PMK SY-5000 Series User manual

4-Quadrant Voltage Amplifier
For Resistive And Inductive Loads
SY-5001 | DC – 5 MHz | 800 V/µs | ±150 V | ±10 A
Probing Solutions.
Made in Germany.
EN
Instruction Manual
SY-5000 SERIES
SY-5002 | DC – 7 MHz | 450 V/µs | ±75 V | ±6 A

SY-5000 Series
2
Manufacturer.....................................................................................................................................
Warranty............................................................................................................................................
Safety Symbols.................................................................................................................................
Declaration of Conformity..................................................................................................................
WEEE / RoHS Directives..................................................................................................................
Safety Information.............................................................................................................................
About 4-Quadrant Voltage Ampliers.................................................................................................
Accessories Supplied........................................................................................................................
Maintenance......................................................................................................................................
Specications.....................................................................................................................................
Operating Elements - Front Panel, Back Panel...............................................................................
Graphs: Frequency and Temperature Dependent Behavior.............................................................
• Maximum Power Dissipation, Maximum Output Voltage | SY-5001...........................................
• Maximum Output Voltage | SY-5000 Series...............................................................................
• Low Signal Gain, Maximum Output Voltage | SY-5002..............................................................
• Maximum Power Dissipation | SY-5002, Output Impedance | SY-5000 Series..........................
Operation and Safety Concept.........................................................................................................
Cooling Concept and Warm-up........................................................................................................
Operating Voltages...........................................................................................................................
Connecting the Signal Source..........................................................................................................
Connecting the Load........................................................................................................................
Operational Startup..........................................................................................................................
Overload and Overcurrent Protection..............................................................................................
Gain, Oset Zero and Output Voltage Ranges | SY-5001................................................................
Remote Control................................................................................................................................
• SY-5001.....................................................................................................................................
• SY-5002.....................................................................................................................................
• Remote Command Examples...................................................................................................
Firmware Update / USB Service Port..............................................................................................
Copyright, Revision..........................................................................................................................
Contents
3
3
3
4
4
5
6
7
7
8
10
14
14
15
16
17
18
18
18
18
18
19
19
20
20
20
23
24
24
28

SY-5000 Series
3
Caution, risk of danger. Refer to manual.
Caution, risk of electric shock.
Earth (ground) TERMINAL.
IEC Safety Symbols
The following symbols may appear on the product or in this instruction manual:
Manufacturer
PMK Mess- und Kommunikationstechnik GmbH
Koenigsteinerstrasse 98
65812 Bad Soden am Taunus, Germany
Internet: www.pmk.de
E-Mail: sales@pmk.de
Phone: +49 (0) 6196 5927 - 930
Fax: +49 (0) 6196 5927 - 939
3
3
3
4
4
5
6
7
7
8
10
14
14
15
16
17
18
18
18
18
18
19
19
20
20
20
23
24
24
28
Warranty
PMK warrants this product for normal use and operation within specifications for a period of
two years from date of shipment and will repair or replace any defective product which was
not damaged by negligence, misuse, improper installation, accident or unauthorized repair
or modification by the buyer. This warranty is applicable only to defects due to material or
workmanship. PMK disclaim any other implied warranties of merchantability or fitness for a
particular purpose. PMK will not be liable for any indirect, special, incidental, or consequen-
tial damages (including damages for loss of profits, loss of business, loss of use or data,
interruption of business and the like), even if PMK has been advised of the possibility of such
damages arising from any defect or error in this manual or product.

SY-5000 Series
4
EC Directives:
WEEE Directive 2012/19/EU - Waste Electrical and Electronic Equipment
RoHS Directive 2011/65/EU - Restriction of the use of certain Hazardous Substances
in Electrical and Electronic Equipment
This electronic product is classified within the WEEE/ RoHS category list as monitoring
and control equipment (category 9) and is compliant to the following EC Directives.
Declaration of Conformity
Your help and efforts are required to protect and keep clean our environment. Therefore re-
turn this electronic product at the end of its life either to our Service Department or take care
of separate WEEE collection and professional WEEE treatment yourself. Do not dispose as
unsorted municipal waste.
WEEE/ RoHS Directives
CEI/IEC 61010-1:2010 - Safety requirements for electrical equipment for measurement,
control and laboratory use
CEI/IEC 61326-1:2013 - Electrical equipment for measurement and laboratory use - EMC
requirements Class A and basic levels
PMK declares the conformity of this product with the actual required safety standards in accor-
dance with the Low Voltage Directive (LVD) 2014/35/EU and EMC Directive 2014/30/EU:

SY-5000 Series
5
To avoid personal injury and to prevent fire or damage to this product or products connected
to it, review and comply with the following safety precautions. Be aware that if you use this
instrument in a manner not specified the protection this product provides may be impaired.
No liability is assumed for consequences which arise from non-observance of the notes and
warnings in this manual.
Do not open housing.
Device coverings may not be removed by operating personnel. Maintenance and
repair work is exclusively reserved to qualified service personnel.
Do not modify the instrument.
Do not operate the product in explosive, wet or damp atmospheres.
Indoor use only.
Do not operate in wet or damp environment. Keep the product dry and clean.
Do not operate with suspected failures.
Refer to qualified service personnel.
Keep away from hazardous live circuits.
Avoid open circuitry. Do not touch connections or components when power is present.
Ground the instrument.
The SY-5000 Series is a device of Protection Class I. In order to exclude the hazard of
electric shocks, the device housing must be grounded and therefore always operated
over the supplied three-conductor mains cable with protective ground conductor. Addi-
tionaly a 4 mm ground socket (protective earth) is available at the front panel.
Only qualied personnel should use this instrument.
Caution - danger to life.
Dangerous voltages capable of causing death might be present when using this
instrument. Use extreme caution when handling, testing and adjusting.
Safety Information

SY-5000 Series
6
About 4-Quadrant Voltage Amplifiers
The SY-5000 are linear, extremely wideband precision power ampliers. They are ideal for all
applications that require fast changing signals at any purely resistive and inductive loads.
Two selectable operating voltages modes are available for high voltage/low current and low voltage/
high current applications. SY-5001 oers in addition switchable gain factors and automatic oset
control. In case of very low impedance loads, choose the low operating voltage mode to reduce
the power loss and increase the output current.
The devices are equipped with a temperature-controlled, low noise fan. Overtemperature protection,
power-loss calculation and absolute-current monitoring guarantee perfect short-circuit and
overload protection.
The ampliers can be operated via operating elements on the front panel, and are also fully remote
controllable via USB. SY-5001 can also be remote controled via GPIB with SCPI commands.
Firmware updates can be installed by customer at any time.

SY-5000 Series
7
Accessories Supplied
Maintenance
●Cleaning the device
Wipe over the test device with a soft, moist cloth. Do not use any chemical cleaners.
●These operating instructions
●Power cable

