ASHA Beam Profile PMB-201 User manual

Cover
ASHA Beam Prole
Thermal Power Meter Bundle
PMB-201

WWW.ASHA-CO.IR2
About Us
In order to design and manufacture programmable optometrical, op-
tomechanical and optoelectronical devices, the science-based compa-
ny, ASHA Beam Prole, was established in 2016.
ASHA has been able to design a wide range of devices, relying on
indigenous capability and knowledge. Also, since being calibratable is
essenal for measuring equipment, clients can carry out their period-
ical calibraon.
Field of activity
The main eld of acvity of this company is the design and produc-
on of laser, opcs and photonics equipment and can be categorized
in the following topics:
»Diode laser driver
»Laser diode stabilizer
»Laser Parameters measuring equipment such as Power Meter, Beam Pro-
ler, and Ultra-fast Detector
»Lithographic masks
»Linear posioner with stepper motor driver
»Calibraon of Power Meters and beam prolers
»Design and producon of soware to facilitate the use of electro-opc
and optomechanical programmable equipment

3WWW.ASHA-CO.IR
Main features
Introduction
Measuring the output of light sources is very important. For exam-
ple, the opcal power of medical lasers is very sensive and can cause
irreparable damage if the funcon of the device goes out of its proper
range. In experimental and research applicaons, it is always neces-
sary to measure the output of laser sources. In addion to the above,
it is crucial to pay special aenon to the necessity of calibraon of
industrial and military laser equipment.
Hence, the thermal power meter is designed for high precision laser
applicaons. Producons of this company can well meet the needs of
the scienc and industrial community in the eld of power metering.
The thermal power meter is designed for high power laser applica-
ons. The products of this company can well meet the needs of the
scienc and industrial community in the eld of power metering.

WWW.ASHA-CO.IR4
Features
»Direct connecon to the mulmeter with a special cable (depends on
order)
»Ability to operate in 3 dierent ranges (depends on order)
»Adjustable for desired light sources in the specied wavelength range
(depends on order)
»Able to have specic Gain for required scale (depends on order)
»Measuring the power of coherent opcal sources
»Measuring the power of connuous light sources
»Possibility to remove background radiaon (depends on order)
»Internal rechargeable baery for portable use
Applications
»Study incident and reected beam power and qualify light sources
»Study the eect of environmental changes on laser sources performance
»Study the variaon of power when passing through (transparent) envi-
ronments
»Examining the quality of lenses and opcal windows
»Measure the intensity of reecon/passing beam from dierent surfaces
Warning! With regard to the items listed in the requirements secon,
to maximize the accuracy of the device, it is strongly recommended to
connect the power meter earth to the standard one.

5WWW.ASHA-CO.IR
Technical Features
Item Min Typ Max Unit
Input Voltage 4.75 55.25 V
Input Current 60 250 500 mA
Baery -<2000 - mAh
Baery Life -20 - h
Charge Time - 5 - h
Wavelength Range 0.19 -20 µm
Opcal Power Working Range/
Display Power Range >10 -2×104mW
Output Voltage - - <2.5 V
Ranges 0.01 ~ 0.1999 0.01 ~ 1.999 0.1 ~ 19.99 W
Measurement Uncertainty ±5% + 100
Counts
±5% + 10
Counts
±5% + 5
Counts

WWW.ASHA-CO.IR6
More Spec. Discripon
Dimenon Detector 50x40x18 mm3
Console 180x165x70 mm3
Weight Detector < 300 gr
Console < 1000 gr
Detector Connecon DB9
Output Connecon BNC
Acve Detector Area 20x20 mm2
Input Aperture Ø10 mm
Display 2000 counts - 7segment
Figure 1.Thermal Power Meter Detector

7WWW.ASHA-CO.IR
Figure 2.Thermal Power Meter Console - Front View
Figure 3.Thermal Power Meter Console - Back View

WWW.ASHA-CO.IR8
Contents
About Us ............................................................................ 2
Field of acvity 2
Main features..................................................................... 3
Introducon 3
Features 4
Applicaons 4
Technical Feature 5
Power Meter Parts ............................................................. 9
Packaging Content 9
Detector Parts 10
Console Parts 11
Setup................................................................................ 13
Safety 13
Preparaon and measurement 14
Requirements 16

9WWW.ASHA-CO.IR
Power Meter Parts
Available Accessories
Please make sure that all accessories are available for the box in your
rst use.
»Thermal Sensor x 1
»Photodiode Power Meter Console x 1
»User Manual* x 1
»5V Power Supply x 1
»Mulmeter Connector Cable x 1
»M4 Screw x 1
»M6 Screw x 1
* Digital version of this user manual is available by scanning the QR-
code undwer the console.

