Best Value Vacs Icarus True Dewax User manual

2 – 5 LB Icarus Operating Manual
12

2 – 5 LB Icarus Operating Manual
12
Table of Contents
Introduction:……………………………………………………………………………………………………………………………………………………….3
i. General Uses and Extraction Information……………………………………………………………………………………………....3
ii. General Safety Information…………………………………………………………………………………………………………………....3
iii. Getting Started……………………………………………………………………………………………………………………………………....4
iv. Manipulating Thermals…………………………………………………………………………………………………….…………………....4
v. Controls and Components……………………………………………………………………………………………………………………...6
Startup and Operation:……………………………………………………………………………………………………………………………………….6
i. Pre-Run Procedure/Testing……………………………………………………………………………………………………………………..7
ii. Adding Solvent to System……………………………………………………………………………………………………………………..10
a. Bottom Flood Input…………………………………………………………………………………………………………………..11
b. Top Flood Input…………………………………………………………………………………………………………………………12
iii. Operating Dewaxing Column…………………………………………………………………………………………………………………12
a. Draining Dewaxing Column/Nitrogen Assist………………………………………………………………………………13
iv. Recovery………………………………………………………………………………………………………………………………………………..14
a. Pre-Recovery Procedure…………………………………………………………………………………………………………….14
b. Recovery to Vapor Port….………………………………………………………………………………………………………….15
c. Recovery to Liquid Port……………………………………………………………………………………………………………..16
Post Run Procedure................................................................................................................................................17
i. Removing the Extract…………………………………………………………………………………………………………………………….17
ii. Cleaning and Gear Maintenance……………………………………………………………………………………………………………17

2 – 5 LB Icarus Operating Manual
12
Introduction:
The Icarus True Dewax closed loop is intended to perform hydrocarbon botanical extraction within a sealed
system. Rack mounting provides ease of work, eliminating heavy lifting associated with operating large
capacity units of the past.
Here are some features that make running the Icarus so user-friendly.
3 stage setup
o2 part modular material column ( 2.5 or 5lb capacity)
oSeparated dewaxing chamber with filters
oSeparated Collection Base for easier recovery
Rack Mounting
Manifold setup to remove and add pressure to system
Polycarbonate sight window on dewax and collection chambers.
General Uses and Extraction Information
Hydrocarbon extraction is performed by passing a hydrocarbon solvent over an organic material to separate
terpenes and other hydrocarbon compounds. Solvent is then distilled to leave the extracted compounds
behind. The Icarus is a passive closed loop, so moving the solvent through the system is dependent upon the
manipulation of temperature and pressure. This system is intended to be run with 100% butane (R-600) or
iso-butane (R-600a). Solvents always seek the lowest pressure in the system. By chilling the receiving vessel
below the boiling point of the solvent used, the liquid solvent will seek that vessel to reduce its pressure.
General Safety Information
When operated and maintained according to the directions in the manual common and safety procedures, the
Icarus system should provide a safe and reliable extraction process. This unit should be run only in extremely
well ventilated areas. If running the unit indoors, it must be operated in areas approved by local fire marshal,
in accordance to local and state laws/ordinances. Always pressure check system prior to every use. Make sure
all gasket seals are cleaned with compatible solvents, and checked for wear before each use.
The Icarus system uses flammable solvents. Use EXTREME caution while operating unit.
ALWAYS OPERATE IN EXTREMELY WELL VENTILATED AREAS

