manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. AquaLab
  6. •
  7. Measuring Instrument
  8. •
  9. AquaLab 4TE Series User manual

AquaLab 4TE Series User manual

1. Getting Started—Chapter 4
• Place AquaLab on a level surface.
• Use AquaLab where environmental temperature remains fairly stable.
• Plug in AquaLab and turn on power (switch in back).
• For best results, let AquaLab warm-up for 15 min.
2. Sample Preparation—Chapter 8
• Fill sample cup no more than half full (-7ml). Do Not Overll!
• Minimum sample amount should cover bottom of sample cup
(small gaps OK).
• Ensure rim and outside of sample cup are clean.
• Make sure sample temperature is not more than 4°C above chamber
temperature.
3. Taking a Reading—Chapter 9
• Place sample cup in sample chamber. Close lid carefully to avoid spillage.
• Use the up and down arrows to select the desired moisture content model.
WARNING: Selecting the incorrect model will cause you to get an
inaccurate moisture content reading.
NOTE: Contact AquaLab if you don’t have a model for a product.
• Move latch left to start measurement.
• AquaLab beeps once when starting awmeasurements.
• First awvalue will display after one to two minutes, but reading is not
complete until save icon appears.
• Sample’s nal awreading, temperature and moisture content will display.
4. AquaLab Series 4TE DUO Menu—Chapter 5
Calibration Menu
• Press Menu Icon on main screen.
• Press Enter twice to start calibration.
AquaLab 4TE DUO Quick Start
Enter Icon Menu Icon
Configuration Data
Measurement
1. Getting Started—Chapter 4
• Place AquaLab on a level surface.
• Use AquaLab where environmental temperature remains fairly stable.
• Plug in AquaLab and turn on power (switch in back).
• For best results, let AquaLab warm-up for 15 min.
2. Sample Preparation—Chapter 8
• Fill sample cup no more than half full (-7ml). Do Not Overll!
• Minimum sample amount should cover bottom of sample cup
(small gaps OK).
• Ensure rim and outside of sample cup are clean.
• Make sure sample temperature is not more than 4°C above chamber
temperature.
3. Taking a Reading—Chapter 9
• Place sample cup in sample chamber. Close lid carefully to avoid spillage.
• Use the up and down arrows to select the desired moisture content model.
WARNING: Selecting the incorrect model will cause you to get an
inaccurate moisture content reading.
NOTE: Contact AquaLab if you don’t have a model for a product.
• Move latch left to start measurement.
• AquaLab beeps once when starting awmeasurements.
• First awvalue will display after one to two minutes, but reading is not
complete until save icon appears.
• Sample’s nal awreading, temperature and moisture content will display.
4. AquaLab Series 4TE DUO Menu—Chapter 5
Calibration Menu
• Press Menu Icon on main screen.
• Press Enter twice to start calibration.
AquaLab 4TE DUO Quick Start
Enter Icon Menu Icon
Configuration Data
Measurement
5. AquaLab Performance Verification—Chapter 7
• AquaLab performance and cleanliness should be checked frequently
(Each day of use is recommended)
• Use AquaLab Verication Standards to adjust linear offset.
• See Chapter 6 for AquaLab cleaning instructions.
6. Error Messages & Troubleshooting—Chapter 13
• Sample temperature greater than 4°C above AquaLab’s sensor causes error.
• If sample’s awis below ~0.03, AquaLab displays a less than (<) symbol.
• For read times longer than 5 minutes, see Chapter 13.
• For cleaning instructions, see Chapter 10.
AquaLab 4TE DUO Quick Start
Cautions
• Do not overll sample cups.
• Never tip or move AquaLab with sample loaded in sample chamber.
• Sample chamber contamination can damage AquaLab or cause
inaccurate results.
5. AquaLab Performance Verification—Chapter 7
• AquaLab performance and cleanliness should be checked frequently
(Each day of use is recommended)
• Use AquaLab Verication Standards to adjust linear offset.
• See Chapter 6 for AquaLab cleaning instructions.
6. Error Messages & Troubleshooting—Chapter 13
• Sample temperature greater than 4°C above AquaLab’s sensor causes error.
• If sample’s awis below ~0.03, AquaLab displays a less than (<) symbol.
• For read times longer than 5 minutes, see Chapter 13.
• For cleaning instructions, see Chapter 10.
AquaLab 4TE DUO Quick Start
Cautions
• Do not overll sample cups.
• Never tip or move AquaLab with sample loaded in sample chamber.
• Sample chamber contamination can damage AquaLab or cause
inaccurate results.
2365 NE Hopkins Ct.
Pullman WA, 99163
[email protected]
www.aqualab.com
©2015 Decagon Devices, Inc.
Printed in USA
13943 | 20-01-15
Call for AquaLab technical support
• 509-332-5601 • 7am – 5pm Pacic Time
To order cups and standards
• www.store.aqualab.com
2365 NE Hopkins Ct.
Pullman WA, 99163
[email protected]
www.aqualab.com
©2015 Decagon Devices, Inc.
Printed in USA
13943 | 20-01-15
Call for AquaLab technical support
• 509-332-5601 • 7am – 5pm Pacic Time
To order cups and standards
• www.store.aqualab.com

Other manuals for 4TE Series

2

Other AquaLab Measuring Instrument manuals

AquaLab Series 3 User manual

AquaLab

AquaLab Series 3 User manual

AquaLab PRE User manual

AquaLab

AquaLab PRE User manual

AquaLab 4 Series User manual

AquaLab

AquaLab 4 Series User manual

AquaLab 4TE Series User manual

AquaLab

AquaLab 4TE Series User manual

AquaLab Pawkit User manual

AquaLab

AquaLab Pawkit User manual

AquaLab 4TE Series User manual

AquaLab

AquaLab 4TE Series User manual

Popular Measuring Instrument manuals by other brands

Huvitz HLM-9000 Service manual

Huvitz

Huvitz HLM-9000 Service manual

Altana BYK byko-test manual

Altana

Altana BYK byko-test manual

CruzPro WSD110 manual

CruzPro

CruzPro WSD110 manual

SF Ratemeter 5135.55 instruction manual

SF

SF Ratemeter 5135.55 instruction manual

Tonghui TH2829 Series Operation manual

Tonghui

Tonghui TH2829 Series Operation manual

Texmate CL-B101D40HZ manual

Texmate

Texmate CL-B101D40HZ manual

Fluke 368 user manual

Fluke

Fluke 368 user manual

TRIALMA II Instructions for use

TRIALMA

TRIALMA II Instructions for use

Silva ADC Wind user manual

Silva

Silva ADC Wind user manual

Dranetz PowerGuide 4400 user guide

Dranetz

Dranetz PowerGuide 4400 user guide

Storch ProDist smart manual

Storch

Storch ProDist smart manual

Bosch GAM 270 MFL Original instructions

Bosch

Bosch GAM 270 MFL Original instructions

Lippert Tire Linc OEM INSTALLATION MANUAL

Lippert

Lippert Tire Linc OEM INSTALLATION MANUAL

Escort iMiniPlus user manual

Escort

Escort iMiniPlus user manual

TE Connectivity Integra Ri3 Installation and operating instructions

TE Connectivity

TE Connectivity Integra Ri3 Installation and operating instructions

Philips Joey Dosimeter Instructions for use

Philips

Philips Joey Dosimeter Instructions for use

Newtons4th PSM1700 Communications manual

Newtons4th

Newtons4th PSM1700 Communications manual

SCENTROID Scentinal SL50 manual

SCENTROID

SCENTROID Scentinal SL50 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.