Ametek VDS 200Q Series Instruction Manual

M a n u a l
f o r O p e r a t i o n
VDS 200Q series
Voltage Drop Simulator pulses 2b, 4
VDS 200Q25.2, VDS 200Q50.2, VDS 200Q100.2,
VDS 200Q150.2, VDS 200Q200.2
Testing of electronic modules in 12V/24V or 48V supply
systems.
The VDS 200Q is four-quadrant linear power amplifier with a
very low source impedance. It simulates the battery power
supply of a vehicle and complex power supply distortions in the
power range up to 12 kW. Many different waveforms are
integrated as standard such as pulse 2b required by ISO 7637
and the starting profile from ISO 16750 as well as requirements
from many manufacturers’ standards like LV 124 and LV 148.
•ISO 7637
•ISO 16750
•SAE J1113
•Manufacturer spec
LV 124, LV 148, GM,
Ford, Chrysler, BMW,
VW, PSA, Renault,
Fiat ....
Version:
1.32.1 / 22.11.2019
the benchmark for emc
Replaces:
1.3 / 18.11.2019
File Name:
UserManual-VDS200Qxx.2-E-V1.3.1.doc
Print Date:
22.11.2019

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 3 / 36
Contents
1. Model Overview.............................................................................................................................................4
1.1. VDS 200Q Models........................................................................................................................................4
2. Operating Functions .....................................................................................................................................5
2.1. Front view .....................................................................................................................................................5
2.2. Rear view VDS 200Q200.2...........................................................................................................................8
2.3. Safety with voltage setting............................................................................................................................9
3. Operation......................................................................................................................................................11
3.1. Description of the menus............................................................................................................................11
3.2. Main Menu ..................................................................................................................................................11
3.2.1. ISO 7637.....................................................................................................................................................13
3.2.1.1. Pulse 4 voltage drop..................................................................................................................................13
3.2.1.2. Pulse 2b......................................................................................................................................................13
3.2.2. ISO 16750-2 WD 03/2000-2.......................................................................................................................14
3.2.2.1. Short voltage drop.......................................................................................................................................14
3.2.2.2. Slow decrease / increase............................................................................................................................14
3.2.2.3. Supply voltage profile..................................................................................................................................15
3.2.2.4. Pulse ’Starting profile’ .................................................................................................................................15
3.2.2.5. Sinus Sweep...............................................................................................................................................16
3.2.2.6. Overvoltage Vmax ......................................................................................................................................16
3.2.3. Functions ....................................................................................................................................................17
3.2.3.1. Sine wave ...................................................................................................................................................17
3.2.3.2. Jump Start...................................................................................................................................................18
3.2.3.3. VDS Externally via the analog input ...........................................................................................................18
3.2.3.4. Pulse 4 ( GM 9105 P) .................................................................................................................................19
3.3. Service........................................................................................................................................................20
3.4. Setup...........................................................................................................................................................21
3.5. Source Settings...........................................................................................................................................22
4. Test Equipment............................................................................................................................................24
4.1. Construction................................................................................................................................................24
5. Technical Data .............................................................................................................................................28
5.1. Test level.....................................................................................................................................................28
5.1.1. Sense.............................................................................................................................................................28
5.2. Trigger.........................................................................................................................................................28
5.3. Input/output.................................................................................................................................................28
5.4. Interfaces ....................................................................................................................................................28
5.5. General .......................................................................................................................................................29
5.6. Environmental conditions............................................................................................................................29
6. Maintenance.................................................................................................................................................30
6.1. General .......................................................................................................................................................30
6.2. Test set-up..................................................................................................................................................30
6.3. Test set-up with software iso.control ..........................................................................................................30
6.4. Example Test setup with AutoWave and VDS 200Q..................................................................................31
6.5. Calibration and Verification.........................................................................................................................32
6.5.1. Factory calibration.......................................................................................................................................32
6.5.2. Guideline to determine the calibration period of AMETEK CTS instrumentation .......................................32
6.5.3. Calibration of Accessories made by passive components only:.................................................................32
6.5.4. Periodic In-house verification......................................................................................................................32
7. Delivery Groups...........................................................................................................................................33
7.1. Basic equipment VDS 200Q200.2..............................................................................................................33
7.2. Accessories and options.............................................................................................................................33
8. Appendix ......................................................................................................................................................34
8.1. Declaration of CE-Conformity.....................................................................................................................34
8.1.1. Declaration of CE-Conformity VDS 200Q200.2..........................................................................................34
8.1.2. Connectors VDS 200Q200.2......................................................................................................................35
8.2. VDS 200Q Menu overview .........................................................................................................................36

