESS AND SAMPLING
CONTINUOUS BACKLIGHT
LOG DATA
VIEW LOG DATA
PRINT LOG DATA
CLEAR LOG DATA CLEAR ALL LOG DATA
DON'T CLEAR LOG DATA
DEFAULT GLOBAL SETTINGS RESET GLOBAL SETTINGS
DON'T RESET SETTINGS
PROGRAM START DELAYED START HHH:MM (0-167.59)
REMOTE & DELAYED START HHH:MM (0-167.59) MAINTAINED CONTACT
PULSED CONTACT
SAMPLE VOLUME ENTER NUMBER (1-9,000)
NUMBER OF SAMPLES ENTER NUMBER (1-99)
EQUAL INTERVALS HH:MM (0-99:59)
HH:MM (0-99:59)
#1 HH:MM (0-99:59)
#2 HH:MM (0-99:59)
#3 HH:MM (0-99:59)
#4 HH:MM (0-99:59)
#5 HH:MM (0-99:59)
VIEW/EDIT TIME BETWEEN SAMPLES UNEQUAL INTERVALS #6 HH:MM (0-99:59)
PROGRAMS #7 HH:MM (0-99:59)
#8 HH:MM (0-99:59)
#9 HH:MM (0-99:59)
#10 HH:MM (0-99:59)
#11 HH:MM (0-99:59)
#12 HH:MM (0-99:59)
# OF CONTACT CLOSURES ENTER NUMBER Only appears if MAINTAINED is not
(0-9999) Default is 0 chosen under PROGRAM START/REMOTE & DELAYED START
DEFAULT PROGRAM RESET PROGRAM SETTINGS
SETTINGS DON'T RESET SETTINGS
USER PROG #1
USER PROG #2
SAVE USER PROGRAM USER PROG #3
USER PROG #4
USER PROG #5

14
NAVIGATION
Looking at the LCD display screen below:
1. The MENU key takes you to the top level of the menu as
shown above.
2. The arrows in the first display column indicates which way
the user can go from each menu selection.
3. Solid box in second column indicates selected menu item,
empty box is an unselected menu item.
NOTE: Only one selection is allowed in each menu option.
4. UP/DOWN arrow keys on the control console move the
solid box to select a menu item and scroll the display if
more menu items are available.
5. The ENTER key accepts the selected item and moves
down to the next menu level. If there is not a lower menu
level, it will return to the top level for the selected item. For
example, if selecting a tubing material, pressing ENTER will
return back to the PUMP HEAD menu selection.
6. CANCEL key cancels any changes that were made to the
current menu selection and returns to previous menu.
7. CANCEL key will back up to previous menu level and exit
the menus when at the top level.
8. Display is 2 lines by 24 characters so only 2 lines of the
menu are seen at one time.
GLOBAL SETTINGS
GENERAL
The GLOBAL SETTINGS are not saved in memory like
multiple sampling programs. The only settings available are
those currently loaded into the non-volatile memory.
Selecting a menu option or setting using the up and down
arrow keys does not load that option or setting in memory
until the ENTER key is pressed.
Settings must be modified as necessary to suit the sampling
site and equipment used. Many of these settings may use the
default values, such as pump head selection, tubing size or
formulation. Other parameters such as lift height and tubing
length or location will vary with the sampling site and will be
changed frequently.
GLOBAL SETTINGS TOUR
The tour of the GLOBAL SETTINGS menu will take the operator
through each of the menu selection display screens. Going
through the entire menu is not necessary each time a change
to one or two GLOBAL SETTINGS parameters is made. It is
recommended that you take the tour to familiarize yourself
with all of the options and screens available.
Select GLOBAL SETTINGS by pressing the ENTER key.
The display will now show the following choices:
DISPLAY CONTRAST
Allows adjustment of the display contrast for readability.