2 – 5 LB Icarus Operating Manual
12
Getting Started
Before you begin, you are going to want to make sure you have the necessary supplies to run the unit. Here is
a list of some things that will be needed to operate the unit.
Large containers for Ice and temperature control fluids
Alcohol or Glycol for dewaxing chamber
Butane
Nitrogen gas cylinder with regulator
Refrigerant scale
Combustible Gas Leak Detector (recommended)
Explosion-proof exhaust fan (recommended)
Cleaning solvent (D-limonene is recommended)
Once the machine is packed with material and assembled, always pressure test the gasket and clamp
connections. Pressurize the entire unit to 90 PSI with nitrogen gas using the manifold setup. Allow pressure
to sit for at least 10 minutes, checking to make sure no pressure is lost. If the unit is sealed, connect vacuum
pump to manifold and pull a full vacuum.
Manipulating Thermals
In order to move the solvent through the system, you must ensure that the supply vessel is of higher pressure
than the receiving vessel. This can be achieved by making sure the receiving vessel is colder than the supply
vessel.
If you refer to (fig. 1), you can see that solvent pressure is directly related to its temperature. As the solvent
increases in temperature above the boiling point (fig.2), pressure will increase. Alternatively, since we are
working in a system that is at a full vacuum, as the solvent gets colder than the boiling point, pressures will
start to go into the negative.
Solvent Boiling
Point
n-Butane (R-600) 30.2 °F
Always tighten clamps evenly on each side. Unit MUST be pressure tested to 90 PSI
before each use. Failure to do so could result in solvent leaks.

2 – 5 LB Icarus Operating Manual
12
(-1 °C)
Isobutane (R-600a) 10.94 °F
(-11.7 °C)
For the best results, it is recommended to chill the solvent. It is important to not chill the solvent below the
boiling point of the chosen solvent, removing any pressure from the supply. Chilled solvent will aid in the
dewaxing process, however, since there is a separated dewaxing chamber, there is no need to go overboard
on chilling the solvent initially.
When distilling the solvent in the collection, it is important to note the boiling point of the most delicate
compound in your extract; i.e. when extracting hops or rosemary, your recovery temperature should not exceed the
boiling point of your lowest boiling terpene, Beta-caryophyllene. It is important to note that boiling points decrease as
the level of vacuum increases. The change in boiling points can be calculated using the Clausius Clapeyron Equation. For
B-caryophyllene,
[246.2°F @760torr (-0.00in.hg) 0% vac]
[180.00°F @100torr (-25.98 in.hg) 87% vac]
[159.99°F @50torr. (-27.95 in.hg) 93.5% vac]
[138.82°F @23.4torr (-29 in.hg) 96.9% vac]
[109.44°F @7.6torr. (-29.62 in.hg) 99% vac]
By keeping the warm water bath below the boiling point of the lowest boiling compound, you are able to
preserve the full spectrum of your extract.
FIG. 2
FIG. 1

2 – 5 LB Icarus Operating Manual
12
Controls and Components
The valve layout and hose configuration is listed in the diagram below (fig.3). Hoses diameters are listed
below.
Start-up and Operation:
In this section, we will cover pre-run procedure and the operation of the Icarus True Dewax system. It is
important to have your tools on hand at all times (wrenches, scales, buckets). When setting up the unit, it is
recommended to position all clamps with the same positioning, if possible. This ensures continuity if
adjustments are needed during operation.
Always make sure your system is positioned on a level surface with adequate air flow. Butane is known to
pool in cool areas, so it is important to ensure no areas of stagnant air flow exist in the workspace. If
operating indoors, it is important that your workspace meets the criteria for a Class I, Division 1 work
environment. Please consult your local fire marshal to ensure workspace is in accordance to local
laws/ordinances.
FIG. 3

2 – 5 LB Icarus Operating Manual
12

2 – 5 LB Icarus Operating Manual
12

2 – 5 LB Icarus Operating Manual
12

2 – 5 LB Icarus Operating Manual
12
Pre-run Testing And Procedure
Assemble each column as pictured in diagrams. Always check that fasteners on rack are tightened before each
use. Make sure all filter screens are in place and filter plates have a coffee or lab filter in place. (Coffee filters
are 20micron). Once unit assembled, pack material column with your chosen organic material.
The Icarus is intended to be packed with dry material. This system capacity is figured at 4.2 g/in3. Capacity
may vary depending on material/packing density. It is recommended to give a firm pack; a loose fill of the column
will allow solvent to pass over the material too easily. Once the material column is packed, install remaining
clamps and connect hoses.
After unit is fully packed and assembled, attach nitrogen cylinder to manifold and perform pre-run pressure
testing.
Every time you assemble your Icarus system, it is vital that the system is pressure
tested to 90 PSI with nitrogen gas. This ensures all clamps/gasket and hose
connections are sealed. Make sure all valves are in open position during testing.
Allow pressure to sit for at least 10 minutes before releasing pressure and pulling
system to a full vacuum.

