ARJAY ENGINEERING HYDROSENSE 4410-LMP User manual

MODEL: HYDROSENSE 4410-LMP
OIL IN WATER MONITOR
USER MANUAL (REV: 1.2)
ARJAY ENGINEERING LTD.
2851 Brighton Road
Oakville (Toronto), Canada
L6H 6C9
Tel: ++1 (905) 829-2418
Fax: ++1 (905)829-4701
North America: 1-800-387-9487
Web: www.arjayeng.com
Email: arjay@arjayeng.com

This page intentionally left blank

HYDROSENSE 4410-LMP Revision 1.2 P age 3|29
TABLE OF CONTENTS
NOTICE
Please read the HydroSense Installation Notes (3.1) prior to locating
and mounting the enclosures.
SPECIFICATIONS....................................................................................................................................5
1.0USE HAZARD INFORMATION.......................................................................................................6
2.0INSTRUMENT OVERVIEW.............................................................................................................7
2.1
FEATURES..................................................................................................................................................... 7
2.2
DESCRIPTION................................................................................................................................................ 7
2.3
INTERFERENCES
AND
AFFECTS
TO
ACCURACY..................................................................................... 8
2.4
ROUTINE
CLEANING
PROGRAM ................................................................................................................. 9
3.0INSTALLATION............................................................................................................................10
3.1
HYDROSENSE
INSTALLATION
NOTES..................................................................................................... 10
3.2
GLOSSARY
OF
SYMBOLS .......................................................................................................................... 11
3.3
UNIT
INSTALLATION ................................................................................................................................... 12
3.4
PERMANENT
POWER
CONNECTION
(AC
POWERED
MODELS
ONLY).................................................. 14
3.5
ELECTRICAL
INSTALLATION...................................................................................................................... 14
4.0STARTUP AND SETTINGS..........................................................................................................15
NOTES
ON
VALUE
ENTRY.......................................................................................................................... 15
PASSWORD
PROTECTION......................................................................................................................... 16
POWERUP
DISPLAY.................................................................................................................................... 16
5.0CONTROL SETUP (<CONTROL> KEY).....................................................................................17
5.1
RELAY
SETTINGS
<1>................................................................................................................................. 17
5.1.1
SETPOINTS <1>............................................................................................................................ 17
5.1.2
TIME DELAY ON <2> .................................................................................................................... 17
5.1.3
RELAY ENABLE / DISABLE <3> OR <4>..................................................................................... 17
5.1.4
RELAY FAILSAFE <5> .................................................................................................................. 18
5.2
M
A
OUTPUT
SETTINGS
<2>........................................................................................................................ 18
5.2.1
4-20 mA OUT OR 0-20mA OUT <2>............................................................................................. 18
5.2.2
mA DIRECT OR INDIRECT <3>.................................................................................................... 18
5.3
TOOLS
<3>................................................................................................................................................... 18
5.3.1
RELAY TEST <1>.......................................................................................................................... 18
5.3.2
mA OUTPUT TEST <2>................................................................................................................. 18
6.0CALIBRATION SETUP (<CALIB> KEY).....................................................................................19
6.1
BEFORE
CALIBRATION............................................................................................................................... 19
6.2
CALIBRATION .............................................................................................................................................. 20
6.2.1
SELECT # OF CALIBRATION POINTS <1>.................................................................................. 20
6.2.2
AUTOMATIC CALIBRATION......................................................................................................... 20
6.2.3
MANUAL CALIBRATION WITH LABORATORY RESULTS.......................................................... 21
6.2.4
LIGHT REFERENCE AND TEMPERATURE SETTING................................................................ 21
6.2.5
CHANGE CALIBRATION LOCATION ........................................................................................... 21
6.2.6
CALIBRATION PERIOD................................................................................................................. 22
7.0SETUP (OPERATION AND DIAGNOSTICS) (<SETUP> KEY)..................................................23
7.1
SYSTEM
SETTINGS
<1>.............................................................................................................................. 23
7.1.1
NET ID <1> .................................................................................................................................... 23
7.1.2
FILTER <2>.................................................................................................................................... 23
7.1.3
UNITS <3> ..................................................................................................................................... 23
7.1.4
PASSWORD <4>........................................................................................................................... 23

