Atmos 51 D User manual

ATMOS
Chair 51 D
2014-08 Index: 08
MedizinTechnik
Operating Instructions
English
503.0460.B
503.0461.B
503.0462.B
503.0463.B
503.0464.B
503.0465.B
503.0466.B

2
ATMOS
MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Str. 16
79853 Lenzkirch
Germany
Phone + 49 (0) 76 53 / 689-0
Fax + 49 (0) 76 53 / 689-190
+ 49 (0) 76 53 / 689-393 (Service Center)
www.atmosmed.de
Seite
1.0 Einleitung
1.1 Hinweise zur Gebrauchsanweisung........................
1.2 Zweckbestimmung..................................................
1.3 Funktion ..................................................................
1.4 Erklärung der Bildzeichen.......................................
2.0 Sicherheitshinweise ............................................
3.0 Aufstellung und Inbetriebnahme
3.1 Bedienelemente......................................................
3.2 Anschlussfeld im Geräteboden...............................
4.0 Bedienung
4.1 Einsetzen / Entfernen des
DDS-Bakterienfi lters / Übersaugstop......................
4.2 Verwendung des DDS-Spritzschutzes....................
4.3 DDS-Behälterdeckel aufsetzen / entfernen.............
4.4 DDS-Behältergriff aufsetzen ...................................
4.5 DDS-Behältergriff verschließen / öffnen .................
4.6 Einhängen des DDS-Sekretbehälters.....................
4.7 DDS-Schlauchadapter einsetzen............................
4.8 Zusammenbau des 3DS Sekretbehälters...............
4.9 Wechseln des 3 DS Innenbeutels...........................
4.10 Einhängen des 3DS-Sekretbehälters......................
4.11 Schlauch anschließen.............................................
4.12 Ein / Ausschalter .....................................................
4.13 Vakuum einstellen...................................................
4.14 Absaugen................................................................
4.15 Prüfen des DDS-Bakterienfi lters /
Übersaugstop..........................................................
4.16 Überprüfung von Teilen die der Abnutzung
unterliegt .................................................................
page
1.0 Introduction
1.1 Notes on operating instructions ............................3
1.2 Intended use .........................................................3
1.3 Function ................................................................3
1.4 Explanation of symbols.........................................4
1.5 Scope of supply ....................................................5
1.6 Transport and storage...........................................5
2.0 For your safety ...................................................6
3.0 Setting up
3.1 Setting up the ATMOS Chair 51 D ........................7
3.2 Height and inclination adjustment of backrest ......7
3.3 Adjusting the position of the backrest ...................8
3.4 Mounting the backrest...........................................9
3.5 Using the backrest resp. elbow rest......................9
3.6 Seat height ajdustment .........................................9
4.0 Cleaning
4.1 Recommended surface disinfectants..................10
5.0 Maintenance ......................................................11
6.0 Technical Specifi cations .................................11
7.0 Inspection..........................................................11
General Standard Terms and Conditions
Table of Contents