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 4 / 36
1.
Model Overview
1.1.
VDS 200Q Models
Standard models
Model Voltage
voltage
current
Max. Inrush
current
Sinus
f max
VDS 200Q25.2
-20 to 80 V
Imax = ±25 A
75A for 200 ms
250 kHz
VDS 200Q50.2
-20 to 80 V
Imax = ±50 A
50A for 200 ms
250 kHz
VDS 200Q100.2
-20 to 80 V
Imax = ±100 A
200A for 200 ms
250 kHz
VDS 200Q150.2
-20 to 80 V
Imax = ±150 A
450A for 200 ms
250 kHz
VDS 200Q200.2
-20 to 80 V
Imax = ±200 A
600A for 200 ms
250 kHz
VDS 200Q200.2 VDS 200Q150.2 VDS 200Q100.2

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 5 / 36
2.
Operating Functions
2.1.
Front view
1Display
2Function keys "F1..F7"
3"TEST ON"
4Knob (Inc/Dec)
5 Cursor keys "" and "→"
6EXIT
7ESC
8BNC CRO Trigger ( for oscilloscope)
9 OUTPUT ACTIVE LED
1 Display
All functions and parameters are displayed (8 lines with max. 40 characters).
2 Function keys "F1 .. F7"
Parameters and functions, displayed in the lowest line, can be selected with the related function key.
3 Test On
By pressing the key"TEST ON" the test procedure is initiated with the preselected parameters. The yellow
button is illuminated and indicates the Test ON status. After “Test OFF” or when no test is started, the output
voltage and the current will be set to zero.
4 Knob (Inc / Dec)
The knob increments or decrements test parameters with a numeric value or selects from a list of parameters.
5 Cursor keys
Parameters and functions can be changed on-line. The selection of these parameters is realized with the cursor
moving to the left or to the right.
6 Exit
Pressing the "EXIT" function will return to the previous menu. This is only possible if no test routine is running.
7 ESC
When pressing the ESC button the user moves back one page in the menu.
8 BNC CRO Trigger
At the BNC connector CRO TRIGGER a signal is available to trigger an oscilloscope.
9 OUTPUT ACTIVE LED
Output to the DUT or other generators.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 6 / 36
Additionally, on the front of the VDS 200Q models, there are the following elements:
Power ON/OFF Switch
Used to power on and off the system. At start, a synchronization and boot process will occur. After a few
moments, the system is ready to be used.
CAUTION:
Do not misuse the power on/off switch.
The VDS 200Q contains multiple controllers and signal generators that must be synchronized at
startup.
For best results, please allow at least five seconds between switching off and back on the
generator.
Output Connectors and Interconnects VDS 200Q150.2 and VDS 200Q200.2
1 Fuse 10 A (5x20 mm) for the PFM mains
2 Mains output to external PFM 200N200
3 Frame Bus out to PFM 200N200
4 Ground connector Minus to GND to
minimize noise, when an isolated output is not
needed
5 Fuse 10 A (5x20 mm) against excess ground
leakage current
6 DC out to DUT or PFM 200N200
7 Sense input

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 7 / 36
Output Connectors and Interconnects VDS 200Q100.2
1 Fuse 10 A (5x20 mm) for the PFM mains
2 Mains output to external PFM 200N100
3 Frame Bus out to PFM 200N100
4 Ground connector Minus to GND to
minimize noise, when an isolated output is not
needed
5 Fuse 10 A (5x20 mm) against excess ground
leakage current
6 DC out to DUT
7 Sense input
8 Measurement or
Low-current output (32A max.)
9 Fuse 32 A for Low-current output

