Agilent Technologies 54657A User manual

User’s Guide
Publication Number 54657-97019
August 2000
For Safety information Warranties and Regulatory
information see the pages behind the index
© Copyright Agilent Technologies 1991-1996 2000
All Rights Reserved
Agilent 54657A, 54658A,
and 54659B
Measurement/Storage Modules

Measurement/Storage Modules
The Agilent 54657A 54658A and 54659B Measurement/Storage
Modules provide additional measurement and storage capabilities to
the Agilent 54600–Series oscilloscopes. The 54657A has a GPIB
interface and the 54658A has a RS-232 interface. The 54659B has a
RS-232 interface plus an additional parallel output connector which
allows the module to be connected to both an RS-232 controller and a
parallel printer at the same time. The main features are:
•Full Programmability.
•Hardcopy output.
•Three additional automatic voltage measurements (amplitude
preshoot and overshoot).
•Two additional automatic time measurements (delay and phase
angle). User defined measurement thresholds of 10%/90%
20%/80% or selected voltage levels.
•Two additional cursor measurements (voltage in percent and time
in degrees).
•Two additional cursor measurement sources (math function 1
and 2).
•Waveform math functions (addition subtraction multiplication
differentiation integration and FFT)
•Time and date tagging of hard copy and and nonvolatile memories.
Three uncompressed nonvolatile trace memories.
•Additional 64K of nonvolatile trace memory (with data
compression) for up to 97 more trace memories..
•Unattended waveform monitoring by use of mask templates.
•Built-in automatic mask generation and mask editing capabilities.
ii

Accessories available
•34810B BenchLink/Scope software package.
•10833A 1 meter (3.3 feet) GPIB cable.
•10833B 2 meter (6.6 feet) GPIB cable.
•10833C 4 meter (13.2 feet) GPIB cable.
•10833D 0.5 meter (1.6 feet) GPIB cable.
•13242G 5 meter (16.7 feet) RS-232 cable for printer/plotter and HP
Vectra 25-pin serial port.
•17255M 1.2 meter (3.9 feet) RS-232 cable for printer/plotter and
HP Vectra 25-pin serial port.
•17255D 1.2 meter (3.9 feet) RS-232 cable for IBM PC/XT 25-pin
serial port.
•92219J 5 meter (16.7 feet) RS-232 cable for IBM PC/XT 25-pin
serial port.
•24542G 3 meter (9.9 feet) RS-232 cable for 9-pin serial port.
•34398A 2.5 meter (8.2 feet) RS-232 cable.
•34399A RS-232 Adapter Kit.
iii

In T is Book
This book is the user’s guide for the Agilent 54657A 54658A and 54659B
Measurement/Storage Modules and contains three chapters.
Installation Chapter 1 contains information concerning installation and
interconnection of the Measurement/ Storage Modules.
Operating the Measurement/Storage Module Chapter 2 contains a
series of exercises that guide you through the operation of the
Measurement/Storage Modules.
Reference Information Chapter 3 lists the reference information
concerning the Measurement/Storage Modules.
iv

Installation
1
Reference Information
3
Operating the Measurement/
Storage Module
2
Index
v

vi

Contents
1 Installation
Oscilloscope Compatibility 1–2
To install the Measurement/Storage Module 1–3
To configure the interface 1–4
2 Operating the Measurement/Storage Module
Math unctions 2–3
unction 1 2–4
unction 2 2–5
T Measurement 2–8
Automatic Measurements 2–14
Setting Thresholds 2–15
To make delay measurements automatically 2–19
To make phase measurements automatically 2–21
To make additional voltage measurements automatically 2–23
To make additional cursor measurements 2–25
Unattended Waveform Monitoring 2–29
To create a mask template using Automask 2–30
To create a mask template using Autostore 2–31
To create or edit a mask using line segments 2–33
To edit an individual pixel of a mask 2–35
To edit the mask to test only a portion of a waveform 2–36
To start waveform monitoring 2–38
To automatically save test violations 2–40
Creating a delay testing mask 2–42
Creating a frequency testing mask 2–44
Creating an overshoot testing mask 2–46
Creating a rise time testing mask 2–48
Testing the eye opening of an eye-pattern signal 2–50
To save or recall traces 2–52
To create a label for a trace memory 2–54
To set real-time clock 2–55
Contents–1