WWW.ASHA-CO.IR10
Detector Parts
Figure 4. Thermal Power Meter Detector - Front View
1. Anodised Aluminum Body
2. M4 screw for vercal aligning
3. M6 screw for vercal aligning
4. Detector Screws(if open, guarantee void)
5. Input aperture

11 WWW.ASHA-CO.IR
Figure 5.Front Parts of Thermal Power Meter Console
1. Power range selector
2. Power range LED
3. 4 digits display for reporng
power
4. Charging status LED
5. ON/OFF Key
6. Input Power
7. Detector connector
8. Oset null volume
(depends on order)
9. Output (BNC)
10. To download datasheet scan
this QRcode
Console Parts
Figure 6. Back Parts of Thermal Power Meter Console

WWW.ASHA-CO.IR12
Pinout
Figure 7.Pinout - Thermal Power Meter Detector

13 WWW.ASHA-CO.IR
Setup
Safety
It should be noted that all safety ps in this manual are only use-
ful when used in accordance with the condions and for use un-
der the menoned situaons. All parts should be connected to
each other with a suitable and shielded cable. If you need to re-
place some parts of Asha’s company with others, you need to in-
form ASHA’s Technical support. If any changes in the compo-
nents of the product are applied, it may lose its funconality.
Due to the body reecon and opcal lter, it is recommend-
ed that you turn o the light source or set it in low pow-
er mode when using in the desired locaon. It is also nec-
essary to have an orthogonal beam on the sensor surface.
This instrumentaonal device only will be calibrated when placed in
the foam and special box according to the original packaging. It is rec-
ommended that you take alternave packaging from ASHA company
if necessary.

WWW.ASHA-CO.IR14
Preparation and Measurement
To set up the device, rst it is necessary to put the sensor perpen-
dicular to the beam axis with the proper connecon at the desired
locaon, and then connect the sensor cable to the device.
The device has a display on its front panel, and it’s possible to directly
display the result on the screen in addion to measuring through the
mulmeter. The output of this device, according to the order, is able
to be calibrated for required ranges, so that the output voltage of the
device will be a coecient of beam power.
The display on the device shows the exact amount of opcal power
according to the selected range. If the voltmeter is used at the range
of 200mW, the measured voltage is ten mes the real power value. In
the 20W range, the measured voltage is one-tenth of real power and
in the 2 W power range, the reported voltage is the same as the real
value of power.
If you want to measure power using a voltmeter, using the supplied
cable, connect the analog output of the device (BNC port) to the cor-
responding BNC cable, then connect the black probe to the COM port
and the red probe to the voltmeter. Then turn on the device by using
the On/O key. If the device has enough baery, the display on the
front panel of the console will turn on. Otherwise, connect the adapter
to the input power connector. In this case, the charging light turns on,
which means that the baery is charging and you can use the console
at the same me. When the baery is fully charged, the red light will

15 WWW.ASHA-CO.IR
turn o and you can unplug the adaptor from the console.
This device can measure 3 ranges specied in the technical table.
To prepare the device to measure the desired range, turn the selector
and select the desired range. In the end, turn on the beam source or
remove the obstacle and measure the beam.
One of the features of this product (according to the order)is the oset
seng, which can be used to remove the background radiaon and
then measure the absolute power of the beam source. To do this, use
the oset knob on the back panel of the device and rotate the knob
to apply the desired oset according to the applicaon. One of the
important applicaons of this feature is the removal of the background
radiaon from the end result, which is very useful in environments
with unwanted background radiaon.

WWW.ASHA-CO.IR16
Requirements
»High precision analog or digital voltmeter (5.5Digits)
»Appropriate Power Supply (Available in Package)
»Standard ground connecon (Due to the high sensivity of
the sensor, if there is no proper ground connecon, the envi-
ronmental noise will greatly aect the sensor performance.)
It is necessary to menon that this power meter is very precise and sen-
sive and due to the precision the device output (listed in the table),
in order to use full capabilies of the device, it is recommended to use
voltmeters with the accuracy beer than 100μV.

17 WWW.ASHA-CO.IR
Drawings
Figure 8. Thermal Power Meter Console - Dimensions

WWW.ASHA-CO.IR18
Figure 9.Thermal Power Meter Detector - Dimensions

19 WWW.ASHA-CO.IR
This page is le blank intenonally.