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 8 / 36
2.2.
Rear view VDS 200Qxx.2
1Emergency Switch
2Fuse or
(additional Mains on some models)
3Earth Connections
4Mains Input 1-ph or 3-ph
5-7Optional AutoWave Out (N/C)
8 AutoWave Out to VDS In
9 VDS Analog In
10 Optional Framebus In
11 Ethernet (to AutoWave)
12 Framebus Out
13 IEEE 488.1
14 USB (to VDS)
1 Emergency Switch
Used to connect the emergency lockout switch.
2 Fuse
See 5.5 General, and 4.2 Fuses
Note: On the 208V models, this is replaced by an additional mains connector for countries where 100-120V mains is the
norm. In this case, be sure to connect this to mains.
3 Reference earth connection
During immunity tests, it may be useful to connect the simulator with reference earth plane of the test set-up.
For complete rack installations, all different test generators shall be grounded at this point.
4 Mains input
This connector depends on the VDS model
5-7 Optional AutoWave Outputs
Can be used to connect the AutoWave with additional sources
8 AutoWave CH1 Output
Not present on some units when connected internally
NOTE: If equipped with a CH1 output, the supplied BNC bridge must be connected to Analog In to the VDS
9 Analog In to VDS
Connection to AutoWave or an external arbitrary waveform generator if needed.
10 Optional Framebus In
The framebus is usually taken from the AutoWave
11 Ethernet connection pass-through for AutoWave (if equipped)
12 Framebus Out
Be sure to use the supplied Framebus terminator 101732
13 IEEE 488 // GPIB
IEEE 488 interface with IEEE connector.
14 USB interface
USB interface “USB B” connector. For data transfer a USB interface is available. The internal RS 232 interface
is converted to USB standard. Therefore, the user must set the same Baudrate in the device and control
software.
Note: USB interface is for controlling the VDS 200Q in stand-alone configuration. See the PFM 200N200 and
AutoWave manuals if alternative configurations will be used. AutoWave, doesn’t support USB communications.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 9 / 36
2.3.
Safety with voltage setting
To ensure a safe operation of the DUT (Device Under Test) some restrictions in the operation of the instrument
are built in. These restrictions are explained within this paragraph.
CAUTION:
The VDS 200Q series, as well as most linear power amplifiers will create spikes at start and stop.
These are limited to 15V normally, but inductive loads can generate high voltages. Do not touch any
conducted parts of the system and be sure to remove the DUT during power up and shut down or
damage may occur.
Do not use the VDS 200Q with a parallel battery.
A four-quadrant source, like the VDS 200Q can supply both positive and negative voltage as well as positive or
negative current in any combination. It can also sink current that may be supplied from external sources. Often,
using a classic 'source' (power supply or 1 - 2 quadrant source) the user felt the need to buffer the voltage in
order to get inrush or to properly simulate the impedance of the battery in the car. This, of course, can no longer
be controlled as the source simply becomes a charger for the battery.
However, when using a four-quadrant source like the VDS 200Q and the source's voltage is set lower than the
battery, the battery can discharge through the source.
This can result in a dangerous condition, not only for the four-quadrant source, but also because you've
effectively short-circuited the battery when setting the voltage to zero (or switching the four-quadrant source off!).
Remember that all sources must have a source impedance of <10 mOhm according to ISO 7637.
Finally, by shorting (or overcharging) the battery, the battery can actually overheat, crack (releasing acid!) or
even explode due to the release of hydrogen.
CAUTION:
Do not use a parallel battery with the VDS 200Q or a dangerous condition can occur by exceeding
the current of the VDS 200Q, and/or overheating the battery.
The VDS 200Q basically is divided into 2 different operation modes, which includes their individual test routines
and supply voltage setting. The different modes can be listed as follows:
DC Source
Within this mode the VDS 200Q is used as a simple DC power source in the range of up to 80V; and an
integrated current limiter.
Arbitrary Waveform Simulator
Within this mode the VDS 200Q generates arbitrary waveforms and signals which are specified in different
standards, such as pulse 4 of ISO 7637.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 10 / 36
All these different test modes are changing generally the voltage supply setting of the generator in a different
way. This would mean a certain risk for the operator to burn out the connected DUT by higher DC supply
voltages as intended. Therefore it is decided to clearly separate the two test modes by the following structure
listed:
Analogue input +10/-10V
The amplifier can be controlled by an external signal generator. The operator therefore shall select the User
Test Routines of the VDS 200Q part and start the menu Extern. The amplifier is then able to be remotely
controlled.
The input signal range is +10 to -10V in the frequency range of 0-250kHz. The output power (EUT test
supply) is capable of 0-40V or 0-80V and a nominal current depending on setting. For example, in the
normal (8x) gain setting, a 2V input signal would result in 16V out. Please note, the 3dB guaranteed
bandwidth of 250 kHz is not guaranteed when controlling the source with the BNC analog input. Typical
bandwidth when using an external control signal is 150 kHz.
1. DC source
When entering the Arbitrary Wave test mode the dc output voltage of the VDS 200Q is automatically set to the
nominal voltage of the DUT. The actual nominal voltage can be defined in the service menu under "Set-up".