HYDROSENSE 4410-LMP Revision 1.2 P age 4|29
7.1.5MORE <5> ..................................................................................................................................... 23
7.1.6mA OUTPUT TRIM <1>................................................................................................................. 24
7.1.7DECIMAL POINTS <2>.................................................................................................................. 24
7.2DIAGNOSTICS <2> ...................................................................................................................................... 24
7.2.1PROCESS VALUES <1> ............................................................................................................... 24
7.2.2CALIBRATION DATA <2> ............................................................................................................. 24
7.2.3OTHERS ........................................................................................................................................ 24
7.2.4SN / HWREV <4>........................................................................................................................... 24
7.3DATE /TIME SETUP <3>.............................................................................................................................. 24
8.0TROUBLESHOOTING..................................................................................................................25
9.0CONTROLLER SETTING SHEET................................................................................................26
10.0SAMPLE PREPARATION (FOR CALIBRATION OR TESTING).............................................27
11.0PERIODIC TESTING AND MAINTENANCE ............................................................................28
11.1PERIODIC TESTING ................................................................................................................................ 28
11.2FLOW PLATE CLEANING ........................................................................................................................ 29
11.3LAMP REPLACEMENT............................................................................................................................. 29
11.4SPARE PARTS LIST................................................................................................................................. 29
TABLE OF FIGURES
FIGURE 1 – SECURE LAMP INSIDE THE LAMP BOX........................................................................12
FIGURE 2 – FLOW PLATE SECURE AGAINST TRAY........................................................................13
FIGURE 3 – BLUE LAMP BOX SECURE AGAINST TRAY..................................................................13
FIGURE 4 - USER INTERFACE.............................................................................................................15

HYDROSENSE 4410-LMP Revision 1.2 P age 5|29
SPECIFICATIONS
Specifications are subject to change without notice:
Specification Details
Power Input: 100 - 230VAC (85-264VAC)*, 50/60Hz (47-63Hz)*, 1.26A – 0.62A, 2A Fuse
OPTIONAL: 24 VDC @ 3.5 A max
*Operational Range
User Interface:
Display Four line LCD with simultaneous display in PPM, current date and calibration
due
Communication Interface: RS-485 Modbus, optional HART or Fieldbus Foundation module for uni-
directional communication of ppm.
Relays / Analog Outputs:
Relay Outputs 6A DPDT Slave relays are available, dry contacts with LED panel indication
R1 & R2: Setpoint Alarm Relays: User settable alarm points and delay time
(0-99 seconds delay on)
R3: Maintenance relay- Maintenance is required (See Troubleshooting
section 8.0)
R4: Fault alarm relay – Service is required (See Troubleshooting section 8.0)
mA Signal Output 4-20 mA DC, 900 Ohms, isolated, field scalable
Instrument Performance:
Measuring Range 0 - 5000 ppm Hydrocarbon in Water
Instrument Accuracy ± 0.1 PPM
Process Accuracy +/- 1.0 ppm typical
The process accuracy is reflected by the site calibration to a known
hydrocarbon concentration and a stable background water. Changes in the
hydrocarbon make-up and variations in the process may affect the
instrument output.
Sensitivity 145 PPB (diesel reference) & 463 PPB (crude oil reference)
Calibration A library of up to 10 calibrations with up to five concentration entry points per
calibration to maximize accuracy
Signal Filtering 20-1000 samples/average
Environmental:
Ambient Temperature 5-55 C
Protect from direct sun or rain. Instrument shelter or indoor use is
recommended. Higher temperatures may be accommodated with air
conditioning.
Relative humidity Up to 90% (non-condensing)
Process Requirement:
Process Sample
Temperature 0-40 C without cooler; above 40 C with cooler.
Inlet Flow Rate Minimum: 1.0 L/M (continuous and stable)
Optimum: 3.0 to 5.0 L/M (continuous and stable)
Inlet pressure Minimum 2 psi, maximum 1000 psi, *minimum 20 psi when equipped with
optional cooler,
*For safety reasons a reducing valve should be used for pressure above 100 psi.

HYDROSENSE 4410-LMP Revision 1.2 P age 6|29
Specification Details
Mechanical Specification:
Enclosure Dimensions 12.0"W x 37.5"H x 9.25"D (305mm W x 953mm H x 235mm D)
Sample Inlet 3/8” NPT female
Sample Outlet 2” NPT male (outfall must be unrestricted gravity to drain)
Weight 33Kg (73 lbs)
Enclosure Rating Type 4X, IP65, 316 Stainless Steel with viewing window
Approval Standards: CSA SPE1000 and Canadian Electrical Code (CEC)
USA National Electrical Code (NEC)
1.0 USE
HAZARD
INFORMATION
CAUTION Indicates a potentially hazardous situation that may result in minor or
moderate injury.
WARNING Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
DANGER Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
NOTICE Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.

