Thermoline TRH-1250 User manual

HUMIDITHERM
TEMPERATURE AND HUMIDITY CHAMBERS
MODELS INCLUDE:
TRH-1250
TRH-2000
WITH STAR-700 CONTROL
Thermoline Scientific Equipment Pty. Ltd.
T/A Thermoline Scientific ABN 80 000 859 129
10-12 Ross Place. Wetherill Park. N.S.W. 2164. Australia.
P.O. Box 6862, Wetherill Park Delivery Centre, N.S.W. 1851. Australia.
Phone: (02) 9604 3911. International: 61 2 9604 3911.
Fax (02) 9725 1706. International: 61 2 9725 1706.
Email: sales@thermoline.com.au Web: www.thermoline.com.au

2 | P a g e
IMPORTANT INFORMATION
This manual contains operating and safety information. The operator must read and understand
the contents of this manual prior to using this equipment. Incorrect operation may cause harm or
damage the severity is classified by the following Alert Boxes. Ensure that this manual is saved for
future reference:
Warning alerts apply when there is a
Possibility of personal injury.
Caution alerts apply when there is a Possibility
of damage to the Equipment .
To maintain safe operation of this equipment be aware of the ALERT BOXES located throughout
these operating instructions.
CONTENTS
IMPORTANT INFORMATION
2
CONTENTS
2
PRODUCT WARRANTY
3
INTRODUCTION
4
THE MANUAL
4
UNPACKING
4
LOCATION & SERVICES REQUIRED
5
ASSEMBLY & INSTALLATION
6
FUNCTION
8
PERFORMANCE
8
PRINCIPLE OF OPERATION
8
EXPLANATION OF CONTROLS
10
STAR-700 CONTROLLER
10
HIGH & LOW SAFETY
10
HUMIDITY TRIP LIGHT
10
OTHER CONTROLS
11
OPERATING THE CABINET
12
STAR-700 OPERATING GUIDELINES
12
SAFE OPERATING VALUE (SOV)
12
PROGRAM SETUP
15
STARTING PROGRAM
20
TREND SCREEN
22
DATA MENU
23
SECURITY SCREENSAVER
24
CALIBRATION SCREEN
27
SYSTEM SETTINGS & CODES
28
DIAGNOSTICS
29
MAINTENANCE
30
WATER-FLOW ADJUSTMENT
31

3 | P a g e
Standard 12 month Warranty
Thermoline Scientific Equipment Pty Ltd ABN 80 000 859 129 (“Thermoline”)
Thermoline warrants to the original purchaser that this product will perform to its product specification for a period of 12
months from date of purchase, provided that the installation of the product has been carried out in accordance with the latest
version of the manufacturer's instructions and further provided that the use of the product complies with that specified in the
relevant specification. Thermoline is not responsible for any loss or damage arising from incorrect usage, usage outside the
suitability of the product as stipulated in the manufacturer's instruction, damage caused by accident, fire, flood, act of God
or failure to properly install, operate or maintain the goods in accordance with the printed instructions provided.
The following statement applies only to product sales that fall within the definition of a Consumer Sale set out in the
Australian Consumer Law contained within the Competition and Consumer Act (Cth) 2012:
‘Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a
replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You
are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not
amount to a major failure.’
Notwithstanding the preceding clause and to the extent permissible by law, the liability of Thermoline is limited, in relation to the
warranted product and at the option of Thermoline to:
replacing the product or the supply of equivalent product;
the repair of the product;
the payment of the cost of replacing the product or of acquiring equivalent product; or
the payment of the cost of having the product repaired.
To the extent permitted by law, all other warranties whether implied or otherwise, not set out in this Warranty are excluded and
Thermoline is not liable in contract, tort (including, without limitation, negligence or breach of statutory duty) or otherwise to
compensate the Purchaser for:
any increased costs or expenses;
calibration/certification services;
any loss of profit, revenue, business, contracts or anticipated savings;
any loss or expense resulting from a claim by a third party; or
any special, indirect or consequential loss or damage of any nature whatsoever caused by Thermoline’s failure in
complying with its obligations or the purchaser’s failure due to accident damage, impact, misuse or negligence.
The benefits given to the purchaser in this Warranty are in addition to other rights and remedies under a law in relation to the
products or services to which this warranty applies.
This warranty applies only to products purchased and installed in Australia and does not cover any consumable items e.g.
filters, light globes, ultrasonic nebulizers. The warranty does not extend to labour and freight costs where the warranted
product is located outside Australia.
To make a warranty claim, contact Thermoline on 02 9604 3911 or [email protected]m.au.

