manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Ohmeda
  6. •
  7. Monitor
  8. •
  9. Ohmeda 5250 User manual

Ohmeda 5250 User manual

5250
Respiratory
Gas
Monitor
Operation
and
Maintenance
Manual
S
G
a
aie
Display
Mi
Release
of
A
BOC
Health
Care
Company
Critical
Care
RIEL
WE
E
YU
Figure
2-1
5250
RGM
Front
Panel
Display......
2-1
Figure
2-2
Display
Release
a
2-2
Figure2-3DisplayRemoval............................
2-2
Figure
2-4
Sample
Filter
Cartridge
Replacement
............
ини
2-3
Figure
2-5
5250
RGM
Back
Panel....................
2-4
Figure
2-65250RGMDbDisplay..........................
2-6
Figure
2-7
5250
RGM
Display
Versions
..........
2-7
Figure
2-8
Setup
Screen
Changes
.................
2-12
Figure
2-9
Waveform/Trend
and
ScaleSelection............................
2-13
Figure
2-10
Changing
Alarmlimits..............
2-15
Figure
3-1
Respiratory
Gas
Zero
and
Span
Calibration...
urine
3-2
Figure
3-2
Calibration
Gas
Connection...........
3-3
Figure
3-3
Patient
Circuit
O,
:
Sensor
Calibration
................
0...
3-4
Figure
4-1
Patient
Circuit
Adapters
(Proximal)
eee
4-1
Figure
4-2
Typical
valid
plethysmographic
WaveforMm...........................
1.11.
4-2
Figure
4-3
Noisy
plethysmographic
waveforr
이
이니
이
아
이
어
아 아 아 아
아
마
아
아
의
4-3
Figure
4-4
Patient
Circuit
Adapter
Connection
Location......................
4-4
Figure
4-5
GMS
Pressure
Gauge
Coupling
Installation
.....................
4-5
Figure
4-6
Flow
Sensor
Connection
................
4-6
Figure
4-7
Sensor
Clip
Replacement
After
Mmtubation...........................
4-6
Figure
4-8
Proper
Sensor
and
Clip
Positioning
in
Distal
Mounting
ーー………
4-7
Figure
7-1
Sample
Filter
Cartridge
Replacement
................
7-3
Figure
7-2
Auto
Zero
Scrubber
Replacement
(Agent
Version
Only)
.…………………………………
7-3
Figure
7-3
Cleaning
the
Sample
Chamber......
7-5
Figure
8-1
Fuse
Replacement
..........................
8-5
Figure
8-2
Signal
Processing
Software
Replacement..........................
8-6
Figure
8-3
Display
Processing
Software
Replacermnent
iene
8-6
Figure
8-4
Internal
Oxygen
Sensor
Replacement
„r
8-7
Figure
8-5
Replacing
the
Sensor Cartridge
..
Figure
8-6
Transducer
cartridge
replacement
8-9
Figure
8-7
Sensor
clip
removal
from
the
breathing
CÍrCuit
„ee
8-9
Figure
A-1
Service
Screen
ーーーーー
A-1
Figure
A-2
Pneumatics
Service
Screen
..........
A-2
Figure
A-3
Service
Calibration
Screens..........
A-3
User
Responsibility
This
Product
will
perform
in
conformity
with
the
description
contained
in
this
operating
manual
and
accompanying
labels
and/or
inserts,
when
assembled,
operated,
maintained,
and
repaired
in
accordance
with
the
instructions
provided.
This
Product
must
be
checked
periodically.
A
defective
Product
should
not
be
used.
Parts
that
are
broken,
missing,
plainly
worn,
distorted,
or
contaminated
should
be
replaced
immediately.
Should
such
repair
or
replacement
become
necessary,
Ohmeda
recommends
that
a
tele-
phone
or
written
request
for
service
advice
be
made
to
the
nearest
Ohmeda
Regional
Service
Center.
This
Product
or
any
of
its
parts
should
not
be
repaired
other
than
in
accordance
with
written
instructions
provided
by
Ohmeda
and
by
Ohmeda
trained
personnel.
The
Product
must
not
be
altered
without
the
prior
written
ap-
proval
of
Ohmeda's
Safety
Department.
The
user
of
this
Product
shall
have
the
sole
respon-
sibility
for
any
malfunction
that
results
from
improper
use,
faulty
maintenance,
improper
repair,
damage,
or
alteration
by
anyone
other
than
Ohmeda.
CAUTION:
Federal
law
in
the
USA
and
Canada
restricts
this
device
to
sale
by
or
on
the
order
of
a
licensed
medical
practitioner.
©The
BOC
Group
Inc
1988,
1989,
1990
All
rights
reserved.
No
part
of
this
publication
may
be
reproduced
in
any
form
without
written
permission
from
Ohmeda.
ェ
ーーー
απ πο
στ
JO
ee
Ne
LA
TT
TIE
this
manual:
e
Warnings
alert
the
user
to
potential
product
use
or
misuse
that
could
harm
the
user
or
others.
e
Cautions
alert
the
user
to
potential
product
use
or
misuse
that
could
harm
the
product,
associated
equipment,
or
the
surrounding
environment.
The
warnings
and
cautions
that
appear
in
the
text
of
this
manual
are
described
in
the
following
sections.
À
Warnings
The
Ohmeda
5250
Respiratory
Gas
Monitor
(RGM)
does
NOT
have
the
ability
to
identify
anesthetic
agents.
If
the
agent
selected
on
the
monitor
is
not the
agent
delivered
to
the
patient
circuit,
the
values
will
be
inaccurate.
If
a
mixture
of
anesthetic
agent
is
delivered,
the
monitor
may
not
indicate
accurate
values.
In
the
presence
of
alcohols,
ketones,
or
other
organic
hydrocarbon
vapors,
the
5250
RGM
will
not
indicate
accurate
readings
of
anesthetic
agent.
This
includes
their
presence
not
only
in
the
sample
line
or
patient
circuit,
but
also
ethyl
alcohol
present
in
a
patient's
bloodstream.
Handle
the
5250
RGM
with
care.
Damage
to
the
RGM
or
inaccurate
operation
may
result
from
im-
proper
handling.
Do
not
operate
the
5250
RGM
unless
it
is
prop-
erly
calibrated.
Inaccurate
patient
parameter
readings
will
result.
Keep
the
circuit
O,
sensor
attached
to
the
unit
to
assure
accuracy
of
O,
readings.
If
detached,
the
sensor
must
be
reattached
and
stabilized
for
an
equivalent
length
of
time
(up
to
a
maximum
of
14
hours)
before
recalibrating.
If
the
RGM
fails
to
respond
as
described
in
the
calibration
procedure,
do
not
use
that
portion
of
the
monitor
until
the
malfunction
is
corrected.
For the
5250
RGM
without
Agent,
use
only
Ohmeda
calibration
gas
or
calibration
gas
of
6%
CO,,
50%
O,,
and
44% N,O
(40.05
volume
%
gravimetric
standard).
For
the
5250
RGM
with
Agent,
use
only
Ohmeda
calibration
gas
of
4%
halocarbon-22,
6%
CO,,
40%
N,0,
and
50%
O,.
Refer
to
Section
6.1
to
order
calibration
gas.
SpO,
data
is
NOT
collected
when
electrosurgical
interference
is
detected.
If
long
periods
of
inter-
ference
exist,
the
values
for
SpO,
and
pulse
rate
are
shown
as
dashes.
Use
only
Ohmeda
probes
with
Ohmeda