3
1.2 Intended use 1.3 Function
With the doctor's chair ATMOS Chair 51 D the exami-
ner can take optimum position to the patient.
The ATMOS Chair 51 D is equipped with a mechani-
cal height adjustment. The seat height can be varied
between 56 bis 74 cm.
The ATMOS Chair 51 D is mobile on 5 castors and
equipped with a backrest which is adjustable in height
and inclination. In addition it is rotatable and can
therefore be used as elbow rest.
Name: ATMOS®Chair 51 D
Main functions:
This doctor's chair enables the optimum positioning of the
doctor with regard to height and access to the patient
Med. indications/ application:
For positioning the doctor at the patient during his treatment
Specifi cation of the main function:
• Height adjustment of the seat by gas compression spring
from approx. 56 cm up to approx. 74 cm
• 360° rotation
• Lockable in every heights position
• Backrest adjustable in height and inclination, rotatable
• Mobile on 5 castors
Application organ: Positioning of the patient
Application time: Temporary (Max. 60 minutes)
Application site:
In clinics and practices for ENT doctors and phoniatrists. The
application of the doctor’s chair must be executed by medi-
cally trained persons only.
Contraindications:
• Max. weight load 140 kg
• May only be used as seating on horizontal surfaces
The product is: □ active X not active
Sterility: Not necessary
Single use product / reprocessing: No single use product
1.0 Introduction
1.1 Notes on operating instructions
These operating instructions contain important notes on how to operate the ATMOS Chair 51 D
safely, correctly and effectively. Their reading helps to avoid risks, and also to reduce repair costs
and down-time. That increases, amongst other things, the reliability and service-life of the device.
These operating instructions serve not only for new operating personnel to be instructed in its use,
but also for use as a reference manual. Any reprint - even in extracts - only after written permission
from ATMOS.
These operating instructions must always be kept available near the device.
Care and safety inspections in conjunction with professional execution provide for operational safety
and readiness for use of your ATMOS Chair 51 D and are therefore a must besides regular cleaning.
Repair work and safety inspections may be carried out only by expert personnel authorised by
ATMOS. By applying only original spare parts you will have the guarantee that operational safety,
readiness for work and the value of your ATMOS Chair 51 D will be preserved.
● The product ATMOS Chair 51 D bears CE marking CE 0124 according to the EC Directive of the
council for medical products 93/42/EEC and meets the basic requirements of Appendix I of this
directive.
● The product ATMOS Chair 51 D complies with all applicable requirements of the directive
2011/65/EC restricting the use of certain hazardous substances in electrical and electronic
equipment (“RoHS”).
● The declaration of conformity can be obtained on our website at www.atmosmed.com.
● The quality management system applied at ATMOS has been certifi ed according to international
standards EN ISO 9001 and EN ISO 13485.
● Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any
possible dangerous situations.

4
1.0 Introduction
!
■
●
→
click
click
Important information
Symbols at the ATMOS Chair 51 D
REF
SN
Observe operating instructions
1.4 Explanation of pictures and symbols
Short cuts / symbols contained in this manual
Please press where
dot indicates
Subnumeration
Numeration
General information
Follow the arrows
whilst proceeding,
sequence
Replace
Check
Please read,
important information
Move, plug ... in this
direction
Engage, check
correct fi t
Turn, shift ... in this
direction
Warning, especial diligent
notice
Pictures contained in this manual
The CE sign shows that this
product meets the appropriate
requirements of the
EC Directives
Serial number
Order number
Creation date

5
1.0 Introduction
1.5 Scope of supply
● Prior to dispatch, the ATMOS Chair 51 D was subjected to an extensive functional test and has been carefully packed.
Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note).
Chair Operating
instructions
1.6 Transport and storage
● The transport of the chair may be effected only in a dispatch
carton upholstered and offering suffi cient protection.
● Please document and report damages in transit immediately.
For complaints or return deliveries, please use the enclosed
form QD 434.
●Ambient conditions:
Transport/Storage: -20...+50°C;
5...95 % air humidity
non-condensing
at air pressure 700...1060 hPa

6
Warranty period for this unit: 2 years. This period is
unaffected by any repair or maintenance carried out under
the terms of warranty. Please also pay attention to our
enclosed General Standard Terms and Conditions.
The warranty will be rendered invalid in case of damages
caused due to the utilization of accessories or consumables
which are not approved by ATMOS for use with this unit.
ATMOS is not liable for personal injury and damage to
property if
• no original ATMOS parts are being used,
• the advice for use in these operating instructions is not
being observed,
• assembly, new settings, alterations, extensions and
repairs have been carried out by personnel not
authorised by ATMOS.
This operation manual corresponds with the construction
of the unit and with the current status of safety-related
standards at the time of printing. Proprietary rights are
existing for all described circuits, processes, names, soft-
ware programs and units.
Before putting the ATMOS Chair 51 D into operation, vi-
sually check it for any signs of damage. Prior to use please
check proper function of the chair.
Dispose of the packaging material, observing the
applicable waste-control regulations.
The ambient conditions specifi ed in section 7.0 must be
strictly observed.
2.0 For your safety
!
For your safety

