Alaris 2090 Manual

Model 2090
Directions for Use
Thermometer
CORE
•
CHECK
®
Tympanic Thermometer System

THIS PAGE LEFT BLANK

INTRODUCTION GETTING STARTED OPERATING
INSTRUCTIONS
TABLE OF CONTENTS
INFORMATION
DISPLAYS
MAINTENANCE
TABLE OF CONTENTS i
INTRODUCTION
ABOUT THE SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
FEATURE SUMMARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Thermometer Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
GETTING STARTED
WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
INSTRUMENT FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
SETUP PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
OPERATING INSTRUCTIONS
PROBE COVER DISPENSE/LOAD INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Probe Cover Dispenser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Load Probe Cover Onto Thermometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
TAKING A TEMPERATURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
INFORMATION DISPLAYS
CORE•CHECK®THERMOMETER LCD DISPLAY SCREENS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
MAINTENANCE
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
THERMOMETER CLEANING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
SERVICE INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Product Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

GENERAL CONTACT INFORMATION
Customer Advocacy
For clinical and technical questions, feedback, and troubleshooting assistance.
Phone, toll-free, within the United States and Canada: (800) 854-7128, Ext. 7812
E-Mail: [email protected]
Technical Support and Service
Center Addresses:
AE
ALARIS Medical Systems Middle
East Office, PO Box 5527
Dubai, United Arab Emirates.
Tel: (971) 4 28 22 842
Fax: (971) 4 28 22 914
AU
ALARIS Medical Australia Pty Ltd,
8/167 Prospect Highway,
Seven Hills, NSW 2147
Australia.
Tel: (61) 2 9838 0255
Fax: (61) 2 9674 4444
BE
ALARIS Medical Belgium B.V.,
Otto De Mentockplein 19,
1853 Strombeek- Bever,
Belgium.
Tel: (32) 2 267 38 99
Fax: (32) 2 267 99 21
CA
ALARIS Medical Canada Ltd,
5975 Whittle Road, Suite # 120,
Mississauga, Ontario L4Z 3N1
Canada.
Tel: (1) 905-507-1131
Fax: (1) 905-507-6664
CN
ALARIS Medical Systems Office,
Suite 401 No. 88 Chang Su Rd.
Shanghai, PC 200040,
China.
Tel: (56) 8621-62488591
Fax: (56) 8621-62482567
DE
ALARIS Medical Deutschland,
GmbH, Pascalstr. 2,
52499 Baesweiler,
Deutschland,
Tel: (49) 2401 604 0
Fax: (49) 2401 604 121
ES
ALARIS Medical España, S.L.,
Avenida Valdelaparra 27,
Edificio, Alcor,
28108 - Alcobendas, Madrid,
España,
Tel: (34) 91 657 20 31
Fax: (34) 91 657 20 42
FR
ALARIS Medical France, S.A.,
95, rue Péreire,
78105 St Germain en Laye
Cedex.
France.
Tél: (33) 1 39 10 50 00
Fax: (33) 1 30 61 22 23
GB- Manufacturer’s
ALARIS Medical UK Ltd,
The Crescent, Jays Close,
Basingstoke,
Hampshire, RG22 4BS,
United Kingdom.
Tel: (44) 0800 389 6972
Fax: (44) 1256 388 411
HU
ALARIS Medical Hungary,
Döbrentei ter 1,,
H-1013 Budapest,
Magyar,
Tel: (36) 14 88 0232
Fax: (36) 12 01 5987
IT
ALARIS Medical Italia S.P.A.
Via Ticino 4,
50019 Sesto Fiorentino,
Firenze, Italia,
Tél: (39) 055 34 00 23
Fax: (39) 055 34 00 24
NL
ALARIS Medical Holland, B.V.,
Kantorenpand “Hoefse Wing”,
Printerweg 5,
3821 AP Amersfoort,
Nederland.
Tel: (31) 33 455 51 00
Fax: (31) 33 455 51 01
NO
ALARIS Medical Norway A/S
Solbréveien 10A, 1383 ASKER,
Norge.
Tel: (47) 66 98 76 00
Fax: (47) 66 98 76 01NO
NZ
ALARIS Medical NZ Ltd.
Unit 14, 13 Highbrook Drive.
East Tamaki, Auckland
New Zeland.
Tel: (64) 9 273 3901
Fax: (64) 9 273 3098
SE
ALARIS Medical Nordic, AB.
Hammarbacken 4B,.
191 46 Sollentuna,
Sverige.
Tel: (46) 8 544 43 200
Fax: (46)8 544 43 225
US
ALARIS Medical Systems, Inc.
10221 Wateridge Circle,.
San Diego, CA 92121 USA.
Tel: (1) 800 854-7128
Fax: (1) 800 854-6179
ZA
ALARIS Medical S.A. (Pty) Ltd,
Unit 2 Oude Molen Business Park
Oude Molen Road, Ndabeni,
Cape Town 7405, South Africa.
Tel: (27) 0860 597 572
Fax: (27) 21 5107567