SY-5000 Series
8
¹ Maximum non-destructive input voltage. ² Change current limit via remote control command. ³ Output stays o after recovery time. ⁴ “Maximum power dissipation“.
⁵ The generated power loss of AC signals depends on signal shape and phase angle between current and voltage.
Specications Specications that are not marked as guaranteed are typical, valid
after 30 minutes warm-up time, mains voltage: 230 V / 50 Hz.
This product comes with 2 years warranty.
SY- 5001 SY- 5002
Signal input BNC
Input impedance 50 Ω ±2 % Input impedance 50 Ω ±2 %
Gain 60 / 30 / 10 / 5 / 1 ±2 % (±100 ppm / °C) Gain 30 ±1 % (±100 ppm / °C)
Input voltage range ±3 V Input voltage range ±3.2 V
Input voltage (max.) ¹ ±3.5 V Input voltage (max.) ¹ ±5 V
Power output BNC OFF position Galvanic isolation from amplier output
and connection to GND. OFF position
Low impedance 0 V output, can be swit-
ched to 100 kΩ to signal ground via remote
command.
Output voltage
(50 Ω load,
max. 3% THD)
HIGH mode
±130 V p e a k (f < 1 MHz)
HIGH mode
±75 V p e a k (f < 1 MHz)
±65 V p e a k (f < 2 MHz) ±37.5 V p e a k (f < 2 MHz)
±43 V p e a k (f < 3 MHz) ±25 V p e a k (f < 3 MHz)
LOW mode
±64 V p e a k (f < 2 MHz)
LOW mode
±37.0 V p e a k (f < 2 MHz)
±42 V p e a k (f < 3 MHz) ±24.5 V p e a k (f < 3 MHz)
Output current
HIGH mode
Maximum
±2.5 ADC
HIGH mode
Maximum
±2.5 ADC
±5.0 A A C (f >10 Hz) ±5.0 A A C (f >10 Hz)
Impulse (< 5 ms) ±7.5 A p e a k Impulse (< 5 ms) ±6.0 A p e a k
LOW mode
Maximum
±5.0 ADC
LOW mode
Maximum
±5.0 ADC
±10.0 A A C (f >10 Hz) ±6.0 A A C (f >10 Hz)
Impulse (< 5 ms) ±15.0 A p e a k - -
Over current
switch o ±6 A ² (initial) Switch back on after 10 s ³ ±6 A Switch back on after 10 s ³
Small signal band-
width (100 mVrms
@ 50 Ω load)
-3 dB > 5 MHz -3 dB > 7 MHz
-1 dB > 3 MHz -1 dB > 3 MHz
Output oset
voltage
±1.5 mV
±0.1 mV/°C
±2 mV
±0.1 mV/°C
±3 mV max. ±5 mV max.
DC oset control automatic, controllable via rmware automatic
Slew rate
(50 Ω load) > 800 V/µs > 450 V/µs
Rise time < 380 ns ±150 V square wave @ 50 Ω load < 270 ns ±75 V square wave @ 50 Ω load
Noise
Gain 60
DC - 20 Hz 100 mV p p
Gain 30
DC - 20 Hz 1.5 mV p p
20 Hz - 10 MHz < 25 mV r m s 20 Hz - 10 MHz < 10 mV r m s
Gain 1 / 5 / 10 / 30
DC - 20 Hz 40 mV p p
-
20 Hz - 10 MHz < 6 mV r m s
Distortion
(50 Ω load)
< 0.1 % 100 kHz / 75 V p e a k < 0.1 % 100 kHz / 65 V p e a k
< 1.0 % 700 kHz / 115 V p e a k < 1.0 % 800 kHz / 70 V p e a k
< 1.0 % 1 MHz / 100 V p e a k < 1.0 % 1 MHz / 65 V p e a k
< 1.0 % 2 MHz / 43 V p e a k < 1.0 % 2 MHz / 30 V p e a k
< 1.0 % 3 MHz / 26 V p e a k < 1.0 % 3 MHz / 15 V p e a k
Output impedance 30 mΩ +0.33 µH 50 mΩ +0.3 µH
Over temperature
switch o > 70 °C Switch back on: < 50 °C ³ > 70 °C Switch back on: < 50 °C ³
Power loss ⁵
(permissible)
DC+ / DC-: 400 W up to 25 °C, linear decrease to 220 W at 70 °C DC+ / DC-: 200 W up to 25 °C, linear decrease to 110 W at 70 °C
AC > 10 Hz: 800 W up to 25 °C, linear decrease to 440 W at 70 °C AC > 10 Hz: 400 W up to 25 °C, linear decrease to 220 W at 70 °C
Specications that are not marked as guaranteed are typical.

SY-5000 Series
9
Power loss
switch o See graph ⁴ Switch back on after 10 s ³ See graph ⁴ Switch back on after 10 s ³
Interfaces USB, GPIB USB
Power input
100 V / 115 V / 220 V / 230 V / 240 V; 50 Hz / 60 Hz
Switchable by mechanical switch with protective plate
100 V / 115 V / 220 V / 230 V / 240 V; 50 Hz / 60 Hz
Switchable by mechanical switch with protective plate
Power consumption max. 800 W Power consumption max. 450 W
Overvoltage category CAT II Overvoltage category CAT II
Pollution degree 2Pollution degree 2
Safety standard IEC61010-1:2010 IEC61010-1:2010
Operating
temperature 10 °C to 40 °C 10 °C to 40 °C
Specications Specications that are not marked as guaranteed are typical, valid
after 30 minutes warm-up time, mains voltage: 230 V / 50 Hz.
This product comes with 2 years warranty.
SY- 5001 SY- 5002
³ Output stays o after recovery time. ⁴ “Maximum power dissipation“.
Specications that are not marked as guaranteed are typical.