oxime-
ters
and
monitors;
otherwise,
patient
injury
or
equipment
damage
may
result.
continued
circulation
distal
to
the
probe
site
after
application.
ーー
ご
ピピ
デー
ぐー
Patient
Safety—Prolonged
monitoring
or
patient
condition
may
require
changing
the
probe
test
site
periodically.
Change
the
probe
site
at
least
every
four
hours
to
reduce
the
risk
of
blistering,
skin
erosion,
or
ischemic
skin
necrosis
(especially
if
the
site
is
poorly
perfused).
Do
not
place
the
airway
pressure
adapter
on
the
expiratory
check
valve
of
the
absorber.
Blockages
in
the
tubing
circuit
can
cause
high
patient
airway
pressures
and
may
not
be
detected
by
the
RGM.
The
Minute
Volume
and
Tidal
Volume
alarms
are
usable
only
if
the
TVX
flow
cartridge
is
installed
in
the
correct
section
of
the
breathing
circuit.
Operate
the
5250
RGM
only
with
the
TVX
flow
cartridge
placed
in
the
expiratory
limb
or
com-
mon
airway
of
the
breathing
circuit.
If
the
flow
sensor
is
placed
in
the
inspiratory
limb,
the
RGM
will
not
provide
exhaled
volume
data.
If
the
5250
RGM
is
used
with
a
hanging
bellows
type
anesthesia
ventilator,
the
monitor
may
register
volumes
in
spite
of
circuit
disconnection.
Water
in
the
TVX
flow
cartridge
will
restrict
the
motion
of
its
internal
vanes
and
cause
erroneous
tidal
volume
and
minute
volume
readings.
Position
the
tubing
so
that
water
drains
away
from
the
cartridge
and
does
not
accumulate
in
it.
Exposure
of
the
sensor
clip
to
a
direct
beam
of
light
may
cause
erroneous
tidal
volume
and
minute
volume
readings.
Shield
the
sensor
clip
with
opaque
material
if
the
reading
is
suspect.
Distal
Mounting:
When
placing
the
TVX
flow
cartridge
on
the
absorber
valve,
be
certain
to
obtain
a
secure
fit,
but
do
not
force
the
cartridge
in
place
as
tightly
as
possible.
Because
the
sensor
cartridge
is
tapered,
a
secure
fit
can
be
achieved
without
excessive
force.
A
properly
placed
cartridge
is
removable
by
hand,
but
it
is
advisable
that
a
tool
(pliers
or
channel
lock
pliers)
be
available
to
remove
the
cartridge
in
case
it
has
been
jammed
on
too
tightly.
Before
every
use,
ensure
that
the
TVX
flow
cartridge
is
operational and
easily
removable.
The
flow
sensor
must
be
correctly
oriented
in
the
breathing
circuit
as
indicated
by
the
markings
on
the
sensor
clip.
The
arrows
should
point
away
from
the
patient.
If
the
clip
is
not
mounted
correctly,
the
volume
monitor
will
not
operate
properly.
Precautions
The
circuit
O,
display
and
alarms
are
only
usable
if
enabled
on
the
Setup
Screen
and
a
calibrated
circuit
O,
sensor
is
installed.
The
RGM
must
be
calibrated
for
the
currently
installed
O,
sensor.
Do
not
insert
the
circuit
O,
plug
into
the
flow
sensor
connector
on
the
back
of
the
machine.
Damage
to
the
monitor
could
result.
Perform
the
checkout
procedure
in
Section
5.1
before
using
the
monitor
on
a
patient.
If
the
monitor
fails
any
test,
it
must
be
removed
from
use
until
it
has
been
repaired
and
checked
for
correct
operation.
Do
not
block
airflow
from
the
air
intake
or
ex-
haust
vents.
Inaccurate
readings
and/or
damage
to
the
RGM
may
result.
Do
not
place
the
RGM
on
surfaces
with
above
ambient
temperatures.
The
presence
of
nebulized
agents
in
the
sample
gas,
such
as
Mucomyst
(Registered
trademark
of
Bristol-Myers),
may
over
extended
periods
of
exposure,
tend
to
obstruct
internal
RGM
filters.
Do
NOT
use
the
5250
RGM
in
the
presence
of
flammable
anesthetics.
Do
not
remove
the
cover
of
the
RGM.
Refer
servicing
to
qualified
service
personnel.
Service
personnel
should
disconnect
the
power
cord
before
servicing
the
RGM.
For
continued
protection
against
fire
hazard,
replace
only
with
the
same
type
and
rating
of
fuse.
A
Cautions
Use
only
the
8-foot
sample
tube
assembly
sup-
plied
with
the
monitor.
A
longer
tube
may
change
the
operating
characteristics
(specifica-
tions)
of
the
monitor,
such
as
degrade
the
re-
sponse
time.
Order
replacement
sample
tubes
as
listed
in
Section
6.1.
Avoid
storing
the
monitor
and
probes
at
tem-
peratures
outside
the
following
range:
-20°
to
60°
C
(-4°
to
140°
F).
When
attaching
the
display
panel,
ensure
that
both
the
top
and
the
bottom
of
the
display
panel
are
fully
seated
into
their
slides.
Failure
to
fully
seat
the
display
panel
may
allow
the
display
panel
to
fall
off
and
be
damaged.
Empty
the
fluid
trap
before
each
patient
or
whenever
the
trap
is
more
than
half
full.
Failure
to
empty
the
trap
may
allow
it
to
fill
while
monitoring
a
patient
and
cause
the
monitor
to
stop
sampling
gas.
Make
sure
the
monitor
voltage
selected
agrees
with
the
local
voltage
available.
Do
not
cover
or
block
the
cooling
fan.
Connect
only
a
high
input
inpedance
device
(10K
ohm
or
higher)
to
the
analog
output
connector.
Improper
loading
will
upset
the
cor-
respondence
between
the
measured
voltage
and
the
intended
output
voltage.
Use
only
the
power
cord
supplied
with
the
RGM.
When
replacing
the
power
cord,
use
only
the
power
cord
specified
for
this
RGM.
Pressure
in
excess
of
10
psi
above
atmospheric
could
damage
the
Paw
inlet
or
sample
inlet
to
the
RGM.
Avoid
storing
O,
sensors
outside
the
following
range:
0º
to
26º
C
(32º
to
80º
F).
O,
sensors
must
have
shorting
clips
or
be
connected
to
the
monitor
when
in
storage
or
not
in
use.
When
in
use,
the
circuit
O,
sensor
should
always
be
facing
downward
to
reduce
moisture
buildup
on
its
sensing
surface.
Do
NOT
autoclave
or
pressure
sterilize
the
5250
Respiratory
Gas
Monitor.
Do
NOT
gas
sterilize
the
5250
RGM.
Never
immerse
the
RGM
in
liquid.
The
elec-
tronic
circuitry
can
be
short-circuited,
causing
permanent
damage.
Use
the
cleaning
solution
sparingly.
Do
NOT
saturate
the
RGM.
Excessive
solution
can
flow
into
the
RGM
causing
damage
to
internal
com-
ponents.
Following
sterilization
of
the
patient
circuit
oxygen
probe
housing
(front
half
only)
or
the
tee
manifold
with
ethylene
oxide,
quarantine
parts
to
allow
dissipation
of
residual
ethylene
oxide
gas
absorbed
by
the
material.
