AirSep ImPulse Elite User manual

ImPulse™Elite Patient Manual
ImPulse™Elite
OCD
System
Patient
Manual
401 Creekside Drive • Buffalo, New York 14228-2085 USA

ImPulse™Elite Patient Manual

ImPulse™Elite Patient Manual
Components ..........................................................................................................1
Features ................................................................................................................2
Introduction............................................................................................................3
Maximizing Delivery with ImPulse Elite.........................................................3
Initial Setup ....................................................................................................3
Safety .....................................................................................................................4
Description of Parts ...............................................................................................6
How to Operate the Unit
Oxygen Cylinder Connection ........................................................................7
Testing and Replacing Battery ......................................................................9
Turning On the Unit for Proper Oxygen Usage ......................................... 10
Turning Off the Unit .................................................................................... 12
Alarm Light Indicators ................................................................................. 12
Switching to Continuous Flow .................................................................... 13
Removing/Replacing the Cylinder ............................................................. 13
ImPulse Elite as a Portable Unit ........................................................................ 14
Proper Care and Maintenance .......................................................................... 15
Troubleshooting Guide ..................................................................................... 16
Equipment Provider Information
Specifications ............................................................................................. 17
Limited Warranty ................................................................................................ 18
Table of Contents
i

ImPulse™Elite Patient Manual

ImPulse™Elite Patient Manual
The complete ImPulse™Elite OCD System includes the following,
as shown in Figure 1 below:
The Ambulatory Oxygen System
for Complete Patient Mobility
Not shown:
• (1) One Alkaline “D” Energizer®Battery
• ImPulse Elite OCD System Patient Manual
Note: The carrying case is available separately.
Figure 1
ImPulse Elite
Oxygen
Conserving
Device
Components
1

ImPulse™Elite Patient Manual
• Easy to operate with just one “D” alkaline battery (Energizer
recommended). Cannula connects in seconds. Patient-accessible
controls.
• Oxygen-conserving technology allows smaller cylinders to last longer.
• With ImPulse Elite, the “D” battery provides the greatest ambulation
time. For most patients, 10–18 months.
• Pulse modes offer up to 6 lpm efficiency.
• Fail-safe electronic circuitry and alarm system provide added
assurance of proper unit function and oxygen delivery.
• Audible alarm when device does not detect inhalation (inspiration), if
cannula is disconnected or tubing kinked, or if the unit is inadvertently
left on when not in use.
• Only minimal maintenance required.
• Designed for maximum patient mobility.
• Economical way for ambulatory oxygen patients to enjoy freedom of
movement.
• Developed to enhance quality of life for patients using supplemental
oxygen.
Features
2

ImPulse™Elite Patient Manual
ImPulse™Elite is an electronic oxygen conserving device (OCD) that provides
convenience and mobility to patients with therapeutic oxygen needs. Designed
and manufactured by AirSep Corporation, ImPulse Elite uses portable cylinders
to provide patients with oxygen delivery at their prescribed dosages.
With this device, actual oxygen delivery to the patient occurs at or near the
beginning of the breathing cycle when the patient starts to inhale.
Because ImPulse Elite does not deliver oxygen during exhalation, which
accounts for approximately two-thirds of the breathing cycle, it wastes little or
no oxygen. This differs greatly from conventional oxygen systems that deliver
oxygen continuously. In addition, many patients are more comfortable with
oxygen conserving devices than continuous flow systems because they have
fewer problems with nasal dehydration that can result from constant flow into
the nostrils.
Maximizing Delivery with ImPulse Elite
ImPulse Elite’s advanced electronics enable it to respond effectively to a
patient’s respiratory cycle. ImPulse Elite senses the start of inhalation. Because
only a small amount of oxygen is actually required to oxygenate the blood
effectively, ImPulse Elite delivers the precise amount when it is most beneficial —
at the beginning of inhalation.
ImPulse Elite quickly delivers a pulse of oxygen compatible with your breathing
rate and automatically adjusts with any change in your breathing cycle —
depending on your prescribed oxygen requirement.
This method of oxygen delivery has been clinically tested and determined to
meet oxygen needs satisfied by comparable flow settings on continuous flow
units up to 6 lpm.
Initial Setup
As part of your initial set up with ImPulse Elite, your clinician may choose to
assess your saturation level through pulse oximetry. This helps to ensure
proper titration while using the oxygen conserving device.
Introduction
3

