RadonAway RadStar RS300 User manual

RadStar RS300
Operating Instructions
RS300 FEAT RES
Indicator Light
Power Jack
Printer Port
• 12 or 24-hour start time
delay option
• Internal memory to retain data
Print Button
Clear Button
ey Switch
INO28 - Rev D
• Accurate measurement in a wide range
of temperatures and relative humidity
• Nominal sensitivity: 16 cph/pCi/L

1.0 SETTING P THE RS300
2.0 T RNING NIT ON
3.0 ERASING STORED DATA
4.0 MONITORING FOR RADON GAS
5.0 OBTAINING A READING FROM THE RS300
6.0 SETTING P THE OPTIONAL PRINTER
7.0 OBTAINING A TEST REPORT PRINTO T
8.0 DOWNLOADING DATA FROM THE RADSTAR
9.0 TRANSPORTING THE RADSTAR TO A PC OR PRINTER
FIG RE A: EXAMPLE RS300 TEST RES LT PRINTO T
Table of Contents
3
3
3
3
4
4
4
5
5
6

2.0 T RNING NIT ON:
2.1 Insert key into unit and turn it one click to the right to the [ eypad On] position.
The unit will beep.
3.0 ERASING STORED DATA:
3.1 Press and hold the [ Hold to Clear ] button. There will be an audible tone coming
from the unit when this is done. Release the [ Hold to Clear ] button after the tone
stops. There will be a short “beep” to signify that the memory has been erased. This is
done to erase any prior test that has been stored in the memory so that the unit will
test from the first hour. If this is not done, the unit will start testing from the hour that it
was shut off on the last test.
3
1.0 SETTING P THE RS300:
1.1 Plug power adapter that is marked “RadStar Power Supply” into any 120VAC
outlet.
1.2 Plug free end of power adapter into left side of RS300-Green LED should come on.
This tells you that the NiCad batteries are charging. Power cord should be left plugged
in at all times during a radon test to ensure that the batteries remain charged. If there
is no power available, such as in new construction, the unit may be run for a typical
48hr test on the batteries alone as long as the unit has been charging with the power
adapter over night.
4.0 MONITORING FOR RADON GAS:
4.1 You may begin monitoring for radon gas by turning the keyswitch to the
[Measure On] position. The key may be removed in this position to prevent tampering
with the unit.
4.2 The red LED next to the [ Measure On ] logo will stay on for forty seconds then go
out. This is a normal condition. Every thirty seconds after there LED will blink. This is a
normal condition.
Note: When the key is turned to the [ Measure On ] position to start a new radon test
and the unit “beeps”, it means that the data needs to be erased from the prior
test. See section 3.0.
Note: est duration per EPA Protocols must be at least 48 hours.

4
5.0 OBTAINING A READING FROM THE RS300:
5.1 Any time during or after a test turn the keyswitch to [ eypad On] position.
Note: o continue testing at anytime when the keyswitch has been turned to the
[ Keypad On ] or the [ Off ] position, simply turn the keyswitch to [ Measure On ]
position. Note that the unit will “beep” to signify continued testing.
Note: When a test is continued by going from [ Keypad On ] to [ Measure On ] the
hour that you were testing in starts over again. Also note that if you try to look at
the readings before the first hour is up the PC or printer will read: “No Data”.
6.0 SETTING P THE OPTIONAL PRINTER:
6.1 Plug the AC/DC adapter (the one marked “Printer Power Supply”) into any 120VAC
outlet. Plug the power jack from the adapter into the back of the printer.
6.2 Connect the D-Sub cable supplied with the unit to the printer and the RadStar.
6.3 Turn the printer on via the rocker switch. The red LED on the printer should come on.
7.0 OBTAINING A TEST REPORT PRINTO T:
7.1 After completing a test turn the keyswitch to [ eypad On ] position.
7.2 Press and hold down [ Print ] button until unit beeps. The unit will print the test
report. See Fig.A (p.6) for an explanation of the printout.
Note: If you try to get a printout within the first hour of testing, the printer will
display “No Data to Print”.
7.3 To obtain multiple printouts, simply press the [ Print ] button again.