Follow
sterilizer
manufacturer's
recommendations
for
procedure
and
aeration
period.
After
sterilization
with
ethylene
oxide,
the
flow
sensor
should
be
quarantined
in
a
well
venti-
lated
area
to
allow
dissipation
of
residual
ethylene
oxide
gas
absorbed
by
the
probe.
Follow
the
sterilizer
manufacturer's
recommen-
dations
for
specific
aeration
periods
required.
Never
insert
cleaning
brushes
or
other
foreign
objects
through
the
flow
cartridge
vanes.
The
precision
movement
may
be
damaged.
Following
flow
cartridge
sterilization
with
eth-
ylene
oxide,
quarantine
the
cartridge
in
a
well
ventilated
area
to
allow
dissipation
of
residual
ethylene
oxide
gas
absorbed
by
the
cartridge
material.
Follow
the
manufacturer's
recommen-
dations
for
specific
aeration
periods
required.
In
some
cases,
aeration
periods
of
seven
days
or
more
may
be
required.
Use
Lue
Levolunenaed
cleaning
solution
spar-
ingly;
do
not
saturate
or
immerse
the
flow
sensor
clip.
Do
not
attempt
to
clean
the
sample
chamber
of
an
RGM
with
Agent.
Cleaning
may
cause
per-
manent
damage
to
the
sensor
and
void
the
warranty.
Follow
the
sterilizer
manufacturer's
recommen-
dations
concerning
procedures
and
aeration
periods.
Maximum
voltage.
No
more
than
5
volts
should
appear
on
any
pin
of
the
analog
output connec-
tor.
Do
not
use
a
strong
alkaline
detergent
to
clean
the
sample
chamber
because
alkaline
detergent
may
corrode
the
aluminum
casing.
Only
competent
individuals
trained
in
the
repair
of
this
equipment
should
attempt
to
service
it.
Detailed
information
for
more
extensive
repairs
is
included
in
the
service
manual
solely
for
the
convenience
of
users
having
proper
knowledge,
tools,
and
test
equipment,
and
for
service
representatives
trained
by
Ohmeda.
Alter
replacing
the
internal
U,
sensor,
perform
the
calibration
procedure
in
Section
3.1
to
verify
the
RGM
is
working
properly.
Allow
at
least
five
minutes
for
the
sensor
to
stabilize
before
calibrating.
After
replacing
the
probe
assembly,
perform
the
calibration
procedure
in
Section
3.2
to
verify
that
the
RGM
is
working
properly.
Allow
at
least
five
minutes
for
the
sensor
to
stabilize
before
calibrating.
After
replacing
the
sensor
cartridge,
perform
the
O,
sensor
calibration
in
Section
3.2
to
verify
that
the
RGM
is
working
properly.
Allow
at
least
five
minutes
for
the
sensor
to
stabilize
before
calibrating.
Never
tamper
with
the
set
screws
in
the
flow
cartridge.
Such
action
will
render
the
cartridge
unusable.
Malfunctioning
flow
cartridges
must
be
de-
stroyed
to
prevent
their
inadvertent
use.
Notes
Bip
LUlaliual
USSULUJES
LILE
Proper
preparation,
operation,
and
maintenance
of
the
Ohmeda
5250
Respiratory
Gas
Monitor
(RGM).
Read
through
the
entire
manual,
with
an
RGM
avail-
able
for
hands-on
experience,
before
using
the
RGM
with
a
patient.
Pay
special
attention
to
the
warnings
and
cautions,
which
appear
throughout
the
manual
and
are
listed
in
the
“Precautions”
section.
Read
the
user
responsibility
statement;
it
describes
what
is
expected
of
the
user
to
maintain
a
safe
and
accurate
instrument.
Read
the
warranty;
it
describes
Ohmeda's
responsi-
bility
in
case
of
a
functional
defect.
Keep
this
manual
available
for
answering
questions
that
may
arise
concerning
the
moni-
tor's
operation,
its
maintenance,
or
in
case
of
failure,
its
repair.
The
maintenance
procedures
in
this
manual
can
be
performed
by
the
opera-
tor.
Always
handle
the
RGM
and
accessories
with
care
to
prevent
physical
damage
to
the
equip-
ment
or
inaccurate
operation.
The
5250
RGM
can
be
ordered
with
a
number
of
options.
Check
the
packing
slip
against
the
invoice
to
ensure
you
have
received
what
was
ordered.
If
possible,
save
the
packaging
materi-
als
for
shipping
or
storing
the
monitor.
neciion
4
describes
the
controls
ana
connectors
for
the
monitor
and
the
software
program
used
for
the
setup
and
operation
of
the
monitor.
Section
3
explains
how
to
calibrate
the
monitor
before
using
it
with
a
patient.
Section
4
describes
how
to
connect
the
patient
to
the
RGM
for
the
parameters
you
are
monitor-
ing.
Section
5
details
a
preoperative
checkout
procedure
that
must
be
performed
before
the
RGM
is
used
with
a
patient.
The
monitor
opera-
tion
procedure
summarizes
the
steps
required
of
the
operator
before
starting
to
monitora
.
patient.
:
Section
6
lists
the
accessories
and
the
specifica-
tions
for
the
5250
RGM.
Section
7
explains
how
to
clean
and
sterilize
the
5250
RGM
and
accessories.
Section
8
details
how
to
perform
routine
opera-
tor
maintenance
on
the
5250
RGM
and
tells
you
what
to
do
if
the
monitor
is
not
working
prop-
erly.
The
Appendix
has
information
that
you
may
need
to
refer
to
occasionally
and
information
about
the
78xx
ventilator
interface.
The
software
features
and
functions
described
in
this
manual
apply
to
Rev
3.00
or
greater
of
the
dis-
play
processor.
The
Service
Screen
shows
your
current
revision
level.
This
manual
covers
the
following
5250
Respiratory
Gas
Monitors:
Bagic
ROGMU
nen
sama
sanan
sanane
sansa
aaa
en
村
6051-0000-009
(Includes
CO,
O,
ΝΟ,
Airway
Pressure,
Tidal
and
Minute
Volume,
Resp.
Rate,
and
Circuit
O,)
Basic
RGM
with
SpO,
and
PulseRate.....................
Basic
RGM
with
Agent...
ses
n
iene
seen
ne
nzio
nec
Basic
RGM
with
Agent,
SpO,,
and
Pulse
Rate.......
ee
6051-0000-010
昌和
6051-0000-011
ee
6051-0000-012
Notes
Ar
MA
NFL
NA
CL
The
Onmeda
5250
Respiratory
Gas
Monitor
(RGM)
provides
in
one
compact
package
a
total
respiratory
monitoring
solution
for
the
operating
room
or
intensive
care
unit.
The
RGM
uses
microprocessor
technology
to
measure
CO,,
O,,
N,O,
and
airway
pressure.
Air-
way
flow,
anesthetic
agent,
SpO,,
and patient
circuit
O,
are
monitoring
options.
All
of
these
parameters
are
displayed
on
a
4"
by
8"
flat
electroluminescent
display
screen
and
are
inte-
grated
with
an
alarm
management
system.
The
monitor
is
controlled
by
infrared
touch
controls,
and
the
display
is
removable
for