SY-5000 Series
10
Operating Elements, SY-5001
Front Panel Elements:
[1]: ON/OFF Switch to turn the amplier from standby to „ON“ or back to standby.
[2]: OUTPUT Power connector for output (±150V / ±10A max.). In „OFF“ state the output is switched to signal ground.
[3]: OUTPUT Switch to turn the output „ON“ or „OFF“. The switch lights up when the amplier is switched on.
[4]: GAIN The gain switches are used to adjust the overall gain of the amplier. The LED of the selected ampli-
cation factor lights up permanently. Changing the gain triggers an automatic oset adjustment.
[5]: INPUT Signal input (±3V input voltage range), input impedance 50Ω, BNC.
[6]: GROUND SOCKET Protective earth connection.
[7]: OFFSET ZERO Switch to trigger, an DC oset adjustment. Lights up permanently after successful oset
adjustment, ashes at approx. 1 Hz in the event of an error during the oset adjustment.
[8]: OUTPUT VOLTAGE HIGH / LOW Switch to change the output voltage mode from HIGH (±150V) to
LOW (±75V) or vice versa. The LED lights up in HIGH voltage mode.
[9]: PROTECTION Lights up continuously at an overtemperature, ashes slowly (1 Hz) when exceeding maxi-
mum current limit or permissible power loss, and ashes quickly (3 Hz) at a hardware defect.
[10] READY Lights up permanently when the amplifier is ready for operation.

SY-5000 Series
11
CAUTION
Back Panel Elements:
[1]: GPIB CONNECTOR IEEE488-2 connector for remote interface.
[2]: USB CONNECTOR TYPE B USB service interface for software updates, and for remote control.
[3]: MAINS SWITCH All-pole mains switch for complete mains isolation of the instrument.
[4]: MAINS SOCKET Socket for a C13 connector for mains supply.
[5]: SELECTABLE MAINS VOLTAGES Selectable AC mains voltages 100V / 115 V / 220V / 230V / 240V.
Please mind the local mains voltage before connecting the device.
Failure to observe this can lead to malfunctions or even destruction of the amplier!
Operating Elements, SY-5001

SY-5000 Series
12
Operating Elements, SY-5002
Front Panel Elements:
[1]: ON / OFF Switch to turn the amplier from standby to „ON“ or back to standby.
[2]: OUTPUT Switch to turn the output „ON“ or „OFF“. The switch lights up when the output is switched on.
[3]: OUTPUT Power connector for output (±75 V / ±10 A max.) with isolated BNC socket. In „OFF“ state
the output is pulled down with 100 kΩ to signal ground.
Reference potential is connected to protective earth via 1MΩ.
[4]: INPUT 50 Ω Signal input (±3.2 V input voltage range) with isolated BNC socket. Reference potential is con-
nected to protective earth via 1MΩ.
[5]: GROUND SOCKET Protective earth connection.
[6]: OUTPUT VOLTAGE HIGH / LOW Change the operating voltage mode:
HIGH (LED on): HIGH operating mode, output voltage up to ±75 V.
LOW (LED o): LOW operating mode, output voltage up to ±37.5 V; recommended with low impedance loads.
[7]: READY / PROTECTION
Green: Lights up when the amplier is ready for operation („READY“).
Red: Protection mechanism is active (OUTPUT „OFF“).
Red LED lights up constantly: Over-temperature protection; the device switches on automatically after
the drop of the temperature.
Red LED ashes slowly (1 Hz): Disconnection because of exceeded permissible power loss.
Red LED ashes rapidly (3 Hz): Hardware fault! Stop usage immidiately! The device needs repair and
should be sent to the manufacturer.

SY-5000 Series
13
Operating Elements, SY-5002
CAUTION
Back Panel Elements:
[1]: Plug for non-heating apparatus with mains switch and fuseholder Fuse 4 AT (5x20 mm).
All-pole mains switch for complete mains isolation of the amplifier.
Socket for a C13 connector for mains supply.
[2]: Selectable mains voltages Selectable AC mains voltages 100 V / 115 V / 220 V / 230 V / 240 V.
[3]: USB-B socket USB service port for connection to a computer (emulated COM-Port) for remote cont-
rol and firmware updates.
Please mind the local mains voltage before connecting the device.
Failure to observe this can lead to malfunctions or even destruction of the amplier!