ImPulse™Elite Patient Manual
ImPulse Elite meets applicable safety and environmental regulations. This
assurance means you can fully enjoy all the benefits of this device. However,
with all electronic devices and applications that require oxygen, normal care
and precautions must be followed.
Before you attempt to use this oxygen conserving device and a portable
oxygen cylinder, read all warning and caution notes in this section of the
manual. Contact your Equipment Provider should you have further questions.
Warnings
Federal (USA) law restricts this oxygen conserving device to sale by, or on order of, a
licensed physician. This device must be used only under the supervision of a licensed
physician.
High purity oxygen promotes rapid burning. Do not allow smoking or open flames within
five feet of this device, the portable oxygen cylinder, or any oxygen-carrying accessory.
If you observe that the oxygen conserving device is not working properly or you feel ill,
contact your Equipment Provider and/or physician immediately.
Although oxygen is a non-flammable gas, it vigorously supports combustion. Use no oil,
grease, or petroleum-based products on or near the device or any oxygen source.
Use this device and the oxygen cylinder only with clean, dry, oil-free hands.
Observe all federal and/or international regulations regarding vehicular travel by air, sea,
or land while using and/or transporting oxygen cylinders.
Safety
4

ImPulse™Elite Patient Manual
Cautions
Maintain a backup supply of oxygen, and periodically check the contents gauge on the unit,
which monitors the consumption of oxygen in the cylinder.
Do not use this device with a humidifier or expose it to water or other liquids.
Do not expose the device to extreme temperatures.
Do not use this device
• with the batteries removed.
• if the oxygen tubing is kinked or obstructed.
• if the device was tampered with or opened by
anyone other than your Equipment Provider.
5

ImPulse™Elite Patient Manual
As you can see in Figure 2 below, all parts are within easy reach and allow you
to adjust your ImPulse Elite unit as necessary. Follow your clinician’s instruc-
tions before making any change to the flow selector setting.
Description of Parts
Securing Knob
Oxygen Outlet
Cylinder
Contents Gauge
Pulse Flow/
Continuous
Flow Switch
Battery Test
Button
Low Battery
(Yellow Indicator)
Pulse Flow
(Green Indicator)
Alarm
(Red Indicator)
Battery
Compartment
(not shown)
Flow Selector
On/Off Switch
Figure 2
6
Brass CGA Sealing
Washer
Figure 2a
Figure 2b
Easily connects to
any standard post-
valve cylinder.

ImPulse™Elite Patient Manual
Oxygen Cylinder Connection
Warnings
High purity oxygen promotes rapid burning. Do not allow smoking or open flames within
five feet of this device, the portable oxygen cylinder, or any oxygen-carrying accessory.
Use no oil, grease, or other petroleum-based products on or near the device or any
oxygen source.
Use this device and the oxygen cylinder only with clean, dry, oil-free hands.
Oxygen in the cylinder is under high pressure. Do not open the cylinder valve without
the ImPulse Elite unit properly connected to the cylinder.
Your Equipment Provider will review proper cylinder care and connection
procedures with you the first time you operate your unit.
Refer to the procedure on the following page should you need assistance
with connecting your ImPulse Elite unit to an oxygen cylinder.
How to Operate the Unit
7

