ARCUS
Led BLEACHING SYSTEM OPERATIONS GUIDE
5
IV. SAFETY PRECAUTIONS
As with all technical devices, the proper function and safe operation of this unit depend on the
user’s compliance with the standard safety procedures as well as the specific safety
recommendations described in this Operations guide.
1. The unit must be used in strict accordance with the operations guide. The manufacturer
accepts no liability for any damage resulting from the use of this unit for any other
purposes.
2. Before start-up of the unit make sure that the operating voltage indicated on the unit
label corresponds to the mains voltage. Operation of the unit at the different voltage may
damage the unit.
Electrical safety is ensured by Protection class against electrical current damages
according to EN 60601-1-1990+A1:1993+A2:1995+A13:1996. The devices must be plugged
in standard electrical network, power voltage 100-240VCA/50-60Hz.
Mains outlet where the device is plugged in must be grounded “SHUKO” type.
ARCUS BLEACHING SYSTEM should be operated in indoor premises in the following
conditions:
- temperature: + 10° –+40°C;
- atm, pressure: 700 –1060 hPa;
- relative humidity: 30 –75%.
The device must not contact liquids, such as water and solvents, aggressive or flammable
substances and their vapors, also pouring or wiping the device with the same.
-Absence of chemically active or flammable substances
3. ARCUS BLEACHING SYSTEM is a source of exceptionally intensive light in blue-green
range, to wich human eye is high sensitive. That is why serious measures must be taken
towards patients, medical personnel and nearby people, animals and plants. As measures
can be used: protection goggles for operator and googles, mask and high factor sun
protection cream for patient.
Irradiation of the eyes bears certain risk for their damage. Therefore the light must not be
directed toward the eyes! Exposure must be restricted to the working area.
ATTENTION! The emitted light is in the wavelength range 430-490 nanometers
During work use protection goggles and/or protection shield!
4. The unit must not be applied to pregnant women and suckling mothers, patients
aged under 17. The unit can be used only doctor’s advice by or in persons with
heart pacemakers implant; persons with photobiological reactions; persons
currently taking photosensitive medication; persons with cataract surgery; persons
with retinal diseases, etc.
5. Solvent, flammable liquids and powerful heat sources must not be kept near the
unit as exposure to one of these factors may damage the plastic parts of the unit.