easy
access
and
viewing.
A
kit
included
with
the
monitor
contains
a
pressure
sensing
tee,
patient
sample
catheters,
airway
adapters,
a
sample
chamber
cleaning
kit
(5250
RGMs
without
Agent),
and
calibration
gas.
Optional
accessories
allow
the
monitoring
of
patient
circuit
O,
and
airway
flow.
The
SpO,
and
anesthetic
agent
monitoring
options
may
be
either
ordered
factory
installed
or
added
later
by
Ohmeda
service
representatives.
The
computer
program
cartridges
are
accessible
through
the
rear
panel
of
the
main
unit
and
through
the
side
panel
of
the
display,
which
makes
software
upgrades
and
future
options
simple
to
install.
Clock
/Calendar
Display
Sample
Inlet
PAW
Inlet
Spo,
Connector
Power
Switch
Display
ВЕН
Display
Release
Alarm
.
Lights
Alarm
Silence
Figure
2-1
5250
RGM
Front
Panel
Lİ
calibration
procedure.
ON.
NANA
NN
m
ZA;
An
RS-232
interface
provides
communication
to
a
printer,
computer,
or to
an
Ohmeda
78xx
ventila-
tor.
Note:
The
"xx"
in
78xx
represents
a
variable,
because
there
are
7800
and
7810
ventilators
in
use.
Seven
analog
output
channels
are
available
on
the
rear
panel
for
connection
to
a
strip
chart
re-
corder.
2.2
Front
Panel
Controls
and
Connectors
Refer
to
Figure
2-1.
Sample
Inlet
The
sample
inlet
connects
the
patient
circuit
to
the
RGM
via
a
small
diameter
gas
sampling
tube.
The
connector
accepts
a
male
luer
lock
fitting.
CAUTION:
Use
only
the
8-foot
sample
tube
assembly
supplied
with
the
monitor.
A
longer
tube
may
change
the
operating
characteristics
(specifications)
of
the
monitor,
such
as
degrade
the
response
time.
Order
replacement
sample
tubes
as
listed
in
Section
6.1.
Paw
Inlet
The
Paw
inlet
allows
attachment
of
the
pressure
sensing
tube
from
the
patient
circuit
to
the
RGM.
Graphic
Display
Alarm
Area
ασια
4/
WCSCLIPULOn
SpO,
Connector
The
SpO,
connector
allows
attachment
of
the
nine-pin
oximeter
probe
to
the
RGM.
CAUTION:
Use
ONLY
the
Ohmeda
SpO,
probes
supplied
for
this
5250
RGM
model.
Check
the
Identification
Number/Serial
Number
tag,
which
is
located
on
the
cable
near
the
connector.
The
model
number
must
read:
MOD
8122-00x
or
8121-00x
(x
represents
a
digit
from
1
through
7).
Power
Switch
The
power
switch
turns
AC
power
to
the
RGM
On
(|)
or
Off
(0).
Display
Release
Depress
this
button
to
slide
the
display
to
the
right
and
permit
observation
of
the
water
trap
or
to
remove
the
display
from
the
chassis.
The
dis-
play
has
an
8-foot
extension
cord
attached.
Press
Display
Release
Figure
2-2
Display
Release
Align
Display
with
White
Line
To
remove
the
display
from
the
RGM,
press
the
Display
Release
push
button
and
slide the
display
to
the
right
until
the
alignment
marks,
in
the
trap
area
and
on
the
upper
right
side
of
the
display,
line
up
as
shown
in
Figure
2-2.
Gently
lift
the
display
straight
up
and
pull
the
bottom
away
from
the
chassis
as
shown
in
Figure
2-3.
To
attach
the
display
to
the
RGM,
first
coil
the
cord
in
the
cord
storage
box,
insert
the
top
of
the
display,
position
the
alignment
marks
at
the
top,
and
gently
push
the
bottom
towards
the
chassis.
Slide
the
display
to
the
left
until
the
latch
clicks.
CAUTION:
When
attaching
the
display,
ensure
that
both
the
top
and
the
bottom
of
the
display
are
fully
seated
into
their
slides.
Failure
to
fully
seat
the
display
may
allow
the
display
to
fall
off
and
be
damaged.
Align
Edge
to
Indicator
Slide
to
Right
Lift
Up
BS
===
FR
D
Lift
Up
Then
Pull
Out
from
Bottom
+
LUAALA
L'ILE
The
fluid
trap,
located
behind
the
display,
collects
fluids
separated
from
the
aspirated
patient
sample.
To
remove
the
trap
bottle
for
emptying,
pull
the
bottle
straight
down.
The
O-
rings
can
be
lubricated
with
lubricant,
such
as
Vac
Kote’
or
Cello
Seal?,
which
are
specified
safe
for
use
in
an
oxygen
enriched
environment.
To
replace
the
bottle
push
the
bottle
into
posi-
tion.
CAUTION:
Empty
the
fluid
trap
before
each
patient
or
whenever
the
trap
is
more
than
half
full.
Failure
to
empty
the
trap
may
allow
it
to
fill
while
monitoring
a
patient
and
cause
the
monitor
to
stop
sampling
gas.
Sample
Filter
Cartridge
Some
units
have
a
gas
sample
filter
cartridge
(Figure
2-4)
located
above
the
fluid
trap.
Re-
place
this
filter
cartridge
when
the
advisory
message
SAMPLE
FILTER
BLOCKED
appears
in
the
alarm
display
area
or
when
the
CO,
wave-
form
response
time
is
degraded.
To
remove
the
cartridge,
pull
it
straight
out
from
the
fluid
trap
assembly.
To
replace
the
cartridge,
place
it
in
position
and
slide
it
in
until
it
seats
properly.
Check
for
leaks
by
occluding
the
sample
inlet
and
noting
that
the
unit
purges
within
five
seconds.
Graphic
Display
The
entire
graphic
display
is
covered
by
a
matrix
of
infrared
targets.
A
controlis
displayed
as
an
option
on
a
screen.
Simply
touch
the
option
to
Figure
2-4
Sample
Filter
Cartridge
Replacement
FU
LIV
OLIO
EER
EEE
AL
Lo
LIUL
LIGUUDDGIV
LU
CAGIL
pressure
on
the
screen
since
the
infrared
beam
is
broken
when
a
finger
nears
the
option.
Clock/Calendar
Display
The
clock/calendar
display
area,
located
on
the
top
left
side
of
the
display
screen,
indicates
the
time
of
day
(24-hour
clock)
in
hours
and
minutes,
the
day
of
the
week,
the
day
of
the
month,
the
month
of
the
year,
and
the
year.
To
set
the
clock/calendar:
1.
Select
MENU
from
the
display
screen.
2.
Select
SETUP
from
the
menu.
3.
Select
CLOCK
SET
from
the
Setup
Screen.
Time
and
date
changes
can
then
be
made
on
the
screens
displayed.
The
clock/calendar
will
main-
tain
the
current
time
and
date
when
the
power
is
switched
off.
Note:
Some
older
monitors
do
not
have
hardware
to
support
the
clock/calendar.
On
these
monitors,
no
clock
appears
on
the
main
screen
and
no
clock
set
option
is
available.
Alarm
Area
The
alarm
area
indicates
alarm
messages
on
the
top
right
side
of
the
display.
Alarm
Lights
The alarm
lights
display
either
a
flashing
red
light
a
flashing
yellow
light,
or
a
continuous
yellow
light
depending
on
the
type
of
alarm.
Refer
to
“Alarm