2 – 5 LB Icarus Operating Manual
12
Release pressure when no loss in pressure is observed. Disconnect nitrogen cylinder from manifold and
connect vacuum pump. Turn on vacuum pump and open both sides of the manifold. Allow unit to be pulled
to a full vacuum, using the multiple pressure gauges to ensure full negative pressure is achieved. When
completed, the vacuum pump can be put away until next use.
Before you begin adding solvent into the system, fill the sleeve on the dewaxing column with a dry ice and
alcohol/glycol mixture. This allows column to be chilled, ensuring pressures cannot build when dumping
solvent from the material column, and begin dewaxing immediately.
Adding Solvent to the System
Before inputting solvent into the system, it is important to pre-weigh solvent in the LP tank to ensure the
system is not overfilled. This also enables solvent to be re-weighed post extraction, ensuring all solvent has
been recovered. It is recommended that solvent be chilled to ~40°F prior to running. This ensures a total
dewaxing of the final product.
Connect LP tank to the solvent input manifold (please refer to assembly diagram) using the short ¼” SS line. It
is recommended to connect the vapor valve (blue handle) to the manifold. This port contains no dip-tube, so
the cylinder must be inverted to empty liquid solvent.
The solvent capacity of n-Butane in the Icarus is 18.9 lbs. This is the
maximum fill of the Dewaxing column.
*Solvent capacity at 80% fill**
It is recommended to use two wrenches when tightening and loosening hoses.
Failure to do so could result in the loosening of fittings. Use one wrench on hose
and one on flare fitting.

2 – 5 LB Icarus Operating Manual
12
a) Bottom Flood Input:
The first pass of solvent is recommended to be done from the bottom flood input (both valve handles on
manifold are in the horizontal position). This ensures that all material is exposed to solvent. The bottom flood
overflow and the dewaxer input (bottom flood) valves are open. As solvent is entering column, place hand on
the 3/8” overflow line. When solvent reaches the top of the material column, the overflow line will vibrate as
liquid flows through. This indicates the entire material column is full of solvent; at this point, close off the
overflow valve and allow a slight saturation of the material.

2 – 5 LB Icarus Operating Manual
12
b) Top Flood Input:
After the bottom fill overflow valve has been closed, switch the solvent input manifold to the top flood
position (valve handles both in vertical positioning). Open the dewaxer input (top fill) valve, then open the top
fill dump valve. Extract bearing solvent will flow from the material column into the dewaxing column. Leave
the dump valve and dewaxer input valve open until both columns return to negative pressure.
Operating the Dewaxing Column:
One of the main features of the Icarus unit is the separated dewaxing chamber. Extract bearing solvent is
dumped into this chamber, which has an outer sleeve that is packed with dry ice slurry (Outer sleeve should be
packed with dry ice slurry prior to beginning the run).
The dewaxing column filter assembly should have coffee or lab filters secured in the filter plates, filter gaskets
in place, as well as the ball bearings to be packed into the elbow and reducer.

2 – 5 LB Icarus Operating Manual
12
Solvent is chilled within the dewaxing chamber, causing pressures to be lowered. As solvent temperature
reaches ~10°F, paraffin waxes will start to coagulate and fall out of the extracted solution. Solvent
temperatures can reach ~-40°F if given time. When this occurs, all extracted waxes, as well as lipids, would
have fallen out and the solution is now ready for filtration.
(Solvent temperatures can be determined with either a thermocouple or by using pressures as a guide; see pressure chart on pg. 5)
a) Draining the Dewaxing Chamber/Using Nitrogen Assist:
Since the solution inside of the dewaxing chamber has been chilled well below the boiling point of either
compatible solvent, the solution will have no pressure. As stated in the introduction, to move solvent through
a passive system, the supply vessel (in this case the dewaxing chamber) must be at a higher pressure than the
receiving vessel.
In order to accomplish this, the nitrogen assist feature of the Icarus must be used. It is also important to have
the collection base in a cold bath. This will prevent pressures from building in the collection chamber.

