
Electrical safety
3
zDo not move the device after the technical support have installed
and debugged the device. Incorrect operation will result in inaccurate
experiment result. If you require to re-position the device, contact the
technical support.
zOnly trained professionals such as doctors, technicians, or laboratory
assistants can operate the device.
zDo not operate the device in the presence of flammable or explosive
liquids, vapors or gases. Otherwise, it might result in device malfunction.
zDo not disconnect the power cord under the power-on status.
zDo not place tubes or plate on the device. Liquids seeping into the device
might cause malfunction.
zDo not reuse the disposable items.
zOnly the peripheral devices and consumables specified by the manufacturer
can be used.
zIf malfunctions related to fluidics lines (for example, bubble) occur during
an experiment, solve the problems before you restart an experiment.
zIf you have maintenance questions that are not mentioned in this manual,
consult the technical support.
zOnly the components provided by the manufacturer can be used for
device maintenance. Unapproved components might damage the device or
degrade performance.
zThe device is verified before delivery. If serious deviation occurs during use,
contact the technical support for calibration.
Electrical safety
zBefore initial use of the device, assess the electromagnetic environment in
which the device will be used.
zDo not use the device in close proximity to the sources of strong
electromagnetic fields, such as unshielded sources of radiated emissions.
Radiated signals can reduce the accuracy of the results.
zEnsure that the device is properly grounded, and the grounding resistance is
less than 4
Ω
. Failure to do so might result in inaccurate experiment results
or even electric shock.
zDo not remove the device cover and expose the inner component outside.
Otherwise, it might cause electric shock.