Types”
in
Section
2.8.
1.Vac
Kote
is
a
registered
trademark
of
the
Ball
Corporation,
2.Cello-Seal
is a
registered
trademark
of
Fisher
Scientific.
Slide
filter
cartridge
out
to
remove
it.
Slide
collection
bottle
down
to
remove
it.
£/
LJESCrIPTION
Alarm
Silence
The
Alarm
Silence
function
is
an
infrared
touch
function
located
in
the
lower
left
corner
of
the
monitoring
screen.
Touching
this
area
mutes
the
audio
for
the
indicated
length
of
time.
À
new
alarm
during
this
mute
period
resets
the
Alarm
Silence
and
allows
the
audio
alarm
to
sound.
An
additional
touch
during
the
mute
period
will
reset
the
Alarm
Silence
and
allow
a
new
audio
alarm.
À
double
touch
of
an
inactive
alarm,
within
two
seconds,
will
cause
the
ALL
MUTE
message
to
appear
next
to
the
alarm
silence
area,
and
new
alarms
will
not
activate
the
audio
sound
for
the
alarm
silence
interval.
A
third
touch
of
alarm
silence
within
two
seconds
will
put
the
unit
into
permanent
all
mute
mode,
which
mutes
all
alarms
except
LOW
INSPIRED
O,,.
Another
touch
resets
the
alarm
mode.
The
alarm
silence
interval
can
be
changed
on
the
Setup
Screen.
Note:
If
the
78xx
ventilator
interface
is
being
used,
the
mute
period
is
limited
to
30
seconds.
The
All
Mute
function
is
inhibited
if
the
78xx
ventilator
is
selected
as
the
RS-232
device
on
the
Setup
Screen.
A
second
touch
clears
alarm
silence.
2.3
Back
Panel
Controls
and
Connectors
Refer
to
Figure
2-5.
Voltage
Selector
The
voltage
selector
switch
allows
the
operator
to
match
the
available
local
voltage
to
the
mon-
itor.
Voltage
selections
include
100V,
120V,
220V,
and
240V.
Once
the
switch
is
properly
adjusted
no
further
resetting
should
be
re-
quired.
CAUTION:
Make
sure
the
monitor
voltage
se-
lected
agrees
with
the
local
voltage
available.
n
nell
-i
i
Second
touch
[Third
First
touch
after
two
touch
after
alarm
seconds,
or
new
alarm,
alarm
or
alarm
time
[time
Second
touch
expires.
expires
within
two
seconds
7
ALL
MUTE
88
|
~
Third touch
within
two
seconds
Fourth
Touch
Alarm
Silence
Diagram
は
DE
Use
Only:Hospital
Grade
ed
Receptacl
Cooling
Fan
The
cooling
fan
provides
constant
air
circulation
for
the
RGM
to
protect
heat
sensitive
electronic
components.
CAUTION:
Do
not
cover
or
block
the
cooling
fan.
Sample
Exhaust
Connect
the
sample
exhaust
line
(1/8"
barbed
connector)
to
a
waste
gas
scavenging
system
to
properly
eliminate
the
gas
sample.
If
a
closed
circuit
is
required,
return
the
gas
sample
to
the
patient
circuit.
Software
Module
The
RGM's
program
is
stored
in
the
software
module.
Contact
Ohmeda
Customer
Service
as
listed
on
the
back
page
to
obtain
the
latest
soft-
ware
upgrade.
For
software
module
replace-
ment,
refer
to
Section
8.4.
External
Ground
Connection
The
external
ground
connection
allows
equi-
potential
grounding
for
the
monitor
when
re-
quired.
Power
Cord
Receptacle
The
power
cord
receptacle
allows
the
monitor
to
be
connected
to
the
local
power
system.
CAUTION:
Use
only
the
power
cord
supplied
with
the
RGM.
When
replacing
the
power
cord,
use
only
the
power
cord
specified
for
this
RGM.
RS-232
Connection
The
nine-pin
isolated
RS-232
connector
provides
printer,
computer,
or
78xx
ventilator
interface.
O,
Sensor
Input
The
O,
sensor
input
provides
the
connection
for
the
O,
sensor
to
the
8-pin
modular
jack
on
the
RGM.
When
inserting
the
plug,
be
certain
that
the
connector
plug
is
fully
seated
in
the
jack
and
a
locking
click
is
heard
during
insertion.
A
slight
tug
on
the
cord
will
ensure
that
the
plug
is
fully
seated.
Flow
Sensor
Input
The
flow
sensor
input
provides
the
connection
for
the
flow
sensor
to
the
6-pin
or
8-pin
modular
jack
on
the
RGM.
When
inserting
the
plug,
be
certain
that
the
connector
plug
is
fully
seated
in
the
jack
and
a
locking
click
is
heard
during
in-
sertion.
A
slight
tug
on
the
cord
will
ensure
that
the
plug
is
fully
seated.
Analog
Output
The
analog
output
provides
an
analog
signal
proportional
to
the
real-time
N,O
waveform,
CO,
waveform,
and
other
values.
The
8-pin
DIN
analog
output
connector
provides
0
to
1V
full-
scale
sensitivity
with
approximately
a
100-ohm
source
impedance
(see
"Connector
Pinouts"
in
Section
F
of
the
Appendix).
CAUTION:
Connect
only
a
high
input
imped-
ance
device
(10K
ohm
or
higher)
to
the
analog
output
connector.
Improper
loading
will
upset
the
correspondence
between
the
measured
voltage
and
the
intended
output
voltage.
2.4
Displays
A.
CO,
Displays
Refer
to
Figures
2-6
and
2-7.
CO,
Digital
Display
The
previous
breaths
inspired
value
is
dis-
played
to
the
left
of
the
slash
(/)
and
the
maxi-
mum
end
tidal
CO,
value
over
the
last
three
breaths
is
displayed
to
the
right
of
the
slash.
The
selected
unit
of
measure
(%,
kPa,
or
mmHg)
is
displayed
at
the
bottom
right
of
the
display.
The respiratory
rate
(RR),
derived
from
CO,
values,
is
displayed
near
the
top
right.
Note:
0-15%
is
approximately
in
the
0-110
mmHg
range
at
sea
level.
The
mmHg
and
kPa
ranges
are
dependent
upon
local
barometric
pressure.
CO,
Waveform
and
Trend
Display
The
capnogram
is
displayed
with
a
scale
on
the
right
side.
Seventy
minutes
of
end
tidal
and
inspired
trend
at
one
point
every
two
minutes
are
displayed
to
the
left
of
the
capnogram.
The
average
inspired
and
expired
values
for
each
two-minute
period
is
displayed
in
the
trend
area.
CO,
Trend
Display
Eight
hours
of
end
tidal
trend
data
at
one
point
every
two
minutes
are
displayed.
The
available
display
ranges
are
0-20,
0-40
(default),
or
0-80
mmHg;
0-2.5,
0-5,
or
0-10%;
0-3,
0-6,
or
0-12
kPa.
CO,
Breath-by-Breath
Trend
Display
If
selected
on
the
Setup
Screen,
a
200-breath
trend
of
inspired
and
expired
values
is
dis-
played.
Tick
marks
denote
five-minute
intervals
in
this
trend.
Lal
LPP
AB
Ae
Waveform
Alarm
Message
Area
Area
Clock/Calendar
Area
Waveform
Title
———>
70-Minute
§——»
Waveform
Trend
Gas,
SpO
Area
P
2
|
Pressure,
and
Volume
Areas
Scale
Area
—
Alarm
Alarm
Limits
Calibrate
Wave,
Silence
,
Selections
and