7
1
3.0 Setting up
3.1 Setting up the ATMOS Chair 51 D
1. Insert the gas compression spring including cover ()
into the star base ().
2. Attach the seating () to the gas compression
spring ().
3. Fix the backrest cushion to the backrest.
Fig 1.
3.2 Height and inclination adjustment
of the backrest
Height and inclination of the backrest can be adjusted with
the screw (, Fig 2) using a hexagonal wrench.
Fig 2. Screw for the backrest
height adjustment
1
2
3
4
3

8
1
3.0 Setting up
3.3 Adjusting the position of the backrest
The position of the backrest can be adjusted in height and
inclination. In addition it is rotatable and can therefore be
used as elbow rest (, Fig 3).
Fig 4. Adjusting the backrest position
upwards and downwards
Fig 3. Adjusting the
backrest position
Fig 5. Adjusting the backrest position
forward and back

9
1
1
2
3.6 Adjusting the seat height
By pulling the lever (, Fig 9) upwards, the height adjust-
ment is released (for upward and downward movement). The
gas operated compression spring provides for the upward
movement. The height can be reduced by simultaneously
pressing on the seat. When releasing the lever the chair
remains arrested in the desired position.
Fig 9.
Fig 8.
Fig 7.
Fig 6.
3.4 Mounting the backrest
Loosen the screws (and , Fig 6). Attach the backrest to the
relevant drillings and refi x the screws. The backrest is rotatable
and can therefore be used as elbow rest as well. There are 3
mounting positions possible: right, centre, left.
3.5 Using the backrest respectively
the elbow rest
3.0 Setting up
right
centre
left

10
4.0 Cleaning
Do not use any abrasive cleaning agents. For the upholstery usual dry foam may be used.
Treatment with a commonly used care product for artifi cial leather is recommended once a week in order to keep the
upholstery soft and smooth.
!
4.2 Cleaning the unit's surfaces and upholstery
●The surfaces of the ATMOS Chair 51 D are resistant against all the recommended surface disinfectants stated in chapter 4.3
and 4.4. Nevertheless after any length of time discolourations could possibly develop.
● Substances such as blood need to be removed immediately to prevent stains on the cover material of the upholstery.
●The unit itself can be wiped off with a moist (not wet) cloth.
The described measures to clean and to disinfect do not
rule out regulations valid for the particular company!
● In chapter 4.3 Recommended disinfectants“ all sur-
face and instrument disintectants are stated which are
suitable for disinfection.
Always observe the concentration specifi cations and
instructions by the respective manufacturer.
●Do NOT use
- disinfectants containing concentrated organic
or anorganic acids or bases, since these may cause
corrosion damages.
- disinfectants containing chloramides, phenol
derivatives or anionic tensides, since these may
cause stress cracks in the plastics used.
4.1 General information on cleaning and disinfection
Before cleaning
Medical devices must always offer a maximum in safety and function.
4.3 Recommended surface disinfectant
● The surfaces of the ATMOS Chair 51 D can be cleaned/wiped with disinfectants containing the following active ingredients:
• QAV (quarternary ammonium compounds)
4.4 Recommended disinfectants for the upholstery
Disinfectant Ingredients (in 100 g) Manufacturer
ATMOS Green & Clean SK Alkyldimethyl
(application solution) benzylammonium
chloride < 1 g Metasys, Rum (Austria)
Dialkyldimehty
lammoniumchloride < 1 g
Alkyldimethylethyl
benzylammoniumchloride < 1 g