HYDROSENSE 4410-LMP Revision 1.2 P age 7|29
2.0 INSTRUMENT OVERVIEW
2.1 FEATURES
Fluorescence technology
Non-contact sensor
Continuous On-Line monitoring
Instantaneous readings
Multi-point automatic or manual calibration
Temperature and light degradation compensation
4 relays (DPDT 6A contacts) (R1 and R2 for PPM alarms, R3 and R4 for maintenance alarm),
Isolated 4-20 mA output, RS-485 Modbus output, optional HART and Fieldbus Foundation
LCD display in PPM
No moving parts, complete maintenance without turning off the sample stream
2.2 DESCRIPTION
The HydroSense Oil in Water Monitor from Arjay Engineering Ltd. has been designed for municipal and
industrial applications to measure PPM levels of petroleum hydrocarbons in aqueous solutions. Typical
applications include PPM trace amounts of oil in effluent water from storm water runoff, oil in cooling
water, produced water, and oil/water separators. Other measurements and mediums can be monitored
on request (i.e. colorants in fluids, etc.).
A continuous sample is directed into the sample chamber using a pumped or process pressure source.
A stable flow rate is required. The sample is released from the sample chamber by a gravity flow to a
drain or sump.
The sample flow is dispersed evenly down an 88 mm x 200 mm flow plate. This dispersion
accomplishes two favorable results. Firstly, the sample is spread over a wide area, providing a large
surface area for the ultraviolet light to penetrate. This results in an increased excitation of the oil
molecules. Secondly, the gravity flow against the flow plate minimizes the sample depth against the
flow plate. This reduces the effect of suspended solids interference. Fewer oil molecules can ‘hide’ from
the light source.
After the sample leaves the flow plate, it is allowed to gravity flow to the outlet port. The sample outfall
must not be obstructed from gravity flow. For purge application a U-Trap would be required.
The ultraviolet light source is positioned directly in front of the sample flow. The receiver is positioned at
an angle to the flow direction. Both the emitter and the receiver are equipped with precision light filters
to control the wavelengths of the ultraviolet light being emitted and the fluoresced light being received.
A relationship between the measured fluoresced light and the amount of oil in the sample is
mathematically predictable over the measurement range of the instrument. The precision light filters
maximize the predictability over alternative non-filtered methods of measurement.
The sample tray is easily accessed for any necessary cleaning of the flow plate. The ultraviolet light
source is easily accessible for replacement as required.
For periodic testing and calibration, a 3-way valve in the inlet line is provided to manually input a fresh
water source to confirm the instrument response, zero and clean the instrument.

HYDROSENSE 4410-LMP Revision 1.2 P age 8|29
2.3 INTERFERENCES AND AFFECTS TO ACCURACY
The UV fluorescence technique monitors the intensity of light emitted from the passing stream at a
selected wavelength band.
This technique can be quite selective by eliminating the light affect of compounds in the water that do
not share the same fluorescence characteristics of hydrocarbons.
1. When chemical compounds in the water are excited with light energy, only certain compounds will
emit the light back out of the water at a higher wavelength than excited with. These are referred to
as fluorescing compounds. The HydroSense does not respond to most chemicals because it only
responds to fluorescing compounds, of which aromatic hydrocarbons are included.
2. The light used to excite the compounds is filtered to 254 nm +/-. Of all the fluorescing compounds
only certain ones will respond to this wavelength. Some respond to higher and some to lower
wavelengths. This filter narrows the HydroSense response to only those that fluoresce from 254 nm
+/-.
3. This limited number of compounds that do fluoresce from 254 nm light may emit light at any number
of wavelengths such as 290nm, 310 nm, 350 nm 480nm, etc. Aromatic hydrocarbons happen to
fluoresce at approximately 350 nm. By filtering the light sensor from all light except 350 nm +/-, only
compounds that emit light at 350 nm +/- are indicated at the receiver.
4. Oil and Grease in water may be made up of hundreds or thousands of different hydrocarbon
compound structures. The aromatic compounds are fluorescing compounds. The proportion of
aromatics within the total hydrocarbons is generally consistent in a product or process. The
aromatics are therefore used as a tag to correlate the monitor to total hydrocarbons in water.
Changing Oil Types and Sources
Different oils have a different make-up of compounds and the fluorescing strength may vary between oil
types. For instance, diesel fuel may fluoresce much stronger than transformer oil. If the HydroSense is
calibrated using 100 PPM of diesel, 100 PPM of transformer oil may only give a display reading of 50
PPM.
Crude oil may vary from one well to another, lubricating oils from different manufacturers may vary in
their make-up; oils may be dissolved or free, and so on.
The calibration is therefore site selective and should be done using actual process water or with
samples of oil that are to be targeted by the monitor.
The calibrated accuracy relies on the oil type and conditions being consistent. The HydroSense will
respond positively to aromatic hydrocarbons but the display accuracy may be affected by variations in
the types and sources of these hydrocarbons.
Other Chemicals in the Water
The light sensor is selective to compounds in the water that emit light at 350 nm when excited from 254
nm light. If there is a background chemical in the water that fluoresces at these wavelengths, the
HydroSense will respond to them.
If this background chemical concentration is consistent, this interference will be zeroed out during
calibration. Calibration is recommended using process water so that any background interferences are
zeroed out.

