manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Miele professional
  6. •
  7. Rack & Stand
  8. •
  9. Miele professional Insert E 905/1 User manual

Miele professional Insert E 905/1 User manual

Popular Rack & Stand manuals by other brands

Dynex DX-MP3MT Quick setup guide

Dynex

Dynex DX-MP3MT Quick setup guide

Konig & Meyer 24185 manual

Konig & Meyer

Konig & Meyer 24185 manual

ECG LD 326501 instruction manual

ECG

ECG LD 326501 instruction manual

Philips SB7S19S/00 quick start guide

Philips

Philips SB7S19S/00 quick start guide

Woodline KIT-250 lb instructions

Woodline

Woodline KIT-250 lb instructions

ricoo S1222 installation instructions

ricoo

ricoo S1222 installation instructions

Kanto P301 user manual

Kanto

Kanto P301 user manual

VidaBox VB MNT FFS2 installation guide

VidaBox

VidaBox VB MNT FFS2 installation guide

PROEL PLWIN60C user manual

PROEL

PROEL PLWIN60C user manual

AllMounts PLB-43S instruction manual

AllMounts

AllMounts PLB-43S instruction manual

VMP VH-001 instruction sheet

VMP

VMP VH-001 instruction sheet

Nederman SYSTEM 920/400 instruction manual

Nederman

Nederman SYSTEM 920/400 instruction manual

Glasdon ORBIS POST MOUNTING KIT installation guide

Glasdon

Glasdon ORBIS POST MOUNTING KIT installation guide

Songmics LGR105 manual

Songmics

Songmics LGR105 manual

Grape GM401 user manual

Grape

Grape GM401 user manual

FourStarGroup 11225296 manual

FourStarGroup

FourStarGroup 11225296 manual

CHIEF FLR100 installation instructions

CHIEF

CHIEF FLR100 installation instructions

FEC XDCAM EX PMW-EX30 manual

FEC

FEC XDCAM EX PMW-EX30 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

M.-Nr. 07 699 890 / 00 en
Insert E 905/1 for short MIS instruments
,Please read the Warning and Safety instructions in the operating Instructions for your G 7827/28 or PG 8527/28 Washer-disinfector carefully. You should only
process instruments/medical devices in this insert if they have been declared as suitable for machine processing by the manufacturer. Please also observe the
manufacturer's care instructions. The programme selected must be suitable for the materials from which the medical devices to be processed are made. Where
possible the programme should include thermal disinfection. The final rinse should, if possible, always be carried out with fully demineralised water.
If using automatic mobile unit recognition, please make sure that the coding is correctly set.
Miele Company Ltd., Fairacres, Marcham Road, Abingdon, Oxon, OX14 1TW, Tel 0845 365 0555,
Fax 0845 365 0777, Internet: http://www.miele.co.uk / !Alteration rights reserved / 3709
Items supplied:
Insert E 905/1, H 40, W 461, D 510 mm, fitted with:
–16 connectors for adapters, injector sleeves and jets
–2 notched rails for positioning the instruments
The insert is supplied with:
–1 E 336 injector sleeve for MIS instruments, 121 mm long, C11 mm
–2 x E 362 blanking screws
–1 E 442 injector sleeve for MIS instruments, 121 mm long, C11 mm,
with cap C6mm
–2 x E 447 female adapters ]for male Luer locks \
–4 x E 448 silicone tubes with screw thread, 300 mm long, C5 mm and Luer
Lock-Adapter ]
–2 x E 449 male adapters \for female Luer locks ]
–4 x E 452 jets, 2.5 x 60 mm, with screw thread
–3 x E 453 jets, 4.0 x 110 mm, with retaining clip and screw thread
–3 x E 454 jets for trocar sleeves, 10-15 mm, with screw thread
–1 x E 464 holder for E 454 jet
–1 x E 472 clamp for jets C4mm
–1 x E 907 insert/mesh tray with lid for small components
Areas of application:
Insert E 905/1 can be used in mobile units E 474/4, E 574/1, E 774/1 and E 902/1.
The mobile units are supplied with their own operating instructions.
The E 905/1 is designed for cleaning and disinfecting short MIS instruments, e.g.
those used for arthroscopy or laparascopy.
Instruments up to 190 mm long can be placed in the upper level, and up to 220 mm
long in the lower level.
Before loading the machine and before starting a programme, please check
that:
–the silicone adapters in the injector manifold are intact and complete.
,To ensure adequate water pressure, all positions must always be fitted with
jets or blanking screws.
For validation purposes please make sure that you follow the loading
instructions given on the template.
,All MIS instruments must be dismantled for processing in accordance with
the manufacturer's instructions.
,Long, narrow lumen instruments which cannot be taken apart must be
pre-washed in accordance with the manufacturer's instructions.
,Before connecting a hollow instrument please make sure that all hollow
sections are free of blockages. Otherwise the cleaning solution will not be able
to access all areas properly.
Instruments should not be able to come into contact with one another during the
cleaning and disinfecting process.
,Any hollow instruments that have become disconnected from the adapters
during processing must be reprocessed.
Injector sleeves
Injector sleeve E 442 is for processing hollow instruments and instruments with a
Ø less than 8 mm.
The diameter of the instruments should match that of the opening to the injector
sleeves/caps.
^Insert trocar sleeves with a Ø less than 10 mm into the injector manifold. Connect
to the instrument's cleaning port with hose adapter E 448. Connect the hose
adapter to the injector manifold using a silicone adapter.
^Insert hollow trocars into injector sleeves.
^Connect to the instrument's cleaning port using an E 448 hose adapter, and
insert the hose adapter into a silicone adapter.
^Insert long, narrow instruments which cannot be dismantled into an injector
sleeve. The functioning end must be open, if necessary using an E 472 spring
clamp to keep the handle open.
Connect to the instrument's cleaning port using an E 448 hose adapter. Connect
the hose adapter to the injector manifold using a silicone adapter.
Handles
^Place contaminated handles on a jet, e.g. E 452.
Veress cannulae
^Take the Veress cannula apart.
^Place the outer part of the Veress cannula on jet
E 453 with spring clamp E 472.
^Screw the inner part of the Veress cannula to a
Luer-Lock adapter \, keeping the stopcock open.
Small components
^Place small components such as seals and
stopcocks in the mesh tray with the led E 907/1.
^Close the lid, and hang the mesh tray in the mobile
unit.
Trocar sleeves Ø 10 - 20 mm
^Take the trocar sleeves apart.
Process small components in a mesh tray with a
lid.
^Attach trocar sleeves with Ø 10 - 15 mm to an
E 454 jet.
^For trocar sleeves with Ø 15 - 20 mm, use jets
E 464 and E 454 together.
^Position the trocar sleeve with the spring clamp so
that the water reaches critical areas through the
port in the side of the jet.
^Connect the trocar tube at the side with hose
adapter E 448.
^Insert the hose adapter in a silicone adapter.
^Secure the self-closing valve to an E 454 jet for
trocar sleeves or an E 453 jet with a retaining clip.
^Position the self-closing valve with the spring clamp
or retaining clip so that the water reaches critical
areas through the port in the side of the jet.
Insert E 905/1 for short MIS instruments