4 | P a g e
INTRODUCTION
Thank you for selecting this product from the large range of products manufactured in Australia by
Thermoline Scientific.
The materials incorporated in this design have been carefully selected for their durability, technical
performance, and practicability, the resultant product is one we are proud of.
THE MANUAL
Please take the time to read this manual, it contains information that will allow you to understand
how the equipment operates, what it can and cannot do, how to operate the controls, and how to
maintain the equipment.
Ensure that the manual is kept safe but accessible for current and future operators to refer to.
UNPACKING
Remove the equipment from the packing material, check that the delivery is complete.
Delivery should include the following:
1 x Temperature & humidity cabinet (TRH-1250 or TRH-2000)
6 x Stainless steel, open wire shelves.
24 x Stainless steel shelf clips.
2 x Wall spacing rods.
1 x Hose connection fitting.
1 x Perspex manual holder.
1 x Operating manual.
1 x Set of keys for the door locks.
Optional Extras: These items could be including as part of the packing list if ordered:
Water supply bottle: (25 litre bottle complete with hose and connection fittings).
Circular chart recorder: (Complete with chart paper and chart pens).
Data loggers.
Extra shelves.
1 x 4 pin plug for BMS Connection.
Retain the packaging materials until the equipment has been thoroughly tested.
Notify the detail of any damage without delay to your supplier, or Thermoline
Scientific.

5 | P a g e
LOCATION AND SERVICES REQUIRED:
This section deals with the services that your cabinet will require to function properly, and advice
on where to locate the cabinet.
GENERAL:
This equipment is designed to operate under the following ambient conditions:
Indoor use only.
Ambient temperature range: 10º to 35ºC
Ambient humidity range: Maximum relative humidity of 80% for ambient temperatures up to
25ºC.
Optimum performance of chamber is achieved when the ambient temperature is within a
20ºC to 25ºC range.
LOCATION:
Place the cabinet in a well-ventilated area on a firm level surface.
Allow sufficient space around the back and sides of the cabinet to provide adequate ventilation.
Recommended space of 30cm on the sides and at least 10cm at the rear (Note: with spacers
attached).
Ensure that there is sufficient room to allow the doors to open.
SERVICES:
ELECTRICAL:
Confirm that the power supply is suitable for this equipment.
Refer to the equipment rating plate for details.
Model: TRH-1250 15 Amp 240V 50Hz
TRH-2000 15 Amp 240V 50Hz
Note: This equipment requires a 15 amp power point (otherwise known as a general purpose outlet
or gpo). You will notice that it has a large earth pin and will not plug into a standard power point.
WARNING: Ensure that the electrical outlet is earthed and is of the correct
voltage and is capable of carrying the stated current load.
WATER:
A water supply is required for the humidifier (steam generator). Since the water level in the
humidifier tank is automatically regulated, the water supply can come from either mains water or
from a gravity fed water reservoir.
If the water supply is from the mains, we recommend that it should be filtered to prevent the build-
up of mineral salts and residual deposits frequently found in town water. It is advisable that a
Sediment filter and an Activated Charcoal filter be installed in the water supply line to help keep the
total residual deposits down to a minimum. The humidifier is constantly “on the boil” to provide
steam for humidity, and therefore evaporation will leave mineral deposits behind. Failure to keep
the Total Residual Deposits to a minimum can result in premature failure of the Humidifier Tank.
An alternative method to Mains water is to use a water supply from a gravity fed tank. Thermoline
supply an optional 25 litre water tank complete with flexible hose and connection fittings. This
should be elevated by at least 1 metre and filled with water, the 25 litres would last at least 2 ½
days even under the worst operating conditions.
The gravity fed water supply has a number of advantages over mains water supply:
Firstly, the water in the tank can be “treated water” such as distilled water or deionised
water. The deionised water does have the potential under certain conditions to corrode
stainless steel (Refer to Caring for stainless steel on page 30). Using distilled water will not
leave the mineral salt deposits behind that town water does, this will increase the life of the
humidifier and reduce maintenance.
Secondly, if the automatic water level controls were to fail or if there was a water leak, the
worst spillage will only be 25 litres of water instead of a potential flood.
It is advisable that the water to the Humidifier be filtered to prevent the build-up of residual solids. A
Sediment Filter along with an Activated Charcoal Filter would be ideal.

