SeQual Eclipse 1000 User manual

Model 1000 User Manual


Important Safety Instructions 2
Symbols Used in User Manual and on Oxygen System 2
Indications For Use 5
Contraindications 5
Safety Guidelines 6
Oxygen Therapy Safety 6
Operating Safety 7
Power Cartridge Safety 9
Introduction To Your ECLIPSE Oxygen System 10
Understanding the Controls and Front Panel 11
Description of the ECLIPSE 11
Control Panel Features 12
User Controls 13
Front Panel Icons 13
System Status Indicators 13
Recommended Operating Environments 15
The Proper Location 16
Operating Instructions 17
Before Operating 17
Powering Your ECLIPSE System 18
Operating Your ECLIPSE System 25
Alarm Conditions and Troubleshooting 30
Accessories 32
Cleaning, Care and Routine Maintenance 35
Technical Data 38
Summary of All Warnings and Cautions 40
Electromagnetic Compatibility 48
Disposal of Equipment and Accessories 49
Model 1000 User Manual
Contents

Important Safety Instructions are Shown
Under the Following Headings:
SeQual®Technologies Inc. (SeQual) recommends that this User Manual be given to
each patient receiving an ECLIPSETM Oxygen System. This User Manual describes all
the necessary information in order for a patient or caregiver to operate the ECLIPSE.
WARNING Important safety information for hazards that might cause
serious injury.
CAUTION Important information for preventing damage to the ECLIPSE
Oxygen System.
DO NOT Information to which you should pay special attention.
Important! Clarifying information, specific instructions, commentary, sidelights,
or interesting points.
Symbols Used in User Manual and on Oxygen System
Refer to this section for an explanation of the symbols and the warnings that
accompany them.
“Read Instructions”: This symbol, when used alone or in conjunction
with any of the following symbols, indicates the need to consult the
operating instructions provided with the product. A potential risk
exists if the operating instructions are not followed.
“Attention or Caution”: Consult accompanying documents. This
symbol indicates important information that requires the user to take
special precautions when certain conditions are present.
Important Safety Instructions
2|ECLIPSETM Model 1000 User Manual

“No Smoking or Naked Flames”: The use of oxygen therapy
requires that special care be taken to reduce the risk of fire. Materials
that burn in air, and even some materials that will not burn in air, are
easily ignited and burn rapidly in high concentrations of oxygen. For
safety concerns, all possible sources of ignition must be kept away
from the oxygen concentrator and preferably out of the room in which
it is being used. Smoking in the proximity of an operating oxygen
concentrator is dangerous, and can permanently damage the
device and void the warranty.
“Use No Oil or Grease”: Indicates important information about the
possibility of spontaneous and violent ignition that may occur if oil,
grease or other petroleum based substances come into contact with
oxygen under pressure.
“Drip Proof Equipment-IPX0”: The ECLIPSE enclosure does not
provide protection against the harmful effects of liquid ingress. (IPX0,
per IEC 60529)
“Drip Proof Equipment-IPX1”: The ECLIPSE AC and DC Power
Supplies provide protection against the harmful effects of liquid
ingress. (IPX1, per IEC 60529)
“IEC Icon for DC Power (Input)”
“Type B Equipment”: Equipment providing a particular degree of
protection against electric shock, particularly regarding:
- Allowable LEAKAGE CURRENT;
- Reliability of the protective earth connection.
“Amperes”
IPX
0
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |3
IPX
1

“Alternating Current”
“Audible Alarm”: An audible indicator is provided to announce a
reduction in oxygen concentration or when the power has been
interrupted or lost.
“ON/OFF”: Turns the unit ON and OFF but does not directly
disconnect the mains power.
“Increase Flow Setting”: Increases the flow setting in 0.5
increments each time the button is pushed.
“Decrease Flow Setting”: Decreases the flow setting in 0.5
increments each time the button is pushed.
“Flow Mode”: Activates either Continuous Mode or Pulse Flow
Mode Operation.
“External Power Present Indicator”: Indicates the presence of
external power to operate the concentrator.
“Power Cartridge Status”: Shows the amount of charge remaining in
the Power Cartridge.
“Power Cartridge Symbol”: The symbol on the Power Cartridge
pack.
“Oxygen Output”: The port in which the oxygen flows out and where
the tubing is attached.
Symbols (continued)
4|ECLIPSETM Model 1000 User Manual