ADJUSTING THE DISPLAY
CONTRAST
To adjust the contrast of the LCD display, select DISPLAY
CONTRAST and press the ENTER key. The following menu
will be displayed:
View the display and use the up and down arrow keys to
adjust the contrast. Press the up arrow key to increase
contrast or the down arrow key to decrease contrast. When
you have achieved the desired contrast, press the ENTER
key. This will save the contrast setting you have selected
for the display. The system will return you to the following
display.
DISPLAY CONTRAST
g DISPLAY CONTRAST
cLOCATION NUMBER
gGLOBAL SETTINGS
c VIEW/EDIT PROGRAMS
gDISPLAY CONTRAST
c LOCATION NUMBER
gGLOBAL SETTINGS
c VIEW/EDIT PROGRAMS

15
Use the down arrow to select LOCATION NUMBER, the filled
in box will switch from DISPLAY CONTRAST to LOCATION
NUMBER and press ENTER. The display will change to read:
LOCATION NUMBER
Allows the input of up to a 9 digit number to identify the
sampling site. You do not have to erase or delete the number
(site identifier) that may already be there.
INPUTTING A LOCATION NUMBER
Enter the number and press ENTER. The new site
identification number will be entered and the display will
change to:
Use the down arrow to decrement the menu to and select
PUMP HEAD by pressing ENTER.
The display changes to read:
PUMP HEAD
Multiple option screens which allow the operator to select the
pump head, tubing size and tubing formulation or material.
EASY-LOAD is the default pump head supplied with the
Composite Sampler for your convenience in changing the
tubing when needed. If you intend to use the pump head
supplied with the unit press ENTER.
If you wish to use a PTFE TUBING PUMP proceed to the
description of this menu option on page 16.
The display will change to read:
Size 24 is the default tubing size. The tubing supplied with
the Composite Sampler is L/S®24 peroxide cured silicone
Model No. 96400-24. This tubing size will deliver an average
flow of 1,700 ml/min. @ 0 lift.
Size 15 is an alternate tubing size offered for use with the
Easy-Load®pump head supplied with the Composite
Sampler. This size option is offered for circumstances where
the use of L/S® 24 in certain formulations may not be practical.
The average flow of L/S® 15 tubing is 1,000 ml/min. @ 0 lift.
Select the appropriate tubing size and press the ENTER key.
The display will change to the tubing formulation menu.
NOTE: the display is only 2 lines, so you will only see 2
choices at one time. Remember the arrow in the first column
indicates whether there are choices above or below the ones
you are viewing.
The tubing formulations shown are available in either
SIZE 24 or SIZE 15. See Table 1 Tubing Materials in the
SETUP section of this manual for tubing model numbers.
LOCATION NUMBER
_________(0-999999999)
c DISPLAY CONTRAST
gLOCATION NUMBER
c LOCATION NUMBER
gPUMP HEAD
g EASY-LOAD/STANDARD
cPTFE TUBING PUMP
g SIZE 24
cSIZE 15
cTYGON LFL TUBING
cTYGON LAB TUBING
gSILICONE TUBING
cPHARMED

16
PTFE TUBING PUMP
This pump head is offered as an option for those sampling
applications such as toxic fluids and oily waste water where
only the use of Teflon“ tubing will perform adequately for the
application.
Use the up and down arrow keys to navigate and select the
PTFE TUBING PUMP and press ENTER. The display will
change to show the tubing sizes available for use with the
PTFE TUBING PUMP.
Selection of the tubing formulation for sizes 24 and 15 or the
selection of the PTFE tubing sizes will return the display to
read:
Use the down arrow keys to navigate and select RINSE
CYCLES and press ENTER. The display will change to read:
RINSE CYCLES
This sets the number of rinses (0-5) that the Composite Sampler
will perform prior to the taking of each sample. The default is
zero (0). The purpose of the rinse cycle is to remove or rinse
away any contaminants or residues in the sample intake lines
from a previous sampling site or cycle. The tubing will first be
purged of any fluid. Then fluid from the sampling site will be
drawn into the sample intake lines until the fluid is detected by
the flow sensor. The system will then stop and reverse the drive
to expel the fluid. This completes one rinse cycle.
