User Manual
MEAS LT Level & Pressure Transducer User Manual
Measurement Specialties, Inc. www.meas-spec.com 1-757-766-1500
1000 Lucas Way, Hampton, VA 23666 USA 6 of 31 1-800-745-8008
Chemical resistance of polyurethane:
Potable Water, Wastewater, Borax, Butane, Animal Fat, Carbonic Acid, Citric Acid, Cod Liver Oil,
Corn Oil, Glycerin, Glycol, Mineral Oils, Potassium Nitrate, Potassium Sulfate, Silicone Oils, Stoddard
Solvent, Tannic Acid (10), Tartaric Acid, Turbine Oil.
Chemical resistance of ETFE:
Acetic Acid (Glacial), Acetic Anhydride, Acetone, Aluminum Chloride, Anti- Freeze, Bromine, Calcium
Chloride, Calcium Hydroxide, Chlorine, Copper Chloride, Ferrous Chloride, Hydrochloric Acid,
Ketones, Lacquer Thinners, Sulfuric Acid.
The vented cable termination end is specially prepared at the factory to eliminate the potential for
moisture migration. Where the lead wires emerge from under the jacket, there’s potting material and
a shrink tube “boot”, every effort should be made to leave this feature intact. Should the cable be
longer than needed for the installation, it is recommended that the excess length be accommodated in
a service loop and that the potted end of the cable NOT be shortened.
The cable attached to this instrument is specifically engineered for submersible applications. The
polyurethane outer jacket provides long term reliability under most conditions. The cable should be
handled carefully, however, as the jacket may be subject to cutting should it be “raked” over extremely
sharp edges. To guard against water incursion should an inadvertent minor cut occur, we have
incorporated an exclusive “water block” feature immediately beneath the jacket. The cable is fully
shielded, with the shield connected to the metal housing at the transducer end and terminated in a
drain wire at the termination or user end. The shield should always be terminated to a good earth
ground, unless the transducer is installed in an area where galvanic corrosion is known to be a
serious problem.
Lightning Protection
Lightning and surge protection for VDC and 4-20 mA output are available. This is achieved through
the use of 2 protectors, one is integral to the transducer housing and one is provided for the outside
line located at the surface and grounded to a DIN-Rail. Please remember this option must be factory
installed at the time of order entry or as a factory upgrade. This option cannot be installed in the
field. Lightning protection is also available for unit only. Contact customer service for details.
Featuring quick response and low clamping voltages, these devices protect against fast rising voltage
transients as well as severe current surges associated with lightning discharges up to 20,000
amperes. Following a surge, the protector automatically restores the line to normal operation and
awaits the next surge without having to reset a breaker or replace a fuse. The power supply needs to
be limited to 150 mA to avoid lock up of the gas tube after a suppression event. If a lock-up should
occur, this is reset by disconnecting and reconnecting power to the circuit.
Transducers installed with this option have a two year warranty against damage due to voltage
surge.
Please note: When using the lightning protection option on 4-20 mA products, users should
take into account the additional series resistance of this option when selecting the loop power
supply. This option will increase total loop resistance by 88 Ohms, so the supply voltage must
be at least 12VDC when lightning protection is installed.