Specifications
*Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load).
In order to use the freezer at a stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher
than the indicated lowest temperature. In addition, depending on the usage conditions, it may not be possible to reach the indicated
lowest temperature.
HFC
Model
LCO2backup system
LN2backup system
Backup system mounting kit
Circular recorder
Chart paper
Ink pen
Slip chart recorder
Chart paper
Ink pen
Recorder mounting bracket
Sensor cover
Storage racks
3 drawers (for lower chamber)
SANYO DAQ (
Data Acquisition
)
system
MDF-C8V
CVK-UB4
—
MDF-UBK
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
MTR-C8
—
—
MTR-480
MTR-2000
MDF-U32V
CVK-UB2
CVK-UBN2
—
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
—
MDF-30SC
MDF-30R
MTR-480
MTR-2000
MDF-U53V
CVK-UB2
CVK-UBN2
—
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
—
MDF-70SC
MDF-50R
MTR-480
MTR-2000
MDF-U73V
CVK-UB2
CVK-UBN2
MTR-G85
RP-G85
PG-R
MTR-85H
RP-85
DF-38FP
MDF-S3085
(for MTR-85H)
—
MDF-70SC
—
MTR-480
MTR-2000
MDF-C2156VAN
—
Equipped
—
—
—
—
MTR-155H
RP-155
DF-38FP
MDF-S30150
(for MTR-155H)
—
—
—
MTR-480
MTR-2000
Options
Distributed by:
We design and manufacture products in accordance with ISO9001 & 14001 requirements.
RoHS (European Restriction of Hazardous Substances directives) compliant
• All data shown here are measured based on our company standards. If detailed
data is required, please contact our company.
• If product inspection records are required, please contact our company. (There
will be a charge for the records.) Please note that the records do not include tem-
perature adjustment range or data on temperature distribution inside the unit.
Performance Data
40
20
0
–20
–120
–40
–60
–80
–100
076528341
Temperature (˚C) Temperature (˚C)
Time (hours)
525201510 30 35
20
10
15
5
Time (hours) Time (hours)
Ambient temperature (˚C)
Ambient temperature (˚C)
02010 30 40
20
25
10
15
5
Time (hours)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
02 2046810 181612 14
Time (hours)
Time (hours)
Ambient temperature (˚C)
Temperature (˚C)
525201510 30 35
20
10
15
5
Ambient temp. and maintenance
of –70°C by injection of LCO2(30kg)
525201510 30 35
20
10
15
5
Time (hours)
Ambient temperature (˚C)
525201510 30 35
10
15
20
5
Time (hours)
Ambient temperature (˚C)
MDF-C2156VAN
LN2backup system
operation characteristics
MDF-C8V
MDF-U32V
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO2(30kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO2(30kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –135°C by injection of LN2(50kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
Ambient temp. and maintenance
of –70°C by injection of LCO2(30kg)
Central air temp. in cabinet
Ambient temp. 30°C, No-load
MDF-U73V
MDF-U53V
Temperature (˚C)
Time (hours)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0126543210111987
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0126543210111987
Time (hours)
Temperature (˚C)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0126543210111987
CO2backup system
operation characteristics
Pull-up characteristics
during power failure
Pull-down characteristics
Further details regarding our products can be accessed at http://www.sanyo.co.jp/cmg/biomedical
Model
Temperature range
Maximum cooling performance
Exterior dimensions (W x D x H)
Interior dimensions (W x D x H)
Net weight
Effective capacity
Shelves
Access port
Compressor
Refrigerant
Alarm system
Remote alarm contact
Accessories
Model
Temperature range
Maximum cooling performance
Exterior dimensions (W x D x H)
Interior dimensions (W x D x H)
Net weight
Effective capacity
Access port
Compressor
Refrigerant
Alarm system
Remote alarm contact
Accessories
MDF-C8V
-60 to -80°C (1°C increments)
-80°C (Ambient temp. 30°C)
550 x 685 x 945 mm (21.6” x 27.0” x 37.2”)
405 x 490 x 425 mm (15.9” x 19.3” x 16.7”)
Approx. 67 kg (148lbs.)
84 Liters (3.0 cu.ft.)
17 mm diameter, 2 locations backside, bottom
Rotary type, Output: 400W
High/Low temp. alarm (±5°C to ±20°C),
Power failure alarm, Remote alarm contact:
DC30V, 2A
1 set of keys, 1 scraper
MDF-U73V
1010 x 875 x 2010mm
(39.8” x 34.4” x 79.1”)
870 x 600 x 1400mm
(34.2” x 23.6” x 55.1”)
Approx. 350kg (772 lbs.)
728 Liters (25.7 cu.ft.)
MDF-U53V
–50˚C to –86˚C (1˚C increments)
–86˚C (Ambient temp. 30˚C)
770 x 875 x 1990mm
(30.3” x 34.4” x 78.3”)
630 x 600 x 1380mm
(24.8” x 23.6” x 54.3”)
Approx. 299kg (659 lbs.)
519 Liters (18.3 cu.ft.)
Stainless steel, 3 shelves
17mm diameter, 3 locations (back, bottom left/right corners)
Hermetic type, Output:
1100 W (high stage side), 1100 W (low stage side)
High/low temperature, Power failure, Filter check,
Self diagnostics, Door check,
Part replacement nortification
Allowable contact capacity: DC 30V, 2A
1 set of keys, 1 scraper
MDF-U32V
670 x 867 x 1860mm
(26.4” x 34.1” x 73.2”)
490 x 600 x 1140mm
(19.3” x 23.6” x 44.9”)
Approx. 258kg (569 lbs.)
333 Liters (11.8 cu.ft.)
Hermetic type, Output:
400 W (high stage side),
600 W (low stage side)
Temperature, Power failure,
Filter check, Self diagnostics,
Remote alarm contact
MDF-C2156VAN
-125°C to -150°C (1°C increments)
-150°C (Ambient Temp. 30°C)
1730 x 765 x 1010 mm (68.1” x 30.1” x 39.8”)
760 x 495 x 615 mm (29.9” x 19.5” x 24.2”)
Approx. 319 kg (703lbs.)
231 Liters (8.2 cu.ft.)
40 mm diameter, 1 location
Hermetic type, Output: 1100W (high stage
side), 1100W (low stage side)
HFC
High/Low temperature, Power failure,
Filter check, Self diagnosis, Door check,
Part replacement notification
Allowable contact capacity, DC 30V, 2A
1 set of keys, 1 scraper, Connecting tube for
LN2backup system
SANYO Electric Co., Ltd.
Biomedical Division
©2008 SANYO Printed in Japan 2008.3 MA
SHR119