
3.8. Setting volume and time
If a specific volume is to be infused in a given time the RATE display must be left
at zero. When the syringe has been primed, press the START/STOP key (7) to
confirm. Then press and hold the OPTION key (5) until the RATE and VOLUME
displays flash. The time in hours and minutes may be selected in the RATE
display (9) and the volume selected in the VOLUME display (10). The pump
automatically calculates the infusion rate. Check this carefully in the LCD window
(6) before starting the infusion.
3.9. Pressure system
The Syramed SP6000 has automatic pressure monitoring whereby the pressure
in the system is measured via the syringe plunger. The alarm pressure limit can
be set automatically or manually.
Automatic setting:
If the pump is configured for this mode, the alarm pressure limit is automatically
matched to the set rate, the lower the rate, the lower the alarm pressure limit.
Example (Injectomat):
- 50 ml syringe, 25 ml/h, press. limit 800 mBar, Time to alarm: 160 sec.
- 10 ml syringe, 5 ml/h, press. limit 300 mBar, Time to alarm: 100 sec.
Manual setting:
Press the OPTION key (5) sequentially to display pressure and alarm pressure
limit in the LCD window (6). Hold down the OPTION key (5) until the VOLUME
display (10) flashes "Lxxx". The pressure limit may be manually set using the
VOLUME keys (13) in the VOLUME display (10) and the data in the LCD window
changes accordingly. This can also be done while the infusion is in progress.
NOTE: Manual setting of pressure deactivates the automatic pressure setting, i.e.
the pressure remains at the current level independent of the rate selected.
Example (Injectomat):
- 50 ml syringe, 25 ml/h, press. limit 500 mBar, Time to alarm: 100 sec.
- 10 ml syringe, 5 ml/h, press. limit 200 mBar, Time to alarm: 70 sec.
If the pressure rises beyond the limit set, the pump stops and the stored bolus is
automatically reduced to virtually zero volume. An audible and visual alarm is
activated. Check the IV carefully for the cause of the alarm. Do not restart the
pump until the occlusion is released.
3.10. Setting time and date
Press the OPTION key (5) sequentially to display date and time in the LCD
window (6). Hold the OPTION key (5) down until the display flashes. The time
may be set using the volume keys (13) in the Volume display (10), e.g. h9.45 =
9:45 am. This can also be done while the infusion is in progress.
The syramed has the possibility to automatically adjust the daylight save time
(summer time). The adjustments can be done as per EU, US or Australian
regulations. If the text ‘Clock !’ should appear, replace the Lithium backup battery
on the main PCB.
To set the date, first switch the pump off. Press the VOLUME 0.1 ml DOWN and
VOLUME 100 ml DOWN keys (13) together whilst switching the pump on. This
enables the Service Mode. Select the RATE display (9) according to the
following table using the RATE keys (12). Then select the corresponding data in
the VOLUME display (10) using the VOLUME keys (13). Press the START key (7)
each time to confirm each setting:Page 7