1) Turn Moxi ON
Mini Automated
Cell Counter
Your Moxi Z comes with a partial battery charge
and can be used immediately. Alternatively, you
can charge the Moxi by connecting it to a
computer or power adapter via the supplied USB
cable. Be sure to disconnect the USB cable prior
to running samples.
Before you Begin:
- Moxi Z cassettes and diluent.
- Biological cell sample.
- Pipette and tips for 75μL aliquot.
Required Materials:
Quick Start Guide
-Type S Cassettes. Size range: 3 - 20um avg. dia.
Concentration range: 3,000 to 2,500,000 cells/ml
-Type M Cassettes. Size range: 4 - 25um avg. dia.
Concentration range: 3,000 - 500,000 cells/ml
For Adherent Cells: Aspirate media and detach
cells with appropriate disassociation media
(e.g., Accutase).
Sample Prep Options:
2) Depress Tray and
Insert Cassette
3) Load
75μL
Sample
5) Data Summary
Nov 2011 Rev 3.7
Made in the USA
An electronic copy of the complete Moxi Z user
manual is on the USB ash drive that
was included in the Moxi Z kit.
User Manual:
Done 2-26umGating Delete
Dec 16 14:03:00
Count
0
500
1 305 10 15 20 25
Diameter (um)
TEST: 01
Curve Fit Count Average Size
1.28e+05 Cells/ml 9.218 um (dia)
1.55 pL (vol)
MPI= .99Type M
MPI - Moxi Population
Index
Curve Fit
Avg. Cell Size (dia. & vol.)
Count (within
curve t or gated
area)
Cell Size Scale Population Histogram
Cassette Type
Press for
Gating Mode
Touch
Gating
Dec 16 14:03:00
Count
0
500
1 305 10 15 20 25
Diameter (um)
Gated Count
1.28e+05/mL
DeleteBack
Type M MPI= .99
1.28e+05 Cells/ml 9.218 um (dia)
1.55 pL (vol)
Average Size
TEST: 01
2-26um
Curve Fit
Touch
Curve Fit
Curve Fit Results Gated Results
An electronic copy of MoxiChart is on the
USB ash drive that was included in the Moxi Z kit.
MoxiChart facilitates histogram downloading and
rmware updates via Bluetooth and is compatible
with PC and MAC.
MoxiChart Software:
Touch to Move Gates
System is Ready ...
Dec 16 14:03:00
Touch Screen When Loaded...
Back
Touch Here
Small Particle Mode
for
4) Touch Screen to Start
Touch Here
for Cells
< 8μm Dia.
Press to
Change Scale:
2-34, 2-26,
2-18, 2-10μm
Adjust Count Scale