Troubleshooting
For professional in vitro diagnostic use only.
• For all professional healthcare settings
• Place the meter into the docking/
charging station when it is not in use
• Clean with 70% isopropyl alcohol, 10%
dilution of commercial grade sodium
hypochlorite or commercial surface
decontamination preparations
• Do not spray liquid directly to the
meter or immerse into liquid
Nova StatStrip Glucose and ß-Ketone
Control Solution
• Store control material vial tightly closed under 30°C
• Do not refrigerate or freeze.
• Open vial stability is up to 3 months
or until expiration date.
• Store strips at 15°C–30°C.
• Use strips within expiration day or
glucose test within 180 days of opening
the vial and ß-Ketone strips within
90 days of opening the vial.
Sample material
• Wash patient’s hands or clean with alcohol pad and dry
completely before taking the sample.
• The test strip must be lled in a one go. Do not add more sample.
• For capillary nger stick, venous whole blood and arterial samples.
• Suitable also for neonatal samples and patients receiving intensive
medical therapy.
• Suitable anticoagulants: sodium, lithium, and ammonium heparin.
Nova StatStrip Glucose and ß-Ketone
Hospital Meter
Nova StatStrip Glucose and ß-Ketone
strips
Transfer failed: The meter was removed before
data transfer was complete. Please re-dock
the meter.
Flow error: The specimen was incorrectly drawn
into the test strip due to either insucient or incor-
rect sample application. Repeat the test with a new
strip. If the error code persists, perform test using
an alternate method.
Analysis canceled: The test has been canceled,
repeat the test with a new strip. Leave the test strip
in place until the result is displayed on the screen.
Bad sample: Insert a new strip and rerun the test.
If the error code persists, perform the test using
an alternate test strip vial or alternate method.
Replace strip: Occurs after insertion of strip or
occurs during analysis. Insert another strip and
retest. If the error code persists, perform the
test using an alternate test strip vial or alternate
method.
Temperature: Meter will only work within the
temperature range of 1°C–40°C. Return the meter
to an environment within the specied
temperature range 1°C–40°C.
8127-02EN 04/2021.
Transfer failed: Server refuses to allow dialog with
meter, or Connection to server was broken. Please
check the network settings, status of your network,
or contact your administrator for assistance.