6 | P a g e
DRAIN:
Because we are controlling humidity levels inside the cabinet, there will always be water
condensing and coming out of the drain inside the cabinet. There is also a drain valve on the
humidifier tank for maintenance, and an overflow pipe in case of a malfunction.
Therefore it is essential that the location for the cabinet has a floor drain for the water to flow into.
Water from the cabinet will exit through a 19mm drain pipe at the rear of the cabinet, you will need
to connect this drain pipe to the floor drain (see note below).
IT IS IMPORTANT THAT THE CHAMBER DRAIN OUTLET BE CONNECTED TO
A SUITABLE WASTE OUTLET (DRAIN) AT ALL TIMES. MAKE SURE THAT
THERE ARE NO RESTRICTIONS IN THE DRAIN PIPE, IT WILL NEED TO BE
ABLE TO COPE WITH AN OVERFLOW OF 3 LITRES PER MINUTE IN THE
CASE OF A FAULT.
WE ADVISE THAT THE WATER SUPPLY AND THE DRAIN PIPES SHOULD BE
FLEXIBLE. THIS WILL ALLOW THE CABINET TO BE MOVED AWAY FROM
THE WALL FOR ACCESS TO THE REAR OF THE CABINET DURING SERVICE.
ASSEMBLY & INSTALLATION:
ASSEMBLY:
SHELVES: 4 stainless steel clips are supplied for each shelf. Determine the space required
between each shelf level, and simply hook the clips into the slotted strip attached to the walls.
Place the shelves on top the stainless clips.
PERSPEX HOLDER: This holder is provided so that the operating manual can be kept with the
equipment. The operator can use the 2 x screws mounted on the left hand side of the cabinet to fix
the Perspex holder to the cabinet. (See figure 1 for location).
WALL SPACING RODS: 2 threaded rods are provided that screw into the rear of the cabinet to
prevent it from being pushed too close to a wall. A minimum distance of 100mm is maintained with
the use of these spacers. See Figure 1.
Figure 1: Showing installation of wall spacing rods, location of drain and water-in connections, and location
for the instruction manual holder.

7 | P a g e
INSTALLATION:
Locate the cabinet in a well ventilated area on a firm level surface as described earlier.
Make sure that the 2 x wall spacing rods are screwed into the frame at the rear of the
cabinet (See figure 1). This will ensure that the air flow through the refrigeration condenser
is not impeded.
Leave sufficient room at the rear and sides for air flow around the cabinet as previously
described (30cm each side).
Leave enough room in front of the doors to allow them to open fully when loading the
cabinet.
Connect the drain pipe at the rear of the cabinet (figure 1), to a suitable floor drain with
flexible hose (drain pipe is 19mm diameter).
The cabinet is fitted with castor wheels so that it can easily be moved during installation.
The castors also allow the cabinet to be moved for access to the rear during service. Once
the cabinet is in place it can be immobilised by means of the adjustable feet at the two front
corners of the base. Use a spanner to wind these feet out until they touch the floor.
Connect the water supply to the cabinet, either mains water or gravity fed, to the snap on
connector located at the rear of the cabinet (figure 1).
Note: Gravity fed water tank should be at least 1 metre above the ground, to provide a
suitable head for the water to flow properly.
Before turning on the water supply or the power, make sure that the drain valve on the
humidifier is turned off. The humidifier is located in the area below the doors. Open the
hinged panel below the doors to gain access to the humidifier (see photo below). The drain
valve is shown in the closed position.
Plug the power cable into the power point. Note: this cabinet requires a 15 amp, 240 volt
power point, a standard 10 amp power point is not adequate.
You are now ready to use the cabinet!

