manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. AquaLab
  6. •
  7. Measuring Instrument
  8. •
  9. AquaLab 4TE Series User manual

AquaLab 4TE Series User manual

1. Getting Started—Chapter 4
• Place AquaLab on a level surface.
• Use AquaLab where environmental temperature remains fairly stable.
• Plug in AquaLab and turn on power (switch in back).
• For best results, let AquaLab warm-up for 15 min.
2. Sample Preparation—Chapter 8
• Fill sample cup no more than half full (-7ml). Do Not Overll!
• Minimum sample amount should cover bottom of sample cup
(small gaps OK).
• Ensure rim and outside of sample cup are clean.
• Make sure sample temperature is not more than 4°C above chamber
temperature.
3. Taking a Reading—Chapter 9
• Place sample cup in sample chamber. Close lid carefully to avoid spillage.
• Use the up and down arrows to select the desired moisture content model.
WARNING: Selecting the incorrect model will cause you to get an
inaccurate moisture content reading.
NOTE: Contact AquaLab if you don’t have a model for a product.
• Move latch left to start measurement.
• AquaLab beeps once when starting awmeasurements.
• First awvalue will display after one to two minutes, but reading is not
complete until save icon appears.
• Sample’s nal awreading, temperature and moisture content will display.
4. AquaLab Series 4TE DUO Menu—Chapter 5
Calibration Menu
• Press Menu Icon on main screen.
• Press Enter twice to start calibration.
AquaLab 4TE DUO Quick Start
Enter Icon Menu Icon
Configuration Data
Measurement
1. Getting Started—Chapter 4
• Place AquaLab on a level surface.
• Use AquaLab where environmental temperature remains fairly stable.
• Plug in AquaLab and turn on power (switch in back).
• For best results, let AquaLab warm-up for 15 min.
2. Sample Preparation—Chapter 8
• Fill sample cup no more than half full (-7ml). Do Not Overll!
• Minimum sample amount should cover bottom of sample cup
(small gaps OK).
• Ensure rim and outside of sample cup are clean.
• Make sure sample temperature is not more than 4°C above chamber
temperature.
3. Taking a Reading—Chapter 9
• Place sample cup in sample chamber. Close lid carefully to avoid spillage.
• Use the up and down arrows to select the desired moisture content model.
WARNING: Selecting the incorrect model will cause you to get an
inaccurate moisture content reading.
NOTE: Contact AquaLab if you don’t have a model for a product.
• Move latch left to start measurement.
• AquaLab beeps once when starting awmeasurements.
• First awvalue will display after one to two minutes, but reading is not
complete until save icon appears.
• Sample’s nal awreading, temperature and moisture content will display.
4. AquaLab Series 4TE DUO Menu—Chapter 5
Calibration Menu
• Press Menu Icon on main screen.
• Press Enter twice to start calibration.
AquaLab 4TE DUO Quick Start
Enter Icon Menu Icon
Configuration Data
Measurement
5. AquaLab Performance Verification—Chapter 7
• AquaLab performance and cleanliness should be checked frequently
(Each day of use is recommended)
• Use AquaLab Verication Standards to adjust linear offset.
• See Chapter 6 for AquaLab cleaning instructions.
6. Error Messages & Troubleshooting—Chapter 13
• Sample temperature greater than 4°C above AquaLab’s sensor causes error.
• If sample’s awis below ~0.03, AquaLab displays a less than (<) symbol.
• For read times longer than 5 minutes, see Chapter 13.
• For cleaning instructions, see Chapter 10.
AquaLab 4TE DUO Quick Start
Cautions
• Do not overll sample cups.
• Never tip or move AquaLab with sample loaded in sample chamber.
• Sample chamber contamination can damage AquaLab or cause
inaccurate results.
5. AquaLab Performance Verification—Chapter 7
• AquaLab performance and cleanliness should be checked frequently
(Each day of use is recommended)
• Use AquaLab Verication Standards to adjust linear offset.
• See Chapter 6 for AquaLab cleaning instructions.
6. Error Messages & Troubleshooting—Chapter 13
• Sample temperature greater than 4°C above AquaLab’s sensor causes error.
• If sample’s awis below ~0.03, AquaLab displays a less than (<) symbol.
• For read times longer than 5 minutes, see Chapter 13.
• For cleaning instructions, see Chapter 10.
AquaLab 4TE DUO Quick Start
Cautions
• Do not overll sample cups.
• Never tip or move AquaLab with sample loaded in sample chamber.
• Sample chamber contamination can damage AquaLab or cause
inaccurate results.
2365 NE Hopkins Ct.
Pullman WA, 99163
[email protected]
www.aqualab.com
©2015 Decagon Devices, Inc.
Printed in USA
13943 | 20-01-15
Call for AquaLab technical support
• 509-332-5601 • 7am – 5pm Pacic Time
To order cups and standards
• www.store.aqualab.com
2365 NE Hopkins Ct.
Pullman WA, 99163
[email protected]
www.aqualab.com
©2015 Decagon Devices, Inc.
Printed in USA
13943 | 20-01-15
Call for AquaLab technical support
• 509-332-5601 • 7am – 5pm Pacic Time
To order cups and standards
• www.store.aqualab.com

Other manuals for 4TE Series

2

Other AquaLab Measuring Instrument manuals

AquaLab PRE User manual

AquaLab

AquaLab PRE User manual

AquaLab 4TE Series User manual

AquaLab

AquaLab 4TE Series User manual

AquaLab 4TE Series User manual

AquaLab

AquaLab 4TE Series User manual

AquaLab 4 Series User manual

AquaLab

AquaLab 4 Series User manual

AquaLab Pawkit User manual

AquaLab

AquaLab Pawkit User manual

AquaLab Series 3 User manual

AquaLab

AquaLab Series 3 User manual

Popular Measuring Instrument manuals by other brands

Michell Instruments Optidew Vision user guide

Michell Instruments

Michell Instruments Optidew Vision user guide

JPT LP Series user manual

JPT

JPT LP Series user manual

Endress+Hauser Proline Promass P 300 manual

Endress+Hauser

Endress+Hauser Proline Promass P 300 manual

Hanna Instruments HI83141-1 user manual

Hanna Instruments

Hanna Instruments HI83141-1 user manual

Brookfield TT100 Installation, operation and maintenance instructions

Brookfield

Brookfield TT100 Installation, operation and maintenance instructions

Mollet MSD-420 quick start guide

Mollet

Mollet MSD-420 quick start guide

LI-COR LI-8100A quick start guide

LI-COR

LI-COR LI-8100A quick start guide

Endress+Hauser Liquiphant M with FEL56 Functional safety manual

Endress+Hauser

Endress+Hauser Liquiphant M with FEL56 Functional safety manual

Novasina AW SPRINT Series operating instructions

Novasina

Novasina AW SPRINT Series operating instructions

Fluidwell F Series Operation manual

Fluidwell

Fluidwell F Series Operation manual

CLAS AC 5300 manual

CLAS

CLAS AC 5300 manual

wtw LRD 01 instruction manual

wtw

wtw LRD 01 instruction manual

Reed Instruments Smart Series instruction manual

Reed Instruments

Reed Instruments Smart Series instruction manual

Megger MTO106 user manual

Megger

Megger MTO106 user manual

CYPRES 2 quick guide

CYPRES

CYPRES 2 quick guide

Heath Heathkit SB-644A manual

Heath

Heath Heathkit SB-644A manual

Campbell HMP60 product manual

Campbell

Campbell HMP60 product manual

Morris Products 57276 user manual

Morris Products

Morris Products 57276 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.