Agilent Technologies IMG-300 User manual

DRAFT
vacuum technologies
IMG-300 Inverted
Magnetron Gauge
Manual No. 699908230
Revision A
March 2009
INSTRUCTION MANUAL
(505)872-0037
www.idealvac.com

DRAFT
Copyright 2009
Varian, Inc.
IMG-300 Inverted Magnetron Gauge

IMG-300 Inverted Magnetron Gauge
DRAFT
iii
Warranty
Products manufactured by Seller are warranted against defects in materials and workmanship for twelve
(12) months from date of shipment thereof to Customer, and Seller’s liability under valid warranty
claims is limited, at the option of Seller, to repair, replace, or refund an equitable portion of the
purchase price of the Product. Items expendable in normal use are not covered by this warranty. All
warranty replacement or repair of parts shall be limited to equipment malfunctions which, in the sole
opinion of Seller, are due or traceable to defects in original materials or workmanship. All obligations of
Seller under this warranty shall cease in the event of abuse, accident, alteration, misuse, or neglect of
the equipment. In-warranty repaired or replaced parts are warranted only for the remaining unexpired
portion of the original warranty period applicable to the repaired or replaced parts. After expiration of
the applicable warranty period, Customer shall be charged at the then current prices for parts, labor,
and transportation.
Reasonable care must be used to avoid hazards. Seller expressly disclaims responsibility for loss or
damage caused by use of its Products other than in accordance with proper operating procedures.
Except as stated herein, Seller makes no warranty, express or implied (either in fact or by operation of
law), statutory or otherwise; and, except as stated herein, Seller shall have no liability under any
warranty, express or implied (either in fact or by operation of law), statutory or otherwise. Statements
made by any person, including representatives of Seller, which are inconsistent or in conflict with the
terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an
officer of Seller.
Warranty Replacement and Adjustment
All claims under warranty must be made promptly after occurrence of circumstances giving rise thereto,
and must be received within the applicable warranty period by Seller or its authorized representative.
Such claims should include the Product serial number, the date of shipment, and a full description of
the circumstances giving rise to the claim. Before any Products are returned for repair and/or
adjustment, written authorization from Seller or its authorized representative for the return and
instructions as to how and where these Products should be returned must be obtained. Any Product
returned to Seller for examination shall be prepaid via the means of transportation indicated as
acceptable by Seller. Seller reserves the right to reject any warranty claim not promptly reported and
any warranty claim on any item that has been altered or has been returned by non-acceptable means of
transportation. When any Product is returned for examination and inspection, or for any other reason,
Customer shall be responsible for all damage resulting from improper packing or handling, and for loss
in transit, notwithstanding any defect or non-conformity in the Product. In all cases, Seller has the sole
responsibility for determining the cause and nature of failure, and Seller’s determination with regard
thereto shall be final.
If it is found that Seller’s Product has been returned without cause and is still serviceable, Customer will
be notified and the Product returned at Customer’s expense; in addition, a charge for testing and
examination may be made on Products so returned.
3/1/00

IMG-300 Inverted Magnetron Gauge
This page intentionally left blank.
DRAFT

IMG-300 Inverted Magnetron Gauge
DRAFT
v
Contents
Preface ........................................................................................................................................... viii
Hazard and Safety Information.................................................................................................... viii
IMG-300 General Purpose Inverted Magnetron Gauge .............................................................. 1-1
Description.................................................................................................................................1-1
Application.................................................................................................................................1-1
Technical Specifications .............................................................................................................1-3
Ordering Information..................................................................................................................1-3
Fringe Field Contours..................................................................................................................1-4