8 | P a g e
FUNCTION:
This equipment is used to provide an economical and accurate temperature and humidity
controlled environment for laboratory evaluation and long term storage of products, within the
range shown below in the TRH control envelope.
WARNING: Do not use flammable solvents or combustible materials as fire or explosion
may occur. This equipment uses components which may be a source of
ignition.
PERFORMANCE
The operating range for this equipment is +10°C to 60C and 10% to 95% RH
MAXIMUM TEMPERATURE MAXIMUM HUMIDITY
For this product is 60C For this product is 95% RH.
C-TICK
The equipment referred to here complies with requirement of AS/NZS 1044:1995 electromagnetic
interference (EMI).
PRINCIPLE OF OPERATION:
Read this section in conjunction with figure 2. This will give the operator a guide on how the
cabinet actually controls the conditions within the workspace.
CONTROLS:
Temperature and Humidity is controlled by the STAR-700 touchscreen controller. This is mounted
on the front door of the TRH cabinet.
AIR CIRCULATION:
In order to maintain close temperature and humidity tolerances, good air flow is required. Three
centrifugal fan wheels driven by externally mounted motors, draw the air through a “conditioning
area” below the false floor inside the cabinet. The fan wheels then force the air up the rear
perforated panel to be evenly distributed horizontally across each shelf level. Air then returns to the
“conditioning area” through slots in the floor, for the conditioning cycle to start again.

9 | P a g e
TEMPERATURE:
The air is heated by means of an element below the false floor, and cooled by means of the bare
pipe evaporator. The refrigeration unit is located in the area below the cabinet. The STAR-700
controller cycles the heating and cooling to maintain stable temperature.
HUMIDITY:
The air is humidified by steam from the steam generator located in the area below the main
cabinet. We call the steam generator the “humidifier”. Steam from the humidifier enters the
conditioning area by means of a steam pipe.
The air is dehumidified by the same bare pipe evaporator used to cool the air temperature. The
STAR-700 controller cycles between humidifying and dehumidifying to maintain stable humidity.
Figure 2:

10 | P a g e
EXPLANATION OF CONTROLS
Access: The STAR-700 Controller is located in the front door. On initial power up the STAR-700
displays the screensaver. See STAR-700 Operating Guidelines on how to exit screensaver mode.
Figure 3: Control Panel.
Temperature/Humidity Control: The STAR-700 microprocessor controller allows accurate control
of the Temperature and Humidity with in the cabinet.
High & Low Temperature Alarm: The STAR-700 microprocessor controller has a High and Low
temperature alarm which will become active if the actual temperature deviates the set value by +/-
5.0degC. In the event of a High Temperature Alarm all sources of heating are electrically isolated.
In the event of a Low Temperature Alarm the refrigeration compressor is isolated
Humidity Trip Light: This is located on the Humidifier itself and can only be viewed through the
access door at the front near the floor. This light should only be illuminated when a failure of the
water supply to the humidifier occurs and the floats switch interlock fails, causing an over
temperature condition in the humidifier. To Reset: First correct the water supply fault, then reset
the manual reset safety thermostat at the front of the humidifier (small red button), located behind
the hinged access door at the front of the cabinet. (See figure 4).

11 | P a g e
Figure 4: Humidifier Components.
OTHER CONTROLS:
Element Safety Thermostat: This is a factory set thermostat that will only trip in the event that the
cabinet exceeds its maximum temperature of 60.0degC.
Water Pressure Regulator: Located behind the hinged access panel in the base of the cabinet.
Needle Valve: Located behind the water pressure regulator.
Note: See page 18 on how to adjust regulator and needle valve.
ELECTRICAL - POWER LEAD
A mains power lead with a three pin plug is supplied to fit a 15 Amp General Purpose Outlet
(G.P.O).
OPERATION
WARNING: 1. To avoid injury, do not use flammable solvents in this cabinet, as fire
or explosion may result. This cabinet contains components that may ignite such
materials.

