Taylor Hobson Surtronic Duo 112-3115 User manual

Specification
Traverse length. 5mm (0.2in)
Measurement length 4mm (0.16in)
Traverse speed 2mm/s (0.08in/s)
Gauge range 200µm (8000µin)
Gauge resolution 0.05µm (2µin)
Display resolution Ra. 0.01µm (0.4µin)
Display resolution Rz, Rt*, Rp*, Rv*. 0.1µm (1µin)
Cut off 0.8mm (0.03in) +/- 15%
Uncertainty 0.1µm (4µin) or 5% of result, to 95% confidence level
Results range Ra 40µm (1600µin), see ISO 4288:1996
Results range Rz, Rt*, Rp*, Rv*. 199.9µm(8000µin), see ISO 4288:1996
Stylus Diamond conisphere with 5µm nominal radius, 90º tip angle. Skid
radius: 10mm nominal.
Filter type 2CR, (200:1)
Sample spacing 1um
Stylus Force 10mN (1gf) max at mid range
Batteries Control Unit: One x 3V Lithium 2450. Traverse Unit: Three x 3V
Lithium 2450
Units µm or µin
Operating conditions 10 to 35ºC, 80%RH non-condensing
Size 125mm x 80mm x 38mm (4.92in x 3.15in x 1.5in)
Weight 200g (7oz)
*5 parameter model only
Protected by US Patent No 6629373 and US Design Patent No D439850
Infra-red communications link
The range of the infra-red (IR) link may be reduced in strong lighting conditions or where there is
interference at the specific wavelengths or frequencies used by the IR devices. Sources of such
interference include some fluorescent lighting products. If operation is affected by such conditions,
it may be necessary to reduce the distance between the two units, shield the infra-red ports from
the stray light or use the Duo in its Connected mode.
Disposal of the Surtronic Duo
When disposing of the Surtronic Duo, please remove the batteries as described
above. Dispose of the batteries and the instrument independently in accordance with
local regulations. The instrument should be disposed of as Waste Electrical and
Electronic Equipment (see EU Directive 2002/96/EC) and not put into general waste.
Declaration of Conformity
Manufacturer’s Name: Taylor Hobson Limited
Manufacturer’s Address: 2 New Star Road, Leicester, England, LE4 9JQ
It is declared that the products: Surtronic Duo 112-2916 (2 parameter) and 112-3115 (5
parameter) conform to all applicable requirements of BS EN 50081-1:1992, BS EN 50082-1:1998
and BS EN 61010:2001.
The above product complies with the requirements of the EMC Directive 89/336/EEC as
amended.
The above product complies with the requirements of the EMC Directive 73/23/EEC as amended.
Taylor Hobson Ltd.
PO Box 36,
2 New Star Road,
LEICESTER England LE4 9JQ
Tel: +44 (0) 116 276 3771
www.taylor-hobson.com
© Taylor Hobson – Surtronic Duo Manual 08/2011
Surtronic Duo
112-2916 (2 parameter model) and
112-3115 (5 parameter model)
Operator’s Leaflet K505-54
Issue 3
www.taylor-hobson.com

REFERENCE SPECIMEN
Ra = 5.81 µm
Calibrate using this area
Caution
The pick-up is extremely sensitive to shocks. When positioning the instrument onto the Surface to be
measured, carefully place the pick-up onto the surface to avoid damage.
When positioning the pick-up onto the calibration standards surface, ensure that is positioned over
the surface to measure as shown.
Direction of
p
ick u
p
travel
Direction of pick up travel
Calibration
standard
During the procedure, the display will show the following Sequence.
The control unit must be kept pointing at the traverse unit during this period. Upon successful
completion of the set-up procedure, the display will show a value.
There are five values stored as follows:
Value 1 - gain pot setting
Value 2 - Rz value (21.5+/- 10%)
Value 3 - speed pot setting
Value 4 - Rp value - period of the profile signal (68+/- 2%)
Value 5 - Can be ignored for this procedure.