To accept the default setting press either CANCEL or ENTER.
To change the number of cycles, enter the number of cycles
desired and press ENTER.
The display will change to read:
Use the down arrow key to navigate and select VOLUME
UNITS and press ENTER. The display will change to read:
VOLUME UNITS
Allows the user to select the units in which the sample will be
measured. The default is MILLILITERS. The two options are:
MILLILITERS
Displays the sample volume from 1-9000 milliliters.
GALLONS
Displays the sample volume from 0.001 to 2.377 gallons.
To accept the default press either CANCEL or ENTER. To
change the VOLUME UNITS from MILLILITERS to GALLONS,
use the arrow key to navigate and select GALLONS and press
ENTER. The display will change to read:
Use the down arrow key to navigate and select DISTANCE
UNITS and press ENTER. The display will change to read:
c PTFE 4 MM
c PTFE 6 MM
g PUMP HEAD
c RINSE CYCLES
RINSE CYCLES
0 (0-5)
c EASY-LOAD/STANDARD
g PTFE TUBING PUMP
c RINSE CYCLES
g VOLUME UNITS
g MILLILITERS
c GALLONS
g VOLUME UNITS
c DISTANCE UNITS
g FEET
c METERS

17
DISTANCE UNITS
Allows the user to select English or metric distance measurement
units. The default is feet.
The two options are:
FEET
Displays the VERTICAL LIFT (1-25) and TUBING LENGTH
(1-99) in whole feet.
METERS
Displays the VERTICAL LIFT (0.3-7.6) and TUBING LENGTH
(.3 to 30.1) in tenths of a meter.
To accept the default press either CANCEL or ENTER. To
change the DISTANCE UNITS from FEET to METERS, use the
arrow key to navigate and select METERS and press ENTER.
The display will change to read:
Use the down arrow key to navigate and select LIFT HEIGHT
and press ENTER. The display will change to read:
LIFT HEIGHT
Sets the draw or lift height to which the sample must be lifted
to enter the sample reservoir.
This parameter allows the system to accurately calculate the
sample volume for each revolution of the pump head at a
specific height. Sample volume decreases as lift height
increases. This parameter will be expressed in feet or meters
depending upon your earlier choice of DISTANCE UNITS. The
default is 6 feet or 1.8 meters. Enter the number and press
enter. The acceptable values will be shown in parenthesis on
the 2nd line on the right side of the display.
Once you press enter the display will change to:
Use the down arrow key to navigate and select TUBING
LENGTH and press ENTER. The display will change to read:
TUBING LENGTH
Sets the length of tubing through which the sample volume
must be drawn. Setting the tubing length allows the system
to calculate the correct amount of time for purging the tubing
prior to taking a sample. This parameter will be expressed in
feet or meters depending upon your earlier choice of DISTANCE
UNITS. The default is 12 feet or 3.6 meters. Enter the number
and press enter. The acceptable values will be shown in
parentheses on the second line on the right side of the display.
Once you press enter the display will change to:
Use the down arrow key to navigate and select LCD BACK-
LIGHT and press ENTER. The display will change to read:
NOTE: the display is only 2 lines, so you will only see 2
choices at one time. Remember the arrow in the first column
indicates whether there are choices above or below the ones
you are viewing.
LCD BACKLIGHT
Sets the conditions under which the display is backlit for
optimum readability. There are four choices of backlighting.
LIFT HEIGHT
6 FEET (0-25)
g LIFT HEIGHT
c TUBING LENGTH
g DISTANCE UNITS
c LIFT HEIGHT
TUBING LENGTH
12 FEET (1-99)
g TUBING LENGTH
c LCD BACKLIGHT
c NONE—POWER SAVER
g BACKLIGHT ON KEY PRESS
c KEY PRESS AND SAMPLING
c CONTINUOUS BACKLIGHT

18
NONE—POWER SAVER
The display will not be backlit under any circumstance. This
is the default setting and is used to conserve power. If the
LCD display can be read without difficulty in the sampling
environment, then this is adequate.