12 | P a g e
OPERATING THE CABINET
1) Ensure that the cabinet is located as previously described, refer to the index, LOCATION,
Page 4.
2) Loading/Unloading the Cabinet
Load Distribution
There is a direct relationship between the temperature/humidity uniformity throughout the
workspace and the air distribution.
If the air cannot freely circulate, then the controlled conditions cannot be maintained.
The air enters the workspace through the rear perforated panel flowing in a forward
direction through the stored product before entering the conditioning area via the
perforations at the floor panels.
Product under evaluation should not be placed against the rear perforated panel and side
walls or over the return air grill at the floor of the chamber.
3) Door Opening
When the doors are open, the controlled conditions will begin to deteriorate. To minimise
losses always close the doors as soon as practical. Should the doors be opened for a long
time, then the control may extend the time to return to the set conditions beyond that
normally expected.

13 | P a g e
‘STAR-700’ TOUCHSCREEN OPERATING GUIDELINES.
Safe Operating Value (SOV).
On initial Power up the STAR-700 Touchscreen shows the Screen Saver page. See Below
Image. On Initial power up cabinet controls the ‘Safe Operating Value’ only. The Touchscreen is
fitted with a Security Screensaver to stop unauthorised access to the controlling parameters. This
Security Screensaver becomes active if the Touchscreen has not been touched for more than
5minutes or if power is cycled.
To exit out of the Security Screensaver simply press the ‘Press to Log In & Exit
Screensaver’ Button. The User Access window will appear. To exit out of the Security Screensaver
and enter into the Main Menu Screen use the following passcodes:
User Number: 1
Passcodes: 1111
These passcode should only be given to Authorised Personnel. To enter the User Number
and/or the Passcode simply press the Numeric Window and the Numeric Keyboard will be
displayed. Simply enter the User Number and/or the Password and press enter. Once you have
pressed enter press the screen anywhere above the Login Button and the Main Menu screen will
appear.
When in the Main Menu Screen you will notice that the indicator in the top right corner
states that the Diurnal/Program Cycle is Off. The Values below the heading ‘Safe Operating Value’
are the current set values and the controller will start to control these values. During ‘Safe
Operating Value’ control, the Lights can only be manually operated by using the drop down menu
underneath the heading ‘Lights (SOV)’. The ‘Lights SOV’ is a manual operation of the Lights, in the
ON mode they are continually on and in the OFF mode they are continually off. To operate the
Lights on a Diurnal/Program Cycle the operator will need set up the Diurnal/Program Cycle. To
change the ‘Safe Operating Value’ simply touch the ‘Operating Value’ box. A popup numeric
keypad is displayed. Type in the new value and press enter. The cabinet will automatically control
the new ‘Safe Operating Value’.
The ‘Safe Operating Value’ has been designed to be set by the operator to a safe condition
that will not cause damage to any of the Research, Experimental and/or Plant Material within the
cabinet, in the event of a power failure. If the controller is operating in ‘Diurnal/Program Cycle’
mode and there is a power failure, on resumption of power the ‘Diurnal/Program Cycle’ mode will
stop and the cabinet will be controlled at the ‘Safe Operating Value’.
Security Screensaver (SOV Mode)

14 | P a g e
Security Screensaver with User Access Window.
Main Menu Screen (SOV Mode)
Main Menu Screen with Pop-Up Numeric Keypad.

15 | P a g e
Program/Diurnal Set-Up.
The STAR 700 has a 250 segment Programmer. The 250 segments are split between 25
Programs (10 Segments per program). The Program Set-Up is securely protected by means of a
user name and password. By pressing the Thermoline Logo the Login Screen appears. The
Username is 02 and the Password is 2222. Once the Password is entered the next touch of the
screen needs to be the ‘Program Set-Up’ button. If the screen is touch anywhere but the ‘Program
Set-Up’ button the password will be reset and will need to be re-entered. Once the Program Set-Up
button has been pressed then the Program Set-Up screen is displayed.
Main Screen with Login Screen showing Login Screen
Diurnal/Program Set Up Screen.
The above screen shows the Program set-up page. This consists of 25 programs having a total of
250segments (10segments per program).
Program 1: Segments 1-10
Program 2: Segments 11-20
Program 3: Segments 21-30
Program 4: Segments 31-40 (etc)
Program 25: Segments 241-250.