These values can be viewed by toggling the display using the right-hand Selector button. Should an
error occur during set-up, the procedure will terminate and an E code will be transmitted to the
control unit for display.
“After executing the set-up procedure, the instrument should be calibrated before making any
measurement.”
E codes
The following E codes may appear on the display during the course of using the Surtronic Duo.
E=1. Infra Red communication error. This might be caused by loss of the line of sight between the
two units or other interruption of the infra-red connection. – Repeat the operation.
E=2. Measurement attempted when not calibrated. Calibrate the Traverse Unit.
E=3. Traverse Unit - Oscillator fault - Re-attempt the operation. If the fault persists, return the
Surtronic Duo to a Taylor Hobson Service Centre for servicing or repair.
E=4. Traverse Unit - Motor Positioning fault - Re-attempt the operation. If the fault persists, return
the Surtronic Duo to a Taylor Hobson Service Centre for servicing or repair.
E=5. Traverse Unit - NVRAM Error - Re-attempt the operation. If the fault persists, return the
Surtronic Duo to a Taylor Hobson Service Centre for servicing or repair.
E=6. Profile Over-range. The surface being measured may exceed the measurement range of the
Surtronic Duo. Reattempt the measurement.
E=7. Parameter Over-range. The surface being measured may exceed the measurement range of
the Surtronic Duo. Re-attempt the measurement.
E=8. Traverse Unit - Motor Speed Setup error. Check that the traverse is parallel to the edge of
the calibration standard, then re-try the set-up procedure.
E=9. Traverse Unit - Gain Setup error. Check that the traverse is entirely over the structured area
of the standard. Re-attempt the set up procedure or calibration procedure
E=10. Control Unit - Oscillator fault - Re-attempt the operation. If the fault persists, return the
Surtronic Duo to a Taylor Hobson Service Centre for servicing or repair.
E=11. Traverse Unit - Calibration information download error. Repeat the calibration procedure.
E=12. Traverse Unit - Not set up or set-up information lost. Run the Set up procedure.
E=13. <Reserved>
E=14. Traverse Unit - Electronic Measurement error – Re-attempt the operation. If the fault persists,
return the Surtronic Duo to a Taylor Hobson Service Centre for servicing or repair.

REFERENCE SPECIMEN
Ra = 5.81 µm
Calibrate using this area
Position the Traverse Unit on the surface to be measured.
If using remote mode, ensure that the line of sight between the infra red
transceivers is unobstructed; maintain the line of sight throughout the
measurement.
To begin the measurement press either of the Start buttons. During the
measurement the Control Unit will show the ‘measurement in progress’
display.
Upon successful completion of a measurement the measured
parameter is displayed as shown.
Should an error occur during measurement an appropriate E code will be displayed.
All parameters (not just the selected parameter) are transmitted to the Control Unit at the same
time and temporarily stored. The stored results may then be stepped through using the Select
buttons.
Calibrating the gauge
To ensure accuracy of results, it is recommended that the Surtronic Duo be calibrated periodically
using the calibration specimen supplied with the instrument.
Position the Traverse Unit to measure the Calibration Standard supplied with the instrument ensuring
that the traverse is parallel to the edge of the standard and entirely over the structured surface. DO
NOT use any other standard.
On the Control Unit press the Mode Select buttons until √is indicated on the Mode status.
Press the Start button on the Control Unit. This will initiate the gauge calibration procedure.
At the end of the calibration procedure the Control Unit will display the results of the calibration.
The Ra value should be 5.81µm and the Rz value 21.5µm. If an error occurs an E code is
displayed.
Set Up Procedure
The Surtronic Duo is set up before leaving the factory. The set-up procedure is used to set the
speed of the traverse unit and the gain of the gauge. Normally it is only necessary to repeat the
set-up procedure after changing the pick-up. However, if it is required to set up the unit after
purchase, the following procedure should be used.
Separate the units ready to use them in Remote Mode.
Turn on the traverse unit and position it on the calibration standard as shown above.
Activate the Control Unit.
