BioLAB 300 Series User manual

Operation Manual
Climatic Chamber
Series 300
Thank you for Choosing Biolab products. Please read the “Operating Instructions” and
“Warranty” before operating this unit to assure proper operation.

Notices
●In order to ensure safety, please read this anual carefully
●Make sure put this anual in convenient place for later use
●Our co pany doesn’t provide a safe guarantee if do not follow the instruction anual
●This anual only for user and authorized technician,it should be kept properly
●No notice if any changes because of product i prove ents
www.biolabscientific.com
2
Climatic Chamber BCCL 300 series

Index
1. Safety notices and warnings...............................................................................................................04
2. Safety operation and preventive action.........................................................................................05
3. Instruction ( application, perfor ance, technical para eters).............................................08
4. Structure.....................................................................................................................................................09
5. Installation.................................................................................................................................................12
6. Operation...................................................................................................................................................13
7. Alar and safety function....................................................................................................................21
8. Routine using and aintenance........................................................................................................21
9. Optional using..........................................................................................................................................22
10. Trouble shooting......................................................................................................................................23
11. Specifications.............................................................................................................................................24
12. Wiring layout..............................................................................................................................................25
3
www.biolabscientific.com
Climatic Chamber BCCL 300 series

01 Safety notices and warning
●“Warning” sy bol
●“Attention” sy bol
●The eaning of sy bols
prohibiting
ust follow
●Sy bols on equip ent
AC
protective conductor ther al
power is connected
power is disconnected
warning, attention, caution and danger
www.biolabscientific.com
4
Climatic Chamber BCCL 300 series
It will cause serious har or fatal accident if not co ply with
warning
It will cause injury ,equip ent da age and the loss of relative
property if not co ply with attention

02 Safety operation and preventive action
Warning :
Do not place this equip ent outdoors. if it exposed in the rain, it ay cause short
circuit and electric shock.
Only professional person have qualification to install this equip ent. If not, it
ay cause electric shock or fire.
Should place this equip ent on the fir ground in case of tu ble. If not, it ay
cause injury because it capsizes.
Do not place equip ent in hu id environ ent or a place with dripping water.
Otherwise it ay cause short circuit or electric shock
Do not place equip ent near fla able aterials and volatile substance.
Otherwise it ay cause explosion or fire.
Do not place equip ent in the area where surrounded by acidic or corrosive
gas, Otherwise it ay cause short circuit or electric shock
Please use power supply socket with protective conductor ter inal in case
electric shock. If power socket without protective conductor ter inal, it is necessary to
install it by licensed technician.
Do not connect protective conductor ter inal through gas, water pipe, telephone
line or lighting arrester which will cause electric shock.
Please use specified power supply. If not, it ay cause electric shock or fire.
Do not put volatile and infla able substances in the inner cha ber of
equip ent if it cannot be sealed, or it ay cause explosion or fire.
Do not insert nail or wire and si ilar etal objects into any inlet or outlet of
equip ent, or it ay cause electric shock or injury
Please operate this equip ent in safe area if it stores any toxic ,har ful and
radioactive substances, or it ay do har to hu an and environ ent.
Make sure to cut off power supply before aintaining equip ent in case it
causes electric shock or injury.
5
www.biolabscientific.com
Climatic Chamber BCCL 300 series

