Varex Imaging Claymount AmpMC 3-field Series User manual

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 2/24 www.vareximaging.com
Table of Contents
Table of Figures ........................................................................................................................................3
1. Introduction ...................................................................................................................................4
Contact information..............................................................................................................4
Declaration of Conformance ................................................................................................4
Symbols used in this document ...........................................................................................4
Abbreviations .......................................................................................................................5
General warnings, cautions and notes ................................................................................5
Supplied components ..........................................................................................................5
Accessories..........................................................................................................................6
2. Product description .......................................................................................................................7
Intended use ........................................................................................................................7
Description of the device......................................................................................................7
Principle of operation ...........................................................................................................7
Classifications ......................................................................................................................7
Restrictions on use ..............................................................................................................8
Contraindications .................................................................................................................8
Overview of the device.........................................................................................................8
Specifications .......................................................................................................................9
3. Installation.................................................................................................................................. 10
Installation requirements................................................................................................... 10
General Installation instructions........................................................................................ 10
3.2.1. Check according to the build-in LED’s .............................................................................. 10
3.2.2. Generator Switch off check................................................................................................11
Necessary recurrent testing ...............................................................................................11
4. Mains isolation ........................................................................................................................... 12
5. Service, maintenance and cleaning .......................................................................................... 12
Safety precautions ............................................................................................................ 12
Cleaning ............................................................................................................................ 12
Disinfection........................................................................................................................ 12
Procedure at defects......................................................................................................... 12
6. Device Data ............................................................................................................................... 13
1001, 1006, 1008, – 3-field integrating amplifier............................................................... 13
6.1.1. Generator Interface connections....................................................................................... 14
6.1.2. Sensitivity settings ............................................................................................................ 14
6.1.3. Ramp Polarity settings ...................................................................................................... 15
6.1.4. Inputs reset / exposure settings ........................................................................................ 15
6.1.5. Inputs for measuring field selection .................................................................................. 15
6.1.6. Plug pinout reset / ramp selection..................................................................................... 16
6.1.7. Assignment of measuring fields to the signal inputs......................................................... 16
7. Quality Assurance...................................................................................................................... 16
8. Disposal, ESD and EMC compatibility ...................................................................................... 17
Disposal ............................................................................................................................ 17
ESD................................................................................................................................... 17
EMC compatibility ............................................................................................................. 17
8.3.1. Deviations ......................................................................................................................... 17
8.3.2. Allowances ........................................................................................................................ 18
8.3.3. Precautions ....................................................................................................................... 18
8.3.4. Emissions Compliance...................................................................................................... 18
8.3.5. Immunity Compliance ....................................................................................................... 18

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 3/24 www.vareximaging.com
9. Product label and symbols on the device.................................................................................. 21
Product label ..................................................................................................................... 21
Symbols on the device ...................................................................................................... 21
Table of Figures
Figure 1 Means of protection................................................................................................................. 12
Figure 2 Image of an AmpMC ............................................................................................................... 13
Figure 3 Lay-out Mechanical ................................................................................................................. 13
Figure 4 DIL switch location .................................................................................................................. 13
Figure 5 The connection against GND in active H mode ...................................................................... 15
Figure 6 The connection against GND in active L mode....................................................................... 15
All rights reserved
Although this guide is prepared with utmost care, Varex Imaging Nederland B.V. assumes no
responsibility for errors or omissions.
Varex Imaging Nederland B.V. cannot be held accountable for damages of any nature arising from the
use, and / or use of any options other than original Varex Imaging Nederland B.V. products.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 4/24 www.vareximaging.com
1.Introduction
Contact information
This manual provides all the technical information necessary for the correct installation, application
and maintenance of the AmpMC.
If you need additional information, need support or want to report a problem with the device, please
contact your distributor or Varex Imaging Nederland B.V.:
Manufacturer
Distributor
Name
Varex Imaging Nederland B.V.
Address
Anholtseweg 44
7091 HB Dinxperlo
The Netherlands
Telephone
+31 (0)315 659150
Fax
+31 (0)315 659140
E-mail
Netherlands.CNC@vareximaging.com
Website
www.vareximaging.com
For support and service purposes, please note the following information:
Model name:
Part number:
Serial number:
Declaration of Conformance
Varex Imaging Nederland B.V. hereby declares that this product is in conformity with the essential
requirements and provisions as set forth in European Union Council Directive 93/42/EEC concerning
medical devices (revision 2007-09-27). See the included Declaration document.