HYDROSENSE 4410-LMP Revision 1.2 P age 9|29
If an interfering background chemical changes in concentration, the HydroSense will sense this change.
Consideration to this affect is important for alarms and recording. Filtering of the water, changes to
chemical use, or special light filtering may be required to provide more stable readings.
The periodic introduction of fluorescing chemicals into the water may also affect the reading. During
these conditions, operators should be acknowledged that nuisance alarms may occur. Soap
manufacturers will often include fluorescing dyes in the product for appearance and identification.
Green dyes are typical in industrial degreasers and commercial soaps. Fluorescing chemicals are often
included in detergents to enhance the visual affect of a cleaned product such as clothes.
Not all of these commercial dyes will affect the wavelengths of the HydroSense, however, green dyes
have proven to be a common interference.
Suspended Solids and Turbidity
The unit is calibrated to a passing stream of water. The amount of light fluoresced by the aromatic
hydrocarbons determines the calibration parameters. The light received by a hydrocarbon and then
sent to the receiver is based on a stable light path through the water. If suspended solids or turbidity
block the light getting to the hydrocarbon, light cannot be fluoresced back to the sensor. Readings can
be dampened by an increase in solids or turbidity. When process water is used during the calibration,
the offset affect of solids is taken into account and zeroed out.
The design of the large surface sensing area verses the small sensing depth minimizes the affect of
turbidity in the HydroSense. In effect, the hydrocarbons have little place to hide behind solids. In
circumstances of dramatic changes in turbidity, sample-conditioning techniques prior to the
HydroSense should be considered.
Temperature Compensation
Temperature can affect the light sensor. Temperature compensation is built into the unit. The
temperature compensation coefficient will be preset at the factory.
2.4 ROUTINE CLEANING PROGRAM
The HydroSense relies on a constant flow of water across the sensing plate. Excessive particulates and
algae in the water can build up on the flow plate and in the overflow tray. This will eventually affect the
performance of the unit.
Setting up a Routine Cleaning Program is vital to the successful performance of the unit.
Each application will vary in the frequency of cleaning. Some may require daily wipes and some may
require monthly cleaning. A basic wipe down of the flow plate can be done without having to shutdown
the stream or power. The wipe procedure will take 2 to 3 minutes.
To set up a schedule, it is recommended to program a daily wipe of the flow plate using a clean paper
towel. After two weeks of daily cleaning, determine if every other day may be adequate. If so, set this
program in place for two weeks. Slowly extend the frequency between cleanings until an adequate
program frequency is determined for your individual site conditions.
Cleaning the flow plate and plastic surface should be done with clear non-fluorescing chemicals. An
appropriate flow plate cleaner is Windex brand “anti-fog”. Low concentrate Muriatic acid (18%) or
Isopropyl Alcohol are also effective. Use only chemicals approved for your site, personal safety and
disposal. Be sure to thoroughly rinse the flow plate with clean water before installing.
CAUTION Repeated visual contact with the light source can be harmful. Avoid
looking directly at the ultraviolet light source. Wear UV protection.

HYDROSENSE 4410-LMP Revision 1.2 P age 10|29
3.0 INSTALLATION
NOTICE If any damage to the instrument is found, please notify an Arjay
Engineering representative as soon as possible prior to installation.
NOTICE Qualified Personnel must undertake all installations.
WARNING If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
3.1 HYDROSENSE INSTALLATION NOTES
NOTICE Read these notes before installation.
The system is comprised of two main components, the Sample Chamber and the Controller.
1) The Sample Chamber receives the flow sample from the process and outputs it to drain. The
chamber should be wall or rack mounted on a vertical and horizontal plane to allow a proper flow
through the sensing unit.
2) The Sample Chamber should be located close to the process to reduce the lag time of the sample
to the unit. This will offer more instantaneous readings and real time recording.
3) The outlet gravity flows to drain, and consideration of a close proximity to the drain is important.
Also, mount the unit where it is readily accessible for maintenance and periodic testing by manual
insertion of known samples.
4) If the process flow is not under pressure, the chamber should be mounted below the process level
so the sample can free flow down and through the unit. A pump may be used. The maximum input
pressure is 1000 psi although a reducing valve to less than 100 psi is recommended. The minimum
input flow rate is a continuous and stable 1 liter/minute. 3 – 5 liter/minute is recommended.
5) The inlet connection to the unit is a 3/8” female thread. A barb connection may be threaded to this
when flexible inlet tubing is used. Clear flexible 3/8” or 1/2” inlet tubing is suggested for the inlet
sample. This will provide a visible indication of the sample, as well as an indication of contaminant
build-up. To minimize the contamination in the tubing, Teflon lined tubing may be desirable.
Hazardous Locations or local regulations will dictate materials to be used.
NOTICE Clear tubing should not be used outdoors where algae build-up from
sunlight is increased.
6) An on/off valve at the process is recommended to shut down the system for maintenance and/or
sample tube replacement. (An on/off valve is included within the unit for throttling flow. This can be
used for internal maintenance). A 3-way valve is included at the inlet for a fresh water input for
zeroing, cleaning, and testing.
7) The sample gravity flows out of the Sample Chamber. The outlet tube must only be installed in a
downward vertical or downward graded horizontal direction. Any excessive upward direction will
cause the sample to back up and could flood the sample chamber.
8) The outlet fitting is a 2” male thread. Do not reduce this size. This will cause a restriction and flood
the system.

