DIGITAL MICRO OHMS METER Model: 53C-II
Connect sample conductor in 4 terminal connection as follows:
-I & +I = Current Leads , -V & +V =, Potential Leads.
L= the instrument indicates resistance of conductor in Length L, the distance between Potential leads should be kept 1metre for
convenience.Keep current leads well away from potential leads.
Take care that clip -I should NOT touch-V &+V directly except through conductor.
2. Measurement of Resistance of Transformers, Motors etc.
Connect leads in 2 or 4 terminal arrangement. Switch ’ON’ power. Change range if Required. Wait for a few seconds for a stable
reading. Switch off power and disconnect leads and change over to next measurement. Do not disconnect leads when instrument
is ‘ON’.
3. Calculation of Heater Resistance.
V2/W = R..................................................... where V = Heater operating voltage
W = Wattage of heater
R = Resistance of heater.
4. Measurement of Temperature Rise.
Of copper windings by increase in resistance (IS-9678-1980).
T2+235 R2
T1+235 R1 where
T2= Temperature in Deg. C of winding at the end of the test.
R2= Resistance of winding at the end of the test.
T1= Temperature in Deg. C of winding (cold) at the moment of initial resistance measurement, and
R1= Initial resistance of winding.
5. Instrument Adjustment & Calibration.
Test instruments required:
1. Std. 200mV DC meter of 4 ½ digit of known accuracy.
2 .Std. resistance of 1m Ohms,10m Ohms, 100m Ohms, 1 Ohm,10 Ohms, 1K Ohm & 10K Ohms
Procedure:
Connect 10 Ohms std resistor to the instrument. Switch ‘ON’ Power. Clip Std voltmeter across clips -V & +V. Adjust test current by
preset P5 so that Std Voltmeter reads 75mV. Adjust instruments indication by preset on DPM to read value of std 10 Ohms resistor.
Change range to 1.999 Ohms, Connect 1 Ohm std resistor. Adjust test current by adjusting preset P4 so that the instrument
displays value of std 1 Ohms resistor. Set all other ranges likewise. All ranges must be adjusted by connecting resistors in 4
Terminal connection. Please recheck calibration using the same resistors. Please send this manual with the instrument whenever
it is sent to a laboratory for re calibration. Recommended re calibration cycle is 1 Year.
6. Operational Safety.
Extreme care must be taken while calibrating the instrument as there are dangers of high voltage inside the instrument.
-I -V +V +I