Rion UV-16 User manual

3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan
http://www.rion.co.jp/english/
INSTRUCTION MANUAL
Two-Channel Charge Amplier
UV-16


i
Organization of this manual
This manual describes the features, operation, and other aspects of the Two-Chan-
nel Charge Amplier UV-16. If the unit is used together with other equipment to
congure a measurement system, consult the documentation of all other components
as well. The section starting on page iii contains important information about safety.
Be sure to read and observe these precautions in full.
This manual contains the following sections.
Outline
Gives basic information about the unit and contains a block diagram.
Names of Parts and Functions
Briey identies and explains the front panel controls, rear panel connec-
tors and all other parts of the unit.
Linking
Explains how to congure multiple units as a system.
Power Supply Connection
Explains how to make connections for powering the unit.
Accelerometer Selection and Connection
Explains the types of accelerometers that can be used, how to connect them,
and how to set them up for measurement.
Output Connections
Explains how to use the output connectors on the rear panel.
Measurement
Explains the steps for setup and measurement.
Performance Characteristics
Contains charts that show the characteristics of the high-pass lter and low-
pass lter as well as velocity (VEL) and displacement (DISP) frequency
response characteristics.

ii
* Company names and product names mentioned in this manual are
usually trademarks or registered trademarks of their respective
owners.
Specications
Lists the technical specications of the unit.
Reference Material
Provides representative data for accelerometer inherent noise. Maintenance
parts are also listed in this section.

iii
FOR SAFETY
In this manual, important safety instructions are specially marked as shown
below. To prevent the risk of death or injury to persons and severe damage
to the unit or peripheral equipment, make sure that all instructions are fully
understood and observed.
Caution
WARNING
Important
Disregarding instructions
printed here incurs the risk of
death or severe injury to per-
sons.
Disregarding instructions
printed here incurs the risk of
injury to persons and/or dam-
age to peripheral equipment.
Disregarding instructions
printed here incurs the risk of
damage to the product.
Mentioned about the tips to
use this unit properly. (This
messages do not have to do
with safety.)
Note

iv
To conform to the EU requirement of the Directive
on Waste Electrical and Electronic Equipment,
the symbol mark on the right is shown on the
instrument.
The product described in this manual is in conformity with the
following standards;
Electrical equipment for measurement control and laboratory
use.
RION Co., Ltd.
3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan
BNC-BNC
BNC-BNC
AB
AB
AB
AB
AB
AC adapter
NC-99 series
Accelerometer
Core
Core
Core A: ZCAT2749-0430 (TDK)
Core B: ZCAT2436-1330 (TDK)
Equivalent product
This product can be used in any areas including residential
areas.

v
Precautions
•Operate the unit only as described in this manual.
•Do not disassemble the unit or attempt internal alterations.
•Observe the following precautions before using the unit:
- Make sure that all connections are properly and safely established.
- Make sure that the unit is operating normally.
•The permissible ambient temperature range for operation of the unit is
−10 to +50°C. Relative humidity must be 90% or below.
•Do not use or store the unit in locations which
- may be subject to strong magnetic elds or strong radiation, or
- may be subject to high levels of dust or splashes of water, or
- may be subject to gases or air with high salt or sulphur content, or
are in the vicinity of stored chemicals, or
- may be subject to high temperature, humidity, or to direct sunlight,
or
- may be subject vibrations or shock.
•Always switch off the power after using the unit.
•When disconnecting cables, always hold the plug and do not pull the
cable.
•Use only the specied AC adapter or other specied power source.
•This is a precision device. Take care not to drop the unit and protect it
from shocks.
•The LCD panel of the unit can easily become scratched. Do not tap the
panel with a pointed object such as a pencil, screwdriver, etc.
•In case of malfunction, do not attempt any repairs. Note the condition of
the unit clearly and contact the supplier.
•When disposing of the unit, follow national and local regulations regard-
ing waste disposal.

vi
Contents
FOR SAFETY............................................................................... iii
Precautions......................................................................................v
Outline............................................................................................1
Names of Parts and Functions.........................................................3
Front Panel ................................................................................3
Rear Panel .................................................................................5
Display screen............................................................................6
Bottom Panel .............................................................................8
Top Panel...................................................................................8
Linking ...........................................................................................9
Linking procedure .....................................................................9
Removing the link connector ................................................... 14
Linking when using the Battery Unit BP-17.............................15
Using a single UV-16 unit ........................................................16
Power Supply Connection..............................................................17
Using an AC adapter................................................................18
Using the Battery Unit BP-17...................................................19
Accelerometer Selection and Connection ......................................21
Accelerometer connection........................................................22
Accelerometer Mounting..........................................................23
Output Connections.......................................................................25
Measurement.................................................................................26
Operation mode .......................................................................26
Default settings...................................................................26
Accelerometer setup mode.......................................................27
Setup mode..............................................................................29
Measurement mode.................................................................. 31
Bar graph indication...........................................................33
Output signal......................................................................33
Overload indicator LED .....................................................34
Calibration (CAL) mode ..........................................................35