HYDROSENSE 4410-LMP Revision 1.2 P age 11|29
9) For non-purged/pressurized models the outlet of the tube should be open to air, not submerged in
water or a process which would cause a backpressure. This could result in a restriction of the
effluent flow and spillage within the lower chamber.
For purged/pressurized models, a slight and constant backpressure is required to maintain a
cabinet pressure. To accomplish this, submerge the effluent tubing approximately 10 cm below the
water surface or run the outlet tube with a slight upward grade or drain trap prior to the downward
outfall. Be sure the tube never rises higher than the bottom of the chamber outlet.
10)The Sample Chamber must be mounted indoors or in a heated housing when sample freezing may
occur. The inlet and outlet tubing must not be exposed to freezing environments. For outdoor
installation, the unit must be sun and rain shielded.
11)The controller operates using 100-230VAC (85-264VAC)*, 50/60Hz (47-63Hz)*, 1PH, 1.26A –
0.62A or 24 VDC as ordered. (*Operational Range)
12)The monitor provides LED indication of the relay status. The relays are dry contacts and will accept
AC or DC inputs. Setpoint alarm levels are on Relay 1 & 2. Relay 3 is maintenance alarm, which
indicates offset drift and that cleaning, or re-calibration is necessary. Relay 4 indicates a Lamp
Replacement requirement or controller failure.
13)A 4-20 mA DC output signal proportional to the PPM level is provided. This is a signal capable of
driving 900 ohms. Remote indicators, receiving devices and their distances should be considered
when choosing a location for the Arjay Controller.
14)A RS-485 Modbus output is provided. This can be used to link to computers and other equipment.
Optional HART and Fieldbus Foundation are available.
15)Shielded wiring is required for the output alarms and signals to avoid EMI and RFI interference from
other equipment near the sample unit.
16)The Monitor is housed in a Type 4X 316 Stainless Steel enclosure. Extremes in temperature and
humidity should be avoided. Indoor or an environmentally mounted instrument shelter is
recommended.
3.2 GLOSSARY OF SYMBOLS
Attention, consult accompanying documents
Attention, veuillez consulter les documents ci-joints.
Protective Earth
Terre de protection Fuse
Coupe-circuit; fusible
Direct Current (DC)
Courant continu Normally open relay contacts
Contacts travail
Normally closed relay contacts
Contacts Repos Power off
ArróÕ (mise hors tension)
Power on
Marche (mise sous tension) LLive
Sous tension
NNeutral
Neutre GGround
Terre

HYDROSENSE 4410-LMP Revision 1.2 P age 12|29
3.3 UNIT INSTALLATION
1) Locate an area that is environmentally protected from wide variances in temperature and
humidity. Indoor installations are recommended.
When selecting the location, consider that regular maintenance and testing is desirable for the
proper and accurate operation of the instrument.
If the sample input is not from a pumped source, locate the instrument in a position that will
receive a continuous representative sample from the process stream. The farther from the
stream, the greater the lag time of readings vs. actual process variance.
2) Mount the enclosure on a vertical wall or support of minimal vibration. The sample will be
flowing over a Flow Plate. A bubble level is provided in the sample chamber. Be sure the unit is
mounted level. This is necessary for a proper flow across the Flow Plate.
3) Open the Blue Lamp Box and secure the new lamp into place with the metal band and knurled
screw. Make sure the lamp is seated between two tabs of metal band as per picture below.
Figure 1 – SECURE LAMP INSIDE THE LAMP BOX
4) Plug the lamp terminal into the socket. Close the Blue Lamp Box.
TABS
SEATED BETWEEN TABS
KNURLED
SCREWS
METAL
BAND