Safety operation and preventive action
Warning :
Do not touch any electric co ponents or switch with wet hand, or it ay cause
electric shock
Make sure wear ask when aintaining the equip ent to prevent any har ful drug
substance and airborne particle.
Do not splash water onto the equip ent, or it ay cause electric shock or short
circuit .
Do not place container which is filled of water on the top of equip ent, or it ay cause
short circuit or electric shock.
Do not drag, twine or bind power cord. Do not da age power plug, or it ay cause
electric shock or fire hazard.
Do not use loose power plug, or it ay cause fire or electric shock
Do not dis antle, repair or refit equip ent without authorization and guidance fro
our co pany.It ay cause fire or injury due to the i proper handling.
Please unplug the power if equip ent is alfunctioning. It ay cause fire or electric
shock if it continues.
Press power plug instead of pulling the power cord when you want to unplug the
power fro power socket, or it ay cause electric shock or fire hazard because of
short circuit.
Should unplug the power before oving equip ent. Do not da age power cord.
Da aged cord ay cause electric shock or fire.
Should unplug power plug if it’s not used for long period, or it ay lead to electric
shock, leakage or fire because of wear and tear of insulator.
Keep out of reach of children and the door unsealed if the equip ent is not
supervised or not used for a long period.
Should infor authorized technician when you dispose the equip ent. Should
dis ount the equip ent door to prevent suffocation and such accident.
Keep out of reach of children with the wrapping plastic.
www.biolabscientific.com
6
Climatic Chamber BCCL 300 series

Safety operation and Preventive measure
Attention :
Please clean the dust on the power plug and then insert it into power socket properly,
or it ay cause over-heating or strike sparks
Check te perature, hu idity, seg ent and ti ing and other setting value when
reboot the equip ent after been short circuited or cut off by power supply.
Otherwise ay cause da age lost of products stored inside.
Please place equip ent in ventilated place and dry place if not used for long period
after purchase, or it ay lead to equip ent alfunctioning when use.
Should arrange proper carrying-tools or qualified person when oving equip ent.
Prevent tu bling when oving equip ent, it ay cause da age of equip ent or
hu an injury.
Ensure enough space when oving equip ent. If you need to carry it to the second
or higher floors, ake sure the elevator has enough space for the equip ent and
working personal.
Do not put acidic, alkaline or corrosive substance in the inner cha ber if the container
is not sealed. Otherwise it will cause corrosion or da age to the co ponents of
equip ent.
Attention
www.biolabscientific.com
Climatic Chamber BCCL 300 series

03 Instruction(application ,perfor ance, technical para eters )
Application :
BCCL Cli ate Cha ber is ther ostatic equip ent with heating, cooling,hu idification and
lighting function, highly precise and advanced.
Widely used in plant cultivation, breeding test, bacteria, ould, icrobial cultivation &
preservation, BOD deter ination and so on. Which is also used in biological genetic
engineering, edical treat ent, health and epide ic prevention, edical test, agriculture and
ani al husbandry, aquatic and other scientific research institutions.
Performance :
BCCL series Cli ate Cha ber transfers actual te perature and hu idity detected fro te p.
sensor and hu idity sensor into signal. Through icroco puter control to heater towards
required te perature and hu idity.
Technical parameters:
1. Volu e:150L.250L.400L;
2. Te p range :0~65C;(with lighting10~50C, without lighting and with hu idity5~50C);
3. Te p fluctuation range: ±0.5C(10C~40C);
4. Te p unifor range: ±1C(10C~40C );
5. Hu idity range: 40-95%RH(10C~40C );
6. Hu idity fluctuation: ±2%;
7. Illu inanation: 15000LX(5 levels)
8. Power voltage: 220V/50Hz;
9. Input power:1080W(150L) 1100W(250L) 1350W(400L);
10. Working a bient :a bient te p 10~30C relative hu idity70% below ;
11. refrigeration:R134;
12. Equip ent class:class I
Notice :this equip ent has low te perature auto-defrost function, it is nor al that if there is
fluctuation when low te perature auto-defrost
www.biolabscientific.com
8
Climatic Chamber BCCL 300 series

04 Structure
Parts :
1. storage water tank
2. wheel
3. co pressor part
4. outlet pipe
5. overflow pipe
6. inlet pipe
7. water switch
8. hu idifier
9. esh board
10. evaporator
11. te perature and hu idity
sensor
12. axial fan
13. water tank
14. glass door
15. outer door
16. la p on outer door
17. internal door button
18. door handle
19. controller
20. la p on side door
21. side door button
22. control panel
23. power supply switch
24. cool switch
9
www.biolabscientific.com
Climatic Chamber BCCL 300 series