Symbols used in this document
To ensure adequate and clear understanding of the information provided in this manual, the symbols
listed below are used to indicate warnings, cautions, actions and notes that are important for correct
and safe use of the device.
WARNING:
Warnings are directions which, if they are not followed, can cause fatal or serious injuries
to a user, engineer, patient or any other person or can lead to a mistreatment.
CAUTION:
Cautions are directions which, if they are not followed, can cause damage to the device
described in this manual or any other equipment or goods and can cause environmental
pollution.
NOTE:
Notes provide advice and highlight unusual points. A note is not intended as an
instruction.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 5/24 www.vareximaging.com
Abbreviations
Term
Definition
AEC
Automatic Exposure Control
EMC
Electromagnetic compatibility
ESD
Electro Static Discharge
ME
Medical Equipment
This document contains terminology and definitions based on (international) standards. The
terminology and definitions are formatted in capital letters (e.g. INTENDED USE). Terminology and
definitions from the following standards are used:
Reference
Title
IEC 60601-1
Medical electrical equipment – Part 1: General requirements for basic safety
and essential performance
IEC 60601-1-2
Medical electrical equipment – Part 1-2: General requirements for basic safety
and essential performance – Collateral Standard: Electromagnetic disturbances
– Requirements and tests
IEC 60601-1-3
Medical electrical equipment – Part 1-3: General requirements for basic safety
and essential performance – Collateral Standard: Radiation protection in
diagnostic X-ray equipment
IEC 60601-2-54
Medical electrical equipment – Part 2-54: Particular requirements for the basic
safety and essential performance of X-ray equipment for radiography and
radioscopy
General warnings, cautions and notes
WARNING:
To avoid the risk of electric shock, this equipment must only be connected to a system
isolated from supply mains according to IEC-60601-1
WARNING:
Do not modify this equipment without authorization of the manufacturer.
WARNING:
The device contains sensitive electronics. Ensure that ESD protective measures are in
place when the device is installed or serviced to prevent damage to the device.
Supplied components
The device that you have purchased is packed in a transport box appropriately designed to ensure the
integrity of the device. Please ensure that the contents of the package you received is intact and that
there are no traces of moisture or visual damages. Otherwise, you should immediately contact your
distributor or Varex Imaging Nederland B.V.. The package contains the following components:
Amount
Description
Reference
1
AmpMC
For Model and Part number see the product label on
the AmpMC.
1
Technical Manual
TM20514-10
1
ECDoC
Declaration of conformance (CE)

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 6/24 www.vareximaging.com
Accessories
The following accessories can be ordered separately.
Description
Purpose
Extension cables, for use between SolidStateMC and AmpMC
EXTENSION CABLE, SSMC, 2.5M
To connect all 3-field SolidStateMC models to AmpMC
model 1001 + 1006 + 1008.
EXTENSION CABLE, SSMC, 5M
EXTENSION CABLE, SSMC, 10M
EXTENSION CABLE, SSMC, 45FT
EXTENSION CABLE, SSMC, 15M
EXTENSION CABLE, SSMC, 18M
EXTENSION CABLE, SSMC, 20M
EXTENSION CABLE, SSMC, 25M
Extension cables, for use between the AmpMC and HV generator
EXTENSION CABLE PREAMP, 10M
SubD 9-pin male - SubD 9-pin Female
EXTENSION CABLE PREAMP, 15M
SubD 9-pin male - SubD 9-pin Female
EXTENSION CABLE PREAMP, 20M
SubD 9-pin male - SubD 9-pin Female
EXTENSION CABLE, PREAMP HMC
SubD 9-pin female - AMP 10-pin female
EXTENSION CABLE, PREAMP, 5M
SubD 9-pin male - SubD 9-pin Female
Various
Female connector 6 pin
Connector without cable, solder version. To connect to
SolidStateMC
Male connector 6 pin
Connector without cable, solder version. To connect to
AmpMC

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 7/24 www.vareximaging.com
2.Product description
Intended use
The device is intended to be used in medical diagnostic applications with restrictions to human
diagnostics. The device is intended to be used as accessory of an X-ray system in a professional
healthcare facility environment. The device is intended to amplify and integrate the output signal of a
SolidStateMC to the AEC-module of an X-ray generator. The limitations of use are specified in §2.8.
The device is not intended to be used in fluoroscopy applications; near active HF Surgical Equipment
or in the RF shielded room of a system for magnetic resonance imaging, where the intensity of
Electromagnetic Disturbances is high.
