Startup
The AFA500/BSC must be eld-calibrated once
the room air supply and exhaust is balanced.
When the unit is powered up, the following
sequence of events occurs:
1. The alarm performs a self-test of its
functions, LEDs and audible alarm
(approximately 2 seconds) and then initiates
a delay timer of 30 seconds to allow the
airow sensor to stabilize.
2. During the 30-second stabilizing period, all
alarms and relay outputs are deactivated
and the red and green LEDs remain on.
3. At the end of the delay, the unit will do one
of two things:
a. If the monitor has been calibrated,
the unit enters normal operating mode
(solid green light for safe velocity, red light
and audible alarm if low velocity).
b. If the unit has not been calibrated,
the red and green LEDs will ash, the audible
alarm will be muted.
Installation &
Calibration Guide
Airow Monitor
AFA500/BSC
Calibration - Single Point (default)
1. The calibration process establishes an alarm point at 80% of the
ow rate captured during calibration. That is, the alarm value will
be 20% less than standard airow.
2. Press and hold the ENTER button for 5 seconds to enter Calibration
Mode. This is indicated by both red and green LEDs ashing with
the audible alarm beeping.
3. To initiate calibration, press and hold the ENTER and SET buttons
at the same time. The unit will sample the airow for 5 seconds,
during which time the green LED goes o and the red LED ashes.
The audible alarm continues to sound during the air sampling.
4. If calibration is successful, the monitor will give a two-tone beep at
the end of the air sample, and then automatically enter run mode.
5. If the ENTER or SET button is released during the air sampling
period, or if the airow is uctuating too much, the alarm will emit
a lower-frequency buzzing for a short period and then re-enter
calibration mode. If this occurs, press the ENTER and SET buttons
again to repeat the airow sampling.
6. If, during normal operation after a successful calibration, the airow
drops below the alarm point, the unit will go into alarm condition
(red LED ashing, audible horn beeping). Push the enter button to
temporarily mute the horn.
7. The horn can be permanently disabled by pressing and holding the
SET button for 10 seconds. The horn will sound 3 times to indicate
that it has been disabled. In Safe mode the Green LED will ash if
the horn is disabled or be solid to indicate that the horn is enabled.
To enable the horn press and hold the SET button for 10 seconds,
the horn will sound 3 times to indicate that it has been enabled.
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com
Installation &
Calibration Guide
Connection Details
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com
Operation
Horn Red LED Green LED Operation
Beep* 1 On On Power up 30 second time period to allow sensor stabilization
O Flashing Flashing Requires Calibration - follow calibration procedure
O O On Air SAFE - audible alarm enabled
O O Flashing Air SAFE - audible alarm permanently disabled
On On O Air Fail
O On Flashing Air Fail - Setback mode (remote horn mute activated)
On Flashing O Visor Open
Installation &
Calibration Guide
Monitor Panel Dimensions
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com
Monitor Panel Cutout Dimensions
(NOT shown to scale)
Airow Sensor Installation Diagram
For complete manual and product information, log on to www.tel-uk.com
5.83”
(148mm)
3.46”
(88mm)
1.57”
(40mm)
1.57”
(40mm)
4.65”
(118mm)
3.15”
(80mm)
4.96”
(126mm)
2 xing holes
2.5 mm diameter
for 2 #4 self-
tapping screws,
provided.

Other TEL Monitor manuals

TEL AFA 500 Operating instructions

TEL

TEL AFA 500 Operating instructions

TEL AFA500 MK2 Use and care manual

TEL

TEL AFA500 MK2 Use and care manual

TEL AFA1000 / 2 - Mk2 User manual

TEL

TEL AFA1000 / 2 - Mk2 User manual

Popular Monitor manuals by other brands

Panasonic WV-LC1700 operating instructions

Panasonic

Panasonic WV-LC1700 operating instructions

NEC MultiSync WD551 Setup manual

NEC

NEC MultiSync WD551 Setup manual

ViewSonic ve510b-3 Service manual

ViewSonic

ViewSonic ve510b-3 Service manual

Samsung SyncMaster 913BM user manual

Samsung

Samsung SyncMaster 913BM user manual

AOC U32N3C user manual

AOC

AOC U32N3C user manual

Dell UltraSharp 34 user guide

Dell

Dell UltraSharp 34 user guide

Philips LDH2180 Instructions for use

Philips

Philips LDH2180 Instructions for use

Konica Minolta ACIES brochure

Konica Minolta

Konica Minolta ACIES brochure

AOpen DT2462M user manual

AOpen

AOpen DT2462M user manual

Philips 190V6 user manual

Philips

Philips 190V6 user manual

Sony FD Trinitron HMD-A420 Service manual

Sony

Sony FD Trinitron HMD-A420 Service manual

LG 42VX30ms Specifications

LG

LG 42VX30ms Specifications

Samsung T24E391AK user manual

Samsung

Samsung T24E391AK user manual

Babies R' Us 3927006 user guide

Babies R' Us

Babies R' Us 3927006 user guide

LG 32QN650-B owner's manual

LG

LG 32QN650-B owner's manual

BenQ G2010WP Service manual

BenQ

BenQ G2010WP Service manual

ViewSonic VS16340 user guide

ViewSonic

ViewSonic VS16340 user guide

Barco MDSC-8358 Series user guide

Barco

Barco MDSC-8358 Series user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.