HYDROSENSE 4410-LMP Revision 1.2 P age 13|29
5) Insert the flow plate as shown below. The flow plate has a special reflective core and must be
placed in the Tray with the etched "UP" side facing out (toward you). Snug the knurled screws to
hold the flow plate in place.
Figure 2 – FLOW PLATE SECURE AGAINST TRAY
6) Place the Blue Lamp Box against the Tray and secure it with the retaining latch.
Figure 3 – BLUE LAMP BOX SECURE AGAINST TRAY
7) Be sure the 3 way valve with grab sample take off valve (supplied loose) is installed as per
drawings at the back of this manual. Make sure the inlet sources (process sample and clean
water) are valved for future maintenance and testing operations. Connect the inlet sources to
the 3/8” inlet of the 3 way valve. DO NOT TURN ON THE INLET SOURCES.
Connect the outlet source to drain or a sump.
LATCH
TRAY
LAMP BO
X
KNURLED
SCREWS
KNURLED
SCREWS
ETCHED
“UP” SIGN
FLOW PLATE SEAT
ON TRIANGULAR
PIECES

HYDROSENSE 4410-LMP Revision 1.2 P age 14|29
For non-purged/pressurized models the outlet of the tube should be open to air, not submerged
in water or a process, which would cause a backpressure. This would result in an overflow
inside the chamber.
For purged/pressurized models, a slight and constant backpressure is required to maintain a
cabinet pressure. To accomplish this, submerge the effluent tubing approximately 10 cm below
the water surface or run the outlet tube with a slight upward grade prior to the downward outfall
such as a U-Trap. Be sure the tube never rises higher than the bottom of the chamber outlet.
8) Note that the outlet is a gravity feed line only. The outlet tube must NOT rise above the bottom
of the lower chamber or be connected to a pressure process line.
See video on Arjay Enginnering Ltd. website (www.arjayeng.com) for more information in using
HS4410-LMP Oil in Water Montior.
3.4 PERMANENT POWER CONNECTION (AC POWERED MODELS ONLY)
1) Connection to the building wiring system shall be in accordance with the Canadian Electrical
Code (CEC), Part 1 in Canada, the National Electrical Code, ANSI/NFPA 70 in the USA, or the
local electrical codes of the country where the equipment is being installed.
2) An external mains switch or external over-current protection / circuit breaker device is required
as a disconnect device. This mains disconnect device shall be specified as complying with the
requirements of IEC 947-1 and IEC 947-3.
3) The external mains switch shall be in close proximity to the equipment and within easy reach of
the operator. The switch shall be marked as the disconnecting device for the equipment and
include the symbols to its “ON” and “OFF” positions using the following symbols:
Power Off Power On
4) The wiring for AC power should be 14 – 16 AWG / 300V or as required by local / country codes.
5) After field wiring, the primary wires must be secured to the enclosure by tie-wraps to maintain
the separation from the signal wires.
6) The equipment is suitable for connection to a 15A protected branch circuit.
7) Wiring diagram for permanent connection: See drawings at the back of this manual.
8) Use copper conductors only.
3.5 ELECTRICAL INSTALLATION
See drawings at the back of this manual.

HYDROSENSE 4410-LMP Revision 1.2 P age 15|29
4.0 STARTUP
AND
SETTINGS
NOTICE Calibration must be performed after installation and any lamp
replacement.
Figure 4 - USER INTERFACE
NOTES ON VALUE ENTRY
When entering in numeric values, the cursor can be backspaced to correct mistakes by pressing the
DISPLAY key. This is only true if the cursor is not at the beginning of the displayed value, in which case
the DISPLAY menu is entered.
The decimal point is the dot ( ) key.
Values may be entered with any number of decimal places.
If the entered value is out of the allowed limits, the system displays the limiting value for 2 seconds. For
example, if the alarm time delay value is entered as 5000 seconds, then MAX. 99 is displayed for 2
seconds, and then entry is allowed again. The current value is not changed unless the entered value is
within limits. During value entry, the oil concentration is constantly updated in the background.

HYDROSENSE 4410-LMP Revision 1.2 P age 16|29
PASSWORD PROTECTION
A password must be entered to access any of the 3 value entry menus (CALIB, CONTROL, SETUP> 1-System)
from the normal operating display menu. The factory default password is 2000. A customer password may be
added from the setup menu as described in section 7.1.4 Settings. The prompt for entering the password is
always 9999 regardless of the actual password.
POWERUP DISPLAY
After mechanical and electrical installation of the sample chamber and the controller have been
successfully completed, power up the unit. The LCD should show a similar screen (Normal Operating
Display):
O i l i
n
W
a
t
e
r
M
o
n
i
t
o
r
0
p
p
m
2 0 1 5
/
0
9
/
1
5
1
6
:
2
8
h
r
D a y s
t
o
C
a
l
:
1
8
0
NOTE: The shown values are for example only.
The 2nd line shows the oil concentration value in PPM.
The 3rd line shows the current date and time
The 4th line displays the amount of days left until next calibration.
The main power supplies in upper chamber should also show a Green power light.

