
Cleaning the cuvette of the BMT 964, Rev. 07/2012
page 2 of 2
IN
OUT
cuvette block
bottom view
Cleaning the cuvette
The cuvette window, spacer and seals can be cleaned with water (or with household detergent or alcohol). If
the cuvette is soiled badly, check and clean the cuvette block (you can clean the 12mm bore with a "Q-Tip" or
similar), the inner walls of the PTFE tubing, the filter holder and the fittings. Dry all the parts, and remove
dust with a paint-brush, particularly from the seals. If necessary, use new (quartz) cuvette windows and O-
rings (FFKM).
5. Re-assembling the cuvette
The assembly of the cuvette has to be done in
reverse order of its disassembly:
insert the upper O-ring (if it was removed)
insert the upper cuvette window
insert the spacer-window combination
The etched spacer is bonded to the lower cuvette
window. The assembly must make sure that the
radial opening is facing the outlet and the almost
tangential opening matches the inlet and that the
spacer is located between the
two windows.
For the 50g/Nm3range, a wire
spacer is used, which consists
of two thick rods and two thin
wires which join them. During
assembly, one of the two
thicker rods is placed
diagonally between the gas inlet and outlet (but
without blocking them!).
For special measuring ranges, where a PVC spacer is used, there is only one possible orientation - whereby the
openings of the spacer are lying adjacent to the cuvette block's holes.
Next:
insert the lower cuvette window (if not combined with the spacer)
insert the O-ring
insert the black aluminium ring with the conical face pointing towards the sealing O-ring
insert the spring
Join the circuit board with the cuvette block and fasten it with the four screws, first tightening the two screws
near the cuvette. Whilst tightening the screws, press the circuit board and the cuvette block together (against
the force of the spring) so that the spacer in the cuvette stays in its correct orientation.
6. Re-connecting the tubing
Re-connect the tubing to the fittings on the backplane and tighten the knurled nuts by hand. Now, a leak test is
obligatory! If dust is not removed from the O-rings in the cuvette, fittings and tubing, this could result in an
ozone leakage!
7. Closing the casing
Slide the circuit board into the casing's guide slots and fasten the rear panel with the four Allen screws.
8. Functional test
After thightening the screws holding the rear panel connect the instrument with power and allow warmup for
a several minutes.
Zero the instrument and check the cuvette status value as described in the manual. Factory setting (=clean
cuvette) is 0.0%.
If cleaning was unsuccessful, only a replacement of the cuvette windows will help, now.
IN
OUT