When starting the related test routines the output voltage will be generated as per the setting shown up in the
display for each individual test routine.
- The operator can accept this setting and start the test immediately.
- The operator can first change the parameters and than start the test.
When leaving the test mode DC source the output voltage is automatically reset to the nominal supply voltage of
the DUT. The previous voltage setting will be stored in the test file.
2. Arbitrary Wave Simulator
When selecting the Arbitrary Wave mode, the output voltage will be set automatically to the nominal voltage of
the DUT. The actual nominal voltage can be defined in the service menu under "Set-up".
When starting the related test routines, the output voltage will be generated as per the setting shown up in the
display for each individual test routine.
- The operator can accept this setting and start the test immediately.
- The operator can first change the parameters and then start the test.
When leaving the test mode Arbitrary Wave Simulator, the output voltage is automatically reset to 0V. The
previous voltage test parameters will be stored and can be used for the next test.
The consequence of the structure is that between the different test modes
the DUT supply is automatically switched off.
Be sure to select the appropriate bandwidth (Capacitive, Standard, High Freq.) in the Service
menu appropriate for the necessary test frequency.
While the full voltage and frequency can be programmed, bandwidth is typically measured at
the -3dB point. Therefore, please keep in mind that there will be some attenuation at the
maximum frequencies of each range that may have to be compensated for.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 11 / 36
3.
Operation
3.1.
Description of the menus
The simulator VDS 200Q is operated by an easy menu control system.
Seven function keys are available to select parameters and functions. All
functions are indicated on the display; max. 8 lines and 40 characters.
The selected parameter is blinking and can be changed by turning the knob (incr./decr.).
➔ : The digit to be changed can be selected with the cursor (➔ ).
- Set values are directly indicated on the screen.
- Status on the bottom lines shows the desired status after pressing the function key.
ESC : ESC will take you back to the previous level in the menu and set the displayed
values. The latest settings are stored automatically and will be recalled when the
menu is selected again.
EXIT : The firmware will reset to the main screen.
AMETEK CTS
VDS 200Q100.2
Voltage Drop Simulator
V 1.00.00
Int. V1.00.00
Waveforms
SWN: 001234
The serial number and the version number SWN are
used for traceability reasons. These numbers are listed
in the factory test reports and calibration certificates.
These numbers also are listed within the test reports
generated by the iso.control software
Start-up display example VDS 200Q200.2
3.2.
Main Menu
MAIN MENU
F1 : Source Mode
F2 : Wave Simulator
F6 : Source Settings
F7 : Service
F1
F2
F3
F4
F5
F6
F7
F1 Source Mode
A mode that allows the user to set immediate voltage for quick tests with a DUT.
F2 Wave Simulator
In this mode, the internal signal generator is used to generate arbitrary waveforms as required in different
standards.
The operator can use the generator in this mode as a,
- DC power supply source,
- powerful power arbitrary wave generator with integrated test routines
- or simply as a battery simulator
The amplifier can be remote controlled by any external arbitrary generator. External generators shall be
connected at the rear part of the equipment. Any waveform can be generated up to the upper bandwidth of the
unit.
F6 Source Settings
Set-up, self-test, source settings and addresses of AMETEK CTS can be selected and displayed.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 12 / 36
F7 Service
Set-up, self-test, source settings and addresses of AMETEK CTS can be selected and displayed.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 13 / 36
3.2.1.
ISO 7637
3.2.1.1.
Pulse 4 voltage drop
This pulse simulates supply voltage reduction caused by energizing the starter-motor circuits of internal
combustion engines, excluding spikes associated with starting.
tf [Vb-Va1] < 5ms
Limits depends VDS voltage range
Input restrictions
0.0 V <= Vb + Va1 <= 30.0V (60.0V)
0.0 V <= Vb + Va2 <= 30.0V (60.0V)
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (+ 60.0V)
Va2
- 30.0V (- 60.0V)
-
30.0V (+ 60.0V)
t1
0.1s
-
99.9s
t7
5ms
-
999ms
t8
5ms
-
999ms
t9
0.1s
-
99.9s
t11
5ms
-
999ms
Va
0.0V
-
30.0V (+ 60.0V)
I
1A
-
(Imax)
tri
Auto / Manual
3.2.1.2.
Pulse 2b
This pulse simulates transients from DC motors acting as generators after ignition is switched off.
tr, tf ( 10/90%) =1ms 50%
Limits depends VDS voltage range
Input restrictions
Va1 0.0V
0.0 V Vb + Va1 30.0V (60.0V)
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
0.0V
t1
0.1s
-
99.9s
t6
1ms
-
999ms
td
5ms
-
9999ms
int
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 14 / 36
3.2.2.
ISO 16750-2 WD 03/2000-2
3.2.2.1.
Short voltage drop
This test is to simulate the effect of a classical fuse actuation in another circuit.
tr, tf = 10ms
Limits depends VDS voltage range
Input restrictions
0.0 V Vb + Va1 30.0V (60.0V)
0.0 V Vb + Va2 30.0V (60.0V)
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (60.0V)
Va2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
99.9s
t2
0.1s
-
99.9s
t3
0.1s
-
99.9s
int
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)
3.2.2.2.
Slow decrease / increase
This test is to simulate a gradual discharge and recharge of the battery.
Parameters:
V1
0.0V
-
30.0V (60.0V)
V2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
99.9s
t2
0.1s
-
9999.9s
I
1A
-
(Imax)
Remarks:
WD 03/2000-2 Voltage change rate = ( 3 0.1) V per minute
ISO 16750-2 Voltage change rate = ( 0.5 0.1 ) V per minute