3 Reference Information
Operating Characteristics 3–3
Index
Contents
Contents–2

1
Installation

Installation
This chapter provides you with the information necessary to install
the Measurement/Storage Module on the oscilloscope. Information
required to connect and configure the module to the desired external
devices (such as printer, plotter, computer prior to local or remote
operation is given in the Interface Modules for Agilent 54600-Series
Instruments I/O Function Guide shipped with your module.
Oscilloscope Compatibility
54657A and 54658A These modules are compatible with all Agilent
54600-series oscilloscopes except 54600A, 54601A, and 54602A
oscilloscopes with an operating system version lower than 2.2. If your
54600A, 54601A, or 54602A oscilloscope has an earlier operating system,
it can be updated using upgrade kit HP part number 54601-68702. The
version is briefly displayed on screen when the Print/Utility key
is pressed.
54659B The 54659B in NOT compatible with the 54600A, 54601A,
54602A, or 54610A Oscilloscope. This module is compatible with all
other Agilent 54600-series oscilloscopes with an operating system
version 1.2 or higher or with an operating system version number in the
form of A.XX.XX. If your 54600B, 54601B, 54602B, or 54603B
Oscilloscope has an earlier operating system, it can be updated using
upgrade kit Agilent part number 54601-68704. If your 54610B
Oscilloscope has an earlier operating system, it can be updated using
upgrade kit Agilent part number 54610-68704. The version number is
briefly displayed on screen when the Print/Utility key is pressed.
1–2

To install the Measurement/Storage Module
1Turn off the oscilloscope.
2Install the module as shown below.
The oscilloscope is reset after installation. The installed module is reflected
in the message displayed when you turn on the oscilloscope.
Installing the Measurement/Storage Module
Figure 1–1
Installation
To install the Measurement/Storage Module
1–3

To configure the interface
The Measurement/Storage Module can be connected to a printer, a plotter, or
a computer through the interface. The 54657A has an GPIB interface and
the 54658A has an RS-232 interface. The 54659B has an RS-232 interface
plus an additional parallel output connector which allows the module to be
connected to both an RS-232 controller and a parallel printer at the same
time.
Connect the Measurement/Storage Module to a printer, plotter, or computer
through a suitable cable. The following table shows the Agilent part numbers
of the proper cables.
Interface Cables
Agilent Interface
module Cable function
(Instrument to ..) Module
connector Printer/plotter/
controller
connector
Agilent part number Cable
Length
54657A (GPIB) Printer/plotter/
controller HP-IB HP-IB 10833A
10833B
10833C
10833D
1 m (3.3 ft)
2 m (6.6 ft)
4 m (13.2 ft)
0.5 m (1.6 ft)
54658A (RS-232) Printer/plotter/
controller 25-pin F 25-pin F 13242G
17255M 5 m (16.7 ft)
1.5 m (4.9 ft)
Controller 25-pin F 25-pin M 92219J
17255D 5 m (16.7 ft
1.5 m (4.9 ft)
Controller 25-pin F 9-pin M 24542G 3 m (9.9 ft)
54659B1
(RS-232 and
parallel output)
RS-232 controller 9-pin M 25-pin M 34398A 2.5 m (8.2 ft)
RS-232 controller 9-pin M 9-pin M 34398A 2.5 m (8.2 ft)
RS-232 printer/plotter/
controller 9-pin M 25-pin F 34398A + 34399A
adapter kit 2.5 m (8.2 ft)
Parallel printer parallel parallel C2950A
C2951A 2 m (6.6 ft)
3 m (9.9 ft)
1 The 54659B is not compatible with the 54600A, 54601A, 54602A, and 54610A.
Installation
To configure the interface
1–4

54658A Serial Connections
The signals for the RS-232 port on the 54658A are listed below.
Pin Number Signal
2 Transmit Data
3 Receive Data
4 Request to Send
5 Clear to Send
6 Data Set Ready
7 Signal Ground
8 Data Carrier Detect
20 Data Terminal Ready
SHELL Protective Ground
Pin out of 54658A RS-232 port looking into DB25 female connector
The following figures show the pin outs of the suggested RS-232 interface
cables used with the 54658A 25-pin connector.
Figure 1– 2
Installation
To configure the interface
1–5

Pin out of 13242G/17255M RS-232 cable
Pin out of 24542G RS-232 cable
Pin out of 92219J/ 17255D RS-232 cable
54645b09.cdr
13242G/17255M
Cable
Printer/plotter/
controller
54658A
Module
25-pin
female
25-pin
male
25-pin
female
25-pin
male
1
2
3
4
5
6
7
8
12
11
19
20
1
3
2
8
20
7
4
19
11
12
5
6
Figure 1– 3
54657b10.cdr
24542G
Cable
Controller
54658A
Module
25-pin
female
9-pin
female
9-pin
male
25-pin
male
1
2
3
4
5
6
8
7
4
2
3
5
6
7
20
8
Figure 1– 4
54645b08.cdr
25-pin
female
25-pin
female
25-pin
male
25-pin
male
92219J/17255D
Cable
Controller
54658A
Module
1
2
3
5
6
7
20
1
3
2
20
7
5
6
Figure 1– 5
Installation
To configure the interface
1–6

