ACKERMANN 16-2083 User manual

SHAVER SYSTEM
Ackermann
www.ackermannsurgical.com
• Membrane Suction-Irrigation Pump #16-2083
User Manual
Membrane Suction-Irrigation Pump

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 2 // 20
CONTENTS
1. Front and rear panel 3
3. Intended use
6. Installing the bottle holders 7
7. Spike 7
7.1. Handling of spike 7
7.2. Reprocessing of spike 8
8. Secretion container 8
8.1. Handling of secretion container 8
8.2. Reprocessing of secretion container 8
9.1. Elements of tubing set 9
9.2. Reprocessing of reusable tubing set 9
9.3. Illustration of tubing set 9
9.4. Connecting the tubing set 11
9.5. Removing the tubing set 11
9.6. Using the device in operating room 11
10. Footswitch 12
11.1. Functional buttons 13
11.1.1. Power switch 13
11.1.2. On, off 13
11.1.3. Start, stop 13
11.1.4. Up, down 13
11.2. Start-up, self test 14
11.3. Functional test 14
11.4. Operation of the device 14
15.1. Checking the pneumatic control 16
15.2.Checkingtheowrate 17
16. Accessories 18
17. Appendix 18
18. Contact details 18

Ackermann
Page: 3 // 20
SHAVER SYSTEM
1.
1 LED green START
2 Button START/STOP
3 LED yellow STOP
LED yellow STANDBY
Button ON/OFF
6 LED green ON
7 LL-connector (male) IRRIGATION
8 Buttons PRESSURE UP/DOWN (IRRIGATION) with
irrigation liquid (separately
9 Display PRESSURE (IRRIGATION) available)
10 Display VACUUM (SUCTION)
11 Buttons VACUUM UP/DOWN (SUCTION)
12 LL-connector (female) SUCTION connectors and
overowvalve
13 Connector footswitch (START/STOP)
Power switch
Fusebox
16 Mains connection
17 Equipotential bonding pin
18 Label next check
19Modelidenticationplate
20 Bottle holder
21 Bottle
22 SPIKE IRRIGATION
23 Secretion container with tube

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 4 // 20
Parameter Membrane suction-irrigation pump
Power consumption Trafo-Model =>30 VA
Power supply-Model=>40 VA
Supply voltage Trafo-Model =>230 V/AC-Power fuses T400mA L 250A
Power supply-Model=>100-230 V/AC-Power fuses T3,15A L 250V
Power frequency 50/60 Hz
Classicationacc.toEUdirective93/42/EEC Class IIa
Type acc. to DIN EN 60601-1:2005 BF
Protection class acc. to DIN EN 60601-1:2005 I
Manufactured acc. to DIN EN 60601-1 2005
EMC DIN EN 60601-1-2 2007
Irrigation capacity 3 l/min
Suction capacity 3 l/min
Pressure range 750 mm Hg
Vacuum range - 525 mm Hg
Operating conditions Temperature: +10°C to +40°C
Humidity: 30% to 70%
Storage and transport conditions Temperature: -40°C to +70°C
Humidity: 10% to 90%
Case protection IPX1
Tubing sets Silicone - autoclavable
Dimensions without bottle holders 304 x 115 x 345 mm (WxHxD)
with bottle holders ca. 560 x 115 x 345 mm (WxHxD)
Weight ca. 9 kg
Optional Footswitch
Certication CE0483
16-2083 isa uidpumpfeaturing pressure limitation andis used inminimallyinvasive surgery.Itcanbe applied in all
endoscopicsurgicalprocedureswherethebodycavities arenotcompletelylledwithliquid andwherethevolumeof
irrigation liquid is small compared to the volume of the anatomical cavity.
The pump system is ideally suited for use in the abdomen, thorax and retroperitoneum.