BACKLIGHT ON KEY PRESS
The LCD will be backlit for 10 seconds when any key is
pressed. After 10 seconds the display will return to its normal
unlit state. This setting can be useful during programming
since it improves the readability of the display without wast-
ing power and provides backlighting during programming or
selecting menu options in low ambient light conditions.
KEY PRESS AND SAMPLING
The LCD will be backlit for 10 seconds when any key is
pressed or whenever a sample is to be taken. Ten (10)
seconds after the last key is pressed or after the sampling
cycle is completed the display will return to its normal unlit
state. This setting can be useful during programming since it
improves the readability of the display and allows monitoring
of the sampling process in low ambient light conditions.
If the Composite Sampler will be left unattended during the
sampling process this setting is not recommended.
CONTINUOUS BACKLIGHT
The LCD will always be backlit as long as the display is active.
HINT: The system's automatic power conservation feature blanks
the LCD display if no key has been pressed for 10 minutes.
Blanking of the LCD also occurs 10 seconds after the system
has taken a sample or processed an internal instruction while
running a program. See the FAMILIARIZATION WITH
SYSTEM FEATURES section on page 7.
It is recommended that you experiment with the various
backlighting options to determine your preference. You can
always return to the menu option by pressing enter once you
have made your selection and pressed enter, because the
display will change to read:
Proceeding down the menu, use the down arrow key to
navigate and select the next menu option LOG DATA and
press ENTER. The display will change to read:
HINT: the display is only 2 lines, so you will only see 2
choices at one time. Remember the arrow in the first column
indicates whether there are choices above or below the ones
you are viewing.
LOG DATA
The system logs all events, such as: sample location, cycle
date and time, sample volume, and any faults which may
occur during the sampling process. The system is capable of
logging 300 events. This log is stored in non-volatile memory
outside of the DSP. This log may be viewed or printed.
Viewing or printing the log does not erase or clear the log
from memory. The log may only be erased through the use of
the CLEAR LOG DATA menu option.
NOTE: should you forget to clear the logged data, the system
will begin to overwrite the oldest data first. This occurs only
once the capacity of the non-volatile memory log has been
exceeded.
VIEW LOG DATA
Allows the user or operator to view the data log. This does
not erase or clear the data log.
This is the default option for LOG DATA.
PRINT LOG DATA
Allows the printing of the data log. This menu option requires
the use of the RS 232 cable accessory catalog no. 07580-65.
See the ACCESSORY section for details. Printing of the data
log does not erase or clear the data log.
c LCD BACKLIGHT
c LOG DATA
c VIEW LOG DATA
c PRINT LOG DATA
c PRINT LOG DATA
c CLEAR LOG DATA

19
CLEAR DATA LOG
Allows the operator to clear the data log. This is the only way
the data log may be cleared or erased. Once you have
selected the menu option and pressed ENTER, you will be
taken to a confirmation screen as shown below:
The default selection will not clear the data log. You must use
the up arrow to select "CLEAR ALL LOG DATA" and press
ENTER. Once the ENTER key has been pressed, the log will
be erased and the display will change to read:
Use the down arrow key to navigate and select DEFAULT
GLOBAL SETTINGS and press ENTER. The display will
change to read:
DEFAULT GLOBAL SETTINGS
Allows the operator to return all of the GLOBAL SETTINGS to
their default values.
If you have made changes to the GLOBAL SETTINGS and
lost track of the changes you have made or would just like to
start your changes from some common point, this option will
prove to be helpful. This option allows resetting of the
GLOBAL SETTINGS without having to track your changes or
remember what the default was.