Use, other than above, is identified as abnormal use.
The device is intended as an internal component of an X-ray system. The intended user is defined as
the SERVICE PERSONNEL of an X-ray system.
Description of the device§§
The purpose of the Automatic Exposure Control (AEC) of an X-ray system is to obtain the correct
image contrast by measuring the radiation quantity striking the film or detector.
To result in a high-quality image, the AEC’s exposure end switch will stop generating X-rays
automatically when the demanded radiation dose on the detector is reached.
Principle of operation
The AmpMC main functions are:
•The conversion of electrical current from a SolidStateMC to a voltage ramp. The voltage ramp
signal is proportional to the received dose of the SolidStateMC.
•To receive field-select signals (from operator via console and generator) and use the correct
field of the SolidStateMC as the incoming signal.
A SolidStateMC serves as a measuring device for X-ray radiation with semiconductor components
(photodiodes) as actual sensors. The ionizing effect of X-ray produces a small electrical current in the
photodiodes and this current is lead to an AmpMC.
The AmpMC converts the small current to a voltage signal that is proportional to the X-ray dose rate,
this signal is integrated and presented as a ramp signal to the AEC controller board inside the
generator and therefore represents a value for the image density.
•There is no signal available for external use about the status of the device.
Classifications
Subject
Classification
Reference
CE
IIB
93/42/EEC
Electrical safety
None
IEC 60601-1
Electromagnetic Compatibility
intended environment
Professional
healthcare facility
environment
IEC 60601-1-2
Mode of operation
Continuous
IEC 60601-1
Ingress protection
IP2X
IEC 60529
Protection
No
IEC 60601-1 table 6
Not intended for use in Oxygen Rich environment.
Not suitable for Sterilization.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 8/24 www.vareximaging.com
Restrictions on use
The AmpMC can only be used in combination with a SolidStateMC and in a Radiography X-ray system
that complies with the IEC60601-1 standard, applicable at date of manufacture.
The supply lines must be limited / fused to 15W max. or the AmpMC must be mounted inside a fire
enclosure.
The AmpMC is intended to be installed inside a cabinet of the X-ray system.
Depending on the final assembly, additional measures may need to be taken to comply with EMC
regulations.
Contraindications
The relevant contraindications for the X-ray system continue to apply (see the documentation for the
X-ray system). The AmpMC does not add new contraindications on top of these.
Overview of the device
Table 1 AmpMC specifications
Model
Dimensions
(mm)
Support information
Lay-out
Remarks
Outside
# of
fields
Connections
SolidStateMC
Generator
W
L
H
Cable
(m)
Connector
Cable
(m)
Connector
1001
42
222
21
3
None
6 pin Female
0.93
Sub-D 9p male
§ 6.1
3-field SolidStateMC preamp
with single voltage ramp output
1006
3-field SolidStateMC preamp
with single voltage ramp output
and modified reset circuitry
1008
0.8
AMP Header 10p female
/ GND terminal
3-field SolidStateMC preamp
with single voltage ramp output
Configuration of connections
Connector type at generator side
Cable length at generator side
Connector type at SolidStateMC side
Cable length at SolidStateMC side
# of fields that can be controlled
Height of AmpMC
Length of AmpMC
Width of AmpMC

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 9/24 www.vareximaging.com
Specifications
Description
Reference
Exposure time range
≤ 4 ms to 5 s.
Fine-tuning with potentiometer
- 60% to + 150%
Input range
0.1 – 12.000 nA.
Operating voltage
+/- 12 VDC +/-5% to +/- 15 VDC +/-5%
Measured at the AmpMC.
Supply current
0.1 A @ + 12 VDC 0.1 A @ - 12 VDC
0.1 A @ + 15 VDC 0.1 A @ - 15 VDC
Output signal Ramp
Positive or negative ramp
0 to 10 V @ 12 V
0 to 12 V @ 15 V
Max. Load 47 kΩ
Applicable standards
IEC60601-1:2005 edition 3.0 + C1 + C2 + IS1 + IS2 + A11+ A1
IEC60601-1-2:2014
IEC60601-2-54:2009+C1+C2 + A1
Housing Material
Sheet metal
Weight
< 0.5 kg
Operation environment
Ambient temperature: +10°C to +40°C
Relative humidity: 35% to 85% non-condensing
Atm. Pressure: 860hPa to 1060hPa
Storage
Ambient temperature: -20°C to +60°C
Relative humidity: 35% to 85% non-condensing
Atm. Pressure: 860hPa to 1060hPa

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 10/24 www.vareximaging.com
3.Installation
NOTE:
Allow system to level with room temperature before installation.