Trend
and
Area
Setup
Scale
Menu
Selection
Spo,
Respiratory
Signal
Rate
Indicator
Inspiratory
CO,
.
Exni
. F
t
mmHg/kPa/%
i.
Expiratory
Inspiratory
OX
ory
İ
CO
mmHg/kPa/%
0,%
2
0,%
Display
.
2
Minimum
CO,
Display
Airway
Pressure
Positive
cm
H,0
Inspiratory
—
Mean
Pressure
Paw
Pulse
k
cm
HO
cm
HO
Indicator
>
na
——
Pulse
Rate
Spo,
SPO,
Display
Tidal
Volume
mL/min
:
Flow
Volume
-—
Activity
Bar
RA
Minute
Volume
L/min
TV
and
MV
Display
with
78xx
-
Patient
Circuit
0,%
Patient
Circuit
0,%
Display
with
78xx
η
να."
Pa
MIN
Pressure
МАХ
Ргеззиге
—
Ргеззиге
Plateau
Paw
Display
with
78xx
Selected
Agent
.
Expiratory
Inspiratory
~
一
一
Agent
%
Agent
%
Display
Agent
Display
—
—
N,0%
N,0%
Display
Non-agent
Analyzer
Non-agent
Analyzer
with
SpO,
Option
Agent
Gas
Analyzer
Only
Je
|
95.
:
32/12
HALQOTHANE
,
an
Agent
Gas
ο
λα.
pump
|
Analyzer
with
SpO,
Note:
All
display
screens
show
optional
flow
sensor
Figure
2-7 and
optional
patient
circuit
O,
sensor
installed.
5250
RGM
Display
Versions
ff
LIGSCE
ALP
LAUER
СО,
Alarms
Alarm
Message
Default
Range
Type
HIGH
Et
CO,
OFF
OFF,
0-15%
Emergency
LOW
Et
CO,
OFF
OFF,
0-15%
Emergency
CO,
APNEA
30
sec
20-30
sec
Emergency
HIGH
Fi
CO,
OFF
OFF,
0-15%
Emergency
B.
O,
Displays
O,
Digital
Display
Inspired
and
expired
O,are
measured
from
a
single
galvanic
sensor.
A
microcomputer
con-
trolled
switch
shunts
only
inspired
gas
(based
on
the
CO,
breath
detection)
to
the
sensor
for
20
seconds,
to
obtain
a
stable
O,
concentration.
The
inspired
gas
is
then
measured
for 15
seconds,
and
the
concentration
appears
to
the
left
of
the
slash.
The
expired
gas
is
then
sup-
plied
to
the
sensor
for
20
seconds
and
meas-
ured
for 15
seconds.
The
expired
value
is
displayed
to
the
right
of
the
slash.
This
cycle
repeats
continuously
as
long
as
breaths
are
detected.
O,
Trend
Display
Eight
hours
of
inspired
O,
concentration
trend
data
with
one
point
every
two
minutes
are
dis-
played.
The
available
display
ranges
are
10-70%
(default),
or
10-100%.
O,
Breath-by-Breath
Trend
Display
If
selected
on
the
Setup
Screen,
a
200-breath
trend
of
inspired
and
expired
values
appears.
Tick
marks
denote
five-minute
intervals
in
this
trend.
O,
Alarms
Alarm
Message
Default
Range
Type
LOW
INSPIRED
O,
18%
18-100%
Emergency
HIGH
INSPIRED
O,
OFF
OFF,
18-100%
Emergency
Note:
The
LOW
INSPIRED
O,
alarm
may
not
be
disabled.
C.
SpO,
Displays
(Optional
Feature)
Spo,
Digital
Display
The
SpO,
digital
value
appears
in
a
large
font
size.
The
pulse
rate
appears
above
the
SpO,
value
in
a
smaller
font
size.
A
beating
heart
icon
appears
for
each
detected
pulse.
An
SpO,
beep
tone,
with
a
pitch
proportional
to
the
SpO,
value,
occurs with
each
detected
pulse.
SpO,
Waveform/Trend
Display
The
plethysmograph
appears
in
a
window
with
a
signal
strength
indicator
on
the
right
side.
The
plethysmograph
may
be
either
autoscaled
(de-
fault)
or
fixed
scale
as
selected
from
the
Setup
Screen.
Seventy
minutes
of
SpO,
trend,
one
point
every
two
minutes,
appears
with
an
SpO,
scale on
SpO,
Trend
Display
Eight
hours
of
5pO,
trend
appears
with
one
point
every
two
minutes.
Available
display
ranges
are
80-100%
(default)
or
50-100%.
SpO,
Breath-by-Breath
Trend
Display
If
selected
on
the
Setup
Screen,
a
200-breath
trend
of
the
SpO,
value
is
displayed.
Tick
marks
denote
five-minute
intervals
in
this
trend.
spO,
Alarms
Alarm
Message
Default
Range
Type
LOW
SpO,
90%
OFF,
50-100%
Emergency
LOW
PULSE
RATE
OFF
OFF,
40-150
bpm
Warning
HIGH
PULSE
RATE
OFF
OFF,
80-250
bpm
Warning
HIGH
Spo,
OFF
OFF,
70-99%
Warning
О.
NO
Displays
N,0
Digital
Display
Mean
N,O%
over
a
five-second
interval.
N,O
Trend
Display
Eight
hours
of
N,O
trend,
with
one
point
every
two
minutes.
Display
ranges
are
0-100%
(default),
or
40-80%.
N,O
Breath-by-Breath
Trend
Display
If
selected
on
the
Setup
Screen,
a
200-breath
trend
of
inspired
and
expired
values
appears.
Tick
marks
denote
five-minute
intervals
in
this
trend.
N,O
Alarms
Alarm
Message
Default
Range
Type
HIGH
N,O
80%
OFF,
1-99
Emergency
E.
Airway
Pressure
Displays
Airway
Pressure
Digital
Display
The
peak
pressure
over
the
previous
breath
is
displayed
to
the
left
of
the
slash.
The
minimum
pressure
over
the
previous
breath
is
displayed
to
the
right
of
the
slash.
The
mean
airway
pressure
over
the
previous
breath
appears
in
parentheses
beneath
the
slash.
When
interfaced
to
the
78xx
ventilator,
the
wave-
form
can
be
received
by
connecting
a
tee
into
the
pressure
line.
Data
from
the
ventilator
is
MAX
pressure,
MIN
pressure,
and
PLAT
(pressure
plateau).
Airway
Pressure
Waveform/Trend
Display
The
pressure
waveform
appears
with
the
scale
on the
right
side.
Seventy
minutes
of
trend,
one
Airway
Pressure
Alarms
Note:
When
interfaced
to
the
78xx
ventilator,
pressure
alarm
limits
cannot
be
selected.
These
alarm
limits
are set
on
the
ventilator.
Alarm
Message
Default
HIGH
SUSTAINED
Paw
(15
sec)
30
cm
H,0
SUB-ATMOSPHERIC
Paw-
15
cm
H,0
HIGH
Paw
80
cm
H,0
LOW
SUSTAINED
Paw
(15
sec)
OFF
point
every
two
minutes,
appears
on
the
left
side.
Note:
The
airway
pressure
waveform
will
not
be
in
phase
with
the
CO,
waveform
due
to
CO,
sample
transit
time
delay.
Airway
Pressure
Trend
Display
Peak
pressure
is
the
top
value
and
minimum
pres-
sure
is
the
bottom
value.
Eight
hours
of
trend
with
one
point
every
two
minutes
appears.
Available
display
ranges
are
-10
to
20
(default),
-25
to
50,
and
-40
to
80
cm
H,0.
Airway
Pressure
Breath-by-Breath
Trend
Display
If
selected
on
the
Setup
Screen,
a
200-breath
trend
of
peak
and
minimum
pressure
appears.
Tick
marks
denote
five
minute
intervals
in
this
trend.
Range
Type
OFF,
10-30
cm
H,O
Emergency
fixed
Emergency
OFF,
1-120
cm
H,O
Warning
OFF,
-10-20
cm H,O
Warning
al
Bd
Cr
wD
ho
BB
Be
λα
F.
Tidal
and
Minute
Volume
Displays
Minute
Volume
Trend
Display
Tidal
and
Minute
Volume
Digital
Display
Eight
hours
of
minute
volume
trend,
with
one
point
every
two
minutes,
appears.
Available
The
tidal
volume
in
milliliters
appears