ImPulse™Elite Patient Manual
3. Push the alignment pins into the holes on
the neck of the cylinder, and hold the unit
in place against the cylinder with one
hand while you turn the unit’s securing
knob clockwise with your other hand until
the unit is secured firmly on the cylinder
(Figure 6). Figure 6
Figure 3
1. Loosen the ImPulse Elite
securing knob. To ensure a leak-
tight seal between the cylinder
valve and the unit, make sure the
sealing washer is present in the
center of the unit before
installation (Figure 3).
2. Align the two parallel alignment pins
(Figure 4) on the unit with the holes
on the neck of the cylinder
(Figure 5).
Sealing Washer
Figure 4
Alignment
Pins
Figure 5
Alignment
Holes
Cylinder Valve Toggle
8
To connect the ImPulse Elite to any post-valve cylinder:

ImPulse™Elite Patient Manual
Testing and Replacing Battery
ImPulse Elite requires just one 1.5 volt “D” size alkaline battery (Energizer
recommended) for full operation. Battery life is dependent upon the predeter-
mined unit settings, your breathing rate, and the length of daily ambulation time.
Most patients experience 10–18 months of battery life. AirSep strongly recom-
mends that you maintain a spare battery in your carrying case for replacement
in this unit as necessary.
To install or replace the battery, follow the steps below:
1. Make sure the ON/OFF power switch is set to the OFF position. (It
is not necessary to disconnect the cannula or remove the ImPulse
Elite from the cylinder, although you may find it easier to replace the
battery within the battery compartment door if the unit is first
removed from the cylinder.)
2. If the unit is not connected to the
cylinder, simply hold the unit with one
hand while you press in and lift up on
the battery compartment latch with your
other hand. (Figure 7).
If the unit is connected to the cylinder,
steady the unit by grasping the
neck of the cylinder just below the
unit with one hand while you press
in and lift up on the battery
compartment latch with your other
hand (Figure 8).
3. Insert a fully charged battery into
the inside door of the battery
compartment. Make sure
the battery is properly
positioned to match the
correct positive and
negative polarity markings
(Figure 9).
Figure 7
Figure 8
Figure 9
“+”Positive
Polarity
“-” Negative
Polarity
9

ImPulse™Elite Patient Manual
Figure 10
To test the battery to ensure that there is proper power available to operate
the unit:
1. Make sure there is a battery in the unit’s battery compartment and it
is properly positioned to match the correct
positive and negative polarity markings.
2. Make sure the ON/OFF power switch is set to
the ON position.
3. Press and hold down firmly the center of the
BAT. TEST button with your finger. (Figure
10.) (You can test the battery regardless of
whether or not a cannula is connected to the
oxygen outlet.)
The green PULSE FLOW indicator
illuminates to indicate a good battery. If the battery requires
replacement, the green PULSE FLOW and yellow LOW BATTERY
indicators illuminate.
Turning On the Unit for Proper Oxygen Usage
Note
Do not use a mask with the ImPulse Elite unit.A mask may not conform to the face
properly enough for the unit to fully sense inhalation.
Do not use pediatric or low flow cannulas with the ImPulse Elite unit as their
smaller tubing diameters negatively affect delivery volumes.
Use ImPulse Elite during sleep only under the recommendation of a qualified
clinician.
10