Ackermann
Page: 5 // 20
SHAVER SYSTEM
WARNING! The application of the membrane suction-irrigation pump for small and closed body cavities (arthroscopy and
hysteroscopy) is strictly forbidden, because irrigation process may lead to uncontrolled and dangerous pressure rise.
Check device and accessories immediately on receipt for completeness and transport damage. In case of damages please
inform your authorized dealer immediately.
If the unit has to be returned use only original box and attach the following information:
•Nameandaddress
•ModelNo./SerialNo.
•Failuredescription
•Errormessage
ATTENTION! Contaminated devices are not accepted for check or repair.
The unit consists of:
Pos. Article no.
1Membrane suction-Irrigation pump 16-2083 1
2Test report n/a 1
3User manual n/a 1
4Silicone suction-irrigation tubing set, autoclavable 16-2083-102/-103 1
5Main cable Euro 2 m 16-2083-104 1
6Fuses 16-2083-105 2
7Secretioncontainerwithoverowvalve,autoclavable 16-2083-101 1
8Bottle holder (1 litre) + 2 screws for each 16-2083-106 2
9Spike, autoclavable 16-2083-109 1
10 Hydrophobicsterilelter,sterilepackage,disposable 16-2083-108 2
Please make yourself familiar with this unit according to description in the user manual.
Please note:
•Thepumpmustonlybeusedinendoscopicoperations.Useofpumpforotherthanmentionedproceduresisnotallowed.
Otherwise operator is responsible for any kind of risk.
•Thetypeofirrigationliquidmustbechoseninaccordancewiththeselectedsurgicalprocedure.
•Thedevicemustonlybeusedbytrainedspecialistsatmedicalinstitutions.
•Foryourandyourpatient‘ssafetyuseonlyoriginalaccessories.
•Serviceandmaintenancehavetobecarriedoutonlybyauthorizedpersons.
•Thedeviceisnotexplosion-proofandmust

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 6 // 20
not be used in the vicinity of combustible gases.
•Themanufacturerisresponsibleforthesafety,reliabilityandperformanceofthesystemonlyif:
- Assembly, expansions, new settings, changes and repairs are carried out by authorized persons.
- Electrical installation of the operating room corresponds to the national standards for medical institutions.
- The device is used in compliance with the operating instructions and with its associated accessories.
- The device was switched on with all connecting elements fully connected and the system accessories
were checked for operability before every use.
•Weacceptnoresponsibilityorliabilityfordamageofanykindduetomisuse,abuse,violationofoperatinginstructions,
omission of required maintenance, equipment modications and service by unauthorized persons! Use of other than
originalaccessoriesshallvoidwarrantyandliability!
•Equipotentialbondingpinistobeusedtogetanadditionalpotentialequalizationforthepatient.PleaseconsiderEMC
notes.
The device complies EMC standards DIN EN 60601-1-2 Ausgabe 2007. Nevertheless, EMC interference may lead to
malfunction. If malfunction is observed, ensure that the device has been installed and is operated in accordance with the
EMCguidelines.Notethatportableandmobilehighfrequencycommunicationequipmentmayinuencetheoperation
of this device. If the device is used in the vicinity of or stacked together with other devices, the device or system must
be monitored to ensure standard operation in accordance with these guidelines. If necessary, refer to manufacturer for
further details.
WARNING! The application of pump for small and closed body cavities (arthroscopy and hysteroscopy) is strictly forbidden,
because irrigation process may lead to uncontrolled and dangerous pressure rise.
CAUTION! To avoid risk of electric shock device must be connected to public mains with protective earth conductor only.
IMPORTANT! In case of failure switch off the device by its main power switch. Defective device or parts are to dispose to
the national standards. When operating lifetime has expired manufacturer will take back device for recycling.
description and description description
Follow the instructions for use Type BF applied part Year of production
Legal manufacturer
mdcmedicaldevicecerticationGmbH,
Kriegerstraße 6,
70191 Stuttgart, Germany
Equipotential
(equalising potential)
C a u t i o n!
Hazardous voltage Batch number Article number
Expiring date Non-sterile, reusable Protective earthing
Product contains phthalate Sterilized with ethylene oxide Footswitch connector
Do not re-use Do not use if package is
damaged
Spray protection (protection
against vertically falling drops)
Thisproductshouldbeusedonlybyqualiedmedicalprofessionals!
Notexplosionproof.Donotusethedevicewithindemarcatedhazardzones(explosiveGases)!
Beforeopeningthecoverdisconnectmainsplug!
0483