11
5.0 Maintenance
Only ATMOS technicians or experts authorized by ATMOS
are allowed to open the ATMOS Chair 51 D resp. to repair or
maintain it (with the exception of the cleaning work described
in these operating instructions). Pay attention to technical
and hygienic precautionary measures.
Country-specifi c legislation is to be observed.
ATMOS neither guarantees error-free function nor is ATMOS
liable for personal injury and damage to property if
• no original ATMOS parts are being used,
• the advice for use in these operating instructions is not
being observed,
• assembly, new settings, alterations, extensions and
repairs have been carried out by personnel not authorised
by ATMOS.
The warranty will be rendered invalid in case of damages
caused due to the utilization of accessories or consumables
which are not approved by ATMOS for use with these chairs.
Height adjustment by gas compression spring
Seat diameter 40 cm
Seat height 56-74 cm
Ambient conditions
Transport/storage
-20...+50°C
5...95 % air humidity without condensing,
air pressure 700...1060 hPa
Dimensions HxWxD 105 x 45 x 57 cm
Weight 14 kg
Classifi cation acc. to Appendix IX
EC Directive 93/42/EEC
I
CE marking CE
Max. weight load 140 kg
UMDNS-Code 10-794
GMDN-Code 38448
REF 503.0460.0
Canadian classifi cation Physical Medicine
Device group 89INN
PNC 1
Risk Class Description CHAIR, ADJUSTABLE, MECHANICAL
Issue of the technical specifi cations: 29.06.2011
6.0 Technical Specifi cations
7.0 Inspection
A regular and thouroughly cleaning and disinfection of the ATMOS Chair 51 D according to these operating instructions is taken
for granted.