Control Panel :
1)LCD display ;
2)alar indicator, it is on when equip ent running abnor ally
3)refrigeration indicator is on when co pressor running nor ally
4)hu idity indicator is on when hu idifier running nor ally
5)heat indicator is on when heater running nor ally key
6) key, long press this key for two seconds to start or stop
7) key ,in the ain interface, press this key then backlight; in setting ode, press this key
to increase value or shift para eter, or turn over the above page
8) key, in the ain interface ,press this key then light,long press anual defrost; in setting
ode, press this key to decrease value or shift para eter, or turn over the below page
9) key, decrease key, in setting ode, press the key to decrease setting value ;
10) key, in the ain interface ,press this key to shift display easured value between the
value of running ;long press this key to access setting ode. In setting ode ,press to shift
value, long press to log out setting ode;
Notice: When beeper beeping, press any key to ute
www.biolabscientific.com
10
Climatic Chamber BCCL 300 series

Equipment installation
In order to ake sure the equip ent can run nor ally, please place equip ent as following:
Attention: a bient te perature 10~30°C; relative hu idity less than 70%
1 Avoid exposure to the sunlight
Do not place it in direct sunlight, or it won’t reach predicted perfor ance
2 An efficient ventilative place
If you operate this equip ent in a narrow and concealed roo , it ay lead to
over-heating and alfunctioning. Mini u safe distance between equip ent and wall is
10CM
d) Keep away from heat source
Don’t install the equip ent near heating source. External excess heat will affect perfor ance
of the equip ent and ay cause alfunctioning
e) Flat and firm ground
Make sure to install it in flat and fir ground. Uneven surface or leaning installation ay
da age equip ent or injure people. Proper installation can avoid shaking and noise
f) Avoid humid place
Install the equip ent in a place where hu idity is less than 70%. Otherwise it ay cause
creepage or electric shock.
Warning
Do not place this equipment outdoors If it exposed in the rain, it may cause short
circuit and electric shock Do not place equipment in humid environment or a place with
dripping water Otherwise it may cause creepage or electric shock
13. Avoid place with fla able or corrosive gas.
Do not place equip ent near fla able aterials and volatile substance. Otherwise it ay
cause explosion or fire. Do not place equip ent in the place where has acidic and corrosive
gas, or corrosion will cause creepage, electric shock or equip ent da age.
11
www.biolabscientific.com
Climatic Chamber BCCL 300 series

05 Installation
1. Unpacking
Re ove packing aterials ,open the door for ventilation . Please use neutral detergent to
clean if the shell and panel is dirty. then wipe with wet cloth and at last with dry and clean
cloth
2. level equip ent
Fix equip ent with the front brake-wheel after installation in case equip ent oves
3 Earthing
Warning
Please use power socket that has protective conductor terminal in case of electric shock
If it is not connected, has to install protective conductor terminal by licensed technician
Do not connect protective conductor terminal through gas, water pipe, telephone line
or lighting arrester which will cause electric shock
4. Idle equip ent
Before setting equip ent aside , e pty water in the hu idifier and re ove internal, oisture
thoroughly. Be sure the inner cha ber is dry and cool before closing the door .
5. Move equip ent
Before oving equip ent ,e pty water in the hu idifier .or it will cause creepage or electric
shock because of overflow water or splashed water
Preparation before operation
When equip ent running in the first ti e, please operate as belows:
1. Take out the shelves or other accessories
2. Clean the inner wall with gauze which is soaked by alcohol and then use dry cloth to wipe-
dry
3. Put the shelves into inner cha ber according to your experi ent of require ent
4. Put water tank on the top of equip ent before using ,please connect water pipe with water
inlet and overflow outlet which in the left of equip ent (refer to co ponent picture ) ,and put
storage water tank under the overflow pipe and water outlet for spare use
5. Pour enough pure water into water tank
www.biolabscientific.com
12
Climatic Chamber BCCL 300 series