Installation requirements
WARNING:
Installation and initial operation may only be carried out by an expert who has been
trained in the field of medical diagnostic X-ray equipment.
WARNING:
Modifications to the product are not allowed.
WARNING:
When the enclosure is opened, ESD protective measures must be taken to prevent
damage to the electronics.
WARNING:
Always use shielded cable, the total length of this cable shall not exceed 30 meters.
WARNING:
Improper grounding can cause incorrect functioning of the AmpMC
WARNING:
Adjust the potentiometers with great caution. Using excessive force or over adjustment
will result in improper functioning of the AmpMC
General Installation instructions
Position the AmpMC always outside of the active image area. The position of the AmpMC is not critical
and can be mounted anywhere on the system. If the cable lengths are inadequate see the accessories
list in §1.7.
Connect the AmpMC to the control electronics in the generator by means of the connector, see
chapter 6. The connection diagram and plug pin lay-out of this cable are found under chapter 6.
Correct functioning of the AmpMC is guaranteed only if the cable as well as the SolidStateMC
shielding is properly connected.
After functional control, the automatic exposure control is set to the correct dose, checked in all kV-
ranges and for all film-foil-combinations (if applicable) and put into operation. This procedure must be
done according to the manual of the x-ray system.
Check according to the build-in LEDs
LED + 12 / 15 VDC: The LED is on, when the positive supply voltage is present.
LED - 12 / 15 VDC: The LED is on when the negative supply voltage is present.
LED – Reset: The LED is on at reset, off at exposure.
LED’s field I, field II and field III: The corresponding LED’s are on when the field is switched active.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 11/24 www.vareximaging.com
Generator Switch off check
Without a patient or phantom, a radiation exposure with 80kV, 100mA, 21 s is released. The Automatic
Exposure Control must terminate the exposure in less than 100 ms.
Necessary recurrent testing
WARNING:
Before granting the Automatic Exposure Control for use on patients, check the
functionality of all AEC fields with a phantom.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 12/24 www.vareximaging.com
4.Mains isolation
Figure 1 Means of protection
The AmpMC always needs to be connected to an X-ray system that complies with the required
regulations and standards.
5.Service, maintenance and cleaning
Refer service to a qualified service technician only.
AmpMC models do not require maintenance and will last during the lifetime of the X-ray system. For
calibration see chapter 7.
In case of malfunction of the AEC system, the AmpMC can be checked according to the described test
procedure §5.4.
Safety precautions
When there is structural damage to the housing or cable of the device, label the device as “out of
order” and have the device repaired prior to further use.
Cleaning
Cleaning with a damp cloth is recommended. Use generally available alcohol-based cleaning agents
and do not soak the device with liquid.
Disinfection
Disinfection, when required, with a damp cloth with Isopropyl alcohol is recommended. Before using a
disinfectant, check at a spot on the bottom of the device if the disinfectant will not damage the plastic
and coated metal surfaces.
Do not soak the device with liquid.
Procedure at defects
- Exchange the extension cable (if present).
- Exchange the AmpMC
Secundary
circuit
AmpMC
SolidStateMC

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 13/24 www.vareximaging.com
6.Device Data
Figure 2 Image of an AmpMC
1001, 1006, 1008, – 3-field integrating amplifier
Figure 3 Lay-out Mechanical
Figure 4 DIL switch location

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 14/24 www.vareximaging.com
Generator Interface connections
Designation (Standard)
Pin Numbers for
Sub D
9 pole male
Connector
10 pins
Not connected
1
1
Field II select
2
2
Field I select
3
3
Reset / exposure select
4
4
Ramp output signal
5
5
Field III select
6
6
- 12 VDC to - 15 VDC
7
7
+ 12 VDC to + 15 VDC
8
8
GND (Shield)
9
9
Not connected
N/A
10
Sensitivity settings
By means of DIL-switches the configuration of the following functions is performed.
Definition of setting the DIL switches:
ON Switch closed
OFF Switch open
X Any position
The position of the DIL-switches on the PCB board is shown in Figure 4.
The sensitivity step is selected by means of DIL-switch S1. Fine adjustment in the range of - 60 % to +
150 % is performed by means of potentiometer sensitivity.