IMG-300 Inverted Magnetron Gauge
This page intentionally left blank.
DRAFT

Varian, Inc.
declare under our sole responsibility that the product,
erklären, in alleniniger Verantwortung, daß dieses Produkt,
déclarons sous notre seule responsabilité que le produit,
declaramos, bajo nuestra sola responsabilidad, que el producto,
verklaren onder onze verantwoordelijkheid, dat het product,
dichiariamo sotto nostra unica responsabilità, che il prodotto,
Declaration of Conformity
Konformitätserklärung
Déclaration de Conformité
Declaración de Conformidad
Verklaring de Overeenstemming
Dichiarazione di Conformità
to which this declaration relates is in conformity with the following standard(s) or other normative documents.
auf das sich diese Erklärung bezieht, mit der/den flogenden Norm(en) oder Richtlinie(n) übereinstimmt.
auquel se réfère cette déclaration est conforme à la (auz) norme(s) ou au(x) document(s) normatif(s).
al que se refiere esta declaración es conforme a la(s) norma(s) u otro(s) documento(s) normativo(s).
waamaar deze verklaring verwijst, aan de volende norm(en) of richtlijn(en) beantwoodt.
a cui se rifersce questa dichiarazione è conforme alla/e sequente/I norma/o documento/I normativo/i.
Frederick C. Campbell
Operations Manager
Vacuum Technologies
We
Wir
Nous
Nosotros
Wij
Noi Lexington, MA, 02421-3133 USA
121 Hartwell Avenue
Vacuum Technologies
Lexington, Massachusetts, USA
Varian, Inc.
IMG-300 Inverted Magnetron Gauge
EN 55011
1991 . . . . . . . . . . . . . . . . . . . . . . . . Group 1 Class A ISM emission requirements
EN 61010-1
1993 . . . . . . . . . . . . . . . . . . . . . . . . Safety requirements for electrical equipment for measurement, control, and
laboratory use incorporating Amendments Nos 1 and 2.
EN 50082-2
1995 . . . . . . . . . . . . . . . . . . . . . . . . EMC heavy industrial generic immunity standard
October 2003

IMG-300 Inverted Magnetron Gauge
viii
Preface
Hazard and Safety Information
This product must only be operated and maintained by trained personnel.
This manual uses the following standard safety protocols:
WARNING Warnings indicate a particular procedure or practice, which
if not followed correctly, could lead to serious injury.
CAUTION Cautions indicate a particular procedure or practice, which if
not followed, could cause damage to the equipment.
NOTE Notes contain important information.
Before operating or servicing equipment, read and thoroughly understand all operation/
maintenance manuals provided by Varian. Be aware of the hazards associated with this
equipment, know how to recognize potentially hazardous conditions, and how to avoid
them. Read carefully and strictly observe all cautions and warnings. The consequences of
unskilled, improper, or careless operation of the equipment can be serious.
In addition, consult local, state, and national agencies regarding specific requirements and
regulations. Address any safety, operation, and/or maintenance questions to your nearest
Varian office.

IMG-300 Inverted Magnetron Gauge
ix
Grounding the IMG-300 Inverted Magnetron Gauge Controller
Be certain that the IMG-300 Inverted Magnetron Gauge Controller and vacuum system are
separately connected to a common ground.
WARNING ❑Do not place a ground wire between the vacuum chamber
and the controller chassis; large continuous currents could
flow through it.
❑Personnel can be killed by high voltages (3 kV may be
present in an improperly grounded system).
❑Make absolutely sure that the vacuum system is grounded as
shown in Figure 1.
❑Test the system ground to be sure that it is complete and
capable of supporting at least 10 A.
Figure 1 Ion Gauge and Vacuum System Connections
Use with Combustibles and Mixtures
WARNING As with all ionization gauges, this device is not intrinsically
safe. Exercise extreme care when using this vacuum gauge
while pumping or backfilling a system or in any other system
condition which contains combustible gases or mixtures. The
filament, the grid of a hot filament ion gauge and the high
voltage discharge of a cold cathode gauge can be ignition
sources.
When such a gas or mixture is present, never turn on any such
vacuum gauge and disconnect it from the controller. Failure to
follow this instruction could result in serious injury to personnel
and damage to equipment.
Gauge Controller Vacuum System
Power Distribution
Breaker Box
Power Common or
Safety Ground Lead
Do not make direct chassis to
vacuum system Connection!
Note:
are not shown.
Other power leads

IMG-300 Inverted Magnetron Gauge
x
Vacuum Equipment and Cleanliness
Cleanliness is vital when servicing any vacuum equipment.
CAUTION Do not use silicone oil or silicone grease.
Use powder-free butyl or polycarbonate gloves to prevent skin
oils from getting on vacuum surfaces.
Do not clean any aluminum parts with Alconox. AIconox is not
compatible with aluminum and will cause damage.
NOTE Normally, it is unnecessary to use vacuum grease. However, if it
must be used, do not use silicone types, and use it sparingly.
Apiezon L grease is recommended (Varian Part Number
695400004).