vii
Performance Characteristics..........................................................37
Specications ................................................................................39
Reference Material ........................................................................44
Lower measurement limit ........................................................44
Maintenance parts ...................................................................45

viii

1
Outline
The UV-16 is a two-channel charge amplier for vibration measurements using
piezoelectric accelerometers or accelerometers with integrated preamplier.
By linking the chassis of up to 16 units, a multi-channel conguration with
up to 32 channels is possible. The main unit is equipped with an LCD panel,
alarm LEDs, operation panel, input connectors (micro-dot), output connec-
tors (BNC), DC IN jack, and link connector.
Power is supplied by the optional AC adapter NC-97 series (up to 10 UV-16
units), or NC-99 series (up to 16 UV-16 units) (CE mark). In congurations
with a low number of channels (up to 3 units), the dedicated Battery Unit
BP-17 (option) with 8 IEC R14 (size C) batteries can also be used.

2
Outline
UV-16 Block Diagram
Channel A
input
Channel A AC output
Charge amplier
-20 dB
Charge amplier
0 dB
HPF OFF
HPF 10
Hz
HPF 20
Hz
HPF 50
Hz
Integrator
AMP 2
AMP 1
AMP 4
AMP 3
LCD
CPU
Sensitivity calibration amplier
Memory I/O
Channel B AC output
Power
supply
+12
V
-12
V
+3.3
V
+5
V
+24
V
9 V to 15
V DC
From link connector
Amplier -20 dB
Amplier 0 dB
Average detector
ACC (m/s2)
VEL (mm/s)
DISP (mm)
LPF 1
kHz
LPF 3
kHz
LPF 10
kHz
LPF OFF
OVER
OVER
OVER
Constant-current power
supply 24 V, 4 mA
Integrator
Channel B
input
Charge amplier
-20 dB
Charge amplier
0 dB
Amplier -20 dB
Amplier 0 dB
Constant-current power
supply 24 V, 4 mA
HPF OFF
HPF 10
Hz
HPF 20
Hz
HPF 50
Hz
ACC (m/s2)
VEL (mm/s)
DISP (mm)
LPF 1
kHz
LPF 3
kHz
LPF 10
kHz
LPF OFF
OVER
Integrator
Integrator
AMP 2
AMP 1
AMP 4
AMP 3 Average detector
Sensitivity calibration amplier

3
Names of Parts and Functions
Front Panel
Channel A overload indicator LED
Lights up in red when saturation occurs in the internal amplier circuitry
for channel A.
While the LED is lit, correct measurement is not possible.
Channel B overload indicator LED
Lights up in red when saturation occurs in the internal amplier circuitry
for channel B.
While the LED is lit, correct measurement is not possible.
SETUP key
Pressing this key switches the unit to the following setup modes.
ACC (acceleration)/VEL (velocity)/DISP (displacement) setup
High-pass lter (HPF) setup
Low-pass lter (LPF) setup
CHARGE
CCLD SENS
m/s2
mm/s
mm
Hz
kHz
CHARGE
CCLD SENS
m/s2
mm/s
mm
HPF
LPF
Hz
kHz
HPF
LPF
A
B
A
PICKUP
B
2 CHANNEL CHARGE AMPLIFIER
UV-16
PICKUP SETUP
RANGE RANGE
Channel A
overload indicator LED
Channel B
overload indicator LED
Channel B
PICKUP connector
Channel A
PICKUP connector
SETUP key
LCD panel
PICKUP key
Channel A RANGE / keys Channel B RANGE / keys

4
Names of Parts and Functions
Channel B RANGE / keys
Serve to select the range for channel B.
The key is also used to activate the CAL (calibration) mode.
Channel B PICKUP connector
Serves for connection of the piezoelectric accelerometer or accelerometer
with integrated preamplier.
Channel A PICKUP connector
Serves for connection of the piezoelectric accelerometer or accelerometer
with integrated preamplier.
Channel A RANGE / keys
Serve to select the range for channel A.
The key is also used to activate the CAL (calibration) mode.
PICKUP key
Pressing this key switches the unit to the accelerometer setup modes
below.
Setup for piezoelectric accelerometer/accelerometer with integrated
preamplier
Sensitivity setup
LCD panel
Shows various setup parameters and comprises two bar graph displays.
(For details, see page 6.)