The position of the potentiometer is shown in the drawing in Figure 3.
DIL Switch S1
Sensitivity Indication
Setting
code
S1.1
S1.2
S1.3
S1.4
SolidStateMC
nA/µGy/Sec.*
Signal V/µGy
OFF
OFF
OFF
x
1.2
1.5
2.4
0.64
0.8
1.3
0
ON
OFF
OFF
x
1.2
1.5
2.4
0.32
0.4
0.65
1
ON
ON
OFF
x
1.2
1.5
2.4
0.16
0.2
0.33
2
Standard version
ON
ON
ON
x
1.2
1.5
2.4
0.08
0.1
0.16
3
* depends on which SolidStateMC is used. Refer to the manual of the used SolidStateMC.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 15/24 www.vareximaging.com
Ramp Polarity settings
The polarity of the output voltage ramp is selected by means of switches 1 and 2 of DIL-switch S2.
S2.1
S2.2
Ramp polarity
Setting
code
OFF
OFF
Output open
0
ON
OFF
Positive ramp
1
Standard version
OFF
ON
Negative ramp
2
ON
ON
Output Shorted, not allowed
-
Inputs reset / exposure settings
The selection of active L or active H mode for reset / exposure input is performed by means of switch
3 on DIL-switch S2.
S2.3
Interface
type
Function reset / exposure
Setting
code
OFF
Active H
H=Exposure
L=Reset
0
ON
Active L
L=Exposure
H=Reset
1
Standard version
Inputs for measuring field selection
The choice of active L or H mode for the field selection inputs is performed by means of switch 4 in
DIL-switch S2.
S2.4
Interface
type
Function Field selected
Setting
code
OFF
Active H
H=Field selected
L=Field off
0
ON
Active L
L=Field selected
H= Field off
1
Standard version
Figure 6 The connection against GND in active L mode
Figure 5 The connection against GND in active H mode

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 16/24 www.vareximaging.com
Plug pinout reset / ramp selection
The plug pinout on the SUB-D plug for reset and ramp may be changed by means of DIL-switch S3.
S3.1
S3.2
S3.3
S3.4
Ramp
Reset
Setting
code
ON
OFF
OFF
ON
Pin 5
Pin 4
0
Standard version
OFF
ON
ON
OFF
Pin 4
Pin 5
1
Assignment of measuring fields to the signal inputs
The plug pinout on the SUB-D plug for the measuring field selection may be assigned by means of
DIL-switch S3.
S3.5
S3.6
S3.7
S3.8
Field I
Field II
Field III
Setting
code
ON
OFF
OFF
x
Pin 3
Pin 2
Pin 6
0
Standard version
OFF
ON
OFF
x
Pin 3
Pin 6
Pin 2
1
ON
ON
OFF
x
Pin 2
Pin 3
Pin 6
2
OFF
OFF
ON
x
Pin 6
Pin 3
Pin 2
3
ON
OFF
ON
x
Pin 2
Pin 6
Pin 3
4
OFF
ON
ON
x
Pin 6
Pin 2
Pin 3
5
7.Quality Assurance (QA)
An AmpMC makes part of the X-ray systems performance requirements and the adjustment
procedures for the complete X-ray system are mandatory.
There are no additional QA procedures for using the AmpMC.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 17/24 www.vareximaging.com
8.Disposal, ESD and EMC compatibility
Disposal
This device contains substances that can be hazardous to the environment and care should be taken
when disposed of.
The device is marked with the following symbol:
Follow local regulations regarding disposal of devices that contain electronic parts.
ESD
WARNING:
The device contains sensitive electronics. Ensure that ESD protective measures are in
place when the device is installed or serviced to prevent damage to the device.
EMC compatibility
The device conforms to IEC 60601-1-2:2014 for EMC compatibility and must be installed and put into
service according to the EMC information provided in this manual.
WARNING:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the AmpMC 3-field, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
WARNING:
Not taking EMC measures into account on the wiring may result in increased
EMISSIONS or decreased IMMUNITY. IEC60601-1-2:2014 must be followed for being
complaint with EMC guidelines.
WARNING:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
WARNING:
Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased Electromagnetic
emissions or decreased Electromagnetic immunity of this equipment and result in
improper operation.
NOTE:
The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
Deviations
No deviations from IEC60601-1-2:2014 are applied.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 18/24 www.vareximaging.com
Allowances
No allowances from IEC60601-1-2:2014 are used.