INTRODUCTION
INTRODUCTION
INTRODUCTION 1
About the System
This manual is a guide for using the CORE•CHECK
®
Tympanic
Thermometer System Model 2090. The CORE•CHECK
®
Thermometer is designed to measure infrared energy emitted
from the tympanic membrane. Infrared energy is generated by
the vibration and movement of molecules in all matter. As
temperature increases so does molecular activity, causing an
increase in the radiation of infrared energy.
Infrared energy emitted from the tympanic membrane in the ear
canal provides an excellent measurement of the body's core
temperature. The blood supply for the tympanic membrane is
common with the hypothalamus, the area of the brain that
regulates body temperature.
The CORE•CHECK
®
Thermometer is positioned in the external
auditory canal. The thermometer senses infrared radiation
emitted from the surface of the tympanic membrane and
surrounding tissue. This reading is quickly processed through
the CORE•CHECK
®
Thermometer technology to provide an
accurate measurement of core body temperature.
The CORE•CHECK
®
Tympanic Thermometer System includes
the thermometer, Probe Cover Dispenser - Model 2992, Home
Base - Model 2993, and Probe Cover Bracket - Model 2994. The
thermometer uses ALARIS Medical Systems
®
Disposable Probe
Covers, Model P90.
For additional information, contact your ALARIS Medical
Systems
®
representative.

Feature Summary
PROBE COVER
BRACKET
Attaches to the wall and
holds a box of five probe
cover packets.
HOME BASE
Holds thermometer and
dispenser, also resets the
anti-theft feature.
THERMOMETER
DISPENSER
Holds and dispenses P90
probe covers. Combines
with thermometer to
provide a hand-held unit.
DISPOSABLE
PROBE COVERS
Reduce the risk of cross-
contamination and easily
loads onto thermometer
probe tip.
• Accurate, non-invasive determination of core temperature
• Temperature measurement within 3 seconds
• Anti-theft feature
• Durable by design and construction
• User-selectable Celsius or Fahrenheit temperature readings
• Display backlight for low-light conditions
• Powered by a single 9-volt battery
• Home base holds thermometer and dispenser
• Disposable probe covers and probe cover detectors reduce the risk
of cross-contamination
• Dispenser conveniently stores and dispenses probe covers
• Thermometer performs a self-test
Thermometer Features
2INTRODUCTION
R
R R
R R
RR R
RRRR
R R R
R R RR
R R R R
RR R
RR
R R
R
Disposable
Probe Cover

External Auditory Canal
Infrared Energy
Molecular Activity
Tympanic Membrane
CLOSE AND LOCK
PRESS TO SLIDE
OPEN
Definitions
INTRODUCTION 3
INTRODUCTION
U S E O N L Y
9V ALKALINE
ENSURE
LENS
IS CLEAN
BEFORE USE
Cleaning Advisory.
Alkaline battery used.
Anti-Theft notification.
Instructions for probe cover dispenser.
A narrow, tube-like passage through which sound enters the ear,
also named an external auditory canal.
Relating to the range of invisible radiation wavelengths from 750
nanometers. Generating, using, or sensitive to infrared radiation.
Temperature is a measure of molecular motion. Pressure is a
measure of molecular activity. The higher the pressure, the faster
the molecules are moving and greater energy equals greater temperature.
The membrane in the ear that vibrates to sound.
Conformance to the Medical Device Directive 93/42/EEC
Manufactured Date; expressed in four digits as the year and two digits as
the month.
Attention: Refer to accompanying documentation.
Indicates degree of protection, fluid ingress.
IPX1
MM-Y YYY