16 | P a g e
Each program can be set as an independent program or can be linked to create one large
or multiple large programs. This is achieved by not using an End segment in a particular program.
For example if your program requires the use of 15 segments we can use the 10 segments of
Program 1 and the first 5 segments of Program 2 making Segment 15 (5th Segment of Program 2)
an End Segment type. We would then star program 1 and as the program runs through the first 10
segments of program 1 and because there is no end segment in Program 1 it continues to run
program 2 until it find the end segment at segment 15.
When linking multiple Programs, to use more than 10 segments, the programs need to be
sequential.
By pressing on Program 1 the following screen is displayed. The below example is a Diurnal cycle
of 40degC and 40% day and 10degC and 80% Night. Assuming that 40degC is at midday this
program has been written to start at 25degC and 60%RH assuming that the day condition of
40degC and 40%RH is Midday then this program would Start at 6am.
Program 1 Segment 1-5
Program 1 Segment 6-10
The following is a description of what each of the items in the Program are and what
function they perform.

17 | P a g e
Segment Type: The segment type can either be a Time, Jump To, or End
type.
Time: This determines that the segment is a period of
time. The time is set in Hours, Minutes and
Seconds.
Jump To: This allows the program to repeat a
predetermined set of segments. By selecting
the ‘Jump To’ parameter in segment type and
then entering a segment value at the ‘Jump To’
parameter, will tell the program were to jump
back to. At ‘Jump Cycle’ enter how many times
it jumps back to or repeats the predetermined
set of segments. See Example on next page.
End: This tells the program when to end. There has
to be an end segment. Failure to enter an end
segment will cause the controller to indefinitely
dwell at the last segment.
Jump To: The operator would enter the segment number that the program will jump
back to. This value is ignored if the segment is set to either a Time or End
Segment.
Jump Cycle: The operator would enter a value here as to how many times the ‘Jump To’
repeats itself. A value of 0 will continually repeat the Jump To segments until
the operator ends the Program Cycle manually. This value is ignored if the
Segment Type is set to either a Time or End Segment.
Temperature: The operator would enter the Temperature they require the cabinet to
achieve. This value is ignored if the segment’s Step Type is Jump To.
Humidity: The operator would enter the Humidity they require the cabinet to achieve.
This value is ignored if the segment’s Step Type is Jump To.
Co2: The operator would enter the Co2 they require the cabinet to achieve. This
value is ignored if the segment’s Step Type is Jump To.
Hours: The operator would enter the required length of time for the segment in
Hours. This value is ignored if the segment Step Type is set to either Jump
To or End.
Minutes: The operator would enter the required length of time for the segment in
Minutes. This value is ignored if the segment Step Type is set to either Jump
To or End.
Seconds: The operator would enter the required length of time for the segment in
Seconds. This value is ignored if the segment Step Type is set to either
Jump To or End.
End Action: This value determines what will happen when the program ends. It can be
set to SOV, which stops the program from operating and controls the Safe
Operating Values. Or it can be set to dwell, in which case it will continue to
control the Temperature, Humidity and Co2 of the last Time segment.
G’Soak Temp: This is the Guaranteed Soak Parameter and is only visible if the Guaranteed
Soak is enabled. This parameter is generally not used on the Climatron
range of Plant Growth Cabinets as it can extend the time of a Diurnal
Program effectively making the Diurnal Cycle longer than 24hours. The
Guaranteed Soak allows the Program to maintain its Programmed
Temperature Value for the duration of the Segment Time by automatically
placing the Program into a holding pattern and not allowing the program to
proceed if the Actual Temperature Value deviates the Programmed
Temperature Value by a predetermined value. Once the Actual Temperature
Value is within the predetermine range then the Program will automatically
continue.