IMG-300 Inverted Magnetron Gauge
xi
EMC Warnings
EN 55022 Class A Warning
This is a Class A product. In a domestic environment this product may cause radio
interference. In such cases, the user will be required to correct the interference at his
own expense.
FCC
This device complies with Part 15 of the FCC rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesirable
operation.
NOTE The equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to Part 15 of the
FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment
is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential
area is also likely to cause harmful radio communications
interference in which case, the user will be required to correct
the interference at his own expense.
Contacting Varian
In the United States, you can contact Varian Customer Service at 1-800-8VARIAN.
Internet users:
❑Send email to Customer Service & Technical Support at
vpl.customer[email protected]
❑Visit our web site at www.varianinc.com/vacuum
❑Order on line at www.evarian.com
See the back cover of this manual for a listing of our sales and service offices.

IMG-300 Inverted Magnetron Gauge
This page intentionally left blank.

IMG-300 Inverted Magnetron Gauge
DRAFT
1-1
IMG-300 General Purpose Inverted Magnetron
Gauge
Description
The IMG-300 is a 1.6" diameter inverted magnetron cold cathode ionization gauge that
measures vacuum pressure from 10-3 Torr (10-3 mbar) down to 10-11 Torr (10-11 mbar).
Advantages include:
❑An inverted magnetron design that ensures fast, reliable starting and good linearity
throughout the pressure range as compared to traditional cold cathode/magnetron
type gauges.
❑Minimized outgassing at high vacuum due to temperature increases because there is
no hot filament.
❑Greater ruggedness with the IMG-300 than Bayard-Alpert type gauges because there
is no delicate grid assembly and filament to break or burnout due to high-pressure
exposure.
❑An all-welded stainless steel shell that is virtually unbreakable compared to that of
typical glass Bayard-Alpert gauges.
❑A lack of excess electrons, photons, and heat which can be important is some
applications.
❑Extremely rapid response.
A standard SHV connector and coaxial cable provides for a safe, easy connection to a
Multi-Gauge. Attachment to the vacuum system is made using a ConFlat flange. All
materials exposed to the vacuum system are high vacuum compatible. No elastomers are
used in IMG-300 construction.
Application
Use the IMG-300 wherever an extremely rugged, fast response UHV vacuum gauge is
desired or where the presence of stray electrons and photons can cause problems in the
vacuum system. While the inverted magnetron design ensures good linearity as compared
to a standard cold cathode type gauge, the IMG-300, like other cold cathode type gauges, is
generally not as stable and linear as a well-maintained Bayard-Alpert type.
The IMG-300 with CFF handles bakeout temperatures of up to 250 °C, with the cable
unplugged.

IMG-300 Inverted Magnetron Gauge
1-2
DRAFT
Due to the IMG-300 start time, no specific starting accessory is required under normal
conditions. However, very cold temperatures and contamination build-up can result in
extended starting times, especially at high vacuum.
As with all cold cathode devices, the magnet generates a field external to the device.
When installing the IMG-300, follow these guidelines (Figure 1-1):
❑Do not mount the IMG-300 closer than a few inches to any magnetic material.
❑If no magnetic force from the IMG-300 can be felt, it is sufficiently far away.
❑Exercise care in mounting and handling to prevent the gauge from contacting other
objects, especially steel hand tools as these are strongly attracted by the magnet.
❑Avoid rough handling or prolonged contact with the metal surface which can cause a
decrease in magnet strength and result in a calibration change.
Figure 1-1 Outline Drawing, IMG-300 Inverted Magnetron Gauge
The calibration of the IMG-300 may be maintained under the STARRS service
program. STARRS allows simple, periodic service and calibration for all of your
Vacuum Technologies gauging equipment. Contact Varian Customer Service at
1-800-8VARIAN for more information on the STARRS program.