To reset the GLOBAL SETTINGS you must use the up arrow
key to navigate and select RESET GLOBAL SETTINGS and
press ENTER. This confirmation of your selection to return
the settings to their default value is done to prevent accidental
or unintended changes to the global settings. Once you press
the enter key your previous changes will be lost and cannot
be recovered.
HINT: To exit the menu and return to a system ready prompt
use the cancel key.
HOT KEY
One option of the GLOBAL SETTINGS menu is also accessible
via a "Hot Key" located on the control console keypad. The
key is round, blue in color and labeled LIFT HEIGHT. This key
allows the operator to access the lift height setting in the
GLOBAL SETTINGS without going through the menu.
VIEW/ EDIT PROGRAMS
GENERAL
There are a total of 6 programs supplied with the Composite
Sampler. One program is a fluid transfer program called
CONTINUOUS PUMPING, this program cannot be modified.
The other 5 programs are fluid sampling programs. These
programs are numbered 1 through 5. All programs can be
accessed through the menu system of the Composite
Sampler. The 5 sampling programs can be edited to create
your own sampling programs. Edited or modified programs
must be saved before they can be run. Modifying the factory
supplied programs and saving them as modified programs
will temporarily overwrite the factory program. The factory
programs can be individually restored to their original default
settings upon demand.
Only (1) one program can be loaded and used at a time. A
program is always loaded into memory of the Composite
Sampler.
PROGRAM MENU TOUR
Operation of the system is menu driven. To access the menu,
if not presently displayed, press the MENU button on the
control console. The LCD display will show the following.
Use the down arrow key to navigate and select VIEW/EDIT
PROGRAMS and press ENTER. The display will change to read:
HINT: the display is only 2 lines, so you will only see 2
choices at one time. Remember the arrow in the first column
indicates whether there are choices above or below the ones
you are viewing.
c CLEAR ALL LOG DATA
gDON’T CLEAR ALL LOG DATA
g LOG DATA
c DEFAULT GLOBAL SETTINGS
c RESET GLOBAL SETTINGS
g DON’T RESET SETTINGS
g GLOBAL SETTINGS
c VIEW/EDIT PROGRAMS
g PROGRAM #1
c PROGRAM #2

20
Navigate further into the program options by using the ⁄ down
arrow key.
LOADED PROGRAMS
One of the available programs is always loaded into the
memory for use. How do you verify which program is loaded
for operation of the Composite Sampler? To verify which
program is loaded for operation of the unit, access the menu
by pressing the MENU key and selecting VIEW/EDIT
PROGRAMS, then press the ENTER key.
The display will change to show 2 program options. One of
these options will show a darkened or filled in box next to the
program name. This is the program which is loaded for operation.
See the example below.
In this case PROGRAM #3 is loaded for operation as shown
in the above example.
CONTINUOUS PUMPING
This program is for fluid transfer in the field. The pump will
start when the START/STOP button is pressed or the REMOTE
CONTACT CLOSURE is closed. The pump will stop when
the same button is pressed again, the REMOTE CONTACT
CLOSURE opens or the fluid level sensor in the sample
reservoir causes the system to stop because the reservoir
is full. This program cannot be modified in any manner.
This program selection is not visible when you first access
the program menu. Select VIEW/EDIT PROGRAMS and press
ENTER. The display will change to read as follows.
The arrow at the far left of the display on line 1 shows a
bi-directional arrow, indicating there is an available option
above PROGRAM #1.
Use the up arrow key to navigate upward in the menu. The
display will change to read:
You may now select CONTINUOUS PUMPING by pressing
the ENTER key.
c PROGRAM #2
c PROGRAM #3
c PROGRAM #3
c PROGRAM #4
c PROGRAM #4
c PROGRAM #5
g PROGRAM #3
c PROGRAM #4
g PROGRAM #1
c PROGRAM #2
g CONTINUOUS PUMPING
c PROGRAM #1
Table of contents