THE ECLIPSE OXYGEN SYSTEM IS INDICATED FOR THE ADMINISTRATION OF
SUPPLEMENTAL OXYGEN. THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT
NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES.
YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE
SETTING TO MEET YOUR INDIVIDUAL NEEDS.
RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY
UNDER THE ADVICE OF A PHYSICIAN.
WARNING FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE
ORDER OF A PHYSICIAN.
WARNING THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE
SUSTAINING APPLICATIONS, NOR DOES IT PROVIDE ANY PATIENT
MONITORING CAPABILITIES.
WARNING IN CERTAIN CIRCUMSTANCES, THE USE OF NON-PRESCRIBED
OXYGEN CAN BE HAZARDOUS. THIS DEVICE SHOULD ONLY BE
USED WHEN PRESCRIBED BY A PHYSICIAN.
WARNING FOR ADULT USE ONLY. NOT QUALIFIED FOR USE ON PEDIATRIC
PATIENTS.
WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
Contraindications
Indications for Use
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |5
WARNING AS WITH ANY ELECTRICALLY POWERED DEVICE, THE USER MAY
EXPERIENCE PERIODS OF NON-OPERATION AS THE RESULT OF
ELECTRICAL POWER INTERRUPTION, OR THE NEED TO HAVE THE
OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN. AN OXYGEN
CONCENTRATOR IS NOT APPROPRIATE FOR ANY PATIENT WHO
WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE
RESULT OF SUCH TEMPORARY INTERRUPTION.

Safety Guidelines
6|ECLIPSETM Model 1000 User Manual
This section of the manual provides warnings and cautions that apply to hazards or
unsafe practices that could result in personal injury or property damage.
Oxygen Therapy Safety
“Read Instruction Manual”: DO NOT operate this oxygen system
without first reading and understanding this manual. If you are unable
to understand the warnings and instructions, contact your authorized
SeQual dealer/provider before attempting to use this oxygen system.
“Caution or Attention”: In certain circumstances, oxygen therapy can
be hazardous; therefore, it is advisable to seek medical advice before
using the oxygen system.
DO NOT deviate from prescribed flow rates. Adjust the flow only to your prescribed
flow rate, unless otherwise instructed by your physician.
WARNING “No Smoking or Naked Flames”: The use of oxygen therapy requires
that special care be taken to reduce the risk of fire. Oxygen enhances
the combustion of flammable material. For safety concerns, all possible
sources of ignition must be kept away from the oxygen system and
preferably out of the room in which it is being used. Smoking in the
proximity of an operating oxygen concentrator is dangerous, and
can permanently damage the device and void the warranty.
WARNING “Use No Oil or Grease”: A spontaneous and violent ignition may
occur if oil, grease or other petroleum substances come into contact
with oxygen under pressure. Keep these substances away from the
oxygen system, tubing and connections and any other oxygen source.
DO NOT use any petroleum based or other lubricants.

Operating Safety
■Keep oxygen system and cord away from hot surfaces.
■DO NOT use the oxygen system if it has a damaged cord or plug; contact your
dealer/provider.
■Your oxygen system should be located in a well-ventilated area to allow for
adequate air intake and to avoid the intake of airborne pollutants and fumes.
■ONLY use this device with accessories designed or recommended by SeQual
Technologies Inc.
■If your oxygen system has been dropped, damaged or exposed to water, contact
your dealer/provider for examination or repair.
■DO NOT drop or insert any object into any opening.
■DO NOT block the air inlet or the exhaust vent of the ECLIPSE when it is on
a soft surface, such as a bed or couch or car seat.
■DO NOT cover unit with a blanket, towel, quilt, etc.
■DO NOT overfill the humidifier, if used. Fill with water only to the level shown
by the manufacturer.
■When using ECLIPSE in a vehicle, make sure it is properly secured or restrained.
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |7