to
the
display
ranges
are
0-20
L
(default)
or
0-60
L.
right
of
TV.
The
minute
volume
in
liters
per
minute
appears
to
the
right
of
MV.
Both
are
optional
features.
Minute
Volume
Breath-by-Breath
Trend
Display
If
selected
on
the
Setup
Screen,
a
200-breath
trend
of
inspired
and
expired
values
appears.
Tick
marks
denote
five-minute
intervals
in
this
trend.
Flow
Volume
Activity
Bar
Graph
This
is
a
graphical
representation
of
the
instan-
taneous
flow
in
the
expiratory
limb.
Full
scale
is
100
L/min.
Tidal
and
Minute
Volume
Alarms
Note:
When
interfaced
to
the
78xx
ventilator,
tidal/minute
volume
alarm
limits
cannot
be
selected.
These
alarm
limits
are
set
on
the
ventilator.
Alarm
Message
Default
Range
Type
LOW
TIDAL
VOLUME
OFF
OFF,
50-2500
ml
Emergency
HIGH
TIDAL
VOLUME
OFF
OFF,
50-2500
ml
Warning
LOW
MINUTE
VOLUME
]
OFF
OFF,
.2-50.0
L
Warning
HIGH
MINUTE
VOLUME
OFF
OFF,
.2-50.0
L
Warning
REVERSE
FLOW
OFF N/A
Warning
Note:
The
REVERSE
FLOW
alarm
can
be
enabled
on
the
Setup
Screen.
G.
Patient
Circuit
O,
Displays
Note:
When
interfaced
to
the
78xx
ventilator,
Circuit
O,
alarm
limits
cannot
be
selected.
These
alarm
limits
are
set
on
the
ventilator.
Patient
Circuit
O,
Digital
Display
The
value
shown
is
the
patient
circuit
O,
mean
over
five-seconds
value.
This
is
an
optional
feature.
Patient
Circuit
O,
Alarms
Alarm
Message
Default
Range
Type
LOW
CIRCUIT
O,
18% 18-99%
Emergency
HIGH
CIRCUIT
O,
%
OFF
OFF,
18-99%
Warning
Note:
The
Low
Patient
Circuit
O,
alarm
cannot
be
disabled
unless
the
Patient
Circuit
O,
parameter
is
disabled
on
the
Setup
Screen.
The
display
value
for
the
Patient
Circuit
O,
is
also
enabled
on
the
Setup
Screen.
Also
Note:
Permanent
ALL
MUTE
cannot
permanently
mute
the
Low
Circuit
O,
and
VENT
Low
Circuit
O,
alarms.
H.
Anesthetic
Agent
Displays
(Optional
feature
on
the
5250
RGM)
Anesthetic
Agent
Digital
Display
WARNING:
The
5250
RGM
does
NOT
have
the
ability
to
identify
anesthetic
agents.
If
the
agent
selected
on
the
monitor
is
not
the
agent
delivered
to
the
patient
circuit,
the
values
will
be
inaccurate.
If
a
mixture
of
anesthetic
agent
is
delivered,
the
monitor
may
not
indicate
accurate
values.
WARNING:
In
the
presence
of
alcohols,
ketones,
or
other
organic
hydrocarbon
vapors,
the
5250
RGM
will
not
indicate
accu-
rate
readings
of
anesthetic
agent.
This
in-
cludes
their
presence
not
only
in
the
sample
line
or
patient
circuit,
but
also
ethyl
alcohol
The
anesthetic
agent
selected
from
halothane,
isoflurane,
enflurane,
or
no
agent
is
shown.
The
value
of
the
anesthetic
agent
also
appears.
Anesthetic
Agent
Trend
Display
Eight
hours
of
the
selected
anesthetic
agent
trend,
with
one
point
every
two
minutes,
ap-
pears.
Available
display
ranges
are
0-2%
(de-
fault),
0-4%,
or
0-6%.
Anesthetic
Agent
Breath-by-Breath
Trend
Display
If
selected
on
the
Setup
Screen,
a
200-breath
trend
of
inspired
and
expired
values
appears.
Anesthetic
Agent
Alarms
Alarm
Message
Default
Range
Type
LOW
INSP
AGENT
OFF
OFF,
0-10%
Emergency
HIGH
INSP
AGENT
See
notes
OFF,
0-15%
Emergency
LOW
EXP
AGENT
OFF
OFF,
0-10%
Emergency
HIGH
EXP
AGENT
See
notes
OFF,
0-15%
Emergency
Notes:
1.
When
an
agent
is
selected,
the
following
high
limits
are
automatically
set:
High
Insp.
High
Exp.
Halothane
4.5%
1.7%
Isoflurane
5.5%
3.5%
Enflurane
6.5%
4.2%
2.
When
the
user
reviews
or
changes
the
agent
limits,
those
limits
remain
fixed
until
the
unit
is
switched
off.
3.
If
NO
AGENT
is
selected,
the
RGM
will
alarm
if
trace
agent
is
detected.
The
message
AGENT
DETECTED
appears
and
SET
AGENT
will
flash.
2.5
Changing
the
Setup
Refer
to
Figure
2-8.
1.
Select
MENU
from
the
display
screen.
2.
Select
SETUP
from
the
menu.
The Setup
Screen
appears.
3.
Select
the
desired
parameter
from
the
left
side
of
the
screen.
4.
Select
the
desired
option
from
the
right
side
of
the
screen.
5.
Select
VIEW
ALL
to
see
all
parameters
and
the
current
options
selected.
6. If
preset
programmed
options
are
desired,
press
DEFAULT.
7.
Select
REC
CAL
to
set
up
a
strip
chart
re-
corder
connected
to
the
analog
output.
8.
Select
EXIT
from
the
Setup
Screen
to
return
to
the
display
screen.
The
following
list
describes
the
parameters
on
the
Setup
Screen:
Agent
lets
you
select
the
current
agent
type
being
delivered
to
the
patient.
The
menu
on
the
right
shows
the
agent
selection.
At
powerup,
the
No
Agent
option
is
selected.
Alarm
volume
lets
you
select
the
volume
for
alarms.
The
higher
the
number,
the
louder
the
volume.
Note:
If
the
monitor
is
powered
down
with
the
alarm
volume
set
to
one,
the
alarm
volume
upon
next
powerup
will
be
set
to
two.
Circuit
O,
lets
you disable
the
circuit
O,
alarm
and
the
display
of
the
circuit
O,
value.
Display
labels
shows
additional
descriptive
text
(I/Et,
MAX/MIN)
for
the
parameters.
After
you
know
the
parameters,
you
can
turn
off
the
extra
labels
for
a
less
cluttered
screen.
Note:
If
78xx
is
selected
for
the
RS-232
device,
the
only
available
mute
period
is
30
seconds.
Language
lets
you
select either
English,
French,
German,
Spanish,
or
Italian
for
your
screens.
If
you
change
your
screens
to
a
foreign
language
that
you
don't
speak
and
you
want
to
change
back
to
English,
the
translated
Language
para-
meter
is
as
follows:
English,
Language;
French,
Langue;
German,
Sprache;
Spanish,
Idioma;
and
Italian,
Lingua.
Mute
period
lets
you
change
the
length
of
time
all
of
the
alarms
will
be
muted.
Your
options
are
30, 60,
90,
and
120
seconds.
Note:
If
78xx
is
selected
for
the
RS-232
device,
the
only
available
mute
period
is
30
seconds.
N,0
display
lets
you
eliminate
the
display
of
nitrous
oxide.
Patient
type
lets
you
adjust
the
scale
factors
for
the
flow
sensor
to
the
patient
type
(adult
or
pediatric).
Use
the
adult
setting
for
a
tidal
vol-
ume
that
is
greater
than
250
ml/breath.
Pleth
scaling
(with
SpO,)
when
set
to
auto,
automatically
scales
the
plethysmogram
to
full-
scale.
When
set
to
non-auto,
it
fixes
the
scale
at
the
current
value.