Precautions
Precautions to be taken to prevent adverse events to the PATIENT and the OPERATOR due to
Electromagnetic Disturbances are listed in the column “Electromagnetic environment – guidance” in
the tables below.
Emissions Compliance
Guidance and manufacturer’s declaration – Electromagnetic emissions
The AmpMC 3-field is intended for use in the Electromagnetic environment specified below. The customer or user of the
AmpMC 3-field should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions group
CISPR 11
Group 1
The AmpMC 3-field uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions class
CISPR 11
Class A
The EMISSIONS characteristics of AmpMC 3-field make it
suitable for use in industrial areas and hospitals (CISPR 11 class
A). If it is used in a residential environment (for which CISPR 11
class B is normally required) this equipment might not offer
adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as
relocating or re-orienting the equipment.”
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not Applicable
Immunity Compliance
Guidance and manufacturer’s declaration – Electromagnetic immunity – ENCLOSURE PORT
The AmpMC 3-field is intended for use in the Electromagnetic environment specified below. The customer or user of the
AmpMC 3-field should assure that it is used in such an environment.
Immunity test
Compliance Test level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV
air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3
See next table below
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches)
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Not applicable
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 19/24 www.vareximaging.com
Guidance and manufacturer’s declaration – Electromagnetic immunity –
Compliance Test Levels for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
Band
Servicea)
Modulation
Maximum
power
Distance
IMMUNITY
TEST LEVEL
(MHz)
(MHz)
(W)
(m)
(V/m)
385
380 –390
TETRA 400
Pulse modulation b)
1,8
0,3
27
18 Hz
450
430 – 470
GMRS 460;
FRS 460
FM
± 5 kHz deviation
2
0,3
28
1 kHz sine
710
704 – 787
LTE Band 13,
17
Pulse Modulation b)
0,2
0,3
9
745
780
217 Hz
810
800 – 960
GMS 800/900;
TETRA 800;
iDEN 820;
CDMA 850;
LTE Band 5
Pulse Modulation b)
2
0,3
28
870
930
18 Hz
1 720
1 700 – 1
990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse Modulation b)
2
0,3
28
1 845
1 970
217 Hz
2 450
2 400 – 2
570
Bluetooth;
WLAN
802.11 b/g/n;
RFID 2450;
LTE Band 7
Pulse Modulation b)
2
0,3
28
217 Hz
5 240
5 100 – 5
800
WLAN 802.11
a/n
Pulse Modulation b)
0,2
0,3
9
5 500
217 Hz
5 785
a) For some services, only the uplink frequencies are included.
b) The carrier is modulated using a 50 % duty cycle square wave signal.

Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09 20/24 www.vareximaging.com
Guidance and manufacturer’s declaration – Electromagnetic immunity – Power and Signal PORTs
The AmpMC 3-field is intended for use in the Electromagnetic environment specified below. The customer or user of the
AmpMC 3-field should assure that it is used in such an environment.
Immunity test
Compliance Test level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV AC and DC power
ports
± 1 kV signal ports
100kHz repetition frequency
Mains power quality should be that of a typical commercial or
hospital environment.
Surges
Line-to-line
IEC 61000-4-5
Not Applicable
The AmpMC 3-field is intended to be supplied by a secondary
IEC60601-1 compliant AC or DC power supply.
Surges
Line-to-ground
IEC 61000-4-5
Not Applicable
Conducted
disturbances
induced by RF fields
IEC 61000-4-6
3 V a)
0,15 MHz – 80 Mhz
6 V a) in ISM bands between
0,15 MHz and 80 MHz b)
80 % AM at 1 kHz
Field strengths from fixed RF transmitters, as determined by an
Electromagnetic site survey c), should be less than the compliance
level in each frequency range.
Interference may occur near equipment marked with the following
symbol:
Voltage dips
IEC 61000-4-11
Not Applicable
The AmpMC 3-field is intended to be supplied by a secondary
IEC60601-1 compliant AC or DC power supply.
Voltage interruptions
IEC 61000-4-11
Not Applicable
a) r.m.s. before modulation is applied
b) The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553
MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15
MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to
10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to
29,7 MHz and 50,0 MHz to 54,0 MHz.
c) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the Electromagnetic environment due to fixed RF transmitters, an Electromagnetic site survey should be
considered. If the measured field strength in the location in which the AmpMC 3-field is used exceeds the applicable RF
compliance level above, the AmpMC 3-field should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the AmpMC 3-field.
NOTE
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people
This manual suits for next models
3