Ackermann
Page: 7 // 20
SHAVER SYSTEM
There are two separate bottle holders included to support bottle with irrigation liquid and secretion container. Please
installbothbottleholdersleftandrighttothebottomofpumpusingthefourxingscrewsprovided(seeillustration).
Insert containers into the holders and connect tubing set (see chapter 9.4.).
Illustration of bottle holders:
7. SPIKE
The spike is needed to connect tubing set to the bottle with irrigation liquid. Push spike completely into the bottle with
irrigation liquid. To ensure proper function venting hole must be placed inside the bottle but above liquid level. Tip of spike
must always be completely immersed in the liquid. Connect tubing set IRRIGATION (see chapter 9.4.). Place bottle with
insertedspikeonlyinuprightposition!Usebottleholderonpump.Onlyoriginalspikeensuresproperfunctionofpump.
CAUTION! Useonlyoriginalspike!Cleanandsterilizespikebeforeinitialapplicationandanyfurtheruse.

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 8 // 20
7.2. REPROCESSING THE SPIKE
Read all instructions carefully before reprocessing the spike. For pre-cleaning/cleaning we advise to use Neodisher FA
(manufactured by Dr. Weigert). Pre-cleaning prevents the transmission of germs and stops residues from drying on. It
therefore needs to be performed immediately after the operation. Rinse spike below the surface of the water. Use cleaning
gun to rinse connectors and cannula at 4 bar / 3 min with cold water. Use a plastic brush to remove heavily Encrusted
residues.
WARNING! Donotuseanyxatives.Donotusehotwaterforpre-cleaning.
CAUTION! Do not use abrasive brushes or sponges. Avoid contact with hydrogen peroxide (H2O2).
8. SECRETION CONTAINER
Theprovidedsecretioncontaineristobeusedforcollectingtheaspiredorganicuidandhasamaximumcapacityof2
litres.Itisequippedwithpressurelid,mechanicaloverowprotectionwithintegratedoatandavolumescale.
Close thelid by pressingitontothe secretioncontainer.Ensurethat integratedoatof overowprotectioncanmove
freely.
Connect tubing set SUCTION to the connectors „VACUUM“ and „PATIENT“ (see chapter 9.4.). Both connectors must not be
mixedup!Inversionofconnectionscouldleadtocontaminationoftheoperatorand/orthepump.
Placesecretioncontainerwithoverowprotectiononlyinuprightposition!Usebottleholderonpump.
Whenoverowvalveintervenes,switchoffpumpimmediately.
CAUTION! Clean and sterilize secretion container and its components before initial application and any further use.
Operatesecretioncontainerinuprightpositiononly!Donotoperatewithoutoverowprotection!
8.2. REPROCESSING OF SECRETION CONTAINER
Read all instructions carefully before reprocessing the secretion container. First of all disassemble secretion container
into its single components (see illustration).
Wash all surfaces using hot water (max. 60°C) and dry them carefully. All components are to sterilize by steam sterilisation
(saturated vapour) at 121°C / max.15min. Therefore place the container upside-down.
Carefully check components of secretion container for cracks and brittle parts after each cleaning cycle.
ATTENTION! Donotusesolvents,disinfectantsorabrasiveproductsforcleaning!Obeysterilizationcondition!Dispose
after30cleaningandsterilizationcycles!
IMPORTANT! Makesuretore-assemblethesealscorrectlyandensurethatintegratedoatofoverowprotectioncan
move freely after reprocessing.