Notice :Don’t use NaCl or other Halide solution to clean equip ent ,or it will
cause rust
06 Operation
Display :
1. In the general ode, screen displays ain interface;
2. In aster interface, press key to check interface,after 60 seconds , it will return aster
interface auto atically;
3. In aster interface, long pressing key 4 seconds ,get into setting interface,long pressing
key 4 seconds ,it will return aster interface and save setting value; after 60 seconds without
press any key, it will return aster interface auto atically ,but the setting value is not saved;
13
www.biolabscientific.com
Climatic Chamber BCCL 300 series
lighting:class 6
lighting:class 6
Set lighting:class6

4. In aster interface ,long press key 4 seconds to access para eter password interface ;
different para eter have different password , input correct password, press to access
relative para eter interface; if password is incorrect, press key ,it will return ain interface
5. In para eter password interface , input correct password ,press key, enter relative
para eter interface ; in this interface ,long press for 4 seconds ,it will return ain interface
,and save set value; after 60 seconds without pressing any key, it will return ain interface
auto atically ,but the setting value is not saved;
6. In ain interface press and key for 4 seconds , to internal para eter password
interface ,input correct password, press key to relative para eter interface; if password is
incorrect, press key ,it will return ain interface
www.biolabscientific.com
14
Climatic Chamber BCCL 300 series

Operation Method
Operation steps
For exa ple:after power is connected , the equip ent will start working 120 inutes later.
Set 10 cycles, two seg ents each cycle. in the first seg ent, it needs to work for 720 inutes,
the te perature controlled is at 30 C, light intensity is class 6 ,hu idity controlled is at 60%
RH; in the second seg ent , it need to work for 1200 inutes, the te perature controlled is at
10 C, light intensity is class 4, hu idity controlled is at 50% RH.
a) Press key for 4 seconds to enter the setting interface;
b).press the key, switch the cursor to the value of the total cycle, press and key ,
odify the total cycle to 10; re-press key, the cursor stays at the value of total seg ents,
and odify the total seg ents to 2 ; re-press key, the cursor stays at the value of
appoint ent ti e, odify the appoint ent ti e to 120 inutes ;
c)Then press the key to access the setting interface of the first seg ent, the cursor stays
at the value of te perature , odify the te perature to 30 °C; re-press , the cursor stays
at the value of hu idity, odify the hu idity to 60% RH; then re-press key to odify
light intensity to 6 class ; re-press key to odify ti e to 720 inutes.
d) Re-press to setting interface of the second seg ent, odify the te perature,
hu idity, light intensity and ti e to 10°C, 50%RH, class4 and 120 0 inutes according to
above steps; re-press key for 4 seconds, then log out setting interface , turn back to
ain interface, setting is finished.
e) Long press the key for 2 seconds , the controller will start ti ing according to
appoint ent ti e, when running, it displays : ti ing………, after 120 inutes, the progra will
proceed with the first seg ent
f) After the progra enters the first seg ent, if the ti ing way is 0 , it will start ti ing, and it
will display : ti ing…..:; if the ti ing way 1, it displays : running ….., when te perature reaches
to setting value, start ti ing; if the ti ing way is 2, it will start ti ing after te perature and
hu idity reach setting value
g) It will enter the second seg ent after the first seg ent is over, then it will enter the first
seg ent of the second cycle after the second seg ent of the first cycle is over , then go next
and so on; running stops after the second seg ent of the tenth cycle is over
h) Notice : when hu idity is 0, no hu idification and dehu idification function, this
equip ent can be used as bioche ical incubator
15
www.biolabscientific.com
Climatic Chamber BCCL 300 series