ImPulse™Elite Patient Manual
Figure 11
1. Check the energy level of the battery each time you use the unit.
Refer to “Testing and Replacing Battery” for step-by-step instructions.
2. Check the oxygen cylinder setup and the cylinder contents gauge.
If you need to change cylinders, which includes removal and
reconnection of the ImPulse Elite unit, refer to Oxygen Cylinder
Connection for step-by-step instructions.
3. Make sure the PULSE FLOW / CONTINUOUS FLOW switch is moved
completely right (for pulse flow setting).
4. Set the ON/OFF switch to the ON position.
All three lights illuminate for less than 1 second. The
ImPulse Elite unit automatically calibrates and reaches
sensitivity within approximately 15 seconds. Then the
green light shuts off, and the unit emits a short burst of
oxygen. This indicates ImPulse Elite is ready for use.
5. Connect your cannula to the oxygen outlet. (You may
use up to a 25 ft [7.5 m] cannula, if desired.) (Figure 11.)
6. Turn the cylinder valve slowly one full turn counter-
clockwise to begin oxygen flow.
7. Make sure the cylinder contents gauge on the ImPulse Elite unit registers
within the green shaded/green zoned areas.
This indicates an adequate oxygen supply in the cylinder. (If the cylinder
contents gauge indicates the red REFILL zone, the
cylinder needs to be replaced.)
8. Position the nasal cannula with the prongs inserted
into your nostrils.
9. Set the flow selector between 1 and 6 depending on
your prescribed dosage (Figure 12).
Figure 12
11

ImPulse™Elite Patient Manual
Turning Off the Unit
1. Turn the oxygen cylinder valve fully clockwise to stop oxygen flow.
2. Set the ON/OFF power switch on the ImPulse Elite unit to the OFF
position.
Note
If you do not set the unit’s ON/OFF power switch to OFF when the unit is not in use, the
alarm sounds until the battery fully discharges.
Alarm/Light Indicators
Setting Light Alarm Indicates
OFF No No Unit is not on, battery life is preserved.
ON +
(Flow
Selection)
Green No Pulse flow.
Yellow &
Green
Yes
(Intermittent)
Battery almost discharged. Prepare to replace
battery within 20-30 minutes.
Red Yes
(Continuous)
No inspiration detected for approximately 30
seconds (cannula may be disconnected, tubing
kinked, or unit remains on when not in use).
BAT.
TEST
Green No Sufficient battery life.
Yellow &
Green No Battery almost discharged. Prepare to replace
battery within 20–30 minutes.
No No No battery or discharged battery in unit.
Alarm Light Indicators
The following chart provides an explanation of what each alarm light
indicates:
12

ImPulse™Elite Patient Manual
16
Switching to Continuous Flow
If your unit is equipped with a continuous flow switch:
For very short time periods, this PULSE FLOW/
CONTINUOUS FLOW switch at the base of the unit can
be moved completely left for CONTINUOUS FLOW to
force the oxygen conserving unit to perform as a
continuous flow unit (Figure 13).
A typical use for the CONTINUOUS FLOW setting occurs
when the battery can no longer provide power. This setting
provides 2 lpm of continuous flow oxygen.
Before setting ImPulse Elite to CONTINUOUS FLOW,
make sure you set the ON/OFF power switch to the OFF position to preserve
battery life as this is a manual override and no audible or visual alarm
indicators are available in this “temporary” setting.
Caution
The continuous flow setting uses substantially more oxygen than the normally used pulse
flow setting. Make sure the cylinder contents gauge is observed often when operating the
unit as a continuous flow device.
13
Figure 13
Removing/Replacing the Cylinder
To easily remove the ImPulse Elite unit from a cylinder:
1. Close the oxygen cylinder valve by rotating the valve completely clockwise.
2. Press the BAT. TEST button OR temporarily set the PULSE FLOW/ CON-
TINUOUS FLOW switch to the CONTINUOUS FLOW setting until no
oxygen can be heard venting from the unit.
3. Turn the unit’s securing knob counterclockwise to lift the unit off the cylinder.

ImPulse™Elite Patient Manual
The ImPulse Elite unit and its accessories are light weight and portable. This
means that ImPulse Elite can go virtually anywhere you go.
Cautions
Do not expose the device to, or submerge it in, water.
Do not expose the device to extreme temperatures.
The universal-style carrying case, as shown in Figure 14, can be used with the
unit. During use, ImPulse Elite should remain in your carrying case. This
ensures added protection for the unit and
allows discreet use. Simply check periodi-
cally to ensure that your cannula connec-
tion is secure and that the cylinder shutoff
valve remains in the open position (one full
turn counterclockwise) to enable oxygen
flow.
17
17
ImPulse Elite as a Portable Unit
Figure 14
14