SY-5000 Series
14
Voltage [V]
Frequency [Hz]
10
100
100k 1M 10M
SY-5001 High Vp
SY-5001 Low Vp
Power [W]
Maximum power dissipation per amplifier side
Temperature [°C]
Maximum Output Voltage, SY-5001
Maximum Power Dissipation, SY-5001
Graphs: Frequency and Temperature Dependent Behavior S-5001

SY-5000 Series
15
Voltage [V]
Frequency [Hz]
10
100
100k 1M 10M
SY-5002 High Vp
SY-5002 Low Vp
Maximum Output Voltage, SY-5002
Graphs: Frequency and Temperature Dependent Behavior S-5001

SY-5000 Series
16
Low Signal Gain, SY-5002
Graphs: Frequency and Temperature Dependent Behavior SY-5002
Maximum Output Voltage, SY-5002
Maximum Amplitude [V]
Frequency [Hz]
High Voltage Mode
Low Voltage Mode
k k kM M
Low Signal Gain [dB]
Frequency [Hz]
S11 / Ohm
Network Analyser HP8751A (S.-No.: 3315J01756)
PMK SY-5002, Low Signal Gain
Input Signal: 100 mV, Load: 50 Ohm
k k
kMM

SY-5000 Series
17
Graphs: Frequency and Temperature Dependent Behavior SY-5002
Output Impedance [Ω]
Frequency [Hz]
High Voltage Mode
k
k k M M
Maximum Power Dissipation, SY-5002
Output Impedance, SY-5000 Series
Power [W]
Temperature [°C]

SY-5000 Series
18
Cooling Concept and Warm-up
Operation and Safety Concept
Operating Voltages
Provide sucient space behind and in front of the amplier, that air can circulate through the
device unobstructed. Cool air is sucked through the slots at the left and right side of the front
panel. Allow the amplier a warm-up phase of 30 minutes after switching on for stabilization of
its operating points.
Before the amplication begins, some considerations related to the selection of the operating
voltage mode are recommended for eective and safe operation.
The amplier modes have the following operating voltages:
• HIGH mode for high output voltages and low load currents
• LOW mode for low output voltages and high load currents
In order to keep the power loss of the amplifier low, the operating voltage should always be
selected according to the load.
Connecting the Signal Source
Connect the BNC input to a suitable 50Ω source, e.g. the signal output of your IWATSU B-H
analyzer. To fully drive the amplier use an input voltage of ± 2.5V at 50Ω. Use high quality co-
axial cables and connectors to avoid signal problems such as loose contacts.
The input is connected with its reference potential to the protective conductor. Make sure
short signal lines and close sockets do not create any disturbing hum loops when using
an earthed source.
SY-5000 series power ampliers are equipped with various safety mechanisms in order to protect them-
selves. If the amplier self-protection is activated check your application setup, all connections and wiring.
Connecting the Load
The power output of the amplier has a BNC connector. The reference potential of the BNC output is
connected to protective earth. Ensure sucient contact protection of the cabling to your load. Life-threa-
tening voltages can occur.
The output line is part of the load of the amplier. The inductance of this supply line is added to the output
impedance of the amplier, i.e. the source impedance becomes higher at higher frequencies. Keep the
output lines as short as possible in order to feed as much current as possible into your load.
The SY-5001 is optimized for resistive and inductive loads. Long coaxial lines at the output can
therefore be problematic, as they represent considerable output capacities and make the amplier
unstable. All capacities greater than 300 pF can become critical.