THIS PAGE LEFT BLANK
4INTRODUCTION

GETTING STARTED
GETTING STARTED 5
GETTING STARTED
Warnings and Cautions
A warning is an alert to potential serious outcomes (death,
injury or serious adverse events) to the patient or user.
A caution is an alert to take special care for the safe and
effective use of the device.
CAUTION
WARNING
Definitions
Do not autoclave or immerse the CORE•CHECK®
Thermometer as damage will occur.
Cautions
Operating the CORE• CHECK®Thermometer near equipment
which radiates high energy frequencies (for example
electrosurgical/cauterizing equipment), may affect temperature
readings.
CAUTION
CAUTION
If unusually low or high temperature readings are obtained,
remove the probe cover, verify the probe lens is clean, shiny,
dry, undamaged, and free of debris. Attach a new probe cover
and repeat measurement.
CAUTION
To ensure proper performance of the CORE•CHECK®
Thermometer and to reduce potential injury, observe the
following precautions.
Keep lens clean. Lens should be cleaned with a dry-lint free
scrub or tissue. Replace lens if it remains cloudy.
CAUTION
Damage may occur if alcohol or acetone is used.
CAUTION

6GETTING STARTED
PROBE LENS
PROBE TIP
INFORMATION
DISPLAY
Displays temperature
for approximately four
minutes or until next
measurement is
taken.
GREEN INDICATOR
LIGHT
Flashes once to
indicate temperature
measurement is
complete.
PROBE COVER
DETECTOR
Assures a new
temperature is not
displayed unless probe
cover is replaced after
each measurement.
(Use P90 probe
covers).
BACKLIGHT BUTTON
Lights the display for
use in low light
conditions.
TEMPERATURE
SWITCH
Initiates temperature
measurement.
BATTERY
COMPARTMENT
THERMOMETER
RECEPTACLE
SLIDER
PROBE COVER
LEVEL
INDICATOR
Figure 1
CORE•CHECK®Tympanic Thermometer System
DOOR
Instrument Features
Figure 2
Probe Cover Dispenser

GETTING STARTED 7
GETTING STARTED
1. Carefully unpack the CORE•CHECK
®
Tympanic Thermometer
System and ensure the thermometer, dispenser, and home base
are undamaged. Packaging will include a battery, battery
compartment cover, extra probe tips and a probe cover wall
bracket.
2. Set Temperature mode: Set thermometer to Celsius (°C) or
Fahrenheit (°F) by sliding the recessed switch located on left
side of the battery compartment. Use a small screwdriver or pen
tip to slide the switch into position (°F = toward the outside; °C =
toward the center).
3. Install Battery: Place a 9-volt alkaline battery into the compartment
as shown in Figure 3 and snap connector onto battery terminals.
4. Attach battery compartment cover. Orient the cover's arrow as
shown in Figure 3. Slide cover over opening until cover locks in
place. If necessary to remove the cover, begin by firmly pressing
center of cover with thumb causing a corner to lift, then using the
other hand, push cover at loosened corner and slide off. DO NOT
PRY COVER WITH SHARP OBJECT.
5. Self Test: Press and hold probe cover detector flush against
outer case (Figure 4). While holding, press and release
temperature switch once to initiate thermometer self-test to verify
both the display and thermometer's internal electronics are in
working order.
• The display is in working order if all segments momentarily
appear (Figure 5). If all segments do not appear, do not use
thermometer and refer to Service Information section of this
manual.
• The thermometer's internal electronics are in working order if a
temperature measurement or a "lo" message appears on the
display. If a numbered ERROR message appears on the
display, the thermometer has detected an internal malfunction.
(Refer to ERROR message in the Information Display section of
this manual.)
6. After a successful self-test routine, the thermometer is ready for
temperature measurement.
7. Home Base: The home base should be mounted to a convenient
location using either adhesive tape or screws supplied with the
instrument. Always store thermometer and dispenser in the
home base when system is not in use, which resets the Anti-
Theft feature.
Setup Procedure
Set °C/ °F switch mode
Install battery
Self-test
Mount home base
Q u i c k G u i d e

8GETTING STARTED
Figure 5
Information Display
CORE
RE P L ACE B ATT
COV E R C A L
RETURN TO BASE
ERROR
LOW
Figure 3
Battery Compartment
°C/°F Switch
Battery
Terminals
Battery
Setup Procedure (Continued)
Figure 4
Probe Cover Detector
Press
Battery Cover