Operating Safety (continued)
WARNING DO NOT remove the cover, no serviceable parts inside. Qualified
service personnel should only remove the cover.
WARNING DO NOT operate device or accessories on wet surfaces or in standing
water and DO NOT submerse or expose to water. The ECLIPSE
enclosure does not provide protection against the harmful effects of
liquid ingress. Electrical shock or damage to the unit may result.
WARNING Protect electrical cords from sharp edges. Electrical shock and serious
physical injury may result.
WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied
with your ECLIPSE Oxygen System. Use of a non-SeQual Power
Supply may be hazardous, cause damage to your oxygen system
and will void the warranty.
CAUTION Availability of an alternate source of oxygen is recommended in
case of power outage or mechanical failure.
CAUTION DO NOT expose unit to temperature outside of specified operating
or storage temperatures as this may damage the unit.
CAUTION Locate oxygen tubing and Power Supply cords to prevent tripping
hazards.
CAUTION DO NOT place ECLIPSE in a small, enclosed space, such as a
bathroom or closet, with cannula leading out of the enclosed space.
Safety Guidelines (continued)
8|ECLIPSETM Model 1000 User Manual

Power Cartridge Safety
WARNING This Power Cartridge may explode and cause potential injury
if exposed to or disposed of in a fire.
WARNING DO NOT short-circuit the Power Cartridge’s metal contacts with
metallic objects, such as keys or coins. It may cause sparks or
excessive heat.
WARNING DO NOT disassemble, puncture, or crush the Power Cartridge. Power
Cartridge electrolytes may be toxic if swallowed and can be harmful
to skin and eyes. Keep the Power Cartridge away from children.
WARNING Use of a damaged Power Cartridge may cause personal injury.
WARNING Exposing the Power Cartridge to water or other liquids may cause
personal injury.
CAUTION DO NOT expose the Power Cartridge to temperatures above
140°F (60°C), such as in a vehicle parked in the sun or on a hot day.
CAUTION Charging the Power Cartridge below 50 F° (10° C) or above
104° F (40° C) may degrade the performance.
CAUTION DO NOT drop the Power Cartridge or expose it to mechanical shock.
CAUTION Only use the Power Cartridge for its intended purpose.
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |9

Introduction to Your ECLIPSE Oxygen System
10 |ECLIPSETM Model 1000 User Manual
This User Manual will inform you about the
use and care of the ECLIPSE Oxygen
System and its accessories. Thoroughly
read all the information in this manual in its
entirety before operating your ECLIPSE.
Contact your dealer/provider for proper
training in the use of this device.
YOUR PHYSICIAN HAS PRESCRIBED
SUPPLEMENTAL OXYGEN AS PART
OF A TREATMENT PLAN. LONG TERM
OXYGEN THERAPY IS NOW WIDELY
ACCEPTED AS THE STANDARD
TREATMENT FOR CHRONIC HYPOXIA
CAUSED BY CHRONIC OBSTRUCTIVE
PULMONARY DISEASE. OXYGEN
THERAPY HAS BECOME MORE
VERSATILE IN RECENT YEARS DUE
TO THE DEVELOPMENT OF OXYGEN CONCENTRATORS. THE ECLIPSE MAKES
IT EASIER FOR PHYSICIANS TO SELECT A MODE OF OXYGEN THERAPY THAT
SUITS THE PATIENT’S LIFESTYLE.
THE ECLIPSE MAY BE USED IN HOME, INSTITUTION, AND MOTOR VEHICLES.
USE OF THE ECLIPSE IN HOSPITALS SHOULD INCLUDE THE USE OF A
“HOSPITAL GRADE” POWER CORD.
YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE
SETTING TO MEET YOUR INDIVIDUAL NEEDS. THIS FLOW RATE IS DETERMINED
BY THE AMOUNT OF OXYGEN YOU NEED. RECOMMENDED OXYGEN FLOW
RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN.