Popular Monitor manuals by other brands

Lenovo L27m-28 user guide

Lenovo

Lenovo L27m-28 user guide

Acer 34e user guide

Acer

Acer 34e user guide

Acer XB280HKb user guide

Acer

Acer XB280HKb user guide

BenQ BL902TM user manual

BenQ

BenQ BL902TM user manual

iSMA CONTROLLI PA-LED user manual

iSMA CONTROLLI

iSMA CONTROLLI PA-LED user manual

Barco SLite 10 XP Brochure & specs

Barco

Barco SLite 10 XP Brochure & specs

Lenovo D222 user guide

Lenovo

Lenovo D222 user guide

Kenwood SM-230 manual

Kenwood

Kenwood SM-230 manual

Minarik DLC600 user manual

Minarik

Minarik DLC600 user manual

Hanns.G HH191DP Specification sheet

Hanns.G

Hanns.G HH191DP Specification sheet

Philips 247E6 user manual

Philips

Philips 247E6 user manual

Clarion VMA 5894  VMA5894 VMA5894 Owner's installation guide

Clarion

Clarion VMA 5894 VMA5894 VMA5894 Owner's installation guide

Samsung SyncMaster 323T user manual

Samsung

Samsung SyncMaster 323T user manual

Iiyama ProLite TF1634MC user manual

Iiyama

Iiyama ProLite TF1634MC user manual

ViewSonic GA655 user guide

ViewSonic

ViewSonic GA655 user guide

LG E2042T owner's manual

LG

LG E2042T owner's manual

Fermax VEO DUOX quick start guide

Fermax

Fermax VEO DUOX quick start guide

HP LD4200 user guide

HP

HP LD4200 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.