NUCLEONIX SYSTEMS PVT. LTD
CHAPTER - III
INSTALLATION
A. CONTROLS / CONNECTORS :
EHT (MHV socket) :High Voltage connector, at which the EHT supply is provided
to connect to the X-ray proportional counter / gas proportional
/ He / BF3 / Neutron detector for detector biasing.
Output (BNC) :BNC Co-axial connector at which output from the Pre -
Amplifieris generated for connecting to shaping amplifier.
LV (socket) :Five pin circular connector at which the LV supply voltages
are provided from the NIM instrumentation bin & power supply
unit. Power up the PA320 is fed.
Detector (MHV socket) :MHV connector, to which detector is connected preferably
through as short RG 59 cable. Cable end has to match with
the appropriate connector on the detector.
Test input :This is a BNC socket at which a -ve test tail pulse can be test
injected for testing the pre-amplifier functionality.
B. EXTERNAL CONNECTIONS :
Having connected the Pre-amplifier to the detector as explained earlier, proceed to make
the connection as follows : (Typical x-ray proportional Counting Systems).
1. Using the HV cable connect the OUTPUT from High Voltage Unit, type HV502 /HV503 to
the HV connector (MHV socket) on PA320.
2. Using Co-axial signal cable connect from the OUTPUT BNC connector on PA320 to the
INPUT of any Pulse shaping Amplifier typically LA520 or AF521or Spectroscopy amplifier.
3. Using LV - I/O, 5 core cable connect to the "LV" connector on PA320 to the LV connector
on Instrumentation bin & power supply. If your instrumentation bin has a 9-pin D-connector,
then you wire up one end with 9 pin D-connector & the other end with 5 pin I/O connector
to mote with corresponding connector on pre-amplifier.