Parameters :
a)user parameters:Password 3
Parameter
na e Para eter Function Range (factory
default)
Address
Connected to the co puter, the ailing address of this
equip ent, if several equip ents connect to the co puter,
ailing address should not be sa e;
1~16(1)
Power failure
protection
1 Turn off the power failure protection;
2 Turn on power failure protection, after power is back, the
progra will proceed with running in accordance with the
power cycle, the nu ber of seg ents and ti e before
power disconnect
0~1(0)
Control
ode
1.The syste will tell co pressor whether is nor ally open
or disconnection according to nor ally open te perature
+a bient te perature , nor ally open hu idity ;
When in disconnection ode, open and closing point will
be calculated auto atically according to a bient
te perature
2. The syste will tell co pressor whether is nor ally open
or disconnection according to nor ally open te perature ;
When in disconnection ode, start and closing point will
be calculated auto atically according to a bient
te perature
3.when in disconnection ode, the co pressor will
proceed with cooling according to the setting value of start
and close refrigerant ;
Proceed with dehu idifying according to the setting value
of start and close dehu idification
0~3 (1)
www.biolabscientific.com
16
Climatic Chamber BCCL 300 series

b) Temperature parameters:Password4
Para eter na e Para eter Instruction Range (factory
default)
Over-te perature
alar
When easuring te perature> setting
te perature + over-te perature alar ing, alar
indicator is on, disconnect heating protective relay,
start cooking , beeper beeps
5.0~50.0 (20.0) C
Zero adjust ent Zero adjust ent = ercury ther o eter reading -
Display te perature; -99.9~99.9(0) C
Full-scale
adjust ent
Modify the error argins of actual te perature
,full-scale adjust ent = 1000 * ( ercury
ther o eter reading - Display te perature) /
display te perature;
-999~999(0)
Proportional Ti e proportional adjust ent. 0.0~90.0(15.0)
Integration ti e Integral acting in regulating. (1~9999S) 600
Derivative ti e Differential acting in regulating. (0~9999S)200
Cycle ti e Heating control cycle. 0~60 (5)
Disconnect
heating
When easuring te perature> = setting
te perature + disconnect heating te perature,
turn off the heating output;
-50.0~50.0(2.0) C
Nor ally open
te perature
When the control ode of co pressor is 0, if the
setting te perature> =a bient te perature +
nor ally open te perature, the co pressor is
auto atically switched to disconnection ode and
calculate start and closing points according to the
a bient te perature
When the control ode of co pressor is 1, if the
setting te perature> = nor ally open
te perature, the co pressor is auto atically
switched to disconnection ode and calculate
start and closing points according to a bient
te perature
-50.0~50.0 (50.0)
C
Start
refrigerating
When the control ode of co pressor is 2, the
start point of refrigerant is: easuring
-20.0~20.0(0.5) C
1
www.biolabscientific.com
Climatic Chamber BCCL 300 series

te perature> =setting te perature + start
refrigerant
Close
refrigerating
When control ode of co pressor is 2, the closing
point of cooling is easuring te perature <=
setting te perature + Close refrigeration; ( the
closing point of refrigeration should be less than
start refrigeration point)
-20.0~20.0(0.2) C
Refrigerating
delay
Co pressor delay protection ti e, the interval
ti e between last stoped and reboot > = delay
ti e;
0~60(3)Min
c) humidity parameters:Password5
Para eter
na e Para eter Instruction Range (factory
default)
Over hu idity
alar
When the the easuring hu idity> = set the
hu idity + over hu idity alar ing, disconnected
the hu idification protective relay
0.0~50.0 (20.0) %RH
Hu idity zero Zero adjust ent = hygro eter reading - display
hu idity; -99.9~99.9(0)%RH
Hu idity full
scale
Modify the error argins of actual hu idity ,full-
scale adjust ent = 1000 * (hygro eter reading-
display hu idity) / hu idity;
-999~999 (0)
Proportional
band Ti e proportional adjust ent. 0.0~90.0(15.0)%RH
Integration
ti e Integral acting in regulating. (1~9999)200S
Derivative ti e Differential acting in regulating. (0~9999)30S
cycle ti e Hu idification control cycle. 0~60 (5)S
Off
hu idification
When easuring hu idity> = setting hu idity +
te perature of closing hu idification, disconnect
hu idification output
-50.0~50.0(-2.0)
%RH
Nor ally open
Hu idity
When the control ode of co pressor is 0, if the
setting hu idity <= nor ally open hu idity, the
co pressor is auto atically switched to
disconnection ode
00.0~90.0 (10.0)
%RH
Start When the control ode of co pressor is 2, the -20.0~20.0(5) %RH
www.biolabscientific.com
18
Climatic Chamber BCCL 300 series