5
Names of Parts and Functions
Rear Panel
Channel A AC OUT connector
Supplies an AC signal for channel A.
Channel B AC OUT connector
Supplies an AC signal for channel B.
Name plate
Shows the serial number and other information about the unit.
DC IN jack
Serves for connection of the optional AC adapter or an external DC
source.
AC OUT
ch.A
ch.B
DC IN
9-15V
+
15
No. 12345678
RION CO.,LTD.
MADE IN JAPAN
DC IN jack
Channel A
AC OUT connector
Channel B
AC OUT connector
Name plate

6
Names of Parts and Functions
Display screen
The display serves for both the A and B channel, but the items are the same,
so the following explanation covers them together.
The illustration below shows all possible functions of the display. In actual
operation, not all indicators will light up simultaneously.
CHARGE/CCLD indicator
Shows the accelerometer setting. When using a piezoelectric acceler-
ometer, select CHARGE. When using an accelerometer with integrated
preamplier, select CCLD.
SENS indicator
When the range/sensitivity readout shows the accelerometer sensitivity,
the SENS indicator lights up.
Range/sensitivity readout
Normally, this shows the range setting.
In calibration mode, the indication "CAL" is shown.
In accelerometer setup mode, the sensitivity setting is shown.
CHARGE
CCLD SENS
m/s2
mm/s
mm
Hz
kHz
CHARGE
CCLD SENS
m/s2
mm/s
mm
HPF
LPF
Hz
kHz
HPF
LPF
A
B
SENS indicator
Channel A section
Bar graph
CHARGE/CCLD
indicator
Range/sensitivity readout
Unit indicators
HPF indicators
LPF indicators
CHARGE/CCLD indicator
SENS indicator
Range/sensitivity readout
Unit indicators
HPF indicators
LPF indicators
Bar graph
Channel B section

7
Names of Parts and Functions
Unit indicators
Show which unit is selected.
ACC (acceleration): m/s2
VEL (velocity): mm/s
DISP (displacement): mm
HPF indicators
Show the high-pass lter setting (OFF, 10 Hz, 20 Hz, 50 Hz). OFF is
indicated by "--".
LPF indicators
Show the low-pass lter setting (1 kHz, 3 kHz, 10 kHz, OFF). OFF is
indicated by "--".
Bar graph
Shows the level of the accelerometer signal as a bar graph indication. The
refresh rate of the display is 100 milliseconds.
Note
The display backlight is always on when the unit
is powered from an AC adapter. When the unit
is powered from batteries using the Battery Unit
BP-17, the display backlight is activated for 10
seconds when any key is pressed.

8
Names of Parts and Functions
Bottom Panel
To access the link connector for linking units, remove the cover plates on
the bottom of the unit. The connector is used to supply power to multiple
units.
Use the supplied link plate to join multiple units.
Top Panel
The top panel of the unit has two hooks which can be used to join multiple
units together.
Removing the cover plates gives
access to the link connector.
Bottom view Cover plates
Link hookLink hook Top view

9
Linking
Multiple UV-16 units can be linked to form a multi-channel system.
The maximum number of units that can be linked is 16 (resulting in
32 channels).
When using the Battery Unit BP-17 to power the units, the maximum number
of units that can be linked is three.
Required tool: 1 Phillips screwdriver (JIS B 4633:1998, type H, No. 2)
Important
Precautions for linking units
•
Make sure that power is OFF before starting:
•Disconnect AC adapter.
•When using Battery Unit BP-17: Remove
batteries
•Do not use any screws other than the ones
supplied with the product.
•Do not insert any objects (pieces of metal,
pencil lead, etc.) into the unit.
•Use only the specied screwdriver (Phillips
screwdriver, JIS B 4633:1998, type H, No.
2). Otherwise the cross-shaped slots on the
screw heads may be damaged.
Linking procedure
1.
Remove the 4 screws holding the cover plates on the bottom of the unit, and
remove the cover plates. Do this for all units that are to be connected.
Put the screws and cover plates in a box or other suitable container, to
make sure that you do not lose any parts. The parts will be required
later for reassembly.
Unit bottom
Screws (4)
Cover plates

10
Linking
2. Remove the two link hooks on the top panel. Do not remove these
hooks for the unit that will be the rightmost unit as seen from the
front.
Put the screws and hooks in a box or other suitable container, to
make sure that you do not lose any parts. The parts will be required
later for reassembly.
3. For the leftmost unit as seen from the front, roll up the link cable
around the connector and push it into the empty space in the unit.
3M
J3490-3
For leftmost unit, roll up
link cable along with
connector and tag,
and push into gap.
Link hook and
fastening screws
CHARGE
CCLD
SENS
m/s
2
mm/s
mm
HPF
LPF
Hz
kHz
CHARGE
CCLD
SENS
m/s
2
mm/s
mm
HPF
LPF
Hz
kHz
A
B
A
PICKUP
B
RANGE RANGE
PICKUP SETUP
UV-16
Table of contents
Other Rion Amplifier manuals