1
Insert test strip
Meter turns on automatically.
Important: Before using a
new pack of strips, insert the
code key (enclosed in the
pack of strips) in the side of
the meter.
2
Confirm the code chip
Confirm the code chip
number (displayed) and
strip lot number (on side
of test strip container) are
identical, then press the M
button. Meter will warm up
for approximately 30 seconds.
Flashing hourglass displayed.
3
Obtain a large drop of blood
When the blood drop symbol
flashes and the 180 second
countdown begins obtain a
large drop of blood from the
side of the finger. Gently
milk your finger, DO NOT
SQUEEZE.
4
Apply blood to the test strip
Touch and hold the first large drop of blood to the transparent area
on the test strip, either to the semicircular well or to the side of the
sample application area. Blood will automatically be drawn into
the test strip and the test will begin. Note: Blood must be applied
within 15 seconds of lancing finger. CoaguChek XS will beep when
sufficient blood is present.
1
Removing used lancet
Pull off the Softclix cap. Eject
the used needle by sliding the
ejector forward.
2
Insert new lancet
Insert new lancet as far as it
will go into the lancet holder.
Twist off the protective cover.
3
Replace cap
Push the cap onto the lancing
device and check the comfort
dial is set to your level.
4
Prime lancing device
Press the priming button until
it clicks.
5
Press release button
Place Softclix against the side
of fingertip and press the
yellow release button.
TESTING WITH COAGUCHEK XS
HOW TO USE THE COAGUCHEK SOFTCLIX
For questions, phone toll-free 0508 69 5433
The CoaguChek Softclix is not suitable for use in hospitals or
for testing different persons as there is a risk of contamination. For more information visit: www.coaguchek.co.nz
KEY WARNING MESSAGES
The CoaguChek automatically performs a system check to prevent
inaccurate results. If the meter detects a malfunction it displays the
“Error” message and appropriate symbol.
Error 3: Test Strip
Test strip is beyond it’s expiry date. Check correct date and time is set on
meter. If correct, repeat test with new lot of test strips and code chip.
Error 4: Test Strip Unusable
Often caused by applying the drop of blood before the countdown has
begun. Retest making sure you wait for the 180 second countdown before
pricking the finger and applying the drop of blood.
Error 5: Blood Application
Insufficient blood is applied to test strip. Repeat with new test strip.
Error 6: Measurement
Knocking the meter or strip after applying the blood. Repeat with a new test
strip.
Error: Temperature symbol flashing
The meter is too cold (< 15˚C) or warm (> 32˚C) to measure correctly.
CHANGE OF DATE AND TIME
When using the CoaguChek XS for the first time, or following a battery
change, the date and time may need to be set.
Note: The meter will not work if the date and time are not set.
When the instrument is turned on it will automatically enter the set menu (if this
does not occur, press the set button on the left side of the monitor).
The set menu displays each setting option in the following sequence:
• Date Format select: 31/12/00 (DD/MM/YY) Press set
• Date (year) Press M until current year is displayed Press set
• Date (month) Press M until current month is displayed Press set
• Date (day) Press M until current day is displayed Press set
• Time format select: 12 hour Press set
• Time (hour) Press M until correct hour is displayed Press set
• Time (minute) Press M until correct minute is displayed Press set
ALWAYS wash your
hands in warm water
and dry hands before
handling the test
strips.
Use the side of the
finger to obtain blood
- it’s less painful.
After pricking the
finger gently milk it,
don’t squeeze or press
as this may falsify the
coagulation values.
Apply blood to the clear
window on the test strip,
either side application or
top dosing.
Check the expiry date.
Don’t use the strips after
the expiry date.
Always store strips in
the original container
and replace the lid
immediately.
Change your code key
each time you open a
new vial of strips.
TOP TESTING TIPS
CoaguChek XS PT Test Strips
Some Points to Remember to
Ensure Accuracy of
Your INR Test Result
Roche Diagnostics NZ Ltd
PO Box 62 089, 15 Rakino Way
Mt Wellington, Auckland
Phone: 0508 MY LIFE (0508 69 5433)
Fax: 09 276 8917
www.coaguchek.co.nz
Error messages
74
Error messages overview
Error message (always
prefaced by
error
)
Brief description Refer
to
page
Test strip
symbol
flashing
Test strip in monitor or 76
test strip unusable or the test
strip is not a CoaguChek XS PT
test strip
80
Temperature symbol
flashing
Monitor too warm or too cold 76
Battery symbol flashing Battery level too low 78
Cover symbol
flashing
Measurement chamber cover
open
78
IR symbol
flashing
Infrared download error 90
QC symbol flashing Quality control: test strip
unusable
86
Code symbol
flashing
Code chip error 80
QC
code
ccxs_user_EN.book Seite 74 Montag, 11. Juli 2005 7:11 19

Other manuals for XS System

3

Popular Blood Glucose Meter manuals by other brands

Accu-Chek Inform II Guide

Accu-Chek

Accu-Chek Inform II Guide

philosys Gmate On user guide

philosys

philosys Gmate On user guide

GE GE200 Getting started guide

GE

GE GE200 Getting started guide

Bioland G-427B owner's manual

Bioland

Bioland G-427B owner's manual

Bayer HealthCare Breeze2 Quick test guide

Bayer HealthCare

Bayer HealthCare Breeze2 Quick test guide

Accu-Chek Mobile user manual

Accu-Chek

Accu-Chek Mobile user manual

iSens CareSens N Pop manual

iSens

iSens CareSens N Pop manual

Medel Family instruction manual

Medel

Medel Family instruction manual

Gmate VOICE user manual

Gmate

Gmate VOICE user manual

iSens CareSens N Voice owner's booklet

iSens

iSens CareSens N Voice owner's booklet

FOR-A Diamond CUFF BP owner's manual

FOR-A

FOR-A Diamond CUFF BP owner's manual

GlucoRx TD-4277 owner's manual

GlucoRx

GlucoRx TD-4277 owner's manual

TysonBio EZ Smart-608 user manual

TysonBio

TysonBio EZ Smart-608 user manual

Holtex TysonBio TB100 user manual

Holtex

Holtex TysonBio TB100 user manual

GlucoRx TD-4277 Quick start user guide

GlucoRx

GlucoRx TD-4277 Quick start user guide

Pogo Automatic quick start guide

Pogo

Pogo Automatic quick start guide

Bayer HealthCare CONTOUR NEXT LINK user guide

Bayer HealthCare

Bayer HealthCare CONTOUR NEXT LINK user guide

Clarity BG1000 Procedures manual

Clarity

Clarity BG1000 Procedures manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.