MedizinTechnik
This document is copyrighted. Duplication, translations, microlming and savings on electronic systems, particularly for commercialpurposes
are illegal without prior agreement of the manufacturer. All compiled data are based on manufacturers instructions. All logos,
product names and designations used in this document are property of the respective manufacturer.
We do not take over any warranty and liability in the case of missing inscriptions. Subject to modications and amendments.
1. General:
Our General Standard Terms and Conditions apply exclusively. Client’s
terms and conditions which are contrary to or deviate from our General
Standard Terms and Conditions are not recognised unless their validity
is explicitly confirmed in writing. Our General Standard Terms and
Conditions also apply even if we deliver to clients without reservation,
in the knowledge of the client’s contrary terms and conditions. Our
General Standard Terms and Conditions also apply to all future business
with that client.
2. Proposal - Order Confirmation
Our proposals are subject to change without notice unless otherwise
stated in our order confirmation. Each order is only accepted by us
following our written order confirmation.
3. Orders
Every order requires an exact description of all of our product’s details.
We assume no liability for errors and damage caused by inaccurate or
incomplete ordering details.
4. Prices
Unless otherwise stated in the order confirmation, our prices in the
order confirmation are ex factory prices and exclude packaging and
value added tax. Packaging is charged separately at cost price in the
invoice. Value added tax is charged separately in the invoice according
to the legal rate on the invoice date. We reserve the right to change
prices appropriately should price reductions or increases, especially
due to wage settlements, changes in the price of materials or currency
fluctuations, be incurred. Proof of such changes will be provided for the
client on request.
5. Payment Conditions - Balancing
Unless otherwise stated in the order confirmation, our invoices
are payable with a 3% discount within 10 days (except for repair and
assembly services) or within 21 days from the invoice date net cash;
money receipts is decisive for complying with this term. We are entitled
to charge interest after the due date at a rate 2% above the relevant
basic interest rate of the German Federal Bank. Should the client have
payment arrears, we are entitled to charge interest on arrears at a rate
5% above the relevant basic interest rate of the German Federal Bank.
Should we be able to prove higher damages due to arrears, we are also
entitled to claim these. The client only has the right to balance invoices
against its own claims should such claims be confirmed in a court of
law or recognised by us. The client does not have the right of retention
due to disputed counterclaims.
6. Delivery Periods
Fulfilment of our delivery duties requires the punctual and proper
fulfilment of the client’s duties. The right to defense on the grounds of an
unfulfilled contract is reserved.Should the client default in accepting the
goods delivery or breach other cooperation duties, we are entitled either
to withdraw from the contract or claim compensation for any increased
costs incurred up to that time without setting a further deadline. The right
to make further claims is reserved. Furthermore, in such cases, the risk
of coin-cidental destruction or a coincidental deterioration in the quality
of the delivered goods is transferred to the client in the case of default
in accepting such goods or payment arrears. Acts of God or stoppages
(due to insufficient supplies of material, industrial disputes etc.) entitle
us either to demand an appropriate extension of delivery periods or to
partly or entirely dissolve the delivery contract. This does not give the
client the right to claim damages. We have fulfilled delivery periods if the
delivery goods have left our factory or the client has been informed of
the goods’ readiness for delivery within such delivery periods. Delivery
periods stipulated by the client are not recognisedby us unless they
form part of our order confirmation. We adhere to legal terms and
conditions in cases where, as a result of an undue delay in the delivery
for which we are liable, the client is entitled to claim that his interests
in a continued fulfilment of the contract have ceased. We also adhere
to legal terms and conditions should a delay in delivery be caused by
deliberate or grossly negligent action by us or our representatives for
which we are responsible. We are also responsible for such actions by
our representatives or agents. Should the delivery delay not be caused
by our deliberate infringement of contractual duties for which we are
responsible, our liability is limited to damage which is regarded as typical
for that case. We are liable according to the legal terms and conditions if
and in so far as the delivery delay for which we are responsible is caused
by an infringement of a substantial contractual duty. In such cases, our
liability is also limited to damage which is regardedas typical for that
case. Should the delivery delay be caused by a culpable infringement
of non-substantial contractual duties, our client is also entitled to claim
a one-off damage compen-sationworth 3 percentage points of the
delivery value of the goods for each week’s delay, up to a maximum
which is no higher than 15 percentage points of the delivery value of
the goods
7. Delivery - Familiarisation
In the case of the delivery of devices for the medico-technical industry
which require assembly and/or familiarisation for the final customer using
specialist trade personnel (such as Ear, Nose and Throat Apparatus and
Suction Units), we reserve the right to deliver the goods exclusively to
the relevant specialist traders. Should the trader not carry out assembly
and/or familiarisation for the final customer, this is carried out by us. In
such cases, we reserve the right to charge the client for the additionally
created costs. Our specialist traders operate a recording system so
that, if necessary, our products can be traced to the final customer. The
specialist trader undertakes to immediately report to us all events and
risks which must be reported in connection with our products.
8. Passage of Risk - Packaging
Unless otherwise stated in our order confirmation, delivery is agreed
ex factory. The risk of the goods’ damage or loss is therefore transferred
to the client as soon as the goods leave the factory or the client is in
default of acceptance of the goods. This also applies to cases where we
confirm prepaid carriage. Transport packaging and all other packaging
according to the packaging regulations is not returnable. Our client is
responsible for disposing the packaging at its own cost. Our deliveries are
insured by us at the client’s expense unless explicitly otherwise agreed.
No insurance is arranged in the case of goods which are collected by