IMG-300 Inverted Magnetron Gauge
DRAFT
1-3
Technical Specifications
Table 1-1 lists the gauge technical specifications.
Ordering Information
Table 1-2 lists the part numbers for the IMG-300.
Table 1-1 IMG-300 Technical Specifications
Specification Definition
Operating voltage (HV range) 3 kV DC
Sensitivity 2.5 A/Torr + 100%; -5% @ 5x 10−6 Torr
Measurement Range 10-3 to 10-11 Torr
Maximum Operating Temperature 250 °C, CFF metal gasket; with magnet
Electrical Connector SHV style, 5kV rated
Feedthrough/UHV Ceramic brazed
Ignition Time < 5 seconds at pressures > 1E-6 Torr, typically
20 min @ 1E-10 Torr
Response Time < 10 milliseconds for increasing pressure
Material Exposed to the Vacuum 304L Stainless steel, Nickel, 2% Thoriated
Tungsten, and ceramic
Table 1-2 IMG-300 Parts
Part Number Description
R0343301 IMG-300 2.75 CFF
R0040301 Maintenance kit

IMG-300 Inverted Magnetron Gauge
1-4
DRAFT
Fringe Field Contours
Figure 1-2 and Figure 1-3 represent fringe field contours.
Figure 1-2 Fringe Field Contour

IMG-300 Inverted Magnetron Gauge
DRAFT
1-5
Figure 1-3 Fringe Field Contour

IMG-300 Inverted Magnetron Gauge
This page intentionally left blank.
DRAFT

IMPORTANT
Instructions for returning products
Dear Customer,
Please follow these instructions whenever one of our products needs to be returned:
1. Complete the Request for Return form on our website (www.varianinc.com) and send it to
Varian (see below), taking particular care to identify all products that have pumped or been
exposed to any toxic or hazardous materials.
2. After evaluating the information, Varian will provide you with a Return Authorization (RA)
number via email or fax, as requested.
Note: Depending on the type of return, a Purchase Order may be required at the time the
Request for Return is submitted. We will quote any necessary services (evaluation, repair,
special cleaning, e.g.).
3. Important steps for the shipment of returning product:
❑Remove all accessories from the core product (e.g. inlet screens, vent valves).
❑Prior to shipment, drain any oils or other liquids, purge or flush all gasses, and wipe off
any excess residue.
❑If ordering an Advance Exchange product, use the packaging from the Advance
Exchange to return the defective product.
❑Seal the product in a plastic bag, and package the product carefully to avoid damage in
transit. You are responsible for loss or damage in transit.
❑Varian, Inc. is not responsible for returning customer provided packaging or containers.
❑Clearly label package with the RA number. Using the shipping label provided ensures
the proper address and RA number are on the package. Packages shipped to Varian
without an RA clearly written on the outside cannot be accepted and will be returned.
4. Return only products for which the RA was issued.
5. Product being returned under an RA must be received within 15 business days.
6. Ship to the location specified on the printable label, which will be sent, along with the RA
number, as soon as we have received all of the required information. Customer is
responsible for freight charges on returning product.
7. Return shipments must comply with all applicable Shipping Regulations (IATA, DOT, etc.)
and carrier requirements.
Return the completed request for return form to your nearest location:
North America: Europe: Pacific Rim:
FAX: 1-781-860-9252 FAX: 00 39-011-9979125
Please visit our website for
individual office information.
Toll Free: 800-8VARIAN Fax Free: 00 800 345 345 00
(800-882-7426 Toll Free: 00 800 234 234 00
vacuum

IMG-300 Inverted Magnetron Gauge
This page intentionally left blank.
DRAFT 4/21/09
Table of contents
Popular Transducer manuals by other brands

W.E. Anderson
W.E. Anderson 2800 Series Installation and operating instructions

Simrad
Simrad 50-200 COMBI C - REV D datasheet

Fairchild
Fairchild T7800 Series Installation, operation and maintenance instructions

Knick
Knick ProLine P50000 user manual

Thyracont
Thyracont VSM72 operating instructions

Johnson Controls
Johnson Controls DP150 Series installation guide