Ackermann
Page: 9 // 20
SHAVER SYSTEM
The tubing set consists of 4 single tubes, which are color-coded (see chapter 9.3).
Tubing set IRRIGATION (blue):
Tube1hasgotonemaleLL-connectorandonefemaleLL-connector.Single-usesterilelterisxedtomaleLL-connector
(„to spike“).
Tube 2 does not contain any components. Tubing set SUCTION (red):
Tube3hasgotonemaleLL-connector.Single-usesterilelterisxedtomaleLL-connector(„topump“).
Tube 4 does not contain any components. There are blue and red color-coded connectors on front of the pump.
IMPORTANT! Sterileltersaresingle-useonly.Disposeafteruse,donotreprocess!
Read all instructions carefully before reprocessing the tubing set. The entire user manual set consists of this product
specicusermanual(deliveredtogetherwiththepump)andtheusermanualofthetubingsettobeusedforthepump.
Furthermore obey RKI guidelines and standards ISO17664 and ISO11607.
Beforereprocessingremoveanddisposesterilelters(seeFig.inchapter9.3).
Refer to user manual of tubing set for details of cleaning and drying.
Carefully check tubing set for visible irregularities after each cleaning cycle.
Tubing set is to sterilize by steam sterilisation (saturated vapour). Note: tubes 2 and 4 are to sterilize at least 20 min at a
temperature of max. 134°C. Tubes 1 and 3 are to sterilize at least 5 min at a temperature of max.134°C.
Disposetubingsetafter20cleaningandsterilizationcycles!
0483

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 10 // 20
LL-connector female
Tube 1 IRRIGATION (blue)
26 LL-connector male
27 Disposablesterilelter
28 Tube 2 IRRIGATION (blue)
29 Tube 3 SUCTION (red)
IMPORTANT! Insert tubing set only when device is switched-on and self test is passed successfully. Do not use tubing set
withoutdisposablesterilelters!Startsurgerywithanew,completelylledbottleofirrigationliquid,prepareaspare
bottle.
ATTENTION! Ifthebottleisempty,sterileairistransportedtopatientwithoutanywarning!Toremoveabottle,rstofall
disconnect tubes from pump. To prevent the instrument from sucking to tissue, the suction instrument / cannula should
have a bypass inlet.

Ackermann
Page: 11 // 20
SHAVER SYSTEM
Tubing set has to be sterile, see chapter 9.2.
Step 1 - Tube 1, blue, irrigation:
ConnectdisposablesterileltertofemaleLL-connectorof
spike. Connect male LLconnector of the tube to disposable
sterilelter.ConnectfemaleLL-connectorofthetubeto
connector IRRIGATION on pump.
Step 2- Tube 2, blue, irrigation:
Connect one side of the tube to tube connector on spike and
the other side of the tube to irrigation connector on instrument.
Step 3 - Tube 3, red, suction:
ConnectdisposablesterileltertoconnectorSUCTIONonpump.
Connect male LLconnector of the tube to disposable sterile
lter.Connecttheothersideofthetubetoconnector„VACUUM“
on secretion container.
Step 4 - Tube 4, red, suction:
Connect one side of the tube to connector „PATIENT“ on secretion
container and the other side of the tube to suction connector on
instrument.
First of all disconnect tubes from pump to release pressure
from the bottles. Separate all other connections then. Remove
anddisposesterilelters.Sterilelterisadisposableproduct
and not suitable For reprocessing. The manufacturer is released
from any product liability if products Intended for single use are
processed and re-sterilized contrary to these instructions.
Reprocessing of tubing set see chapter 9.2.

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 12 // 20
IMPORTANT! Remove tubing set after surgery. Sterilize tubing set, secretion container and instruments, always replace
sterilelters!
IMPORTANT! Do notuse pump without sterile lters!Irrigationbottle must be qualiedforthe overpressureof pump
system!
10. FOOTSWITCH
Optionally the pump can be operated by footswitch. Footswitch has the same functionallity as button START/STOP.
Choose a convenient position for the footswitch. Connect the adapter of control line to connector (13) on front panel of
pump. Ensure correct and tight connection. The control line must not be kinked or twisted.
Illustration of footswitch:
AControl line (tube)
BPedal - to start and stop the pump
CAdapter control line to pump