ImPulse™Elite Patient Manual
Keep the ImPulse Elite clean and free from moisture, dust, and extreme
temperatures. To help prevent possible damage to the unit, keep ImPulse Elite
in its carrying case.
Clean the unit periodically by wiping with a lint-free cloth. Pay special attention
to the oxygen outlet for the cannula connection to make sure it remains free of
dust, water, etc. Do not expose or submerge the ImPulse Elite device in water.
We recommend using a one-part bleach to 99 parts water solution or using
Fantastik®spray cleaner as a safe cleaner for the ImPulse Elite’s outer
housing.
A list of undesirable chemical cleaning agents includes, but is not limited to
the following, according to the plastic manufacturer: alcohol and alcohol-
based products (isopropyl alcohol), concentrated chlorine-based products
(Ethylene Chloride), and oil-based products (Pine-Sol, Lestoil). These are
NOT to be used to clean the plastic housing on the ImPulse Elite OCD. These
chemicals and the cleaners that contain them will crystallize the plastic and
break down the plastic’s molecular properties.
Replace the disposable cannula periodically following normal usage.
Additional supplies are available from your Equipment Provider or AirSep.
Return empty oxygen cylinders to your supplier for replacement. Remove the
unit from the empty cylinder before returning the cylinder to your supplier.
Note
Do not open the electronic compartment to service the device. Open only the battery
compartment when necessary. To prevent a voided AirSep warranty, follow all
manufacturer’s instructions.
20
Proper Care & Maintenance
15

ImPulse™Elite Patient Manual
Problem Possible Cause Solution
No battery in unit, or
battery requires replace-
ment.
Battery not inserted
correctly.
Battery may be dis-
charged.
Brass/CGA sealing washer
is missing from unit.
Worn brass/CGA sealing
washer.
Unit not properly secured
to cylinder, or securing
knob not tight.
PULSE FLOW/CONTINUOUS
FLOW switch in CONTINU-
OUS FLOW position.
Cannula tubing is kinked or
not properly adjusted.
ON/OFF power switch set
to OFF position.
Oxygen cylinder is empty.
Oxygen cylinder valve is
closed.
If your unit fails to operate as described in this manual, first review the following
chart for possible causes and solutions. If necessary, contact your Equipment
Provider.
Troubleshooting Guide
Power switch ON
but no lights
display.
Intermittent power,
intermittent
pulsing, lights flash,
alarm sounds.
Oxygen leaks from
unit.
Constant flow of
oxygen from unit.
Unit does not
deliver oxygen.
Any other unit
problem.
Replace battery.
Insert battery properly by
referencing polarity guide on
inside of battery compartment
door.
Replace Alkaline battery with
new D-cell Energizer battery.
Ensure that the brass/CGA
sealing washer is present at the
oxygen cylinder connection on
the unit.
Replace brass/CGA sealing
washer.
Close cylinder valve and verify
that the unit alignment pins are
correctly fitted to the cylinder
alignment holes and that the
securing knob is tight.
Position switch to PULSE FLOW
position.
Check or replace tubing as
necessary.
Set ON/OFF power switch to ON
position.
Replace cylinder.
Open cylinder valve a minimum
of one turn counterclockwise.
Contact your Equipment
Provider.
16
Table of contents
Other AirSep Oxygen Equipment manuals

AirSep
AirSep visionaire User manual

AirSep
AirSep AS-A User manual

AirSep
AirSep Focus Portable User manual

AirSep
AirSep Onyx User manual

AirSep
AirSep SeQual Regalia User manual

AirSep
AirSep FreeStyle User manual

AirSep
AirSep FreeStyle 5 User manual

AirSep
AirSep Focus Portable User manual

AirSep
AirSep FreeStyle User manual

AirSep
AirSep AS-20 User manual