With the Control Unit pointing at the IrDA window on the Traverse Unit, press and hold both
selector buttons simultaneously and then press the right-hand start button whilst still holding the
selector buttons. This will initiate the set-up procedure.
Once the procedure has started, release the buttons.
Note that the infra-red link must be maintained throughout the procedure, which takes about a
minute.
Direction of pick up travel
Calibration
standard
Traverse Unit
Getting Started
Storage. The Surtronic Duo is supplied in its storage configuration, that is with the Traverse Unit
connected to the Control Unit so as to protect the gauge. Before use, it is necessary to separate the
Traverse Unit from the Control Unit.
Note: In order to prevent damage to the stylus, it is recommended that the unit is stored in the
storage configuration when not in use.
Control Unit
Mode status
Mode selector
Display
Parameter status
Parameter selector
On button
IrDA windows
Stylus
Control Unit
To separate, press
here and tilt the
Traverse Unit to
separate as shown.
Positioning
Position the Traverse Unit
on the surface to be
measured as shown.
Stylus
Release button
Traverse Unit

To return the unit to its storage configuration, reverse the above
procedure.
Inserting the batteries
The Surtronic Duo uses four 3V Lithium 2450 batteries,
one in the Control Unit and three in the Traverse Unit.
To insert the batteries, remove the battery covers as shown.
Insert the new batteries as shown, taking care to observe the correct
polarity. Replace the battery covers.
After replacing the batteries there will be a short delay following switch on
before normal operation resumes.
Removing the batteries
If the Surtronic Duo is not going to be used for an extended period of time, it is recommended that
the batteries are removed from both the Control Unit and Traverse Unit.
To remove the battery from the Control Unit, remove the battery cover and then press on the edge
of the battery at a point adjacent to either of the two cut-outs in the moulding as shown. The
battery can then be tipped out of the Unit. Replace the battery cover.
To remove the batteries from the Traverse Unit, remove the battery cover and then press down on
the edge of the battery closest to the centre-line of the Unit. The top battery can then be slid out
of the battery compartment. The remaining batteries can be removed by tipping them out of the
Unit. Replace the battery cover.
Battery Disposal
WARNING: DO NOT ATTEMPT TO RECHARGE
THE BATTERIES.
USED BATTERIES MUST BE DISPOSED OF
SAFELY IN ACCORDANCE WITH LOCAL
REGULATIONS.
DO NOT INCINERATE.
Traverse Unit
battery low
Control Unit
battery low
Battery low indication
During a measurement, if the battery power in either unit is low, then a low battery indication will be
given. The following displays are used to indicate a low battery condition and to identify the affected
unit.
Switching on
The Control Unit is powered whenever the batteries are in place. The unit
automatically enters a power saving mode after five minutes from the last
operation. During power saving mode the display is turned off. To activate
the unit, press any button. The display will be activated again and the
welcome screen displayed.
The Traverse Unit automatically enters a power saving mode after five
minutes from the last operation. In this mode a minimal current is drawn,
sufficient only to maintain the power control circuitry. To reactivate the
Traverse Unit press the On button. An LED on the side of the Unit flashes to
confirm activation.
Making a measurement
There are two modes for operating the Surtronic Duo:
Connected Mode and Remote Mode.
Operation of the Control Unit is identical, whether the system is in Connected mode or Remote
mode.
On the Control Unit use the Select Parameter button to select the parameter that is required to be
displayed upon completion of the measurement.
Use the Select Mode button to select either metric or inch units as required.
Remote Mode.
In remote mode the
Control Unit is
positioned within 1m of
the Traverse Unit and
data is passed between
the two units over the
IrDA link.
Note: The alignment
between the Traverse
Unit and Control Unit
must be maintained
throughout the
measurement process.
Connected Mode.
In connected mode
the Control Unit is
connected to the
Traverse Unit as
shown before the
measurement is
made.
To connect the
Control Unit to the
Traverse Unit, insert
the Control Unit as
shown and tilt until the
latch en
g
a
g
es.