Ackermann
Page: 13 // 20
SHAVER SYSTEM
IMPORTANT! Footswitchshallbedisconnectedwhendeviceisswitchedoff!
ATTENTION! Footswitchmustnotbeconnectedduringselftest!Incaseofmalfunctionpleaseremoveadapter
(C) from connector (13) on front panel for a short time and connect it again.
11.1.1. POWER SWITCH
The power switch (14) is used to switch the device on and off.
The push button ON/OFF (5) activates / deactivates the device.
To activate device:
Press the button ON/OFF. The ON LED (6) lights up.
To deactivate device:
Press the button ON/OFF. The STANDBY LED (4) lights up.
(For safety reasons button ON/OFF is to press at least for 3 sec).
The push button START/STOP (2) puts the pump into operation
and stops it.
To start pump:
Press the button START/STOP The START LED (1) lights up. The
pump is put into operation and the preset pressure/vacuum are
generated within the bottles.
To stop pump:
Press the button START/STOP The STOP LED (3) lights up and the
pump stops.
Note: pressure / vacuum persists within the bottles until a bottle
connection will be opened.
The push buttons (9, 10) are to adjust preset values of pressure
for irrigation and vacuum for suction function.
Increase / decrease IRRIGATION pressure
Increase / decrease SUCTION vacuum
IMPORTANT! Place the device on a level surface in a dry environment.
Temperature +10 - +40°C
Relative Humidity 30-70%

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 14 // 20
WARNING! EXPLOSIONHAZARD!Theelectricalcomponentsarenotexplosionproof.Donotuseinanareawhereammable
gasesarepresent!
1. Connect the pump to mains. Switch on device with power switch. LED STANDBY (4) will be on. Do not connect tubing set
yet!Donotconnectfootswitchyet!
2. Activate the device by pressing button ON / OFF (5). A short beep sounds.
3. The device enters self test.
Self test:
a. All LEDs and displays are turned on for 1 sec.
b. IRRIGATION displays (9) show program version for 1 sec. Then segments of left IRRIGATION display are turned on as
follows:
c. 1. Segment: test of footswitch sensor
d. 1.-2. Segment: test of pressure sensor IRRIGATION
e. 1.-3. Segment: test of pressure sensor SUCTION
f . 1.-4. Segment: test of pressure generator
g. 1.-5. test of vacuum generator
When self test is passed correctly a triple beep sounds. LEDs ON and STOP will be on and displays show preset pressure and
preset vacuum at last pump start.
4. Connect the tubing set (see chapter 9.4.). The valves of irrigation-suction instrument must be closed.
5. Perform functional check.
The functional check must be carried out prior to each operation.
Irrigation:
1. Push button UP/DOWN IRRIGATION (8) to adjust preset pressure to maximum.
2. Start the pump.
3. Pressure increases to maximum within ca. 2 min. and the pressure generator is switched off. If preset pressure is not
obtained check tubes and connections to be tight.
4.Operateirrigationvalveofinstrumentandllthetubingsetandtheinstrumentcompletelywithliquiduntilliquid
emerges from the instrument.
5. The functional check of IRRIGATION is complete.
Suction:
1. Push button UP/DOWN SUCTION (11) to adjust preset vacuum to maximum.
2. Vacuum increases to maximum within ca. 2 min. and the vacuum generator is switched off. If preset vacuum is not
obtained check tubes and connections to be tight.
3. The functional check of SUCTION is complete. Press the button START/STOP to stop the pump. The STOP LED lights
up. Now the device is tested and ready to be used in the operating room.
1 . After performing the functional check the pump can be started.
2. The pump is activated.
3. Pressure and vacuum generators are operating until preset values are obtained within the bottles. Values can be
adjusted during operation.
4. Operate valves on instrument to activate IRRIGATION or SUCTION according to your need. Pressure and vacuum
generators are switched off automatically if preset pressure and vacuum are obtained within the bottles.