HYDROSENSE 4410-LMP Revision 1.2 P age 17|29
5.0 CONTROL SETUP (<CONTROL> KEY)
To access the control setup settings, you will require a password. The password is “2000”.
C o n
t
r
o
l
S
e
t
t
i
n
g
s
1 - R e l
a
y
S
e
t
t
i
n
g
s
2 - m A
O
u
t
p
u
t
S
e
t
t
i
n
g
s
3 - T o o
l
s
5.1 RELAY SETTINGS <1>
Four relay alarm points are available for remote alarm. Of these, two are general-purpose alarm relays
with user settable alarm points, dead band (differential alarm points), and time delay. The remaining
two relays are to indicate Maintenace alarm (R3) and Fault alarm (R4).
If the general-purpose alarm relays are to be used, press <CONTROL>, then press <1> for Relay
Settings Menu.
1 - S e t
p
o
i
n
t
s
2 - T i m
e
D
e
l
a
y
O
n
3 - E n a
b
l
e
4
-
O
f
f
O
N
5 - R e l
a
y
F
a
i
l
s
a
f
e
5.1.1 SETPOINTS <1>
Press <1> for setpoints. Enter the value in PPM for relay R1 to activate. Press <ENTER>. Now enter
the value for the relay R1 to de-activate and press <ENTER>. A small differential between Relay HI and
Relay LO may be desired to eliminate relay chatter if the PPM concentration is fluctuating at the alarm
level. Similarly, enter the high and low alarm points for the second relay R2 if it is to be used. Note: for
a single point alarm with no differential, enter the LO alarm value identical to the HI alarm value.
5.1.2 TIME DELAY ON <2>
To delay the relay alarm for a preset time (in seconds), press <2> for Enter On delay in the Relay
Setting Menu. This will suppress the alarm to eliminate a spurious momentary alarm that may be
caused by an oil globule or process variable not indicative of an alarm condition.
5.1.3 RELAY ENABLE / DISABLE <3> OR <4>
The relays may be disabled from operating for maintenance purposes. Confirm the relays are activated
on the right side of line 3 in the RELAY SETTINGS menu. If the display reads ON, the relays are
activated. Press <3> to enable the relays or press <4> to disable the relays. As a reminder to turn the
relays back on after maintenance the main display will show relays disabled on line 3.

HYDROSENSE 4410-LMP Revision 1.2 P age 18|29
5.1.4 RELAY FAILSAFE <5>
The relays may be set as Failsafe or Non Failsafe. In failsafe mode the relays are energized under
normal operation.
Failsafe typically means that the relay is held in an energized state when in a normal operating
condition as opposed to an alarm condition. In an alarm condition, the relay de-energizes which is
identical to when the instrument power is shut off. The rationale is that the alarm condition should
match the Power Fail condition. The factory default is Yes for Failsafe.
5.2 mA OUTPUT SETTINGS <2>
Determine the desired operating range of the instrument. This will reflect the 4 to 20 mA output range of
the instrument. For example, a range of 0 to 50 PPM may be desirable. The control signal of 4-20 mA
will represent 0-50 PPM. Press <CONTROL> on the keypad, enter the password at the prompt, then
press <2> for mA output settings. Press <1> for mA Zero / span setup. The display will prompt you to
enter the zero point in PPM. This will typically be 0.0. Enter the value and press <ENTER>. The display
will prompt you to enter the span value (i.e., 50.0). Enter the desired value and press <ENTER>.
S
e
t
m
A
O
u
t
p
u
t
1 = m A
Z
e
r
o
/
S
p
a
n
S
e
t
u
p
O u t : 4
-
2
0
m
A
2
=
0
-
2
0
m
A
m A = D I
R
3
=
I
N
V
5.2.1 4-20 mA OUT OR 0-20mA OUT <2>
The mA output signal can be set for 4-20mA or 0-20mA. The third line of display will show the present
setup (Out: 4-20mA). Press <2> to change the desired setting.
5.2.2 mA DIRECT OR INDIRECT <3>
The mA output signal can be set for direct acting (4-20mA = 0-50ppm), or inverse acting (20-4mA = 0-
50ppm). The lower left display will read the present setup (mA: DIR). Press <3> to change the desired
setting.
5.3 TOOLS <3>
5.3.1 RELAY TEST <1>
This feature allows the user to toggle ON and OFF each relay for maintenance purpose. Press the #
keypad for each corresponding relay. (e.g.1=R1). Press “Enter” To Exit this menu.
5.3.2 mA OUTPUT TEST <2>
This feature allows the user to check the mA output at 4mA <1> and 20mA <2> OR Enter a desired
value <3>. This is for maintenance purpose only.