our clients. In the case of transport damage, claims are only handled if
the client receives confirmation of any damage, reduced weight or loss
by the shipping company before accepting the delivery.
9. Warranty
The client is responsible for examining the delivered goods
immediately after receiving them to determine any eventual deficiencies
or delivery errors, and to report these immediately. Should the client
fulfil this examining and reporting responsibility, and should payment
conditions be fulfilled, we shall be liable to the client within the scope
of legal regulations. Our period of warranty shall in all cases be two
years. Our client can make use of the warranty as follows, so long as
he can provide first buyer proof (in the form of an invoice or delivery
note) and provided that the product still has the original, unchanged
serial number:
a. We choose whether to fulfil our guarantee by providing repair
services free of charge - either on the client’s premises or in our factory
- or replacing the product. We can also provide these guarantee
services through an authorised company;
b. Should a product be returned to us, the client agrees to send
the product in its original or similar packaging, offering the same
protection as the original packaging, to our address or any address
notified by us.
c. Our guarantee ceases to apply if changes of any kind have been
made to our product, unless such changes have been made by us
or a company authorised by us, or have been previously agreed
upon in writing by us. Our guarantee also ceases to apply if third
parties have carried out repairs to our products or replaced parts
thereof. This applies regardless of the fact whether these measures
individually or collectively led to a deficiency of the product;
d. We accept no responsibility for damage defects caused by
- operational wear and tear;
- incorrect installation or incorrect or insufficient maintenance;
- incorrect operation of the product (in contradiction to the operating instructions
delivered with the product); - improper use or operating faults; -
inappropriate or negligent handling and care, especially with respect
to dirt, lime, suction of fluids, inappropriate cleaning and sterilisation;
- using accessories and/or replacementpartswhich are not explicitly
approved;
- incorrect assembly and/or initial operation by the client or third
parties; - the client’s negligence in handling the product; - unacceptable
operating conditions, such as humidity, temperatures, the power supply,
vibrations.
- accidents, acts of God, especially lightening, water, fire, public
unrest and insufficient ventilation. We arenot liable for damage to
other objects apart from our product itself, except in thecase of any
deliberate or grossly negligent actions by us or our representatives or
agents. Should no deliberate breach of contract be claimed, our liability
is limited to damage which is regarded as typical for tthat case. This
also applies in the case of our culpable infringement of substantial
contractual duties The indispensable conditions of German Liability
Law remain unaffected thereby.
- For second-hand equipment, the period of warranty shall be reduced
to a period of twelve months.
10.Reservation of Ownership
We retain ownership of our goods until the receipt of all payments
arising from the business relationship, including all demands arising
from installation orders, subsequent orders, repairs, accessory deliveries
and replacement orders. Should we have agreed upon payment on the
basis of cheque and bill transactions, the ownership reservation applies
until the cheque received byus has been paid in, and does not expire
through our credit upon receiving the client’s cheque. In the case of
a breach of contract by the client, especially payment arrears, we are
entitled to repossess our goods. Repossession of our goods repre-sents
a withdrawal from the contract, unless explicitly declared in writing by
us. We have the right to utilise the product after its repossession, whilst
the income form such use is balanced against the client’s arrears, after
deducting appropriate utilisation costs.The client is responsible for
handling the goods with care. Should maintenance and inspection work
be necessary, the client must carry these out punctually at his own cost.
Our client is entitled to sell the goods he has bought from us in a proper
sale transaction. However, he must immediately assign all outstanding
claims to the value of the final invoice sum (including value added tax)
of our claims to his customers or third parties. The client is entitled to
collect this claim even after such assignment. Our right to collect the
claim ourselves remains unaffected thereby.We undertake to release
the securities to which we are entitled if requested to do so by the
client should the realisable value of the our securities be more than 10
percentage points higher than the outstanding claims. We reserve the
right to choose the securities to be released.
11. Plans and Illustrations
We retain ownership of and copyrights to all plans, illustrations,
calculations and other documents which are attached to our proposals.
The client must receive explicit written permission before passing these
on to third parties. Imitating our legally patented products is forbidden
and will be prosecuted.
12.Jurisdiction and Place of Performance
Our central office is the place of performance for all disputes in
connection with these General Standard Terms and Conditions and
the contracts closed with clients under them. This jurisdiction excludes
other jurisdiction relating to persons or subject-matter. Furthermore, our
client is not entitled to bring charges against us in another court should
he file counter-charges, carry out counterbalancing or declare retention.
We, however, are entitled to bring charges against our client at their
general place of jurisdiction or at another relevant court recognised by
German or foreign law.Unless otherwise stated in the order confirmation,
our central office is the place of performance.
Lenzkirch, September 2008
ATMOS MedizinTechnik GmbH & Co. KG
79853 Lenzkirch/Germany
ATMOS General terms and conditions
Table of contents
Other Atmos Indoor Furnishing manuals
Popular Indoor Furnishing manuals by other brands

Julian Bowen
Julian Bowen Jupiter JUP302 Assembly instructions

Linea Deco
Linea Deco Atanaliya 35306 installation manual

IKEA
IKEA KULLEN Assembly instruction

LTL Home Products
LTL Home Products Reflections Blaze installation instructions

Argos
Argos 719/1792 Assembly instructions

IKEA
IKEA BESTA SHELF UNIT 47X50" instructions

Manhattan Comfort
Manhattan Comfort 27900 Assembly instructions

Universal Furniture
Universal Furniture Explore Home U011B652 instruction sheet

OfficeSource
OfficeSource PBTR66 Assembly instructions

Multimo
Multimo 721-PRIMER HAPPY Assembly instructions

Ashley
Ashley Signature One drawer nightstand Assembly instructions

Romina
Romina 4515 Assembly instructions