Ackermann
Page: 15 // 20
SHAVER SYSTEM
If pressure and vacuum decrease below preset values they are switched on again.
If only one of both (IRRIGATION or SUCTION). Is needed you can switch off refering generator: push buttons UP/DOWN to
reduce preset pressure/ vacuum to zero.
5. Stop process by pressing the button START/STOP or the footswitch.
6. Deactivate device by pressing the button ON/OFF for 3 sec.
7. Switch off the device with power switch.
IMPORTANT! Keep enough irrigation liquid, which is prewarmed to 40°C readily available. Use only intact, originally sealed
bottles with irrigation Liquid.
Operatedbottlesarepressurized!Dangerofbursting!
ATTENTION!Iftheirrigationbottleisempty,sterileairistransportedtopatientwithoutanywarning!
IMPORTANT! Ensuretoexchangesecretioncontainerintime!Integratedoverowvalveshallprotectdevicefrombeing
contaminated by secretion Fluid. If in spite of all security precautions taken contamination occurs, switch off device
immediatelyanddisconnectitfrommains.Refertoqualiedtechnicalservice.
•IncaseofcontinuousbeepandblinkingLEDsduringselftestadeviceerrorwasdetected.
* Press button ON/OFF to deactivate device and activate again after 15s. Ensure that tubing set and footswitch are
disconnectedfromdevice!Shouldtheerrormessagestilloccurrefertoqualiedtechnicalservice.
•IncaseofcontinuousbeepandblinkingLEDsduringoperationadeviceerrorwasdetected.
* Disconnect tubing set and footswitch from device. Press button ON/OFF to deactivate device and activate again after
15s.Shouldtheerrormessagestilloccurrefertoqualiedtechnicalservice.
To preserve the device and ensure its proper function please note as follows:
Cleaning:
Switch off the device with power switch. Disconnect device from mains. Remove power cable from mains connection at
the rear of device.
Usemilddisinfectingsolutiontocleantheexteriorofdeviceandfootswitch.Foruidconcentrationfollowinstructions
fromdisinfectantmanufacturer.Strictlypreventtheingressofliquidintothedeviceandfootswitch!
ATTENTION! Thedeviceandthefootswitchmustnotbesterilized!
The unit must be checked by an authorized service partner once a year.
Repair, calibration and further services are to be carried out by authorized service only. Unauthorized services release
manufacturer from any liability for the operating safety of the pump.

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 16 // 20
Theservicepartnershallissueacerticateontypeandextendofannualinspectionandrepair.Inspectionshallbemarked
with a label „next check“ on the rear panel of device.
Opening of casing, service and any other modications of device and accessories by unauthorized persons releases
manufacturer from any liability for the operating safety of the pump.
Dependingonrespectivecountry-specicregulations,afunctionandsafetyinspectionmustbeperformedbytheuser.
Manufacturer recommends quarterly inspections, see chapter 12.
Each device is carefully tested with regard to function and safety by the manufacturer. Results are recorded in the test
report which is attached to the device.
Information regarding schematics, component lists, descriptions and/ or calibration instructions can be provided by the
manufacturer on demand.
For this test you will need:
•Membranesuction-irrigationpump
•originaldisposablesterilelters(2x)
•Manometerwith1mtubeandLL-connectorfemaleforpressuremeasurement(+)
•Manometerwith1mtubeandLL-connectormalevacuummeasurement(-)
This test will be carried out without liquid.
Proceed as follows:
1. Connect the pump to main switch on device with power switch.
2. Activate the device by pressing button ON/OFF.
3. The device enters self test.
4. Connect manometer for pressure measurement to male LL-connector IRRIGATION and manometer for vacuum
measurement to female LL-connector SUCTION.
A) Checking irrigation pressure:
5. Push button UP/DOWN IRRIGATION (8) to adjust preset pressure to maximum.
6. Start the pump.
7. Pressure increases to maximum and the pressure generator is switched off.
8. The pressure manometer shall display 750 mmHg +/- 10%.
9. The check of IRRIGATION pressure is complete. Press the button START/STOP to stop the pump
B) Checking suction vaccum:
10. Push button UP/DOWN SUCTION (11) to adjust preset vacuum to maximum.
11. Start the pump.
12. Vacuum increases to maximum and the vacuum generator is switched off.
13. The vacuum manometer shall display -525 mmHg+/-10%.
14. The check of SUCTION vacuum is complete. Press the button START/STOP to stop the pump.
Incaseofdeviationscheckconnectionsandrepeatthetest.Ifdeviationisobservedagainrefertoqualiedtechnical
service.