dehu idifying
dehu idification of start point: easuring
hu idity> = to setting hu idity + start
dehu idification;
Close
dehu idifying
when the control ode of co pressor is 2,
dehu idification: easuring hu idity <= setting
hu idity + disconnect dehu idification; ( the
closing point of dehu idification should be less
than the start point of dehu idification)
-20.0~20.0(2) %RH
Low
te perature
protection
when the easured te perature is lower than the
low te perature protection, disconnect
hu idification and dehu idification, the alar
indicator is on
-20.0~60.0(5.0) C
d) Defrost parameters:Password9'
Para et
er na e Para eter Instruction Range (factory
default)
Defrost
ode
0: solenoid valve defrost ode; when defrosting, solenoid
valve is on, the fan stops;
1: heating defrost ode; when defrosting, the solenoid valve
is on, there is heating output, the fan stops, the co pressor
stops;
0~1(0) C
Interval
NO1
The interval ti e of defrost between 0.0-9.0 C, in hour 0~200(3)H
Interval
NO2
The interval ti e of defrost between 9.1-15.0 C, in hour 0~200(3)H
Interval
NO3
The interval ti e of defrost when higher than 15.0 C , in
hour
0~200(6)H
Defrost
NO1
The output ti e of defrost between 0.0-9.0 C , in second 0~300(60)S
Defrost
NO2
The output ti e of defrost between 9.1-15.0 C , in when
higher than 15.0 C
0~300(60)S
Defrost
NO3
The output ti e of defrost when higher than 15.0 C , in
when higher than 15.0 C
0~300(60)S
19
www.biolabscientific.com
Climatic Chamber BCCL 300 series

e) Internal parameters:password 101
Para eter
na e Para eter instruction Range (factory
default)
The upper li it
of te perature The upper li it value of setting te perature 0.0~90.0 (60.0) C
The lower li it
of te perature The lower li it value of setting te perature -40.0~0.0(0.0)C
Controller type
0: artificial cli ate cha ber; 1: constant
te perature and hu idity cha ber; 2: light
incubator
0~2(0) C
Display
Language 0: Chinese; 1: English 0~1(0)
Ti ing way 0:0~9999 in;1:0~99h: 0~59 in 0~1(0)
Ti ing
judg ent
0: start ti ing once equip ent works
1:start ti ing when te perature reaches setting
value
2:start ti ing when te perature and hu idity
reach setting value
0~2(0)
High
Te perature
Control
Setting te perature> high te perature control,
the co pressor starts only when over-
te perature alar ing
0.0~99.9(5.0)
Hu idity
deviation Reservation para eters 0.0~99.9 (2.0)
www.biolabscientific.com
20
Climatic Chamber BCCL 300 series
Table of contents
Popular Climate Chamber manuals by other brands

xtemp
xtemp UXB8401-PG280 operating instructions

Binder
Binder BD 53 operating manual

ESPEC
ESPEC LH-113 user manual

Thermoline Scientific Equipment
Thermoline Scientific Equipment TCH-150 operating instructions

serva
serva BlueVertical PRIME instruction manual

Henkel
Henkel Loctite EQ CL42 LED operating manual