OPERATING INSTRUCTIONS 9
OPERATING
INSTRUCTIONS
• Remove thermometer and dispenser from the home base.
• Press and slide dispenser door up until it locks in open
position.
• Install a package of P90 probe covers into the dispenser, with
arrow pointing up toward opened sliding door.
• Slide door down until locked in closed position, leaving the tab
outside the door.
• Firmly grasp tab and pull it out of the dispenser.
NOTE:
Use only ALARIS Medical Systems
®
P90 probe covers with
the CORE•CHECK
®
Thermometer. The size and thermal
characteristics of the probe cover affect thermometer’s
performance. Inaccurate readings may occur unless recommended
probe covers are used.
• Dispense a probe cover by pushing dispenser slide
completely in and slowly releasing.
• Center probe tip onto probe cover film then press firmly to
engage base of probe. This will actuate the probe cover
detector.
• The clear film will stretch smoothly over the entire probe.
Failure to fully load the probe cover will result in a REPLACE
COVER message.
OPERATING INSTRUCTIONS
Probe Cover Dispense/Load Instructions
Probe Cover Dispenser
NOTE:
If probe cover backing is not fully seated against the base
of the probe, a REPLACE COVER message will appear on the
display when the measurement is initiated. If this occurs, replace
the cover.
The following instructions assist in dispensing and loading probe covers onto the thermometer.
Load Probe Cover Onto Thermometer

10 OPERATIONAL INSTRUCTIONS
• The most important step in obtaining accurate temperatures
is correct placement of the probe tip with a P90 disposable
probe cover in the patient's ear.
• Center the probe tip far enough into the ear canal to firmly
seal the ear canal opening and point in the direction of the
tympanic membrane. Gently restrain the head if necessary.
Press and hold the temperature switch until the green light
flashes and temperature reading is displayed. Correct
positioning of the probe tip is necessary for accurate
readings. (See Figure 6 below).
Figure 6
Positioning of Thermometer
Taking A Temperature
If an unusually low or high temperature
reading is obtained, remove probe
cover, verify the probe lens is clean,
shiny, dry, undamaged, and free of
debris. Attach a new probe cover and
repeat measurement.
CAUTION
Operating the CORE• CHECK
Thermometer near equipment which
radiates high energy frequencies (for
example electrosurgical/cauterizing
equipment), may affect temperature
readings.
CAUTION
Evaluate condition of the patient's ear and
adjacent skin. The thermometer should not
be used on patients with ear drainage,
infected or chafed skin adjacent to the ear,
or on patients with incisions adjacent to the
ear. Such usage may increase the risk of
cross-contamination.
CAUTION
GENTLY
PUSH TO
SEAL EAR
CANAL

OPERATING INSTRUCTIONS 11
• Hold thermometer in place until temperature reading
is displayed (Approximately three seconds). The green
indicator light (below the display) will flash briefly.
• Remove thermometer from patient's ear and pull off
probe cover. The used probe cover must be replaced by a new
one before another measurement can be taken.
• Record the displayed temperature. Another temperature
reading can be taken immediately after replacing the probe
cover. The temperature will remain displayed for approximately
four minutes, unless another measurement is taken.
• Store in home base.
Taking A Temperature (Continued)
NOTE:
Removal of the thermometer from the ear before a
temperature is displayed will result in a low temperature reading.
OPERATING
INSTRUCTIONS

12 GETTING STARTED
THIS PAGE LEFT BLANK

INFORMATION
DISPLAYS
INFORMATION DISPLAYS 13
Self-Test - Displays with initiation of each measurement. Inspect
display to ensure all segments are operating as shown.
CORE - Displays with each temperature measurement.
REPLACE COVER (Replace Probe Cover) - A temperature
reading was initiated without a probe cover, a used probe cover,
or an improperly attached probe cover. Check lens for
cleanliness, attach a new probe cover, and reinitiate the
measurement.
RETURN TO BASE (Anti-Theft Feature) - A limited number of
measurements can be taken until thermometer displays a
continuous RETURN TO BASE message. Place thermometer
into home base to reset the anti-theft feature.
RETURN TO BASE - Displays as a flashing message when the
limit of temperature measurements have been met. Return to
base to reset.
Anti-theft feature has been reset. Thermometer can be used
immediately.
CORE
C
CORE
RE PLA CE
COVER
CORE
RETURN TO BASE
C
RETURN TO BASE
INFORMATION DISPLAYS
CORE • CHECK®Thermometer LCD Display Screens
The following examples are LCD display screens modes.
NOTE:
If this self-test display continues to flash, it indicates the
battery should be replaced.