Description of the ECLIPSE
The key features of the ECLIPSE Oxygen System and control panel are important
items to become familiar with.
Understanding the Controls and Front Panel
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |11
INPUT/OUTPUT CONNECTIONS
Oxygen Outlet Port Extension tubing or a cannula is attached to this port.
Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the
top/back of the unit. This filter prevents dust and debris from entering
the ECLIPSE and should be cleaned regularly.
Service Port This connection allows service personnel to diagnose and monitor
ECLIPSE performance. It is not for patient use.
External Power Receptacle Either the ECLIPSE AC or DC Power Supply output cords may be
connected to this receptacle.
Exhaust Vent Exhaust air from the ECLIPSE is expelled from this vent.
BACK
FRONT
Service Port
(Not for Patient Use)
Air Inlet Filter
Cart Insert Screw
Exhaust Vent
Power Cartridge
Latch
Main/Rating Label
(Under Base)
Power Cartridge
Handle
Control Panel
Oxygen Outlet Port
Simplified User
Instructions
(Under Base)
External Power
Receptacle

Control Panel Features
The ECLIPSE displays important operating information using the icons, numbers and
visual indicators on the control panel. This section will help you understand these features.
Understanding the Controls and Front Panel (continued)
12 |ECLIPSETM Model 1000 User Manual
000.6
No Smoking Icon
ON/OFF Button
ON/OFF Indicator (Green)
Alert Indicator (Yellow )
Flow Rate Setting (2 Digits)
(Green)
Alarm Indicator (Red)
Increase Flow Rate Button
Decrease Flow Rate Button
Power Cartridge Status
Gauge (Green)
External Power Present
Indicator (Green)
Pulse Flow Mode
Indicator (Green)
Flow Mode Button

User Controls
This section describes the control buttons and function for your ECLIPSE Oxygen System.
FRONT PANEL ICONS
ON/OFF Button and Indicator: This button turns the system ON or
OFF. The Green Indicator is lit when the unit is ON.
Increase or Decrease Flow Rate Button: Use these buttons to set the
flow rate to your prescribed level.
Flow Mode Button and Indicator: The ECLIPSE contains a button to
toggle between Continuous Flow and Pulse Flow mode. Pulse Flow
mode allows a significant increase in the amount of time the system
can operate from the Power Cartridge. When the Pulse Flow Mode is
activated, the green Flow Mode indicator goes on and a pulse of
oxygen is delivered upon inhalation. When the pulse of oxygen is
delivered, the indicator briefly goes off. Only change your flow settings
at the recommendation of your physician.
SYSTEM STATUS INDICATORS
The ECLIPSE control panel displays important operating information. This section will
help you understand these features.
Green, Yellow, and Red indicators on the front panel indicate the operating condition
of the system. For more information on the indicators, see the Alarm Conditions and
Troubleshooting table.
ON/OFF (Green) Indicator - System OK:
Indicates that the ECLIPSE is ON and operating properly.
ALERT (Y
ellow) Indicator - Low and Medium Priority Alerts:
When steadily illuminated, this indicates a low priority awareness
condition or Caution. Continue to use your system and refer to the
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |13

Troubleshooting section for the proper response. A flashing yellow
indicates a medium priority alert. Prompt response is needed to deal
with a condition. Refer to the Troubleshooting section.
ALARM (Red) Indicator - High Priority Alarms:
Indicates a high priority alarm condition. Immediate response is
necessary to deal with the condition. Refer to the Troubleshooting
section and contact your dealer/provider.
Buzzer: An audible alarm (or buzzer) is used to alert you of the
condition of the unit, either a warning or failure, and to confirm a valid
key press.
Flow Rate Setting Indicator (Green): This is the main focus on your
control panel. Your dealer/provider will set up your prescribed flow
rates for your Continuous Flow Mode (LPM) and your Pulse Flow
settings, if pulse flow has been prescribed. DO NOT change these
settings unless it is under the order of your physician. Each time you
turn the unit on, the last flow mode and flow setting will be used.
Power Cartridge Status Gauge: This indicator displays the charge
remaining in the Power Cartridge. When the Power Cartridge is fully
charged all 10 bars are illuminated. Each bar represents approximately
10% of the total Power Cartridge charge. When the Power Cartridge is
being charged, the charge indicator bars will blink in a waterfall fashion.
If the Power Cartridge is not installed, or if it is improperly installed, the
Status Gauge will not light.
External Power Present Indicator: When the ECLIPSE system is
properly plugged in using the AC or DC Power Supply, this indicator
will appear on the Control Panel.
Understanding the Controls and Front Panel (continued)
14 |ECLIPSETM Model 1000 User Manual