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 15 / 36
3.2.2.3.
Supply voltage profile
This test is to determine the reset behavior of the device under Test at different voltage drops. This test is
applicable to equipment with reset function.
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (60.0V)
Va2
- 29.9V (- 59.9V)
-
30.0V (60.0V)
Ve
- 30.0V (- 60.0V)
-
30.0V (60.0V)
dVa
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
99.9s
t2
0.1s
-
99.9s
t3
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)
3.2.2.4.
Pulse ’Starting profile’
This test simulates a motor startup including a possible ripple.
tr, tf = 10ms
Ripple = 2Hz
Limits depends VDS voltage range
Input restrictions
0.0 V Vb + Va1 30.0V (60.0V)
0.0 V Vb + Va2 30.0V (60.0V)
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (60.0V)
Va2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
999ms
t7
5ms
-
999ms
t8
5ms
-
999ms
t9
0.5s
-
99.5s
t11
5ms
-
999ms
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 16 / 36
3.2.2.5.
Sinus Sweep
This test simulates a residual a.c. on the dc supply
During Int f3 is applied.
Limitations
Vb + Vp <= 30.0V (60.0V )
Vb - Vp = 0.0V
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Vp
0.0V
-
0.1 - 30.0V1)
f1
0.001Hz
-
150.000kHz2)
f2
0.001Hz
-
150.000kHz2)
f3
0.001Hz
-
150.000kHz2)
t1
0.1s
-
999.9s
t2
0.1s
-
999.9s
int
0s
-
999s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)
1) VDS 200Qnn.1 models: 250kHz; VDS 200Q10, 180kHz
2) VDS 200Q10: 0.1 - 60V
3.2.2.6.
Overvoltage Vmax
This test simulates a high energy load dump pulse.
Limits depends VDS voltage range
Input restrictions
0.0 V V1 + V2 30.0V (60.0V)
Parameters:
V1
0.0V
-
30.0V (60.0V)
V2
0.0V
-
30.0V (60.0V)
t1
0.01s
-
999.99 s
t2
0.01s
-
999.99 s
t3
0.01s
-
999.99 s
int
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 17 / 36
3.2.3.
Functions
3.2.3.1.
Sine wave
Limitations
Vb + Vp <= 30.0V (60.0V)
Vb - Vp = 0.0V
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Vp
0.1V
-
30V1)
t1
0.1s
-
99.9s
f1
0.001kHz
-
250.00kHz
f< 100Hz 1.2Hz
t2
1.0s
-
999.9s
n
1
-
30’000 / endl.
I
1.0A
-
(Imax)
Be sure to select the appropriate bandwidth in the Service menu appropriate for the
necessary test frequency.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 18 / 36
3.2.3.2.
Jump Start
Limitations
V1 + V2 <= 30.0V (60.0V)
V1 + V2 = 0.0V
Parameters:
V1
0.0V
-
30.0V (60.0V)
V2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.01s
-
999.99s
t2
0.01s
-
999.99s
t3
0.01s
-
999.99s
int
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)
3.2.3.3.
VDS Externally via the analog input
The power amplifier can be driven by an external control voltage (-10 to 10V), e.g. from an external waveform
generator.
For this purpose the operator must select VDS Extern by pressing the related function key. The Extern
mode is displayed and the unit can only be operated via the coaxial BNC input ANALOG IN at the rear part.
Parameters:
Input voltage at BNC input :-10...10V dc
Extern
Extern
I = 100A
Voltage Current
13.5 V 14.5 A
I = 100A
Start Change
Stop
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7
Note: The measured values are for indication only. Low current values are indicated with the value “LOW”.