54659B Serial Connections
The signals for the 9-pin RS-232 port on the 54659B are listed below.
Pin Number Signal
1 Data Carrier Detect
2 Receive Data
3 Transmit Data
4 Data Terminal Ready
5 Signal Ground
6 Data Set Ready
7 Request to Send
8 Clear to Send
9Ring
SHELL Protective Ground
Pin out of 54659B RS-232 port looking into DB9 male connector
The following figure shows the pin out of the suggested RS-232 interface
cable used with the 54659B 9-pin connector.
Pin out of 34398A RS-232 cable
Refer to the "Programming over RS-232-C" chapter in the Agilent
54600–Series Oscilloscopes rogrammer’s Guide for additional
information.
Figure 1– 6
Printer/Plotter/
Controller
54659B
Module
34398A
Cable
1
2
3
4
5
6
7
8
9
9-pin
female
9-pin
male
54657b07.cdr
9-pin
male
1
2
3
4
5
6
7
8
9
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
9-pin
female
Figure 1– 7
Installation
To configure the interface
1–7

1–8

2
Operating the
Measurement/Storage Module

Operating the Measurement/Storage
Module
This chapter provides you with the information necessary to use the
additional, or enhanced features that the Measurement/Storage
Module provides. Basic operation for the oscilloscope is covered in
the User and Service Guide for your oscilloscope.
This chapter provides you with practical exercises and detailed
information designed to guide you through operation of the following
functions:
•Math Functions
•utomatic Measurements
•Cursor Measurements
•Mask generation and waveform monitoring
•Trace Storage
2–2

Math Functions
Without the Measurement/Storage module installed, addition and subtraction
are the only math operations provided. In addition to the limited selections,
the single function is performed on the pixel position of the data on the
screen.
With the Measurement/Storage module installed, two functions define up to
six operations that create mathematically altered waveforms (not pixel math.)
•Function 1 will add (+), subtract (–), or multiply (*) the signals acquired
on vertical inputs 1 and 2, then it will display the result as F1.
•Function 2 will integrate, differentiate, or perform an FFT on the signal
acquired on input 1, input 2, or the result in F1; then it will display the
result in F2.
The vertical range and offset of each function can be adjusted for ease of
viewing and measurement considerations. Each function can be displayed,
measured (with cursors), stored in trace memory, or output over the
interface.
Operating the Measurement/Storage Module
Math Functions
2–3

Function 1
1Press ± .
2Toggle the Function 1 On Off softkey to enable math function number 1.
3Press the Function 1 Menu softkey
softkey menu with four softkey choices appears. Three of them are related
to the math functions.
4Toggle the + – * softkey until the desired operation is selected.
esults (F1) are displayed on the screen.
ll operations are calculated on a point-by-point basis.
•plus (+) algebraically sum input 1 and input 2 (input 1 + input 2).
•minus (–) algebraically subtract input 2 from input 1 (input 1 – input 2).
•multiply (*) algebraically multiply input 1 with input 2 (input 1 * input 2).
5Press the Units/div softkey and rotate the knob closest to the
Cursors key to set the vertical sensitivity of the resulting
waveform.
6Press the Offset softkey and rotate the knob closest to the
Cursors key to set the offset (from the center graticule) of the
resulting waveform.
Function waveform (F1) is available for viewing, measurement, or storage.
7Press the Previous Menu softkey.
Function 1 Operating Hints
If channel 1 or 2 are clipped (not fully displayed on screen,) the resulting
displayed function will also be clipped. Once the function is displayed, channel
1 and 2 may be turned off for better viewing.
When multiply is the operation selected, the value displayed for units per
division and offset is (V2).
Offset is the value (in V or V2) assigned to the center graticule for function 1.
Normal screen position is 0 V offset, or at the center graticule (until changed).
See "Making Cursor Measurements", and "Saving and Recalling Traces" in this
chapter for more information.
Operating the Measurement/Storage Module
Function 1
2–4
This manual suits for next models
2
Table of contents
Other Agilent Technologies Storage manuals
Popular Storage manuals by other brands

Seagate
Seagate Cheetah X15 installation guide

Seagate
Seagate SkyHawk AI ST10000VE001 product manual

SanDisk
SanDisk PROFESSIONAL PRO-BLADE ECOSYSTEM user manual

Apacer Technology
Apacer Technology USB 2.0 Flash Drive user manual

Transcend
Transcend StoreJetTM 25P user manual

SimpleTech
SimpleTech SP-U35/250 Specifications