The following information recommends certain boundaries that will allow the
ECLIPSE Oxygen System to operate at its full potential.
Operating Temperature 50º F to 104º F (10º C to 40º C)
Operating Humidity 10% - 95% at an 82.4°F (28°C) dew point
Transport/Storage Temperature -4º F to 140º F (-20º C to 60º C)
Humidity: Up to 95% Non-Condensing
Electrical Use no extension cords or electrical
outlets controlled by a switch.
Altitude 0 – 13,123 feet (0 – 4000 meters)
Placement Do Not block air inlet or bottom exhaust vent
(keep at least 3 inches away from walls,
draperies, furniture, etc.)
Environment Smoke, pollutant, and fume free.
Operating Time Up to 24 hours per day when connected
to external power.
When moving the ECLIPSE from an extreme environment, allow time for your
ECLIPSE to acclimate to the recommend operating environment.
WARNING Operating your ECLIPSE outside the recommended operating
environment may impact performance, cause damage and will void
the warranty.
CAUTION Locate the unit in a well-ventilated room that provides adequate air
flow and check if air is being prevented from entering or exiting the
system by furniture, draperies, bedding, clothing or pets.
Recommended Operating Environments
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |15

Recommended Operating Environments (continued)
16 |ECLIPSETM Model 1000 User Manual
The Proper Location
Select a location that avoids fumes and pollutants and allows the oxygen system to
draw in room air through the air inlet filter on the back of the cabinet and exhaust air
out the vent at the bottom of the case without being restricted.
Locate the unit such that the alarms can be heard.
Route oxygen tubing such that it does not kink or is subject to occlusion.
DO NOT cover unit with a blanket, towel, quilt, or other covering, as the unit may
overheat.
DO NOT locate near flammable materials or cleaning agents or in the direct path of
any heat source, such as a heat register or car heater.
Keep the oxygen system at least five (5) feet (1.5 m) away from hot, sparking objects
or open sources of flame.

Before Operating
Your SeQual dealer/provider or authorized service representative will demonstrate
how to operate the oxygen system. This manual serves as a supplemental reference to
help you operate and maintain the unit. If you have any further questions or problems
call your SeQual dealer/provider or authorized service representative.
Important! DO NOT attempt to operate the ECLIPSE System without first reading
the “Safety Guidelines” section of this manual. Follow all operating
instructions. Observe all warnings on the product and in the operating
instructions. To reduce the risk of fire, bodily injuries, and damage to
the oxygen system, observe all safety precautions.
WARNING Protect the ECLIPSE and AC/DC Power Supplies from fluid spills or
drips to avoid possible shock hazards.
CAUTION Always check to see that the Air Inlet and the Exhaust Vent are not
blocked and the air inlet filter is clean before using your system.
Operating Instructions
Model 1000 User Manual
ECLIPSETM Model 1000 User Manual |17
Air Inlet and
Air Inlet Filter
Exhaust Vent

Powering Your ECLIPSE System
The ECLIPSE is a lightweight, portable oxygen system capable of being operated
directly from three different power sources: (1) an AC Power Supply, (2) a DC Power
Supply, or (3) a rechargeable Power Cartridge.
The ECLIPSE charges the optional Power Cartridge when external power (AC or DC)
is present and when the Power Cartridge temperature is below a safe charging
temperature. When external power is disconnected, the system will automatically
switch over to the Power Cartridge, if installed. When external power is restored (by
plugging into AC or DC outlets) the system will accept power from the external source
and replenish the Power Cartridge while operating, if adequate power is available.
Connect the AC or DC Power Supply by aligning the arrow on the plug locking ring
with the white dot on the receptacle of the unit. Push until the locking ring twists
slightly counter-clockwise, then snaps back into alignment with the dot on the
receptacle. Disconnect by grasping the plug, twist counter-clockwise and pull.
Connect the AC Power Cord to a grounded power outlet.
Operating Instructions (continued)
18 |ECLIPSETM Model 1000 User Manual
(3) RECHARGEABLE
POWER CARTRIDGE
(1) AC POWER SUPPLY
(2) DC POWER SUPPLY