Ackermann
Page: 17 // 20
SHAVER SYSTEM
Setup of irrigation pressure and suction vaccum test:
For this test you will need:
•Membranesuction-irrigationpump
•Footswitch
This test will be carried out without liquid.
Proceed as follows:
1. Connect pump to mains. Switch on device with power switch.
2. Activate the device by pressing button ON/OFF.
3. The device enters self test.
4. Connect footswitch to connector on frontpanel of the pump. Ensure correct and tight connection. The control line must
not be kinked or twisted.
5. Press footswitch to start the pump. LED START lights up and pressure and vacuum generators are operating.
6. Press footswitch to stop the pump. LED STOP lights up and pressure and vacuum generators are switched off.
Incaseofdeviationschecksetupandrepeatthetest.Ifdeviationisobservedagainrefertoqualiedtechnicalservice
IMPORTANT! Footswitchshallbedisconnectedwhendeviceisswitchedoff!
Setup of footswitch test:

MEMBRANE SUCTION-IRRIGATION PUMP
Ackermann
Page: 18 // 20
16. ACCESSORIES
Item # Illustration
16-2083-101 Selectioncontainerwithoverowvalve,autoclavable
16-2083-102 Silicone irrigation tubing set, autoclavable
16-2083-103 Silicone suction tubing set, autoclavable
16-2083-104 Mains cable Euro 2 m
16-2083-105 Fuses
16-2083-106 Bottle holder (1 litre) + 2 screws
16-2083-107 Bottle holder (2 litres) + 2 screws
16-2083-108 Hydrophobicsterileltersterilepackage,disposable
16-2083-109 Spike, autoclavable
16-2083-110 Footswitch
All product codes covered by these instructions are listed below:
•Membrane suction-irrigation pump #16-2083
Ackermann Instrumente GmbH
Eisenbahnstrasse 65-67
78604 Rietheim-Weilheim
Germany
Phone: +49 (0)7461 966 17 - 0
Fax: +49 (0)7461 966 17 - 70
E-Mail: sales@ackermannsurgical.com
Web: www.ackermannsurgical.com
End of document

Ackermann
Page: 19 // 20
SHAVER SYSTEM

Ackermann Instrumente GmbH
Eisenbahnstrasse 65-67
78604 Rietheim-Weilheim
Germany
Copyright by Ackermann Instrumente GmbH. Specifications, design and accessories in this user manual are subject to change without any
notice or obligation on the part of the manufacturer.
Ackermann
0483
Table of contents
Other ACKERMANN Water Pump manuals
Popular Water Pump manuals by other brands

Pentair
Pentair BOOST-RITE Installation and user guide

Pentair
Pentair Sta-Rite WATERPRESS INOX Series instruction manual

Warren rupp
Warren rupp sandpiper ii S15 Service & operating manual

Dupla MARIN
Dupla MARIN OEP 800 Operating instruction

Giant
Giant P540 Operating instructions/ repair and service manual

Graco
Graco Bulldog 221-104 Instructions-parts list

Surtek
Surtek BC603 User manual and warranty

Pfeiffer Vacuum
Pfeiffer Vacuum OKTA 250 operating instructions

Xylem
Xylem Lowara MP Installation, operation and maintenance instructions

EBARA
EBARA thebe 3TSM ci Instruction manual and warranty

SKF
SKF LINCOLN P203 Original assembly instructions

VONROC
VONROC SP509AC Original instructions