2 – 5 LB Icarus Operating Manual
12
To start draining the dewaxer:
1. Open nitrogen cylinder (MUST HAVE REGULATOR). Open the regulator to allow ~25psi of nitrogen to
build. Open the nitrogen assist valve on the dewaxing column lid. Allow dewaxer to build to 25 PSI.
Close off both cylinder and assist valve.
2. Open both the dewaxing column dump and the collection column input valve. Allow solvent to flow.
(Solvent flow can be monitored by the polycarbonate sight windows on each vessels top cap). If flow
stops before dewaxing column has emptied, close dump valve and apply ~10 psi extra nitrogen
pressure. Re-open dump and finish draining dewaxer.
3. Once dewaxer has emptied, close off both dewaxer dump valve and the collection chamber input. It is
important to see that solvent flow has stopped flowing into the collection chamber (collection chamber
should appear to be 80% full), as well as solvent has passed through the filter assembly.
*NOTE: Using the dewaxer sight window alone is not adequate for determining whether or not the dewaxer is drained.
Solvent can still be in the dewaxer once passed the top filter. It is recommended to allow time to ensure the dewaxer has
been drained.**
Recovery:

2 – 5 LB Icarus Operating Manual
12
Once the dewaxer has finished draining, it is now time for recovery. Recovery is the distillation of the solvent
in the collection chamber. By distilling, the impurities in the gas will be left behind (in this case, extract) as the
gas moves to the LP tank.
In order to efficiently recover without damaging the extract, a few factors must be acknowledged.
First, we need to note the boiling point of the solvent. This must be noted as the LP tank must be chilled below
this temperature. It is recommended to get the LP tank as cold as possible, as lower temperatures will ensure
gas is instantly liquefied. Colder temperatures allow faster recoveries.
Second, the compounds being extracted must be acknowledged. It is important that the distillation
temperature does not exceed the boiling point of the lowest boiling compound in the extract. In order to
achieve a full extract, it is important that none of the extracted compounds are evaporated during distillation.
*Please refer to page 5 for information on boiling points in relation to vacuum level**
a) Pre-Recovery Procedure:
Since a nitrogen assist was used, nitrogen pressures MUST be bled off before recovery can happen. .
The temperature range of operation within the unit do not get cold enough to depressurize nitrogen.
If pressures are not bled, recovery can be significantly slowed or even locked.
The following steps will help ensure an efficient recovery occurs
1. Once dewaxer has drained, cover LP tank in dry ice and alcohol/glycol. This ensures tank is brought
down to temperature before recovery begins. (it is important that the entire tank is covered, failure to
do so can result in inefficient recovery rates)
2. Prepare recovery by preheating water bath to ~105°F.
Nitrogen pressures MUST be bled off before recovery. Some flammable gas may escape
while bleeding nitrogen pressures
ALWAYS OPERATE IN EXTREMELY WELL VENTILATED AREAS