5
9.0 TRANSPORTING THE RADSTAR TO A PC OR PRINTER:
9.1 If desired, you may transport the RadStar to a PC or printer to retrieve the data.
Turn keyswitch to [ Off ] position remove key and disconnect power supply. When your
destination is reached, simply insert key into unit and turn to [ eypad On ] and print or
download the data.
8.0 DOWNLOADING DATA FROM THE RADSTAR:
The RadStar comes with a Data Capture Utility Program (Windows XP and Vista
compatible), and a USB/Serial Adapter Cable with driver. The accompanying RadStar
Documentation Library describes the set-up and operation of these items. You may
also view the Documentation and download the Program from the AccuStar Labs
website www.accustarlabs.com.
8.1 Install the RadStar Data Capture Utility Program and cable driver on your PC. In the Program,
setup the correct RadStar model number and the correct COM port. Refer to the RadStar
Documentation Library for specific instructions.
8.2 Connect the PC cable to the Print Port on the RadStar, and to your PC. Turn the
RadStar keyswitch to [ eypad On ] position.
8.3 In the Data Capture Utility Program on your PC, click Download Data. The Set
Delay menu will appear. No Delay, 12-Hour Delay, or 24-Hour Delay must be selected.
Note: o meet EPA protocols minimum test durations are 48, 60 and 72 hours for
No Delay, 12-Hour Delay, or 24-Hour Delay respectively.
On the RadStar, press the [Print] button once. The data will download to the PC.
Note: Clicking Download Data on the PC after pressing Print on the RadStar Unit will
result in incomplete data and an error message when exporting results to the Excel Chart.
Note: If you try to download data from the RadStar within the first hour of
testing, the PC will display "No Data".
8.4 The test data downloaded into the Data Capture Utility can then be exported into a
text file, comma delimited file or Excel chart. README files with the software describe
the set-up and operation of the Data Capture Utility. Data Exported depends on the
Delay selected when the Data was downloaded.
“No Delay” - All data from hour 001 is exported. Hours 001-004 are ignored when cal-
culating Average, Min and Max. Plot on Excel chart starts at hour 005.
“12-Hour Delay” - All data from hour 013 is exported. Hours 013-016 are ignored when
calculating Average, Min and Max. Plot on Excel chart starts at hour 017.
“24-Hour Delay” - All data from hour 025 is exported. Hours 025-028 are ignored when
calculating Average, Min and Max. Plot on Excel chart starts at hour 029.

FIG RE A
EXAMPLE RS300 TEST RES LT PRINTO T
6

Manufactured by
RadonAway
3 Saber Way
Ward Hill, MA 01835

Mode: Continuous Monitor, Passive Diffusion
Principle of Operation: Pulse Mode Ionization Chamber
Minimum Sensitivity: 0.27 cpm/pCi/L (minimum 64 alpha counts per hour @ 4 pCi/L)
Units: picoCuries per liter (pCi/L)
Dynamic Range: 0.5 to 150 pCi/L
Linearity: +/- 10% Best Straight Line
Data Storage (Nonvolatile): Up to 240 hourly readings, wrap on over-run.
Power: 120VAC 60Hz Transformer, 8 days rechargeable battery operation.
Indicators: Green AC Power LED, Red Testing LED
Key Switch: 3 Position, Removable ey [ Off ] [ eypad On ] [ Measure On]
Pushbuttons: Momentary [ Print ] [ Hold to Clear ]
Communications: RS-232 port for printing or computer uploads
amper Protection: Tilt switch, AC power fail detection, key switch lock-out.
Operating Range: Temperature 50 F–90 F Humidity 0-80%, non-condensing
Size:
Height 7 inches (includes Carrying Handle)
Width 8 inches
Depth 4 ½ inches
Weight 2 ½ pounds
Color: Green/Black
Mounting: (2) ¼”x 20 Threaded Tripod Mounts
RS300 Instrument Specifications
Manufactured by
RadonAway
3 Saber Way
Ward Hill, MA 01835
Technica Support, Ca ibration
and Repair Service.
800-767-3703
M-F 8:30am - 5:30pm ET
Table of contents
Popular Monitor manuals by other brands

Steltronic
Steltronic ST-HB32 User and installation manual

Siebert
Siebert Series S102 operating instructions

Sanyo
Sanyo AVL-3210 instruction manual

HP
HP V27c G5 Maintenance and service guide

TSI Instruments
TSI Instruments PRESSURA RPM10 Installation instructions and safety information

Ikan
Ikan MS21 user guide