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 19 / 36
3.2.3.4.
Pulse 4 ( GM 9105 P)
In addition to the ISO pulse 4 a 5Hz ripple of 1Vp-p is superimposed during t9 to Va.
tr, tf = 10ms
Input restrictions
0.0 V Vb + Va1 30.0V (60.0V)
0.0 V Vb + Va2 30.0V (60.0V)
Parameters:
Vb
0.0V
-
30.0V (60.0V)
Va1
- 30.0V (- 60.0V)
-
30.0V (60.0V)
Va2
- 30.0V (- 60.0V)
-
30.0V (60.0V)
t1
0.1s
-
99.9s
t7
5ms
-
999ms
t8
5ms
-
999ms
t9
0.4s
-
99.8s
t11
5ms
-
999ms
tri
Auto / Manual
I
1A
-
(Imax)

AMETEK CTS VDS 200Q Series
Manual for Operation V 1.3.1 20 / 36
3.3.
Service
All service functions are indicated on the display.
SERVICE
F1 : Addresses
F2 : Selftest
F3 : Setup
F5 : Set Voltage
F6 : Source Setting
F1
F2
F3
F4
F5
F6
F7
Addresses
The addresses of the AMETEK CTS GmbH and the
AMETEK CTS GmbH in Germany are shown.
The addresses of all AMETEK CTS sales agencies
are listed on the web site of EM TEST under:
www.emtest.com
Selftest
The operator can initialize a self test procedure to check the operation of the instrument. The software will clearly
explain the selftest procedure.
Set-up
The operator can change the simulator setting as explained on the next chapter.
Source Settings
The source settings menu contains features that are specific to the VDS 200Q series as explained in the next
chapter.
This manual suits for next models
5
Table of contents
Other Ametek Amplifier manuals

Ametek
Ametek 7124 User manual

Ametek
Ametek 7280 User manual

Ametek
Ametek SW 5550AE User manual

Ametek
Ametek 7270 User manual

Ametek
Ametek 7230 User manual

Ametek
Ametek Teseq TWT Series User manual

Ametek
Ametek Signal Recovery 5402 User manual

Ametek
Ametek 5113 User manual

Ametek
Ametek Teseo CBA Series User manual

Ametek
Ametek Milmega AS Series User manual