HYDROSENSE 4410-LMP Revision 1.2 P age 19|29
6.0 CALIBRATION SETUP (<CALIB> KEY)
NOTICE To access the calibration settings, you will require a password. The
password is ‘2000’.
NOTICE Calibration must be performed after installation and any lamp
replacement.
NOTICE Allow a minimum of 24 hours of power up prior to calibration to allow
the lamp to stabilize.
Using actual process stream flows for the calibration is desirable to provide the best accuracy. This will
zero out any background influences that may be present in the process water. At least two points of
entry at different contaminant levels are required. These levels may initially be unknown for calibration
purposes and entered in manual calibration at a later date after laboratory results are returned. One of
these points may be clean process water (0 PPM hydrocarbons).
Prepared samples may be used for calibration and testing although variations in personal blending
techniques, the source of the hydrocarbon and background water, the container used, and the retention
time prior to use will all play a part in the concentration reliability and repeatability.
6.1 BEFORE CALIBRATION
Open the sample chamber door. Lift the Lamp Blue Box off the tray and place it on the door brackets.
Check that the flow plate is resting securely in place against the tray. The flow plate has a special
reflective core and MUST be placed in the tray with the etched “UP” side facing out (toward you).
Slowly open the process stream valve to allow a steady and even flow over the flow plate. Check that
the outfall is draining well and water is not building up in the drain tray. Verify that the flow plate is
evenly and completely covered with the flowing water. The sample flow may not initially cover the whole
flow plate by itself. Stir up the flow plate by rubbing the flow plate so the water sheens evenly and
completely across the flow plate as it flows. Place the Lamp Blue Box back against the tray and secure
it with the retaining latch. Close the chamber door. Make sure the Sample Chamber Door is closed
tightly. The door MUST be closed to eliminate any background light interference during and after
calibrations.

HYDROSENSE 4410-LMP Revision 1.2 P age 20|29
6.2 CALIBRATION
C A L I B
R
A
T
I
O
N
1 - S e l
#
o
f
C
a
l
P
t
s
2 - A u t
o
3
-
M
a
n
u
a
l
4 - C a l
P
e
r
i
o
d
6.2.1 SELECT # OF CALIBRATION POINTS <1>
Press the <CALIB> , enter the password at the prompt then Press <1> for SEL # of cal points. The
fourth line shows the current selection. Enter # of points if change is required then press <ENTER>.
For Calibration, the Hydrosense 4410-LMP can accept up to 5 sample points to draw a calibration
curve. The factory default is 2 points where 0ppm (clean process water) and a known ppm level is
used. A calibration curve using more than 2 points is used because some samples may not be linear as
concentrations increase.
S E L E C
T
C
A
L
P
O
I
N
T
S
E n t e r
#
o
f
C
a
l
P
t
s
M i n P
t
s
:
2
M
a
x
p
t
s
:
5
C u r r e
n
t
P
t
s
:
2
6.2.2 AUTOMATIC CALIBRATION
Press <2> for automatic calibration.
A U T O
C
A
L
C a l p
o
i
n
t
1
o
f
2
E N T E R
=
D
o
C
a
l
P
o
i
n
t
1
1 = A b o
r
t
Press <ENTER> to proceed to 1st calibration. Normally, the 1st point is clean process water (0.0 ppm).
Manually put the handle of the 3 way valve to the “up” position and let the clean process water flow
through the Hydrosense 4410-LMP. Enter the ppm value (0.0). Once the reading (FLR rdng) has
stabilized, press <ENTER> key. The display will now proceed to the second point.
A different concentration of contaminated water is required. If the process stream has a different
concentration than 0 ppm, put the handle of the 3 way valve to “down” position to allow the process
stream to flow through the Hydrosense 4410-LMP.
If the concentration of process stream is close to 0.0 ppm, a prepared sample (refer to Section 10.0)
and a sample pump will be used. Disconnect the inlet of clean water and connect the prepared sample
into the inlet. Pump the prepared sample through the Hydrosense 4410-LMP.
Enter the known or unknown*ppm value of stream concentration (i.e., 30.0 PPM). Once the reading
(FLR rdng) has stabilized, press the <ENTER> key.
Table of contents
Other ARJAY ENGINEERING Monitor manuals