SY-5000 Series
19
The SY-5002 is stable at all complex loads. Since the output impedance of the amplier gets
inductive at higher frequencies, this may lead to a signicant overshoot with predominant capaci-
tive loads. This behavior does not indicate a possible instability of the amplier, but is established
due to an excited series resonant circuit of output and line inductance and connected capacity.
In this case, damping the resonant circuit by series resistors is recommended.
Operational Startup
Operation with wrong mains voltage leads to malfunctions or even destruction of the amplier.
Make sure the mains voltage switch is set to the correct mains voltage before connecting
the mains cable to the amplier. Switch on mains voltage with the switch on the mains
socket. The green mains switch on the front panel illuminates and signals the standby
state. Press this button to switch the amplier on completely, fans start running and the
amplier becomes READY after a short period of time. At startup, the amplier is set
to its initial state, which also can be dened by remote control command. Select best
operating voltage mode for your load and connect it to the signal input of your source.
30 minutes warm-up time is recommended. After warm-up phase press OFFSET to start
oset compensation. Activate the amplier with the illuminated OUTPUT button to start
signal processing.
Factory default setting is a gain of 60 at HIGH operating voltage for SY-5001. The SY-
5001 already carries out an initial oset during the switch-on phase. The OFFSET ZERO
LED will light up permanently and indicate a successful oset adjustment. Press the OUTPUT
button to release the output. The amplier supplies the input signal amplied by the gain factor
as output voltage.
Overload and Overcurrent Protection
The SY-5000 series has several partly interleaved safety mechanisms which protect the amplier against
overtemperature and overload.
At a heat sink temperature of 70°C the amplier switches o, the READY LED turns o and the PRO-
TECTION LED lights up permanently. If this occurs, check air supply and proper operation of the fans.
If the temperature falls below 50°C again, the amplier automatically becomes ready.
The amplier switches o, if the remotely controllable short-circuit current or constant DC or AC
(>10Hz) current is exceeded, or if permissible temperature-dependent power dissipation is exceeded.
The READY LED is o and the PROTECTION LED ashes with 1 Hz. The amplier automatically be-
comes ready after 10 s, but the output is not switched on automatically. In case of a presumed power
dissipation, operation with a low operating voltage should be considered. Further information on which
protection is active can be requested by remote control.
If the PROTECTION LED ashes with about 3 Hz, there is a hardware error. In this case
switch o the device and wait at least 1 minute. If errors still occur after the hardware has
been switched on again or occur again during operation, the amplier should be analyzed
at the manufacturer. Do not continue to use the amplier.

SY-5000 Series
20
Integration into Automated Test Systems, SY-5001
Remote Control
• The commands for the GPIB interface are ASCII commands and follow the IEEE488.2
standard for SCPI commands.
• Entries in []-brackets are optional and can be omitted.
• Commands with ? are queries.
• Each command is valid in its short form (capital letters) as well as in its long form.
The commands are not case sensitive.
The entire functionality of the SY-5001 can be remote controlled via USB, or GPIB interface with SCPI
commands, which makes it ideal in automated test environments. The USB interface is installed as
a virtual COM port at 9600 baud, 8 data bits, 1 stop bit and no parity. For more information see
chapter „USB Service Port“.
Remote adjustment of the amplier settings to the load can greatly facilitate operation with switchable
output, selectable operating voltage and adjustable gain with oset control. Complete amplier monito-
ring with heat sink temperature, power dissipation and detailed error query enable safe automated test
sequences.
Basic Commands
*IDN? Query of device ID, e.g. „PMK, SY-5001, 18901980-0101, V1.6“.
*OPC? Operation Complete query returns „1“ after all previous commands are completely
executed.
*RST Reset into dened basic state (Gain = 60, Input = o, Output
Voltage Range Auto = on ), a possible error list is deleted.
*RCL <0...3> Recall previously saved device states.
*SAV <0...3> Save the state of the device under a dened memory number. Memory number
“0” denes the start conguration when the device is switched on.
Gain, Offset Zero and Output Voltage Ranges, SY-5001
Use the GAIN buttons to select the gain of SY-5001. For maximum output voltage, gain 60 must be set,
as the input voltage range is limited to ± 3 V. The maximum output voltage is set by the gain buttons. Other
gain settings may be useful for matching to the signal source. The output noise reduces slightly when the
gain is less than 60. Change the gain to activate automatic oset adjustment. For rmware V 1.5 or lower
automatic oset adjustment starts when gain is changed from or to 60. During changeover the output
is switched o, and automatically on after the oset adjustment is completed. Oset adjustment can be
triggered at any time by pressing the OFFSET ZERO switch or by remote control command.
In automatic mode, the output voltage mode is adjusted to the gain setting. If gain < 60 is selected, the
output voltage mode is set to LOW. The output voltage mode is set to HIGH when a gain of 60 is se-
lected. Automatic switchover remains active until the output voltage mode is manually set, changed
by remote control, or automatic mode is switched o. Automatic mode can also be reactivated by
remote control.
This manual suits for next models
2
Table of contents
Popular Amplifier manuals by other brands

Marantz
Marantz PM-15S1G Brochure & specs

Premier Builders Guild
Premier Builders Guild Jet Signature owner's manual

Acom
Acom 1200S Installation, operation and maintenance

Tecnare Sound Systems
Tecnare Sound Systems T-48 Series operating manual

Yaesu
Yaesu FL-2100Z instruction manual

Classe Audio
Classe Audio CAV-75 owner's manual