2 – 5 LB Icarus Operating Manual
12
(*This temperature is a recommended recovery temperature when dealing with compounds with low boiling terpenes.
Adjust water bath temperature according to the needs of your chosen extract)
3. Bleed off nitrogen by disconnecting the dewaxer dump line on collection base. The now open valve on
the collection is opened to relieve nitrogen pressures
(NOTE: It is important to take temperature of collection chamber. This will help determine the
pressures of our solvent mixture. When valve opens, nitrogen will escape first. Gas pressure leaving
the bleed off valve will drop when nitrogen has been evacuated. Check pressure in collection base
when completed to verify nitrogen is gone. ALWAYS BLEED NITROGEN IN A WELL VENTILATED AREA.
Some flammable solvent may escape.)
4. Place collection base in warm water bath.
5. Open recovery valves on collection base and LP tank
6. Check completion of recovery via LP tank weight and sight windows.
b) Recovery to Vapor Port (blue):
Recovery to the vapor port is the most straight forward method. Solvent vapors from the collection
chamber are emptied into the headspace of the LP tank. As the vapor is chilled within the tank,
pressures drop until the gas is condensed. In order to efficiently recover using the vapor port, it is
essential that the LP tank be completely covered in dry ice. The ultimate goal is to ensure the top
portion of the tank remains as cold as the bottom. Failure to keep the top half of the LP covered can
result in slow recovery, potentially a vapor lock on the end of recovery.
Vapor port recovery requires less attention to pressures as it is difficult for recovered gas to flow in
reverse.
c) Recovery to the Liquid Port (red):
Recovery to the liquid port tends to give increased recovery speeds. The liquid port has a siphon tube
that will push solvent vapors to the bottom of the LP tank. Once liquid levels reach the siphon tube,
solvent vapors will percolate through the condensed liquid, given the pressures entering the LP exceed
the pressure of the LP tank. As vapor percolates in the condensed liquid, vapor pressures are instantly
eliminated. This method of recovery tends to require less dry ice being that it is not as essential to
keep the very top of the LP covered.
Liquid port recovery is typically faster, however attention must be paid to vapor pressures on both the
collection chamber and LP. It is essential that the collection chamber be at a higher pressure than the
LP. Failure to maintain this pressure gradient will result in solvent flow being reversed.
*Note: Liquid will siphon from the LP into the collection until pressures equalize IF the LP pressures drop below that of the
collection.**

2 – 5 LB Icarus Operating Manual
12
Post Run Procedure:
Once recovery has finished, close recovery and LP valves, then shut down warm water circulation. It is
recommended to allow recovery to happen until collection base hits negative pressure, ensuring that all gas
has recovered.
Before the base can be opened, positive/negative pressure must be relieved. Open the nitrogen bleed off
valve to equalize pressure in collection base. Remove the collection base from the warm water bath. Once
the base is dried, the lid of the collection base is ready to be opened. Carefully loosen each side of the clamp
before removing.
Removing the Extract:
After opening the system, it is time to remove the extract. It is important that the utensil used to remove the
extract is chemically compatible with the solvent used. A PTFE scraper is recommended. If using a metal
scraper, avoid using excessive pressure and scratch the platter.
Being that the base and extract should still be warm, the consistency should still be semi runny. Run the
scraper in a circular motion from the outsides to the middle; the extract should pool in the center of the base.
Scoop the extract out from center.
Cleaning and Gear Maintenance:
Once the extraction process is completed, it is important to break down and clean the gear. This will prevent
contamination of future runs, as well as keep gaskets fresh. We recommend cleaning the gear with d-
limonene, however isopropyl alcohol can be used. It is important to wipe gaskets clean rather than soak them
in solvent. If using alcohol to clean gaskets, it is important to wipe them dry to prevent the gaskets from
breaking down.
The dewaxing column and filters will be covered in lipids. It is important to clean this in between every run to
prevent clogging. Some users save the extracted waxes for further refinement, as oil occlusion in the
extracted wax is possible. Ball bearings can be soaked in cleaning solvent, then dabbed dry with a towel.
DO NOT LEAVE LIQUID SOLVENT IN COLLECTION CHAMBER. FAILURE TO FULLY
RECOVER WILL EXPOSE ATMOSPHERE TO FLAMMABLE GAS.
Table of contents
Popular Ventilation Hood manuals by other brands

Zephyr
Zephyr ZNA-M90BS Use, care and installation guide

Gorenje
Gorenje OK-002/II manual

Franke
Franke FSMS F42 SS/BK MATT installation manual

DeLonghi
DeLonghi S72 Instruction on mounting and use

Novy
Novy 7510/15 Operating and installation instructions

Siemens
Siemens LC68BIT50 User manual and installation instructions