14 INFORMATION DISPLAYS
BATT LOW (Battery Low) - Battery is low and requires
replacement.
REPLACE BATT (Replace Battery) - Battery must be replaced.
(Figure 3).
CAL (Recalibrate Instrument) - Indicates calibration of the
thermometer should be checked. (Refer to Maintenance section
of this manual for information concerning calibration).
ERROR (Accompanied by a number) - Thermometer has
detected an internal malfunction. Remove and reinstall battery.
To perform a Self Test: Press and hold probe cover detector
flush against outer case (Figure 4). While holding, press and
release temperature switch once to initiate thermometer self-test
to verify the display and thermometer's electronics are in working
order. If error message appears again, record the number, and
contact qualified service personnel.
The display messages indicates the ambient (room) temperature
is outside the thermometer's operating range:
lo A - Ambient temperature is below 15.6°C (60°F).
hi A - Ambient temperature is above 43.3°C (110°F).
The display messages indicates the patient's temperature is
outside the thermometer measurement range:
lo - Measured temperature is below 25°C (77°F).
hi - Measured temperature is above 43.3°C (110°F).
CORE
BATT LOW
c
RE PLA CE B ATT
CORE
CA L
c
ERR OR
CORE • CHECK®Thermometer LCD Display Screens (Continued)

MAINTENANCE 15
MAINTENANCE
MAINTENANCE
CORE•CHECK®
Thermometer
Technical Service Manuals are available from ALARIS Medical Systems.
They include routine service schedules, interconnect diagrams, component parts lists and descriptions,
test procedures and/or references, and other technical information to assist service personnel in repair
and maintenance of each instrument’s repairable components. Maintenance procedures are intended
to be performed only by qualified personnel.
Measurement Range : 25°C to 43.3°C (77°F to 110°F).
Accuracy: ±0.1°C (± 0.2°F). Meets ASTM Specifications No. E1112-86, at ambient
temperature range of 18.3 to 32.2°C (65-90°F)
Response Time: Less than 3 seconds
Battery: Approximately 8,000 readings on one 9 volt alkaline battery, under normal use
Size: 7.5 x 1.5 x 3.0 inches (19.05 x 3.81 x 7.62 centimeters)
Weight: 8 ounces with battery (266 grams)
Case Material: Durable, impact resistant plastic
Disposable: ALARIS Medical Systems
®
,P90 Probe Covers
Humidity: 15% to 90% Noncondensing
Storage
Temperature: -35°C to 60°C (-32°F to 140°F)
Anti-Theft Feature: 40 measurements are available before return to base is displayed.
Display Messages: Batt Low, Replace Batt, Return to Base, hi A (high ambient), lo A (low ambient),
lo (low temperature), hi (high temperature and numbered error messages for
internal malfunctions
Specifications

16 MAINTENANCE
• The LCD display is tested before each measurement by
lighting all of its segments for a brief period. It is recommended
to inspect the display during this time to ensure all segments
are operating properly. The display test ends when the
temperature or “lo” is displayed.
• ALARIS Medical Systems recommends the
CORE•CHECK®
Thermometer be calibrated every ten thousand readings or
every six months, whichever comes first. The CAL message
automatically flashes after ten thousand readings.
• Calibration must be performed using the ALARIS Medical
Systems
®CORE•CHECK®
Thermometer Calibrator - Model
9000. (Refer to the calibrator's operating instructions for
calibration procedure).
• If thermometer is dropped, inspect the lens and probe tip
for damage. The lens and probe tip can be replaced without
opening the instrument. (Refer to the Technical Manual for
procedure). Calibrate the CORE•CHECK
®
Thermometer if lens
replacement was required.
• The CORE•CHECK
®
Thermometer can be cleaned with a moist
cloth warm water and a mild detergent (such as dish washing
liquid) or with a variety of other agents. See Figure 7 for
appropriate cleaning methods and acceptable cleaning agents
for the different thermometer components.
•
Do not immerse the CORE•CHECK®Thermometer.
• Frequently check the lens for cleanliness. Incorrect readings
can result from debris or fingerprints. Lens should appear clear,
shiny, and unobstructed from protective probe tip.
Thermometer Cleaning
Do not autoclave or immerse the
CORE•CHECK®Thermometer as
damage will occur.
CAUTION