18 | P a g e
G’Soak Humidity: This is the Guaranteed Soak Parameter and is only visible if the Guaranteed
Soak is enabled. This parameter is generally not used on the Climatron
range of Plant Growth Cabinets as it can extend the time of a Diurnal
Program effectively making the Diurnal Cycle longer than 24hours. The
Guaranteed Soak allows the Program to maintain its Programmed Humidity
Value for the duration of the Segment Time by automatically placing the
Program into a holding pattern and not allowing the program to proceed if
the Actual Humidity Value deviates the Programmed Humidity Value by a
predetermined value. Once the Actual Humidity Value is within the
predetermine range then the Program will automatically continue.
G’Soak CO2:This is the Guaranteed Soak Parameter and is only visible if the Guaranteed
Soak is enabled. This parameter is generally not used on the Climatron
range of Plant Growth Cabinets as it can extend the time of a Diurnal
Program effectively making the Diurnal Cycle longer than 24hours. The
Guaranteed Soak allows the Program to maintain its Programmed CO2
Value for the duration of the Segment Time by automatically placing the
Program into a holding pattern and not allowing the program to proceed if
the Actual CO2Value deviates the Programmed CO2 Value by a
predetermined value. Once the Actual CO2Value is within the predetermine
range then the Program will automatically continue.
Lighting/Event Outputs.
At each segment page there is a Lighting/Events button. By pressing this button the display
shows Eight events that can be turned ON and OFF for each segment. The First Event which is
generally labelled Lights is what turns the lights ON and/or Off. A total number of Eight
Lighting/Event Outputs are available, however only the required hardware is fitted per cabinet.
Program 1 Lighting/Event Segment 1-5

19 | P a g e
Program 1 Lighting/Event Segment 6-10
Program Features.
There are a few other features of the program that cannot be changed but are described below.
Guaranteed Soak: This is a tolerance value, set at +/-2.0degC, and guarantees that the
Program Set Values for Temperature only are reached. If during a
program cycle the Actual Temperature deviates the Program Set
Value (PSV) by more than +/-2.0degC the program will be put into a
hold mode until the Actual Temperature is within the PSV. There is
no Guaranteed Soak on the Humidity Value or the Co2 value. These
values can deviate without the program going into a hold mode.
Program Start Value: The Program starts at whatever the Actual Temperature, Humidity
and Co2 are. If when starting the program the Actual Temperature is
more than +/-2.0degC from Segment 1’s PSV the Guaranteed Soak
value will hold the program until it reaches Segment 1’s target
temperature.
Power Failure: In the event of a power failure the Program Cycle will end. On
resumption of power the cabinet will control the Safe Operating
Values.

20 | P a g e
Starting the Diurnal/Program.
The following screen shows the Diurnal/Program Set-Up screen. Once the desired program
has been set up then the operator will need to enter the desired program number value into Start
Program Number. Once the desire value has been entered the operator can then start the program
by using the drop down menu. The following describes the item and the function of the parameters
in the drop down menu.
End The end parameter is what stops a current running program from operating
Start The Start parameter is what starts the program that is entered into the Start
Program Number.
Pause The Pause Parameter will simply pause the program at its present point. It
will hold this point until the resume parameters is selected. If the Start
parameter is selected while a program is in pause it can restart the program
from the beginning.
Resume The Resume parameter will resume a program after it has been paused.
Diurnal/Program Set Up Screen.
Note: The Program can be Started, Stopped, Paused and Resumed in the Main Menu Screen.
However you cannot select the Program Number in the Main Menu Screen.
When the program is operating the Start Program Number value disappears and the
Program Number Running value appears. The Program Running Number is a read only value just
to give the operator and understanding of which program is operating. To change the Program that
is running the operator will need to End the program to change the Start Program Number. This
can only be performed in the Program Set-Up page.
This manual suits for next models
2
Table of contents
Popular Climate Chamber manuals by other brands

Memmert
Memmert HPPlife SUPPLEMENT TO THE OPERATING INSTRUCTIONS

Binder
Binder MKF operating manual

Folyon
Folyon H300 user manual

flowair
flowair LEO FB 95 Technical documentation operation manual

Hanse
Hanse VTC Series Installation, operation and maintenance instructions

Thermo Electron
Thermo Electron 3920 Operating and maintenance manual